IGG ANTI-ATHEROX TEST KIT
Device Facts
| Record ID | K072032 |
|---|---|
| Device Name | IGG ANTI-ATHEROX TEST KIT |
| Applicant | Corgenix, Inc. |
| Product Code | MSV · Immunology |
| Decision Date | Apr 4, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5660 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K072032 · Apr 4, 2008 | IGG ANTI-ATHEROX TEST KIT | Corgenix, Inc. | Retrospective clinical serum samples from patient cohorts (healthy controls, rheumatoid arthritis, and systemic lupus erythematosus) | Retrospective clinical samples were used to compare the performance of the IgG Anti-AtherOx test kit against the REAADS IgG anti-cardiolipin predicate device to establish clinical agreement. | Retrospective clinical samples; Clinical agreement; Diagnostic performance |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Clinical performance evaluation; Retrospective comparative study | 205 healthy controls, 99 patients with rheumatoid arthritis, and 143 patients with systemic lupus erythematosus; Sample Size: 447 total samples | REAADS IgG anti-cardiolipin test kit | Positive/negative percent agreement and overall percent agreement |
Indications for Use
An enzyme-linked immunoassay (ELISA) for the detection of IgG antibodies to complexes formed by oxidized low-density lipoprotein (oxLDL) with β2-glycoprotein I (β2GPI) in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (antiphospholipid syndrome).
Device Story
The IgG Anti-AtherOx Test Kit is an indirect ELISA used in clinical laboratories to detect IgG antibodies against oxLDL/ß2GPI complexes in human serum. The device utilizes a 96-well polystyrene microtiter plate coated with oxLDL/ß2GPI complex. Patient serum is incubated in the wells, allowing specific IgG antibodies to bind to the immobilized antigen. After washing, HRP-conjugated anti-human IgG is added, followed by a TMB/hydrogen peroxide substrate. The resulting color intensity, measured via spectrophotometer, is proportional to the concentration of IgG anti-oxLDL/ß2GPI antibodies. Results are calculated using a log-log regression analysis against provided calibrators. By identifying these autoantibodies, the test assists clinicians in assessing the risk of venous and arterial atherothrombosis in patients with autoimmune conditions like SLE and APS, potentially aiding in the management of hypercoagulability.
Clinical Evidence
Clinical evaluation compared the device against the REAADS Anti-Cardiolipin test using 448 samples (205 healthy controls, 99 rheumatoid arthritis, 143 SLE). Overall agreement was 90.2%. In SLE patients, positive percent agreement was 76.0% and negative percent agreement was 86.4%. In patients with secondary APS, positive percent agreement was 73.9% and negative percent agreement was 88.7%. Subgroup analysis for arterial thrombosis showed 90.5% overall agreement.
Technological Characteristics
Indirect ELISA; 96-well polystyrene microtiter plate coated with oxLDL/ß2GPI complex; HRP-conjugated goat polyclonal anti-human IgG detection; TMB/H2O2 chromogenic substrate; 0.36 N sulfuric acid stop solution; spectrophotometric readout; high-complexity laboratory test; manual/automated processing.
Indications for Use
Indicated for detection of IgG antibodies to oxLDL/β2GPI complexes in individuals with systemic lupus erythematosus (SLE) and lupus-like disorders (antiphospholipid syndrome). For prescription use only.
Regulatory Classification
Identification
A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).
Predicate Devices
- REAADS Anti-Cardiolipin IgG/IgM Semi-Quantitative kit (k022992)