The Invacare Flyer is intended to be used by patients with respiratory disorders who require supplemental oxygen. It supplies a high concentration of supplemental oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Invacare Flyer can be used in a home, institution, vehicle and various mobile environments. The Invacare Flyer does not nor is it intended to sustain or support life.
Device Story
Portable oxygen concentrator; uses molecular sieve and pressure swing adsorption to extract oxygen from ambient air; delivers pulsed demand flow (settings 1-5) via nasal cannula. Operated by patient in home, institution, or vehicle. Device detects inhalation to trigger oxygen pulse. Provides visual/audible alerts for battery status, breath detection, system temperature, and mechanical faults. Benefits patients requiring supplemental oxygen by providing mobile, high-concentration oxygen source. Healthcare provider uses output to manage respiratory therapy; device does not sustain life.
Clinical Evidence
Bench testing only. Performance data confirms oxygen purity (87-95.6%) and compliance with electrical safety and electromagnetic compatibility standards (IEC 60601-1, IEC 60601-1-2, etc.). No clinical data presented.
Indicated for patients with respiratory disorders requiring supplemental oxygen. Not intended to sustain or support life. Used with nasal cannula in home, institutional, or mobile environments.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
### 510(K) SUMMARY FOR THE INVACARE FLYER, Model XPO100
This summary of 510(k) safety and effectiveness information is being supplied in accordance with the requirements of the SMDA of 1990 and 21 CFR 807.92
The assigned 510(k) number is ________________________________________________________________________________________________________________________________________________
Date: July 10, 2007
Submitted by: Invacare Corporation Registration No. 1525712 One Invacare Way Elyria, Ohio 44035-4190
DEC 1 2 2007
Telephone: 440-326-6356 Fax: 440-326-3458
Contact Person: Mr. Carroll Martin
Trade Name: The Flyer
Model: XPO100
Common Name: Oxygen Concentrator
Classification Name: Generator, Oxygen, Portable per 21 CFR 868.5440
Legally Marketed Predicate Device(s): Inogen One Oxygen Concentrator; K032818, May 13, 2004
Device Description: The Invacare Flyer is used by patients with respiratory disorders who require supplemental oxygen. The device can be used in the home, an institutional environment or in a vehicle or other mobile environment. The device is not intended to sustain or support life. The device is used with a nasal cannula to direct oxygen from the device to the patient.
The Invacare Flyer provides oxygen in pulsed demand flow dosages at settings of 1 through 5. The oxygen concentration level of the output gas ranges from 87% to 95.6%.
Standard power options include an AC to DC switching power supply operating from AC power outlet (120 VAC/ 60 Hertz nominal), a DC to DC switching power supply operating from accessory outlets typically found in a mobile vehicle type environment (12 VDC nominal) and an external rechargeable battery.
The Invacare Flyer uses a molecular sieve and pressure swing adsorption methodology to produce the oxygen gas output. Ambient air enters the device, is filtered and then compressed. This compressed air is then directed toward one of two nitrogen adsorbing sieve beds. Concentrated oxygen exits the
510(k) K071928 Response
Page 18
{1}------------------------------------------------
opposite end of the active sieve bed and is directed into an oxygen reservoir where it is delivered to the patient in specific volumes during the inhalation portion of a detected breath.
The basic technology of the Invacare Flyer is equivalent to other approved oxygen concentrators. The principles of operation are equivalent to the noted predicate device.
Intended Use: The Invacare Flyer is intended to be used by patients with respiratory disorders who require supplemental oxygen. It supplies a high concentration of supplemental oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Invacare Flyer can be used in a home, institution, vehicle and various mobile environments. The Invacare Flyer does not nor is it intended to sustain or support life.
