K071918 · Thermogear, Inc. · DWJ · Nov 9, 2007 · Cardiovascular
Device Facts
Record ID
K071918
Device Name
CHILLBUSTER PORTABLE ELECTRIC BLANKET, MODEL 8002
Applicant
Thermogear, Inc.
Product Code
DWJ · Cardiovascular
Decision Date
Nov 9, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The ChillBuster® Model 8002 has been developed to reduce the effects of hypothermia encountered during the trauma of a surgical procedure or other medical crisis which could result in the onset of a hypothermic condition. Use is limited to whole-body warming in adult humans, free of skin conditions or other impairments where distributed heat application is deemed contraindicated by the responsible physician.
Device Story
ChillBuster® Model 8002 is a portable electric blanket system for whole-body warming. Components include a resistively heated blanket, controller, rechargeable battery, and cabling; supports AC/DC power. Used in field transport, ER, surgery, ICU, and post-op care; operated by clinicians. Blanket is encased in a sterile, single-use cover to prevent cross-contamination. Device provides thermal regulation to mitigate hypothermia during medical crises. Clinicians monitor therapy via the controller; improved temperature monitoring and heat distribution assurance compared to previous models. Benefits include patient warming during transport and clinical procedures, reducing hypothermic risks.
Clinical Evidence
Bench testing only. Testing included compliance with IEC/EN 60601-1, EN/IEC 60601-1-2, IEC 60601-2-35, ISO 14971, and CAN/CSA C22.2 No. 601.1-M90. Blanket durability demonstrated via 50 washing/drying cycles. Submersion testing confirmed normal function. Analytical calculations verified heat density and safety cutoff temperatures.
Technological Characteristics
System includes resistively heated blanket, controller, and rechargeable battery. Safety features include 41 ± 1℃ maximum cutoff temperature. Complies with IEC 60601-2-35. Connectivity includes AC/DC power input. Sterilization via single-use sterile blanket cover.
Indications for Use
Indicated for whole-body warming in adult humans to reduce hypothermia effects during surgical trauma or medical crisis. Contraindicated for patients with skin conditions or impairments where distributed heat application is deemed inappropriate by a physician.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
Predicate Devices
ChillBuster® Model 8001 Portable Electric Blanket (K991684)
Reference Devices
Life-Air 1000® Hypothermic Therapy System (K952329)
Submission Summary (Full Text)
{0}------------------------------------------------
K071918
# 510(k) SUMMARY
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR, Part 807.92, Content and Format of a 510(k) Summary.
### 510(k) Summary of Safety and Effectiveness
#### Submitter Information
ThermoGear™ Inc. (contact: Wayne Fields, PhD) 16337 SW Bryant Rd (PO Box 1545) NOV 0 9 2007 Lake Oswego, OR 97035 (Phone): 503.697.1900; (FAX): 503.697.1907
NOV 0 9 2007
Date Summary Prepared: June 1, 2007 (Rev. A)
#### Device Information
ChillBuster® Model 8002 Portable Electric Blanket, with CDRH formal identity:
| Device Group: | System, Thermal Regulating |
|--------------------|----------------------------|
| Medical Specialty: | Cardiovascular |
| Product Code: | DWJ |
| Device Class: | 2 |
| Regulation No.: | 870.5900 |
#### Predicate Devices
#K991684 ChillBuster® Model 8001 Portable Electric Blanket, CDRH formal identity:
| Device Group: | System, Thermal Regulating |
|--------------------|----------------------------|
| Medical Specialty: | Cardiovascular |
| Product Code: | DWJ |
| Device Class: | 2 |
| Regulation No.: | 870.5900 |
{1}------------------------------------------------
#K952329 Life-Air 1000® Hypothermic Therapy System. Its CDRH formal identity is:
| Device Group: | System, Thermal Regulating |
|--------------------|----------------------------|
| Medical Specialty: | Cardiovascular |
| Product Code: | DWJ |
| Device Class: | 2 |
| Regulation No.: | 870.5900 |
#### Device Description
The ChillBuster® Model 8002 Portable Electric Blanket System features a special electrical resistively heated Blanket, a Controller to manage power and device function, a rechargeable Battery, special cabling, and means to connect to external DC or AC power for recharging the Battery and, at the option of the user, operating the Blanket. During therapy, the Blanket is encased in a Sterile, Single-Use Blanket Cover to prevent cross-contamination of successive patients and to prevent soiling. The device can perform therapy at virtually any interval in the following activities or activity chains: a) initial patient preparation and transport in the field, and at any stage of ER, surgery, and ICU or post-op care for injury victims from the field; b) pre-op, surgery, ICU or post-op care of in-facility patients; and c) special situations involving potential or actual hypothermia (e.g. hemodialysis; peritoneal dialysis; plasmapheresis; patient warming during transport from point to point, etc.).
