ACTIVAPIN, NEXFIX RFS PIN, NEXFIX RESORBABLE FIXATION SYSTEM (RFS)
Applicant
Bioretec, Ltd.
Product Code
HTY · Orthopedic
Decision Date
Aug 31, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ActivaPin™ is indicated for fixation of bone fractures, osteotomies, arthrodeses and osteochondral fractures in the presence of appropriate immobilization.
Device Story
ActivaPin™ is a bioabsorbable fixation device for bone fractures, osteotomies, and arthrodeses. Composed of poly(L-lactide-co-glycolide) (PLGA), the pin degrades in vivo via hydrolysis into metabolizable alpha-hydroxy acids. It maintains mechanical function for at least 8 weeks, with full bioabsorption occurring within two years, eliminating the need for secondary removal surgery. The device is available in diameters of 1.5–3.2 mm and lengths of 20–70 mm. This modification adds a disposable pin applicator with a K-wire to facilitate installation. Used by surgeons in clinical settings, the device provides internal fixation; the gradual loss of strength corresponds to the bone healing process, providing structural support during the critical recovery period.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and microbiological performance evaluation of the modified instrument set compared to the predicate device.
Technological Characteristics
Material: Bioabsorbable poly(L-lactide-co-glycolide) (PLGA). Dimensions: 1.5–3.2 mm diameter, 20–70 mm length. Principle: Mechanical fixation via bioabsorbable pin; degrades via hydrolysis. Includes disposable pin applicator and K-wire for installation.
Indications for Use
Indicated for fixation of bone fractures, osteotomies, arthrodeses, and osteochondral fractures requiring immobilization. Contraindicated for diaphyseal bone fractures/osteotomies, weight-bearing cancellous bone fractures/osteotomies, active/potential infection, or cases where patient cooperation is not guaranteed.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K071863 page '1 of 2
Image /page/0/Picture/1 description: The image shows the logo for Bioretec. The logo consists of a stylized symbol above the word "bioretec" in lowercase letters. The symbol is abstract and appears to be composed of curved, overlapping shapes, creating a sense of movement or flow.
G
# SPECIAL 510(k) SUMMARY
AUG 3 1 2007
For the modification to Bioretec ActivaPin™ (K061164)
#### MANUFACTURER
Bioretec Ltd. Hermiankatu 22, Modulight Building FI-33720 Tampere FINLAND
## Contact person:
Ms. Mari Ruotsalainen Quality Manager Phone: +358 20 778 9514 Fax: +358 3 317 0225 Mari Ruotsalainen@bioretec.com
Date prepared: June 28th, 2007
#### DEVICE NAME
Trade Names: Bioretec ActivaPin™, NexFix™ RFS Pin (Resorbable Fixation System) Common Name: Pin, Fixation
### ESTABLISHMENT REGISTRATION NUMBER
Bioretec Ltd. has been submitted the Establishment Registration form and is waiting for an Establishment Registration Number.
# DEVICE CLASSIFICATION AND PRODUCT CODE
Device Classification Name: Pin, Fixation, Smooth Classification Panel: Orthopedic Regulation Number: 21 CFR 888.3040 Product Code: HTY
## PREDICATE DEVICES
Bioretec ActivaPin™ (K061164)
Special 510(k): Device Modification, Date: June 28th, 2007
G-1 Page
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "Page 2 of 2" written in black ink on a white background. The handwriting is cursive and slightly slanted. The words are clearly legible, and the overall impression is that of a page number designation.
Image /page/1/Picture/1 description: The image shows the logo for Bioretec. The logo consists of a stylized graphic above the word "bioretec" in lowercase letters. The graphic is composed of three curved shapes that resemble stylized leaves or abstract forms, arranged in a cluster. The logo is black against a white background.
## DEVICE DESCRIPTION AND PRINCIPLES OF OPERATION
The modified ActivaPin™ is identical to the currently cleared device except for the modification. The only modification to the initial 510(k) K061164 to be cleared is adding a Disposable Pin Applicator with K-wire for installation of ActivaPin™ and revising labeling accordingly. The change does not affect the intended use or alter the fundamental scientific technology of the device.
The ActivaPin" is indicated for fixation of bone fractures, osteotomies, arthrodeses and osteochondral fractures in the presence of appropriate immobilization. Pins are available in several different dimensions, including diameters of 1.5 - 3.2 mm and lengths of 20 - 70 mm.
ActivaPin™ is made of completely bioabsorbable poly(L-lactide-co-glycolide) (PLGA), and they decrade in vivo by hvdrolysis into alpha-hydroxy acids that are metabolized by the body. As the operated bone fracture or osteotomy gains strength during healing, the ActivaPin"1" gradually loses its strength, however, maintaining its function at least 8 weeks. Bioabsorption takes place within two years thus eliminating the need for implant removal surgery.
## EQUIVALENCE TO MARKETED PRODUCTS
The ActivaPin™ Bioabsorbable Pin and its instrumentation is substantially equivalent to the ActivaPin™ cited as predicate device above.
The Bioretec ActivaPin" with its modified instrumentation has the same intended use, principles of operation and technological characteristic as the predicate device. Adding a Disposable Pin Applicator with K-wire for installation of ActivaPin™ and revising labeling accordingly does not raise any questions of safety and effectiveness.
Microbiological testing determined that the ActivaPin™ and the ActivaPin™ Disposable Pin Applicator with K-wire instrument set has substantially similar performance as compared to its predicate devices.
Special 510(k): Device Modification, Date: June 28th, 2007
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of an eagle.
AUG 3 1 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Bioretec Ltd. % Ms. Mari Ruotsalainen Quality Manager Hermiankatu 22, Modulight Building FI-33720 Tampere FINLAND
Re: K071863
> Trade/Device Name: Bioretec ActivaPin™ Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY Dated: August 1, 2007 Received: August 3, 2007
Dear Ms. Ruotsalainen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Mari Ruotsalainen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807,97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for bioretec. The logo consists of a stylized graphic above the word "bioretec" in lowercase letters. The graphic appears to be an abstract representation of a bird's head or a stylized leaf.
#### F Indications for Use Statement
Submitter: 510(k) Number:
Device Name:
Bioretec Ltd.
K07186
ActivaPin™
Indications for Use:
The ActivaPin™ is indicated for fixation of bone fractures, osteotomies, arthrodeses and osteochondral fractures in the presence of appropriate immobilization.
## Contraindications:
1. Fractures and osteotomies of diaphyseal bone.
2. Fractures and osteotomies in weight bearing cancellous bone.
3. Situations where internal fixation is otherwise contraindicated, e.g., active or potential infection and where patient cooperation cannot be guaranteed.
Prescription Use ___ x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clerbau Brelup
(Division Sig Division of General, Restorative, and Neurological Devices
**510(k) Number** K071863
Special 510(k): Device Modification, Date: June 28th, 2007
Page 4 of 4
F-1 Page
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.