| Features | The Invacare Flyer | Predicate Device<br>Inogen One Oxygen Concentrator |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K032818 |
| Date Cleared | TBD | 05/13/2004 |
| Intended Use | The Invacare Flyer is intended to be used<br>by patients with respiratory disorders who<br>require supplemental oxygen. It supplies a<br>high concentration of supplemental<br>oxygen and is used with a nasal cannula to<br>channel oxygen from the concentrator to<br>the patient. The Invacare Flyer can be<br>used in a home, institution, vehicle and<br>various mobile environments. The<br>Invacare Flyer does not nor is it intended<br>to sustain or support life. | The Inogen One Oxygen Concentrator is<br>used on a prescriptive basis by patients<br>requiring supplemental oxygen. It<br>supplies a high concentration of oxygen<br>and is used with a nasal cannula to<br>channel oxygen from the concentrator to<br>the patient. The Inogen One Oxygen<br>Concentrator may be used in a home.<br>institution, vehicles and various mobile<br>environments. |
| Method by which<br>Oxygen is Released | Molecular sieve (mechanical) | Molecular sieve (mechanical) |
| Process by which<br>Oxygen is Released | Pressure swing adsorption | Pressure swing adsorption |
| Sieve Bed Material | Synthetic zeolite | Synthetic zeolite |
| Software/Hardware | Analog and digital electronics with<br>microprocessor | Analog and digital electronics with<br>microprocessor |
| Flow Control | Microprocessor controlled valves | Microprocessor controlled valves |
| Weight | < 7.0 lbs. ( with non-removable internal<br>battery) | 9.8 lbs. (device 8.3 lbs .; battery 1.5 lbs.) |
| Size | | |
| Height | 10.0" +/- 1" | 12.39" |
| Width | 7.0" +/- 1" | 6.00" |
| Depth | 4.3" +/- 1" | 11.62" |
| Oxygen Purity | 87% minimum at all flow rates | "Approximately 90%" |
### Substantial Equivalence:
510(k) K071928 Response
{2}------------------------------------------------
| Flow Rates | Pulse Flow - 1, 2, 3, 4, 5 | Pulse Flow - 1, 2, 3, 4, 5 |
|---------------|-----------------------------------------------------------------|--------------------------------------------------------------------------------|
| Power Options | AC power supply 100-240 V, 50/60 Hz;<br>DC power supply 11-16 V | AC power supply 100-240V, 50/60 Hz;<br>Mobile Power Charger for mobile DC use. |
| Battery | Li-ion, 14.8 Volt, 5.2 Ah | Li-ion, 14.8 Volt, 6.6 Ah |
As the chart above shows, the Invacare Flyer is comparable to its predicate. Both devices have the same intended use for the same patient population, extract oxygen from air using the same methodology, provide comparable oxygen purity and are powered in the same manner.
### Design Standards: The Invacare Flyer has been designed referencing the following standards:
| Standard or Agency | Title |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM 1464-93 (2005) | Standard Specification for Oxygen Concentrators for Domiciliary Use |
| ISO 8359; 1996 | Oxygen Concentrators for Medical Use - Safety Requirements |
| UL 1097, 5th ed., 2004 | UL Standard for Safety for Double Insulation Systems for Use in<br>Electrical Equipment, UL 1097 |
| UL 1431, 2nd ed., 1996 | UL Standard for Safety for Personal Hygiene and Health Care<br>Appliances, UL1431 (with revisions from 2003) |
| EN 55011:1998 | Limits and methods of measurement of radio disturbance characteristics<br>and medical (ISM) radio-frequency equipment |
| IEC 60601-1; 3rd ed. 2006 | Medical Electrical Equipment - Part 1 - General requirements for basic<br>safety and essential performance |
| IEC 60601-1-2; 3rd ed. 2007 | Medical Electrical Equipment -- Part 1-2 -- Collateral standard:<br>Electromagnetic compatibility |
| IEC 61000-3-2: 2005 | Electromagnetic compatibility (EMC). Limits. Limits for harmonic<br>current emissions (equipment input current up to and including 16 A per<br>phase) |
