VASCULAR OCCLUSION SYSTEM (TRUFILL PUSHABLE COILS AND TRUPUSH COIL PUSHER)
Applicant
Cordis Neurovascular, Inc.
Product Code
HCG · Neurology
Decision Date
Sep 26, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the TRUFILL® Vascular Occlusion System (TRUFILL® Pushable Coils and TRUPUSH® Coil Pusher) is as follows: Pushable Coils may be used to reduce or block the rate of blood flow in vessels of the peripheral and neurovasculature. They are intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulas, and other vascular lesions of the brain, spinal cord, and spine.
Device Story
Vascular Occlusion System consists of TRUFILL® Pushable Coils and TRUPUSH® Coil Pusher. Device used in interventional radiology for embolization procedures in neuro and peripheral vascular systems. Physician loads embolic coil into proximal end of microcatheter; uses Coil Pusher to advance coil through catheter lumen until it exits distal tip. System facilitates mechanical occlusion of vascular lesions; benefits patient by reducing or blocking blood flow to malformations or fistulas. Device is manual; no electronic or software components.
Clinical Evidence
No clinical data. Substantial equivalence supported by in-vitro and in-vivo design verification and validation testing, including mechanical integrity (joint strength, turns to failure), packaging integrity, and functional performance (pushability, retractability).
Technological Characteristics
Manual vascular embolization delivery system. Components include corewire, PTFE sleeve, and solder joints. Materials include PTFE. No energy source, software, or connectivity. Sterilization method not specified. Dimensions defined by proximal/distal joint ODs and taper lengths.
Indications for Use
Indicated for patients requiring reduction or blockage of blood flow in peripheral and neurovascular vessels; specifically for interventional radiologic management of arteriovenous malformations, arteriovenous fistulas, and other vascular lesions of the brain, spinal cord, and spine.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
K071962
1/3
# 0-000011
Johnson-Johnson company
## Page 1 of 3 CFF 10463867 Revision 02 Only the Native File to be Used
SEP 2 6 2007
# 510(k) Summary of Safety and Effectiveness
| Submitter's Name/Contact Person | Amarilys Machado<br>Manager, Regulatory Affairs<br><br>Cordis Neurovascular, Inc.<br>14000 NW 57th Court<br>Miami Lakes, Florida 33014<br><br>Ph. 786 313 6493<br>Fax. 786 313 6480<br>Amachad2@crdus.jnj.com |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name / Common Name | The trade name is :<br>Vascular Occlusion System (TRUFILL® Pushable Coils and TRUPUSH® Coil Pusher)<br><br>The common name is:<br>Artificial Embolization Device |
| Classification | This is a Class II Device, per 21 CFR 882.5950 (84HCG) |
| Performance Standard | The FDA under Section 514 of the Food, Drug and Cosmetic Act has not established performance standards for this device. |
| Intended use | The intended use of the TRUFILL® Vascular Occlusion System (TRUFILL® Pushable Coils and TRUPUSH® Coil Pusher) is as follows:<br><br>Pushable Coils may be used to reduce or block the rate of blood flow in vessels of the peripheral and neurovasculature. They are intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulas, and other vascular lesions of the brain, spinal cord, and spine. |
| Device Description | The TRUPUSH® Coil Pusher is used in coil embolization procedures in both the neuro and peripheral vascular systems. An embolic coil is loaded into the proximal end of a compatible microcatheter and the Coil Pusher is used to advance the device through the catheter lumen until it has exited the distal tip of the catheter. |
{1}------------------------------------------------
**Cordis**®
a Johnson & Johnson company
KO 71962
->/3
Page 2 of 3 CFF 10463867 Revision 02 Only the Native File to be Used
Predicate Devices The predicate devices are listed in the table below:
| Device | Company | Product Code | 510(k) Number | Predicate for:<br>(if multiple predicates) |
|-------------------------|-------------------------------|--------------|---------------|----------------------------------------------------------------------------------------------------------------------------------------|
| TRUPUSH®<br>Coil Pusher | Cordis<br>Neurovascular, Inc. | K983483 | HCG | - Intended Use<br>- Design<br>- Operating<br>Principle<br>- Materials<br>- Sterilization<br>- Shelf Life<br>- Manufacturing<br>Process |
0-000012
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Cordis, a Johnson & Johnson company. The word "Cordis" is written in a bold, stylized font, with a small registered trademark symbol to the right. Below "Cordis" is the text "johnson&johnson company" in a smaller, less bold font. The logo is simple and professional, reflecting the company's brand identity.
KU71962
5/3
Page 3 of 3 CFF 10463867 Revision 02 Only the Native File to be Used
#### Summary of Studies
The in-vitro and in-vivo testing was conducted to support substantial equivalence to the predicate device, addressing the process changes, packaging changes and manufacturing site transfer.
#### Design Verification Testing:
- . Total Product Length
- � Taper Length
- OD of the Proximal Joint
- . OD of the Distal Joint
- . OD of the Proximal End of the PTFE Sleeve
- . Proximal Marker Position
- Distal Placement of the PTFE Sleeve
- . PTFE Sleeve Visual Inspection
- . Distal End of PTFE Sleeve Visual Inspection
- . Overall Visual Inspection
- . Distance from the Proximal End of the PTFE Sleeve to the Corewire Taper
- t Distal Solder Joint Length
- . Distal Solder Joint Visuals
- . Turns to Failure
- . Distal Joint Strength
- . Package Integrity
- . Package Challenge
- Seal Integrity Test (Dye Penetration) .
- Packaging-Seal Strength (Pull Test) .
- . Product Migration
#### Design Validation Testing:
- 3-cavity torque device clip .
- Pushability .
- Retractability .
Summary of Substantial Equivalence
The proposed TRUPUSH® Coil Pusher is similar in its basic design, construction, indication for use, and performance characteristics to the predicate device the current TRUPUSH® Coil Pusher.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Cordis Neurovascular, Inc. % Amarilys Machado Regulatory Affairs Manager 14000 Northwest, 57th Court Miami Lakes, Florida 33014
SEP 26 2007
Re: K071962
Trade/Device Name: Vascular Occlusion System (TRUFILL® Pushable Coils and TRUPUSH® Coil Pusher) Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular embolization device Regulatory Class: II Product Code: HCG Dated: August 27, 2007 Received: August 28, 2007
Dear Amarilys Machado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -- Amarilys Machado
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Ftz
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
0-000009
Image /page/5/Picture/2 description: The image shows the word "Corais" in a bold, stylized font. Below the word "Corais" is the text "a Johnson-Johnson company" in a smaller, cursive font. The word "Corais" is the main focus of the image, with the company name providing context.
Page 1 of 1 CFF 20014 Only the Native File to be Used Intended Use Statement
Page 1 of 1
## Indications for Use
510(k) Number (if known):
Device Name: TRUPUSH® Coil Pusher
### Indications For Use:
The intended use of the TRUFILL® Vascular Occlusion System (TRUFILL® Pushable Coils and TRUPUSH® Coil Pusher) is as follows:
Pushable Coils may be used to reduce or block the rate of blood flow in vessels of the peripheral and neurovasculature. They are intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulas, and other vascular lesions of the brain, spinal cord, and spine.
J
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Prescription Use X (Part 21 CFR 801 Subpart D)
1071962
AND/0B10(k)Number-Counter Use
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.