K071839 · Reichert, Inc. · OBO · Feb 27, 2008 · Ophthalmic
Device Facts
Record ID
K071839
Device Name
SOCT COPERNICUS, MODEL 15205
Applicant
Reichert, Inc.
Product Code
OBO · Ophthalmic
Decision Date
Feb 27, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1570
Device Class
Class 2
Indications for Use
The SOCT Copernicus is an optical coherence tomography system indicated for the invivo imaging and measurement of the retina as an aid in the diagnosis and management of retinal diseases.
Device Story
SOCT Copernicus is an AC-powered optical coherence tomography (OCT) system for retinal imaging. Device uses interferometry of 840 nm superluminescent diode (SLED) light reflected from the fundus to generate cross-sectional retinal images. System includes main unit with optical imaging system, light source, power supply, chin/forehead rest, and computer with software for image processing and data analysis. Operated by clinicians in eye care settings. Device captures 25,000 A-scans/second, providing higher resolution (6 μm depth) than traditional time-domain OCT. Clinicians use cross-sectional images and measurements to assess retinal structures, aiding diagnosis and management of vitreo-retinal disorders. Benefits include improved visualization of retinal pathology compared to previous generation systems.
Clinical Evidence
Clinical evidence includes a precision study (repeatability/reproducibility) with 3 operators and 2 devices on 20 eyes (10 normal, 10 pathological). A prospective case series compared SOCT Copernicus to Stratus OCT in 27 eyes from 21 patients. Scans were analyzed across nine retinal fields. Statistical analysis demonstrated good agreement and no medically significant difference in measurements between the subject and predicate devices.
Technological Characteristics
Spectral Domain OCT; 840 nm SLED light source; 750 μW exposure at cornea; 25,000 A-scan/second; 6 μm depth resolution; 12-18 μm transverse resolution. System includes main unit, chin/forehead rest, and PC-based software for image processing. AC-powered.
Indications for Use
Indicated for in vivo imaging and measurement of the retina in patients to aid in the diagnosis and management of retinal diseases.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: K071839 .
#### 1. Submitter's Identification:
Reichert, Inc. c/o Sandra Brown Quality Regulatory Affairs Manager 3362 Walden Avenue Depew, NY 14043 Tel: (716) 686-4599
Date Summary Prepared: January 24, 2008
#### 2. Name of the Device: SOCT Copernicus
#### 3. Common or Usual Name: Optical Coherence Tomographer (OCT)
#### 4. Predicate Device Information:
- Trade Name: STRATUS OCT .
- 510(k) Number: K030433 .
- Common Name: Optical Coherence Tomographer (OCT) •
- Classification Name: Ophthalmoscope, AC-powered .
- Classification: Class II .
- Classification Panel: 86 .
- Product Code: HLI .
- Regulation Number: 886.1570 .
: 上一 7
{1}------------------------------------------------
#### న్. SOCT Copernicus Device Description:
The SOCT Copernicus is an AC-powered device containing illumination and viewing optics intended to examine the retina of the eye.
# SOCT Copernicus Intended Use:
The SOCT Copernicus is an optical coherence tomography system indicated for the invivo imaging and measurement of the retina as an aid in the diagnosis and management of retinal diseases.
#### 6. Technological Characteristics:
The Reichert SOCT Copernicus and the predicate device have similar technological characteristics. Both devices utilize Optical Coherence Tomography, which relies upon interferometry of superluminescent diode light, reflected from the fundus of the eye, to produce a cross-sectional image of the retina. Both devices have similar hardware components. Like the predicate device, the SOCT Copernicus consists of a main unit, which houses the optical system used for imaging the retina, the light source, and a power supply. In addition, both systems utilize a chin rest/forehead rest assembly for patient interface and a computer with software for data analysis / image processing.