| EN 61000-3-3: 2005 | Electromagnetic compatibility (EMC). Limits. Limitation of voltage<br>changes, voltage fluctuations and flicker in public low-voltage supply<br>systems, for equipment with rated current <= 16 A per phase and not<br>subject to conditional connection |
| IEC61000-4-2; 1.2 ed. 2001 | Electrostatic Discharge Immunity Test w/ Amendments 1&2 |
| IEC61000-4-3; 3rd ed. 2006 | Radiated, RF, Electromagnetic Field Immunity Test |
0(k) K071928 Respon
{3}------------------------------------------------
IEC61000-4-5; 2nd ed. 2005 Surge Immunity Test
| IEC61000-4-6; 1st ed. 2006 | Immunity to Conducted Disturbances Induced by RF fields w/<br>Amendments 1&2 |
|-----------------------------|-------------------------------------------------------------------------------------------|
| IEC61000-4-8; 1.1 ed. 2001 | Power Frequency Magnetic Field Immunity Tests |
| IEC61000-4-11; 4th ed. 2004 | Voltage Dips, Short Interruptions and Voltage Variations Immunity<br>Tests w/ Amendment 1 |
| FDA Reviewer's Guide 93 | Reviewer Guidance for Premarket Notification Submissions |
# Performance Testing: The Invacare Flyer has been tested in accordance with the following:
| Standard or Agency | Title |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| EN 60601-1-2; 2nd ed. 2001 | Medical Electrical Equipment - Sec 1.2 - Collateral standard:<br>Electromagnetic compatibility |
| IEC 60601-1-2; 2.1 ed. | Medical Electrical Equipment - Sec 1.2 - Collateral standard:<br>Electromagnetic compatibility |
| EN 55011:1998 | Limits and methods of measurement of radio disturbance characteristics<br>and medical (ISM) radio-frequency equipment |
| CISPR 11: 2003 | Limits and methods of measurement of radio disturbance characteristics<br>and medical (ISM) radio-frequency equipment |
| EN61000-4-3; 2002<br>IEC61000-4-3; 2006 | Radiated, RF, Electromagnetic Field Immunity Test |
| EN 61000-4-4; 1995<br>IEC 61000-4-4; 2004 | Electrical Fast Transient/Burst Immunity Test |
| EN 61000-4-5; 1995<br>IEC 61000-4-5; 2005 | Surge Immunity Test |
| EN 61000-4-6; 1996<br>IEC 61000-4-6; 2006 | Immunity to Conducted Disturbances Induced by RF fields |
| EN 61000-4-8; 2001 | Power Frequency Magnetic Field Immunity Tests |
{4}------------------------------------------------
| EN 61000-4-11; 1995 | Voltage Dips, Short Interruptions and Voltage Variations Immunity<br>Tests |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| IEC 61000-4-11; 2004 | Voltage Dips, Short Interruptions and Voltage Variations Immunity<br>Tests |
| EN 61000-3-2: 2000 | Limits for Harmonic Current Emissions |
| IEC 61000-3-2: 2005 | Limits for Harmonic Current Emissions |
| EN 61000-3-3: 1995 | Voltage fluctuations and flicker Tests |
| IEC 61000-3-3: 2005 | Voltage fluctuations and flicker Tests |
| IEC 60601-1: 2nd ed. 1998 | Medical Electrical Equipment, Part 1: General requirements for basic<br>safety and performance |
| with Amendment 1: 1991 | |
| with Amendment 2: 1995 | |
| UL 60601-1: 1st ed. 2003 | Medical Electrical Equipment, Part 1: General requirements for basic<br>safety and performance |
| CSA 601.1 M90 | Medical Electrical Equipment, Part 1: General requirements for basic<br>safety and performance |
| Quasi-Static Electric Field | Reviewer Guidance for premarket notification submissions 1993 |
| FDA Reviewer's Guide:1993 Electrostatic Discharge Immunity Test per Sec. m,7,ii,a. | |
| FDA Reviewer's Guide:1993 Reviewer Guidance for Premarket Notification Submissions | |
Performance Data: The performance data found in this submission shows that the Invacare Flyer performs as intended and in a manner that is substantially equivalent to the predicate device.
Conclusion: The data presented in this submission shows that the Invacare Flyer performs as intended and in a manner that is substantially equivalent to the predicate devices.
{5}------------------------------------------------
## Executive Summary
## Device Description
The Invacare Flyer is used by patients with respiratory disorders who require supplemental oxygen. The device is not intended to sustain or support life.
