ZIMMER M/L TAPER HIP PROSTHESIS WITH KINECTIV TECHNOLOGY SYSTEM, MODEL(S) 7848 SERIES( MODULAR NECKS), 7713 SERIES
K071856 · Zimmer, Inc. · KWA · Jul 30, 2007 · Orthopedic
Device Facts
Record ID
K071856
Device Name
ZIMMER M/L TAPER HIP PROSTHESIS WITH KINECTIV TECHNOLOGY SYSTEM, MODEL(S) 7848 SERIES( MODULAR NECKS), 7713 SERIES
Applicant
Zimmer, Inc.
Product Code
KWA · Orthopedic
Decision Date
Jul 30, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3330
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis; disability due to previous fusion; previously failed endoprostheses, and/or total hip components in the affected extremity and acute femoral neck fractures. Hemi-hip replacement for the following: fracture dislocation of the hip; elderly, debilitated patients when a total hip replacement is contraindicated; irreducible fractures in which adequate fixation cannot be obtained; certain high subcapital fractures and comminuted femoral neck fractures in the aged; nonunion of femoral neck fractures; secondary avascular necrosis of the femoral head; pathological fractures of the femoral neck; and osteoarthritis in which the femoral head is primarily affected. This femoral stem is for cementless use only.
Device Story
Modular, wedge-shaped femoral stem for total or hemi-hip replacement; features Kinectiv technology allowing modular neck options. Device implanted by surgeons in clinical/OR settings. Modularity enables independent adjustment of leg length and offset; facilitates soft tissue balancing; optimizes hip joint center of rotation; maximizes joint stability across diverse patient anatomies. Stem coated with commercially pure titanium alloy plasma spray for cementless fixation. Output is physical restoration of hip joint function; aids clinical decision-making by providing intraoperative flexibility to address specific anatomical requirements.
Indicated for patients requiring total or hemi-hip replacement due to arthritis (rheumatoid, osteoarthritis, traumatic, polyarthritis), collagen disorders, avascular necrosis, femoral fractures (nonunion, acute, subcapital, comminuted, pathological), congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, previous fusion disability, or failed prior endoprostheses. Hemi-hip replacement indicated for elderly/debilitated patients where total hip replacement is contraindicated or fixation is otherwise irreducible. Cementless use only.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).
Predicate Devices
Zimmer® M/L Taper Hip Prosthesis with Modular Neck Technology (K063251)
Submission Summary (Full Text)
{0}------------------------------------------------
K071856
1
p. 1/3
# Summary of Safety and Effectiveness
30 Jul 07
| Submitter: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Dalene T. Binkley, RAC<br>Senior Associate, Regulatory Affairs<br>Telephone: (574) 372-4907<br>Fax: (574) 372-4605 |
| Date: | July 3, 2007 |
| Trade Name: | Zimmer® M/L Taper Hip Prosthesis with Kinectiv™<br>Technology System |
| Common Name: | Total Hip Prosthesis |
| Classification Name<br>and Reference: | 1. KWA - Hip joint metal/metal semi-constrained<br>with uncemented acetabular shell, 21 CFR § 888.<br>3330<br>2. JDL - Hip joint metal/metal semi-constrained<br>with cemented acetabular shell, 21 CFR § 888. 3320<br>3. LPH - Prosthesis, hip, semi-constrained<br>metal/polymer porous uncemented, 21 CFR § 888.<br>3358<br>4. LWJ - Prosthesis, hip, semi-constrained<br>metal/polymer uncemented, 21 CFR § 888. 3360<br>5. MEH - Prosthesis, hip, semi-constrained<br>uncemented metal polymer, non-porous, calcium<br>phosphate, 21 CFR § 888. 3353 |
| Predicate Device: | Zimmer® M/L Taper Hip Prosthesis with Modular<br>Neck Technology, manufactured by Zimmer, Inc.,<br>K063251, cleared January 24, 2007. |
{1}------------------------------------------------
# 长071856
:
:
p. 2/3
| Device Description: | The Zimmer M/L Taper Hip Prosthesis with<br>Kinectiv Technology System is a modular, wedge-<br>shaped stem that is coated with commercially pure<br>titanium alloy plasma spray. |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The modular neck options allow for soft tissue<br>balancing and easier restoration of the hip joint<br>center of rotation. The modularity feature will<br>allow surgeons to independently equalize leg length<br>and optimize offset while, at the same time,<br>maximizing joint stability for a variety of patient<br>anatomies. |
