PERIOSCAN, MODEL: D 3496

K072107 · Sirona Dental Systems GmbH · ELC · Feb 14, 2008 · Dental

Device Facts

Record IDK072107
Device NamePERIOSCAN, MODEL: D 3496
ApplicantSirona Dental Systems GmbH
Product CodeELC · Dental
Decision DateFeb 14, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Sirona Dental Systems PerioScan is an ultrasonic scaler that aids in the recognition of calculus and intended for use in the following dental and periodontal applications: - . Aiding in subgingival calculus detection - Calculus removal (subgingival and supragingival) . - Retrograde root treatment . - . Endodontics - Insertion of inlays . - . Micropreparation (cavity preparation) . - Periodontology

Device Story

Table-top ultrasonic scaler; aids in subgingival calculus detection and removal. Input: piezo-ceramic handpiece signals. Transformation: ultrasonic power generation; detection-aid-function processes signals to identify calculus. Output: visual (illuminator ring, operator display) and audible (signal tone) indicators for calculus; ultrasonic energy for scaling. Used in dental clinics by practitioners. Operator controls power, irrigation, and mode via footswitch and panel. Detection-aid-function assists practitioner in diagnosis and therapy planning; provides real-time feedback during procedures.

Clinical Evidence

Clinical studies and testing performed to validate the detection-aid-function for recognition of subgingival calculus. No specific performance metrics (sensitivity/specificity) provided in the summary document.

Technological Characteristics

Table-top unit; piezo-ceramic handpiece; 120V mains power; dual internal irrigant containers or external water source. Features illumination ring, signal LEDs, and audible tone for calculus detection. Footswitch-controlled. Software-controlled ultrasonic power delivery. Handpiece includes minor modifications to SiroSonic L for detection signaling.

Indications for Use

Indicated for dental and periodontal applications including subgingival calculus detection, subgingival and supragingival calculus removal, retrograde root treatment, endodontics, inlay insertion, micropreparation, and periodontology. Prescription use only.

