VERIFY 275 DEGREE F. 3-10 INDICATOR; 10 INDICATOR; 3 INDICATOR

K071895 · STERIS Corporation · JOJ · Nov 14, 2007 · General Hospital

Device Facts

Record IDK071895
Device NameVERIFY 275 DEGREE F. 3-10 INDICATOR; 10 INDICATOR; 3 INDICATOR
ApplicantSTERIS Corporation
Product CodeJOJ · General Hospital
Decision DateNov 14, 2007
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

The Verify® 275F 3 Indicator is a chemical indicator intended for use by health care providers to accompany products being sterilized through a sterilization procedure. The Verify® 275F 3 Indicator is an emulating indicator intended for use in steam sterilization. The indicator changes color from yellow to blue/purple when exposed to 275°F (135°C) pre-vacuum sterilization cycles for 3 minutes.

Device Story

Verify® 275F 3 Indicator is a single-use chemical indicator strip; monitors 3-minute, 275°F (135°C) pre-vacuum steam sterilization cycles. Device consists of a 22 mm x 143 mm polypropylene strip with a chemical indicator ink spot and a reference color block. During sterilization, ink changes from yellow to blue/purple, signaling cycle completion. Used by healthcare providers in clinical settings to verify sterilization parameters for medical products. Visual output allows immediate assessment of sterilization cycle efficacy.

Clinical Evidence

Bench testing only. Performance verified against ANSI/AAMI/ISO 11140-1:2005 requirements for Class 6 emulating indicators using a resistometer per ANSI/AAMI ST44.

Technological Characteristics

Polypropylene strip (22 mm x 143 mm) with chemical indicator ink. Emulating (Class 6) indicator. Sensing principle: chemical color change (yellow to blue/purple) upon exposure to specific steam sterilization parameters (275°F/135°C for 3 minutes).

Indications for Use

Indicated for use by healthcare providers to monitor 3-minute, 275°F (135°C) pre-vacuum steam sterilization cycles. Used as an emulating (Class 6) indicator to accompany products during sterilization.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KO71895 STERIS® # NOV 1 4 2007 510(k) Summary For Verify® 275F 3 Indicator STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (216) 354-2600 Fax No: (216) 639-4459 Contact: Jack Scoville Fellow, Regulatory Affairs Telephone: 440 392 7330 Fax No: 440 397 9198 Submission Date: July 06, 2007 STERIS Corporation = 5960 Heisley Road = Mentor, OH 44060-1834 USA = 440-354-2600 {1}------------------------------------------------ #### 1. Device Name | Indicator: | Verify® 275F 3 Indicator. | |----------------------|-----------------------------------------------------------------------------------------------| | Common Name: | Chemical Indicator. | | Classification Name: | Physical/chemical sterilization process indicator (21<br>CFR 880.2800 (b), Product Code JOJ). | #### 2. Predicate Devices - 3M SteriGage Chemical Indicators (K894399) . - DANA SteriScan Indicators (K012195) . - SteriTec Integraph (Cardinal Steam Integrators' ) (K960441) . - STERIS Verify Integrators (K902958, K002937) . - EZ Test Steam Biological Indicator (K963841) . ### 3. Device Description The proposed Verify® 275F 3 Indicator consist of: - A 22 mm x 143 mm polypropylene strip with one 12 mm chemical indicator . ink spot. The indicator ink spot is located on one end of the strip, adjacent to a reference block exhibiting the endpoint color. The indicator ink on the proposed Verify" 275F 3 Indicator changes from yellow to blue/purple color when the steam sterilization cvclc is complete. The Verify® 275F 3 Indicator can be used to monitor 3 minute 275°F (135°C) prevacuum steam sterilization cycles. ### 4. Intended Use The Verify® 275F 3 Indicator is a chemical indicator intended for use by health care providers to accompany products being sterilized through a sterilization procedure. The Verify® 275F 3 Indicator is an emulating indicator intended for use in steam sterilization. The indicator changes color from yellow to blue/purple when exposed to 275°F (135°C) pre-vacuum sterilization cycles for 3 minutes. The performance of Cardinal is a private label brand produced by Steritec under K960441. {2}------------------------------------------------ ### STERIS ABBREVIATED 510(k) PREMARKET NOTIFICATION VERIFY® 275F 3 INDICATOR the Verify® 275F 3 Indicator meets the requirements of ANSI/AAMI / ISO 11140-1:2005 for emulating [Class 6] steam indicators. ### 5. Description of Safety and Substantial Equivalence The proposed and predicate devices are all single use indicators for use in steam sterilization cycles. The differences between the proposed Verify" 275F 3 Indicator and the predicate device are limited to differences in design, matcrial, and parameters of the sterilization cycles these indicators are designed to monitor. These differences do not raise any new issues of safety and efficacy. ### 6. Performance Testing Performance testing was conducted to verify that the proposed Verify® 275F 3 Indicator mect the requirements for emulating [Class 6] indicators as defined in ANSI/AAMI ISO 11140-1:2005 using a resistometer to ANSI/AAMI ST44. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a row. The profiles are simple and abstract, with flowing lines suggesting hair or movement. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 1 4 2007 Mr. John Scoville Fellow, Regulatory Affairs Sterilization Technology STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060 Re: K071895 Trade/Device Name: Verify® 275F 3 Indicator Regulation Number: 880.2800 Regulation Name: Sterilization process Indicator Regulatory Class: II Product Code: JOJ Dated: October 25, 2007 Received: October 26, 2007 Dear Mr. Scoville: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Scoville Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K071895 Verify® 275F 3 Indicator Device Name: Indications for Use: The Verify® 275F 3 Indicator is a chemical indicator intended for use by health care providers to accompany products being sterilized through a sterilization procedure. The Verify® 275F 3 Indicator is an emulating indicator intended for use in steam sterilization. The indicator changes color from yellow to blue/purple when exposed to 275°F (135°C) pre-vacuum sterilization cycles for 3 minutes as indicated in the following table: | MODEL | TEMPERATURE | STERILIZATION TYPE | TIME | |---------------|---------------|--------------------|-----------| | Verify 275F 3 | 275°F (135°C) | Pre-vacuum steam | 3 minutes | Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sheila A. Murphy, MD logy Center Neal Davis K071895 | Page | 1 of 1 | |------|--------| |------|--------|
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...