| Intended Use: | Total hip replacement for the following: severe hip<br>pain and disability due to rheumatoid arthritis,<br>osteoarthritis, traumatic arthritis, polyarthritis,<br>collagen disorders, avascular necrosis of the<br>femoral head, nonunion of previous fractures of the<br>femur; congenital hip dysplasia, protrusio acetabuli,<br>slipped capital femoral epiphysis; disability due to<br>previous fusion; previously failed endoprostheses,<br>and/or total hip components in the affected<br>extremity and acute femoral neck fractures. |
| | Hemi-hip replacement for the following: fracture<br>dislocation of the hip; elderly, debilitated patients<br>when a total hip replacement is contraindicated;<br>irreducible fractures in which adequate fixation<br>cannot be obtained; certain high subcapital fractures<br>and comminuted femoral neck fractures in the aged;<br>nonunion of femoral neck fractures; secondary<br>avascular necrosis of the femoral head; pathological<br>fractures of the femoral neck; and osteoarthritis in<br>which the femoral head is primarily affected. |
| | This femoral stem is for cementless use only. |
| Comparison to Predicate Device: | The Zimmer M/L Taper Hip Prosthesis with Kinetiv<br>Technology System is packaged, manufactured, and<br>sterilized using the same materials and processes as<br>its predicate. The subject device also has the same<br>intended use and fixation methods as the predicate. |
{2}------------------------------------------------
K07185L
p. 3/3
Performance Data (Nonclinical and/or Clinical):
Non-Clinical Performance and Conclusions:
Non-clinical testing demonstrated that the Zimmer M/L Taper Hip Prosthesis with Kinectiv Technology System met performance requirements and is as safe and effective as its predicate.
{3}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Zimmer, Inc. % Ms. Dalene T. Binkley Senior Associate, Corporate Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708
> K071856 Trade/Device Name: Zimmer® M/L Taper Hip Prosthesis with Kinectiv® Technology System Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: Class III Product Code: KWA, JDL, LPH, LWJ, MEH Dated: July 3. 2007 Received: July 5, 2007
JUL 3 0 2007
Dear Ms. Binkley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
Image /page/3/Picture/9 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's emblem in the center. The emblem consists of a stylized caduceus-like symbol with three intertwined strands, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem in a circular fashion.
Re:
{4}------------------------------------------------
Page 2 - Ms. Dalene T. Binkley
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Barbara BruchuP
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 3 - Ms. Dalene T. Binkley
cc: HFZ-401 DMC HFZ-404 510(k) Staff HFZ-410 DGRND D.O. f/t:JSG:afb:7/30/07 OC Numbers: Division of Enforcement A 240-276-0115 Dental, ENT and Ophthalmic Devices Branch 240-276-0115 OB/GYN, Gastro. & Urology Devices Branch 240-276-0115 General Hospital Devices Branch 240-276-0115 General Surgery Devices Branch 240-276-0115 Division of Enforcement B 240-276-0120 Cardiovascular & Neurological Devices Branch 240-276-0120 Orthopedic, Physical Medicine & Anesthesiology Devices and Radiological 240-276-0120 Devices
{6}------------------------------------------------
071856
## Indications for Use
### 510(k) Number (if known):
Device Name:
Zimmer® M/L Taper Hip Prosthesis with Kinectiv™ Technology System
#### Indications for Use:
Total hip replacement for the following: severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, nonunion of previous fractures of the femur; congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis; disability due to previous fusion; previously failed endoprostheses, and/or total hip components in the affected extremity and acute femoral neck fractures.
Hemi-hip replacement for the following: fracture dislocation of the hip; elderly, debilitated patients when a total hip replacement is contraindicated; irreducible fractures in which adequate fixation cannot be obtained; certain high subcapital fractures and comminuted femoral neck fractures in the aged; nonunion of femoral neck fractures; secondary avascular necrosis of the femoral head; pathological fractures of the femoral neck; and osteoarthritis in which the femoral head is primarily affected.
This femoral stem is for cementless use only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Barbara Medium for MAM
Division of Ge Restorati and Neurological Devices
Page 1 of 1
510(k) Number K071856
021
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.