MODIFICATION TO THERMOFOCUS 0800, 0900, 01500 AND 0700 SERIES
K072108 · Tecnimed S.R.L. · FLL · Aug 31, 2007 · General Hospital
Device Facts
Record ID
K072108
Device Name
MODIFICATION TO THERMOFOCUS 0800, 0900, 01500 AND 0700 SERIES
Applicant
Tecnimed S.R.L.
Product Code
FLL · General Hospital
Decision Date
Aug 31, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Thermofocus 0700, 0800, 0900, 01500 series are infrared thermometers intended for the intermittent measurement of human body temperature in people of all ages.
Device Story
Hand-held, battery-operated infrared thermometers; measure natural infrared thermal radiation from human body (forehead, navel, neck, or axilla). Non-contact operation; provides temperature reading to user. Used in clinical or home settings by clinicians or patients. Modifications from predicate include improved calibration options for faster stabilization, display backlighting, configurable 'home' button (hospital versions), simplified toggle controls, and adjustable display power-save settings. Benefits include non-invasive, rapid temperature assessment.
Clinical Evidence
No clinical data. Bench testing only; devices meet ASTM E1965-98 standard for infrared thermometers. Risk analysis performed to ensure safety and effectiveness of modifications.
Technological Characteristics
Infrared thermal radiation sensing; battery-operated; hand-held form factor. Complies with ASTM E1965-98. Features include adjustable calibration, display backlighting, and configurable power-save modes.
Indications for Use
Indicated for intermittent measurement of human body temperature in patients of all ages.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
THERMOFOCUS 0800, 0900, 01500 AND 0700 SERIES (K033790)
Submission Summary (Full Text)
{0}------------------------------------------------
K072108
# 510(k) Summary
## Administrative Information and Device Identification
| Name and address of the manufacturer and<br>sponsor of the 510(k) submission: | Tecnimed srl,<br>12 P.le Cocchi,<br>21040 Vedano Olona (VA) Italy |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Official contact person for all correspondence: | Francesco Bellifemine<br>E-mail: info@tecnimed.it |
| Date Prepared: | July 17, 2007 |
| Device Name: | Thermofocus 0700 series, Thermofocus<br>0800 series, Thermofocus 0900 series,<br>Thermofocus 01500 series. |
| Generic name of the device: | Clinical Electronic Thermometer |
| Classification of new device: | Class II |
| Classification Panel: | General Hospital |
| Product Code and CFR Regulation Number: | FLL and 21 CFR 880.2910 |
| Predicate Device Name and 510(k) Number: | THERMOFOCUS 0800, 0900, 01500 AND 0700<br>SERIES<br>K033790 |
#### Description of Device:
Thermofocus devices belonging to the 0800, 0900, 01500, 0700 series are hand-held and battery-operated thermometers. They take skin temperature mainly in the middle of the forehead, on the navel, neck or under the armpit. The Thermofocus devices use the principle of surveying the natural emission of infrared thermal radiation from all objects, including the human body. Thermofocus thermometers take temperature at distance, without any contact with the patient.
{1}------------------------------------------------
The modified devices incorporate improvements to the thermometers calibration capability and also include other minor "user-friendly" improvements. Specifically, these modifications include:
## Comparison of Device Technological Characteristics to Precedent Device:
| Precedent Models | New Models |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 20-30 minute time frame for the<br>device to be stabilized in a new<br>environment. | The user can choose between three different calibration<br>options depending on the model to lessen the amount of<br>time for the stabilization of the thermometer. |
| Backlighting of the display was<br>not available. | Backlighting of the display is now available. |
| No option to disable the home<br>button. | In the hospital versions, it is possible to disable the "home"<br>button. |
| Toggle between settings using 2<br>buttons. | Toggle between settings using one button. |
| In the precedent versions, the<br>display of the model 0700 turned<br>off after 20 seconds of no use.<br>While in all the other models the<br>display remained on. | In the new models, it is possible to choose the display<br>always on or always off when not in use.<br>It is also possible to set models 0800, 0900 and 01500 so<br>that the display can turn off after 20 seconds. |
| Protective cap and memory button<br>on some models | Protective cap and memory button available on all models. |
### Intended Use:
Thermofocus 0700, 0800, 0900, 01500 series are infrared thermometers intended for the intermittent measurement of human body temperature in people of all ages.
### Non-Clinical /Clinical Testing
Performance tests were not required as the modified devices are essentially equivalent to previously cleared predicate devices. Certain performance tests were conducted as part of the risk analysis and demonstrated acceptable results for devices safety and effectiveness.
Both devices meet the ASTM E1965-98 Standard for Infrared Thermometer for Intermittent Determination of Patient Temperature, as far as this standard applies to them.
### Conclusion:
We have demonstrated that the new Themofocus 0700, 0800, 0900, 01500 series are as safe and effective as our predicate device based on performance testing results as well as the risk analysis supplied with this submission.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Tecnimed S.R.L. C/O Ms. Maria F. Griffin Official Correspondent Mdi Consultants, Incorporated 55 Northern Boulevard, Suite 200 Great Neck, New York 11021
AUG 3 1 2007
Re: K072108
Trade/Device Name: Thermofocus 0800 Series, Thermofocus 0900 Series, Thermofocus 01500 Series, Thermofocus 0700 Series Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: July 31, 2007 Received: August 2, 2007
Dear Ms. Griffin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Griffin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97); You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
Page __ 1 __ of __ 1
510(k) Number (if known): Not Assigned as of this time
Device Name: Thermofocus Professional 0800 series, Thermofocus 0900 series, Thermofocus 01500 series, Thermofocus 0700 series
#### Indications for Use:
Thermofocus 0800, 0900, 01500, 0700 Series are infrared thermometers intended for the intermittent measurement of human body temperature in people of all ages.
Prescription Use (Per 21 CFR 801 Subpart D)
OR
Over-The Counter Use_ X (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Qurs
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K072108
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.