The basic technology of the Invacare Flyer is equivalent to other approved oxygen concentrators. The principles of operation are equivalent to the noted predicate device. The oxygen concentration level of the output gas ranges from 87% to 95.6%. The oxygen is delivered to the patient through the use of a nasal cannula. When the demand for oxygen is detected, the oxygen is delivered through pulsed flow with pulse flow settings of 1 through 5.
The Invacare Flyer uses a molecular sieve and pressure swing adsorption methodology to produce the oxygen gas output. Ambient air enters the device, is filtered and then compressed. This compressed air is then directed toward one of two nitrogen adsorbing sieve beds. Concentrated oxygen exits the opposite end of the active sieve bed and is directed into an oxygen reservoir where it is delivered to the patient in specific volumes during the inhalation portion of a detected breath.
The Invacare Flyer is capable of operation by the patient in a home environment, in an institutional environment, in a vehicle or other mobile environment. Device standard power options include an AC to DC switching power supply operating from AC power outlet (120 VAC/ 60 Hertz nominal), a DC to DC switching power supply operating from accessory outlets typically found in a mobile vehicle type environment (12 VDC nominal) and an external rechargeable battery.
The Invacare Flyer has several lights and indicators providing the user with information regarding the operation of the device:
- 1. Start-Up Indicator the Flyer will briefly light all LEDs and output a brief audible beep when first turned on to indicate the unit is operational.
- 2. Breath Detect Indicator the Flyer will output a defined pulse of oxygen each time it detects an inhalation. The blue flow setting indicator light will blink each time an inhalation is detected and the Flyer outputs a pulse of oxygen.
- 3. Internal Battery Charging Indicator the battery gauge will be illuminated during charging to indicate the internal batter charge status. When the battery is fully charged, all four battery gauge display LEDs will illuminate continuously.
- 4. Low Battery Warning when the Flyer's internal battery (or external battery, if applicable) charge level falls to 25% capacity, a brief audible beep will sound and the lowest level battery gauge display LED will blink.
- 5. Low Battery Alarm when the Flyer's internal battery (or external battery, if applicable) charge level falls to 15% capacity, a brief audible double beep will sound and the lowest level battery gauge display LED will blink rapidly..
{6}------------------------------------------------
- 6. Battery Discharged Alarm when the Flyer's internal battery (or external battery, if applicable) charge level falls to a minimum capacity, a brief audible triple beep will sound, the lowest level battery gauge display LED will blink very rapidly and the unit will shutdown.
- 7. No Breath Detect Alarm when the Flyer is operating but does not sense breathing within a predetermined time period (e.g 60 seconds), a constant audible alarm sounds and the red alarm indicator light will illuminate continuously.
- 8. Breath Rate Over Capacity Alarm if the breathing rate exceeds the capacity of the Flyer, a rapid audible beep sounds and the red alarm indicator light will illuminate intermittently.
- 9. System Too Hot/Cold for Start Alarm if the Flyer senses temperatures outside factory set levels upon start-up, the unit will alarm with a rapid audible beep, the red alarm indicator light will illuminate continuously, the unit will not operate and the flow setting 1 & 2 blue indicator lights will illuminate.
- 10. System Too Hot/Cold Running Alarm if the Flyer senses temperatures outside factory set levels during operation, the unit will alarm with a rapid audible beep, the red alarm indicator light will illuminate continuously, the unit will not operate and the flow setting 1 & 3 blue indicator lights will illuminate.
- 11. Battery Too Hot/Cold Alarm if the Flyersinternal battery senses temperatures outside a factory defined temperature range while the unit is operating, the unit will alarm with a rapid audible beep, the red alarm indicator light will illuminate continuously, the unit will stop running and the flow setting 1 & 4 blue indicator lights will illuminate.
- 12. Stuck Button Alarm During power up of the unit, if it detects that a button is stuck or being pressed too early, the unit will alarm with a rapid audible beep, the red alarm indicator light will illuminate continuously, the unit will not operate and the flow setting 1 & 5 blue indicator lights will illuminate.
- 13. Operating Alarm If units detects abnormal operating conditions in the unit (such as over and under pressure), the unit will alarm with a rapid audible beep, the red alarm indicator light will illuminate continuously, the unit will not operate and the flow setting 3 & 4 blue indicator lights will illuminate.
