K072105 · Cardinal Health 303,Inc. · FRN · Dec 28, 2007 · General Hospital
Device Facts
Record ID
K072105
Device Name
ALARIS SYSTEM WITH MMS, MODEL 8XXX
Applicant
Cardinal Health 303,Inc.
Product Code
FRN · General Hospital
Decision Date
Dec 28, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The incorporation of the Medication Management System (MMS) with the Alaris® System (a.k.a. Medley™ System) provides wired or wireless communication between the Alaris® System and external devices. This is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps to enter infusion data. All data entry and validation of infusion parameters using MMS is performed by the trained healthcare professional. The Alaris System with MMS is integrated into an existing hospital network infrastructure and allows communications to and from external devices, including personal computers, Personal Digital Assistants (PDAs), hospital monitoring systems and Hospital Information Management Systems (HIMS). Bi-directional communication of data includes infusion parameters, system configuration, history, events, trending, alarms and status. In addition the Alaris System with MMS has the capability to transmit, receive and/or store: features, calibration data, datasets, and libraries from external components or devices to the pump.
Device Story
Alaris System with MMS facilitates wireless/wired communication between infusion pumps and hospital network infrastructure (servers, PCs, PDAs, HIMS). Device acts as a bedside safety net by automating infusion parameter programming, reducing manual data entry steps. Trained healthcare professionals operate the system, performing all data entry and validation. System enables bi-directional data exchange, including infusion parameters, system configuration, history, events, trending, alarms, and status. It also transmits/receives/stores calibration data, datasets, and libraries. By automating programming, the device aims to reduce manual programming errors, potentially improving patient safety during fluid, medication, and blood product delivery.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Infusion pump system with integrated wired/wireless communication modules. Connectivity includes integration with hospital network infrastructure (HIMS, PCs, PDAs). System supports bi-directional data transfer for infusion parameters, configuration, and device libraries. Fundamental scientific technology and materials are identical to the predicate device.
Indications for Use
Indicated for use in professional healthcare facilities for the delivery of fluids, medications, blood, and blood products via infusion devices. Intended for operation by trained healthcare professionals to automate infusion parameter programming based on physician orders.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Predicate Devices
Alaris® System with MMS (a.k.a. Medley™ System with MMS) (K030459)
Submission Summary (Full Text)
{0}------------------------------------------------
CARDINAL HEALTH
SPECIAL 510(K) ALARIS® SYSTEM WITH MMS
# ATTACHMENT D
K072105
# SUMMARY OF SAFETY AND EFFECTIVENESS Cardinal Health 303, Inc. Alaris® System with MMS
| SUBMITTER'S NAME: | Cardinal Health 303, Inc.<br>d.b.a. Cardinal Health<br>10221 Wateridge Circle<br>San Diego, CA 92121-2772<br>(858) 458-7830<br>(858) 458-6114 FAX | DEC 2 8 2007 |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| CONTACT PERSON: | Stacy L. Lewis<br>Principal Regulatory Affairs Specialist | |
| DATE PREPARED: | July 27, 2007 | |
| DEVICE NAME: | Proprietary Name<br>Alaris® System with MMS | |
| | Common Name<br>Infusion Pump | |
| | Classification Name<br>Pump, Infusion and accessories, FRN (880.5725) | |
| PREDICATE DEVICE: | Alaris® System with MMS (a.k.a. Medley™ System with<br>MMS) K030459 | |
#### DEVICE DESCRIPTION
As with the predicate device, the Alaris® System with MMS assists our customers in reducing the number of manual steps needed to program an infusion by allowing wireless communication capability to a server and an existing infusion device (the Alaris® System). The wireless communication capability provides a "safety net" at the bedside to help reduce the number of programming errors at the point of care. It allows the Alaris® System to transmit and receive messages with the Alaris® Server which in turn allows communication capability with external devices, including personal computers, Personal Digital Assistants (PDA;s), hospital monitoring systems and Hospital Information Management Systems (HIMS).
{1}------------------------------------------------
# ALARIS® SYSTEM WITH MMS SUMMARY OF SAFETY AND EFFECTIVENESS Page 2 of 2
#### SUBSTANTIAL EQUIVALENCE
The Alaris® System with MMS is essentially the same as the predicate device in that they have the same intended use, operating principles, method of operation, technology, materials and manufacturing processes. The changes as described in this Special 510(k) pose no new issues of safety or efficacy. The Alaris® System with MMS as described in this submission is substantially equivalent to the predicate device.
#### INTENDED USE
The incorporation of the Medication Management System (MMS) with the Alaris® System (a.k.a. Medley™ System) provides wired or wireless communication between the Alaris® System and external devices. This is intended to provide a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps to enter infusion data. All data entry and validation of infusion parameters using MMS is performed by the trained healthcare professional.
The Alaris System with MMS is integrated into an existing hospital network infrastructure and allows communications to and from external devices, including personal computers, Personal Digital Assistants (PDAs), hospital monitoring systems and Hospital Information Management Systems (HIMS). Bi-directional communication of data includes infusion parameters, system configuration, history, events, trending, alarms and status. In addition the Alaris System with MMS has the capability to transmit, receive and/or store: features, calibration data, datasets, and libraries from external components or devices to the pump.
## TECHNOLOGICAL CHARACTERISTICS
A comparison of the technological characteristics of the Alaris® System with MMS and the predicate device has been performed. The results of this comparison demonstrate that the Alaris® System with MMS is equivalent in technological characteristics and the fundamental scientific technology of the predicate device has not been altered.
$$
\begin{array}{ccc}
\mathsf{k}\mathsf{o}\mathsf{7}\mathsf{z}\mathsf{o}\mathsf{6} & \mathsf{p}\mathsf{4} & \mathsf{z}\mathsf{o}\mathsf{z} \\
\mathsf{k}\mathsf{z}\mathsf{o}\mathsf{z} & \mathsf{p}\mathsf{z} & \mathsf{z}\mathsf{o}\mathsf{z} \\
\end{array}
$$
00768
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines representing the serpent winding around the staff.
DEC 2 8 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Stacey L. Lewis Principal Regulatory Affairs Specialist Cardinal Health 303, Incorporated 10221 Wateridge Circle San Diego, California 92121
Re: K072105
Trade/Device Name: Alaris System with MMS Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: December 3, 2006 Received: December 4, 2006
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Lewis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Shiela A Murphy, MD for Chu Lin, Ph D
Chu Lin, Ph D
Chiu Lin, Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
CARDINAL HEALTH
# Attachment B
# INDICATIONS FOR USE
510(k) Number:
12) 05
(To Be Assigned By FDA)
Device Trade Name:
Alaris® System with MMS
### Indications For Use:
The Alaris® System with Medication Management System (also known as the Medley™ System with MMS and the Alaris® System with MMS) is intended for use in today's growing professional healthcare environment for facilities that utilize infusion devices for the delivery of fluids, medications, blood and blood products.
The Alaris® System with MMS is intended to provide trained healthcare caregivers a way to automate the programming of infusion parameters, thereby decreasing the amount of manual steps necessary to enter infusion data. All data entry and validation of infusion parameters is performed by the trained healthcare professional according to a physician's order.
Prescription Use X
Over-Th
Over-The-Counter Use
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Per 21 CFR 801.109)
Clint Damer
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K472145
00763
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.