Regulatory Classification

Identification

An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 173 K072107 510(k) Summary for Sirona Dental Systems PerioScan # l. SPONSOR Sirona Dental Systems GmbH Fabrikstrasse 31 D-64625 Bensheim Germany 2 mar 1 1 - 1 Contact Person: Fritz Kolle Telephone: 49 6251 16 32 94 Date Prepared: July 25, 2007 ## 2. DEVICE NAME | Proprietary Name: | PerioScan | |----------------------|-------------------| | Common/Usual Name: | Ultrasonic Scaler | | Classification Name: | Ultrasonic Scaler | ## 3. PREDICATE DEVICE Sirona PerioScan Ultrasonic Scaler (K050354) and SiroSonic L Ultrasonic Scaler (K033640). ### 4. INTENDED USE The Sirona Dental Systems PerioScan is an ultrasonic scaler that aids in the recognition of calculus and intended for use in the following dental and periodontal applications: - . Aiding in subgingival calculus detection - Calculus removal (subgingival and supragingival) . - Retrograde root treatment . - . Endodontics - Insertion of inlays . - . Micropreparation (cavity preparation) Sirona Dental 510(k) Summary Perioscan July 25, 2007 . APPENDIX G . Page 1 of 3 {1}------------------------------------------------ K072107 - Periodontology #### 5. DEVICE DESCRIPTION The Sirona PerioScan Ultrasonic Scaler consists of a table-top unit with two detachable irrigant bottles with caps (lids), a footswitch, an ultrasonic handpiece, hoses and connectors, a variety of optional tips for different dental procedures, and a tool for attaching the tips to the handpiece. The PerioScan unit provides ultrasonic scaler therapy as described in the above indications and detection-aid-function for recognition of subgingival calculus. The table-top unit houses the operator input and controls for the supply of irrigant, cooling liquid and light for the handpiece. The device is connected to 120V mains power supply, and it may be connected to a mains water supply. The irrigant for the handpiece is drawn from the two internal liquid containers or from the mains water supply. The handpiece is connected to the table-top unit via a hose and can be deposited in the claw. A footswitch with two single foot pedals is used to control the handpiece. The PerioScan handpiece is an electrically operated scaler driven by a piezo ceramics. The handpiece includes an illumination and two signal LEDs and a Spray water controller, all located in the hose. PerioScan switches between detection and scaling operation mode, if selected, using the footswitch. Detection of subgingival calculus is signaled through an illuminator ring and operator display. Additionally a signal tone may be emitted if desired. # 6. BASIS FOR SUBSTANTIAL EQUIVALENCE The overall design of the Sirona PerioScan Ultrasonic Scaler with detection-aidfunction for recognition of subgingival calculus is similar to the design of the Sirona PerioScan Ultrasonic Scaler (PerioScan Basic) with no detection-aid-function. Both of these devices include a dental handpiece and ultrasonic generator. They all contain software, which controls delivery of the ultrasonic power, and include various tip configurations for the differing dental procedures. On both the proposed and predicate PerioScan Scaler (PerioScan Basic) units, the ultrasonic power can be adjusted on the control unit and delivery of the ultrasonic energy is via a foot control. Both devices allow the operator to select among different therapy operating modes. These modes have different ultrasonic power ranges to Sirona Dental 510(k) Summary Perioscan July 25, 2007 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a handwritten string of characters, which appears to be a combination of letters and numbers. The string reads "KO72107". The characters are written in a cursive style, with some connecting strokes between them. A horizontal line is drawn beneath the string of characters. assist the operator in maintaining the ultrasonic power within an appropriate range for specific applications. Both Systems allow for irrigation using two internal containers (small and large) or an external water source. Irrigation flow is controlled by handpiece, operator's panel and footswitch for the Sirona PerioScan with detection-aid-function as well as for the PerioScan with no such function. The PerioSonic handpiece used with the Sirona PerioScan with detection-aidfunction is the SiroSonic L handpiece with minor alterations for enabling detection capability and signaling. Clinical studies and testing have been performed that validate detection-aid-function for recognition of subgingival calculus for the proposed PerioScan device. Based on the comparison of intended use and technical features, Sirona Dental Systems believes that the PerioScan with detection-aid-function for recognition of subgingival calculus is substantially equivalent to the PerioScan with no such function. The proposed and predicate devices have the same general intended use and principles of operation. The overall design of the proposed and predicate devices is similar. The detection function aids the practitioner in recognizing of subgingival calculus who assesses the diagnosis result and the subsequent therapy. Sirona Dental 510(k) Summary Perioscan July 25, 2007 APPENDIX G . Page 3 of 3 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the symbol. The text is in all caps and is written in a sans-serif font. FEB 1 4 2006 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Fritz Kolle Quality Management/Regulatory Affairs Sirona Dental Systems GmbH Fabrikstrasse 31 D-64625 Bensheim GERMANY Re: K072107 Trade/Device Name: PerioScan Model D 3496 Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: February 1, 2008 Received: February 4, 2008 Dear Mr. Kolle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898, In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Kolle Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Suite y. Michien Ows Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K072107 510(k) Number (if known): Device Name: __________ PerioScan Indications for Use: The Sirona Dental Systems PerioScan Ultrasonic Scaler is an ultrasonic scaler that aids in the recognition of calculus and intended for use in the following dental and periodontal applications: - . Aiding in subgingival calculus detection - . Calculus removal (subgingival and supragingival) - . Retrograde root treatment - . Endodontics - . Insertion of inlays - Micropreparation (cavity preparation) . - . Periodontology Prescription Use _____________________________________________________________________________________________________________________________________________________________ X (Part 21 CFR 801 Subpart D) OR Over-The-Counter Use (21 CFR 807 Subpart C) 8/ (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: KD172107 Sirona Dental Systems 510(k) Perioscan July 25, 2007 Page viii
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...