- 14. Compressor Alarm If units detects abnormal compressor conditions (such as Locked Rotor or RPM error), the unit will alarm with a rapid audible beep, the red alarm indicator light will illuminate continuously, the unit will not operate and the flow setting 4 & 5 blue indicator lights will illuminate.
- 15. System Alarm If units detects abnormal system conditions (such as Watchdog Timer error), the unit will alarm with a rapid audible beep, the red alarm indicator light will illuminate continuously, the unit will not operate and the flow setting 3,4 & 5 blue indicator lights will illuminate.
Intended Use: The Invacare Flyer is intended to be used by patients with respiratory disorders who require supplemental oxygen. It supplies a high concentration of supplemental oxygen and is used with a
510(k) K071928 Response
{7}------------------------------------------------
nasal cannula to channel oxygen from the concentrator to the patient. The Invacare Flyer can be used in a home, institution, vehicle and various mobile environments. The Invacare Flyer does not nor is it intended to sustain or support life.
| Features | The Invacare Flyer | Predicate Device<br>Inogen One Oxygen Concentrator |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K032818 |
| Date Cleared | TBD | 05/13/2004 |
| Intended Use | The Invacare Flyer is intended to be used<br>by patients with respiratory disorders who<br>require supplemental oxygen. It supplies a<br>high concentration of supplemental<br>oxygen and is used with a nasal cannula to<br>channel oxygen from the concentrator to<br>the patient. The Invacare Flyer can be<br>used in a home, institution, vehicle and<br>various mobile environments. The<br>Invacare Flyer does not nor is it intended<br>to sustain or support life. | The Inogen One Oxygen<br>Concentrator is used on a prescriptive<br>basis by patients requiring<br>supplemental oxygen. It supplies a<br>high concentration of oxygen and is<br>used with a nasal cannula to channel<br>oxygen from the concentrator to the<br>patient. The Inogen One Oxygen<br>Concentrator may be used in a home,<br>institution, vehicles and various<br>mobile environments. |
| Method by which<br>Oxygen is<br>Released | Molecular sieve (mechanical) | Molecular sieve (mechanical) |
| Process by which<br>Oxygen is<br>Released | Pressure swing adsorption | Pressure swing adsorption |
| Sieve Bed<br>Material | Synthetic zeolite | Synthetic zeolite |
| Software/<br>Hardware | Analog and digital electronics with<br>microprocessor | Analog and digital electronics with<br>microprocessor |
### Device Comparison Table
{8}------------------------------------------------
| Features | The Invacare Flyer | Predicate Device<br>Inogen One Oxygen Concentrator |
|---------------|-----------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Flow Control | Microprocessor controlled valves | Microprocessor controlled valves |
| Weight | < 7.0 lbs. (with internal battery) | 9.8 lbs. (device 8.3 lbs., insertable<br>battery 1.5 lbs.) |
| Size: | | |
| Height | 10.0" +/- 1" | 12.39" |
| Width | 7.0" +/- 1" | 6.00" |
| Depth | 4.3" +/- 1" | 11.62" |
| Oxygen Purity | 87% minimum at all flow rates | "Approximately 90%" |
| Flow Rates | Pulse Flow - 1, 2, 3, 4, 5 | Pulse Flow - 1, 2, 3, 4, 5 |
| Power Options | AC power supply 100-240 V, 50/60 Hz;<br>DC power supply 11-16 V | AC power supply 100-240V, 50/60<br>Hz; Mobile Power Charger for mobile<br>DC use. |
| Battery | Li-ion, 14.8 Volt, 5.2 Ah | Li-ion, 14.8 Volt, 6.6 Ah |
As the chart above shows, the Invacare Flyer is comparable to its predicate. Both devices have the same intended use for the same patient population, extract oxygen from air using the same methodology, provide comparable oxygen purity and are powered in the same manner. The main differences between the two devices are the weight and physical dimensions. Testing shows that these differences have no effect on safety or effectiveness.