#### Intended Use
The ChillBuster® Model 8002 Portable Electric Blanket has been developed to reduce the effects of hypothermia encountered during the trauma of a surgical procedure or other medical crisis which could result in the onset of a hypothermic condition. This is the same as that of the ChillBuster® predicate and the Life-Air predicate.
For both the ChillBuster® proposed and predicate devices, use is limited to wholebody warming in adult humans, free of skin conditions or other impairments where distributed heat application is deemed contraindicated by the responsible physician. In turn, the Life-Air predicate can also treat younger and pediatric patients.
#### Technological Characteristics
The proposed device is largely the same as the ChillBuster® predicate device. The significant differences are that the proposed device can operate the Blanket while connected to the AC line, and the power output to the Blanket is higher in the proposed versus predicate device (but still low compared to the Life-Air Predicate), and the proposed device has improved Blanket temperature monitoring and assurance of even heat distribution.
{2}------------------------------------------------
#### Non-Clinical Performance Data
The proposed device has been tested to currently valid versions of the following Consensus Standards and other Standards that include:
IEC/EN 60601-1, General Medical Safety; EN/JEC 60601-1-2, Electromagnetic Compatibility; IEC 60601-2-35. Heating Pads and Blankets; ISO 14971, Application of Risk Management to Medical Devices; CAN/CSA C22.2 No. 601.1-M90, Canadian General Medical Safety.
The proposed device Blanket was subjected to 50 washing/drying cycles using household detergent and machines, with no sign of wear or deterioration, and function was demonstrated to be normal in the five sessions in which Blanket heating performance was tested. In separate tests, it was demonstrated that the Blanket functions normally even when completely submerged (except for the Blanket Connector that connects the Blanket to a cable from the Controller).
Analytical calculations were carried out to show that the heat density at the actively heated surface of the ChillBuster® proposed device Blanket was higher than that of the ChillBuster® predicate but only a small percentage of that of the actively heated surface of the Life-Air predicate patient cover. Importantly, the Life-Air predicate specifies a safety cutoff temperature of 43 ± 3℃, whereas the ChillBuster® proposed device Blanket limit is 41 ± 1℃. The latter is the same as that specified for maximum safety cutoff temperature in the International Standard IEC 60601-2-35.
#### Conclusion
The ChillBuster® proposed device is substantially equivalent to the ChillBuster® predicate device in terms of System general structure, arrangement, function, and temperature tolerance, and to the Life-Air Predicate device in terms of operation from the AC line, temperature safety limits, and output power to the patient therapy component (blanket or cover). The proposed device and both predicates have the same basic Intended Use and Indications for Use. The Predicate 8001 and the proposed device place qualifications on patient age and condition.
(4)
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a series of interconnected shapes, possibly representing people or services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 0 9 2007
ThermoGear Inc. c/o Mr. Gregor Dzialas Responsible Third Party Official 12 Commerce Road Newtown, CT 06470
Re: K071918
ChillBuster® Model 8002 Portable Electric Blanket Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II (Two) Product Code: DWJ Dated: July 10, 2007 Received: October 24, 2007
Dear Mr. Dzialas:
We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Gregor Dzialas
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Dma R. Vachner
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
#### K07 19 | 8 510(k) Number (if known):
Devicc Name:
## ChillBuster® Model 8002 Portable Electric Blanket
Indications for Use: The ChillBuster® Model 8002 has been developed to reduce the effects of hypothermia encountered during the trauma of a surgical procedure or other medical crisis which could result in the onset of a hypothermic condition. Use is limited to whole-body warming in adult humans, free of skin conditions or other impairments where distributed heat application is deemed contraindicated by the responsible physician.
| Prescription Use | <b>X</b> (Part 21 CFR 801 Subpart D) |
|------------------|--------------------------------------|
|------------------|--------------------------------------|
AND/OR
| Over-The-Counter Use | (Part 21 CFR 801 Subpart D) |
|----------------------|-----------------------------|
|----------------------|-----------------------------|
Anna B. Vachon
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number: k071918
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.