A brief summary of major technological characteristics for both devices is presented below:
| | Factor | Subject Device<br>Reichert “SOCT Copernicus” | Predicate Device<br>Carl Zeiss “STRATUS OCT” |
|---|---------------------------------|------------------------------------------------------------|----------------------------------------------|
| 1 | Method of<br>Operation | Optical Coherence Tomography:<br>Spectral / Fourier Domain | Optical Coherence Tomography:<br>Time Domain |
| 2 | Exposure<br>Parameters | 750 μW at the cornea | 750 μW at the cornea |
| 3 | Light Source | SLED. scan beam wavelength is<br>840 nm. | SLED. scan beam wavelength is<br>820 nm. |
| 4 | OCT Image | 25,000 A-scan/second | 400 A-scan/second |
| 5 | Depth Resolution<br>(in tissue) | 6 μm | 10 μm |
| 6 | Transverse<br>Resolution | 12 μm<br>18 μm typical | 20 μm |
{2}------------------------------------------------
### 7. Discussion of Clinical Tests Performed:
# SOCT Copernicus Precision Test (Repeatability and Reproducibility)
A structured study was conducted with 3 operators (unaware of patient pathology) and 2 devices to determine repeatability and reproducibility precision values for the SOCT Copernicus. The test population consisted of 10 normal eves (with a wide range of media opacity and other parameters affecting image quality) and 10 pathological eyes including eyes with parameters affecting image quality such as diabetic macular edema, wet AMD, dry AMD, cystoid macular edema, vitreous hemorrhage, epiretinal membrane and/or vitromacular traction, and macular holes.
Statistical analysis of the study data showed good agreement across devices and operators.
# SOCT Copernicus Clinical Comparison
A prospective case series comparing the Carl Zeiss Meditec Stratus™ Time Domain Optical Coherence Tomography (TDOCT) to the Reichert Copernicus™ Spectral Domain OCT (SDOCT) was performed at the Cleveland Clinic Cole Eye Institute.
A total of 27 eves from 21 patients with a variety of vitreo-retinal disorders, including five normal eves, were scanned with both machines and analyzed. All scans were analyzed to obtain values from nine fields: central subfield, temporal inner retina, superior inner retina, nasal inner retina, inferior inner retina, temporal outer retina, superior outer retina, nasal outer retina, and inferior outer retina.
Based on the results of the statistical analysis, no medically significant difference seems to exist between the measurements made by the two devices.
#### 9. Conclusions:
· The SOCT Copernicus and Stratus OCT are both intended for in-vivo imaging and measurement of the retina as an aid in the diagnosis and management of retinal diseases.
• The SOCT Copernicus and Stratus OCT use similar technology with the exception that SOCT Copernicus utilizes Spectral Domain technology whereas the Stratus OCT uses Time Domain technology. Spectral Domain technology results in higher resolution retinal images. In addition, the SOCT Copernicus PC software is different from the Stratus OCT, but has been completely tested and validated.
· The devices are substantially equivalent based on the precision, repeatability and clinical testing described herein.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 27 2002
Reichert, Inc. c/o Ms. Sandra Brown Quality/Regulatory Affairs Manager 3362 Walden Avenue Depew, NY 14043
Re: K071839
Trade/Device Name: SOCT Copernicus Spectral Optical Coherence Tomographer Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO Dated: Undated Received: February 15, 2008
Dear Ms. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. B. Egelston, m.D.
Malvina B. Evdelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): _ K071839
Device Name: Reichert SOCT Copernicus Spectral Optical Coherence Tomographer
Indications For Use:
An optical coherence tomography system indicated for the in vivo imaging and measurement of the retina as an aid in the diagnosis and management of retinal diseases.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| <br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------ | <br><br><br>Ana March<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CALLERS AND LEASE OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF CONSULTION OF<br><br>-------<br>-<br> | The name<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A A A A A<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| (Division Sign-Off) |
|-----------------------------|
| Division of Ophthalmic Ear, |
| Nose and Throat Devises |
| 510(k) Number | K071839 |
|---------------|---------|
|---------------|---------|
| Prescription Use (Per 21 CFR 801.109) | <div></div> |
|---------------------------------------|-------------|
| | <div></div> |
| | Page 1 of / |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.