Performance Testing: The Invacare Flyer has been tested in accordance with the following:
| EN 60601-1-2; 2nd ed. 2001 | Medical Electrical Equipment – Sec 1.2 - Collateral standard:<br>Electromagnetic compatibility |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2; 2.1 ed. | Medical Electrical Equipment - Sec 1.2 - Collateral standard:<br>Electromagnetic compatibility |
| EN 55011:1998 | Limits and methods of measurement of radio disturbance characteristics<br>and medical (ISM) radio-frequency equipment |
| CISPR 11: 2003 | Limits and methods of measurement of radio disturbance characteristics<br>and medical (ISM) radio-frequency equipment |
| EN61000-4-3; 2002<br>IEC61000-4-3; 2006 | Radiated, RF, Electromagnetic Field Immunity Test<br>Radiated, RF, Electromagnetic Field Immunity Test |
| EN 61000-4-4; 1995<br>IEC 61000-4-4; 2004 | Electrical Fast Transient/Burst Immunity Test<br>Electrical Fast Transient/Burst Immunity Test |
| EN 61000-4-5; 1995<br>IEC 61000-4-5; 2005 | Surge Immunity Test<br>Surge Immunity Test |
| EN 61000-4-6; 1996<br>IEC 61000-4-6; 2006 | Immunity to Conducted Disturbances Induced by RF fields<br>Immunity to Conducted Disturbances Induced by RF fields |
510(k) K071928 Response
{9}------------------------------------------------
| EN 61000-4-8; 2001 | Power Frequency Magnetic Field Immunity Tests |
|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| EN 61000-4-11; 1995<br>IEC 61000-4-11; 2004 | Voltage Dips, Short Interruptions and Voltage Variations Immunity Tests<br>Voltage Dips, Short Interruptions and Voltage Variations Immunity Tests |
| EN 61000-3-2: 2000<br>IEC 61000-3-2: 2005 | Limits for Harmonic Current Emissions<br>Limits for Harmonic Current Emissions |
| EN 61000-3-3: 1995<br>IEC 61000-3-3: 2005 | Voltage fluctuations and flicker Tests<br>Voltage fluctuations and flicker Tests |
| IEC 60601-1: 2nd ed. 1998<br>with Amendment 1: 1991<br>with Amendment 2: 1995 | Medical Electrical Equipment, Part 1: General requirements for basic<br>safety and performance |
| UL 60601-1: 1st ed. 2003 | Medical Electrical Equipment, Part 1: General requirements for basic<br>safety and performance |
| CSA 601.1 M90 | Medical Electrical Equipment, Part 1: General requirements for basic<br>safety and performance |
| Quasi-Static Electric Field | Reviewer Guidance for premarket notification submissions 1993 |
| FDA Reviewer's Guide:1993 | Electrostatic Discharge Immunity Test per Sec. m,7,ii,a.<br>FDA Reviewer's Guide:1993 Reviewer Guidance for Premarket Notification Submissions |
The results of the testing have shown that the Invacare Flyer is safe and effective for its intended use and substantially equivalent to its predicate.
{10}------------------------------------------------
Image /page/10/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features an abstract eagle design with three horizontal lines representing the bird's body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" are arranged in a circular pattern around the eagle.
DEC 1 2 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Carroll L. Martin Regulatory Affairs Manager Invacare Corporation One Invacare Way P.O.Box 4028 Elyria, Ohio 44036-2125
Re: K071928
Trade/Device Name: Invacare Flyer, Model XPO100 Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: November 8, 2007 Received: November 9, 2007
Dear Mr. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{11}------------------------------------------------
Page 2 - Mr. Martin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sutite Y. Michau Omd
Chin Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
#### 510(k) Number (if known):
### Device Name: Invacare Flyer, Model XPO100
Indications for Use: The Invacare Flyer (Model XPO100) is intended to be used by patients with respiratory disorders who require supplemental oxygen. It supplies a high concentration of supplemental oxygen and is used with a nasal cannula to channel oxygen from the concentrator to the patient. The Invacare Flyer can be used in a home, institution, vehicle and various mobile environments. The Invacare Flyer does not nor is it intended to sustain or support life.
Prescription Use < (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K071928 510(k) Number: _______________________________________________________________________________________________________________________________________________________________
Page 1 of
510(k) K071928 Response
Page 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.