← Product Code [MWI](/productcode/MWI) · K071802

# TESLA GUARD, MODEL EFT G04 (K071802)

_Mammendorfer Institut Fur Physik Und Medizin GmbH · MWI · Nov 9, 2007 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K071802

## Device Facts

- **Applicant:** Mammendorfer Institut Fur Physik Und Medizin GmbH
- **Product Code:** [MWI](/productcode/MWI.md)
- **Decision Date:** Nov 9, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.2300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The TeslaGuard® Patient Monitor is capable of monitoring: SpO2 (Arterial Oxygen Saturation) ECG (3-Lead) IBP (Invasive Blood Pressure) NIBP (Non-invasive Blood Pressure) CO2 and Anesthetic Agents (with optional multi-gas module) This device will produce visual and audible alarms if any of these parameters vary beyond preset limits and produce timed or alarm recordings. With the optional multi-gas module installed, sampled breathing gases from adults and pediatrics can be displayed. The multi-gas module continuously measures the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture, and communicates real time and derived gas information to the TeslaGuard Patient Monitor. The device is intended to be used in the environment where patient care is provided by Healthcare Professionals, such as physicians, nurses, and technicians, trained on the use of the device, who will determine when to use the device based upon their professional assessment of the patient's medical condition. The device is intended for use in the Adult, Pediatric and Neonatal populations.

## Device Story

TeslaGuard is an MRI-compatible patient monitor; operates in magnetic fields up to 20mT; requires 1.5m distance from magnet core. Inputs: SpO2, 3-lead ECG, IBP, NIBP; optional multi-gas module samples breathing gases (CO2, N2O, O2, anesthetic agents). Device processes signals/gas concentrations; displays real-time data; triggers visual/audible alarms for out-of-limit parameters; generates recordings. Used by physicians, nurses, and technicians in clinical settings. Output informs clinical decision-making regarding patient status during MRI procedures; benefits patient by enabling continuous monitoring in high-field environments.

## Clinical Evidence

Bench testing only. Laboratory bench testing validated performance specifications of the multi-gas module. Simulated use testing established safety and effectiveness under maximum intended MRI conditions.

## Technological Characteristics

MRI-compatible patient monitor; multi-gas module for CO2, N2O, O2, and anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane). Operates in magnetic fields up to 20mT. Connectivity: communicates real-time and derived gas information to the monitor. Software-based alarm and recording system.

## Regulatory Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

## Predicate Devices

- MIPM TeslaGuard® ([K052119](/device/K052119.md))
- Draeger Infinity Gamma XL ([K053484](/device/K053484.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| 510(k) Number: | K071802 |
|----------------|---------|
|----------------|---------|

Date: Date: Date:

Page 1 of 2

# 510(k) Summary

#### Introduction

This summary is intended to comply with requirements of the SMDA and 21CFR§807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.

## 510(k) Submitter

Mammendorfer Institut für Physik und Medizin GmbH (MIPM) Oskar-von-Miller-Strasse 6 82291 Mammendorf, Germany

NOV 0 9 2007

## 510(k) Correspondent

Robert N. Clark, President and Senior Consultant Medical Device Regulatory Advisors, Inc. 13605 West 7th Ave., Golden, CO 80401 USA Tel: +1 (303) 463-0900 / Fax: +1 (303) 234-9413

## Date Prepared

June 25, 2007

#### Trade Name of Device

TeslaGuard®

## Common Name of Device

MRI Compatible Patient Monitor

## Classification

Name: Monitor, Physiological, Patient (without arrhythmia detector or alarms) Product Code: MWI Regulation Number: 21CFR§870.2300 Device Class: II

## Device Description and Intended Use

The TeslaGuard design allows monitoring of intensive care patients while in an MRI-scanner. During use, the unit must be positioned in a way that the maximum field strength is not higher than 20 mT, and the distance to the magnet core is at least 1.5m.

This 510(k) has been filed to establish substantial equivalence for an optional multi-gas module. With the multi-gas module installed, sampled breathing gases from adults and pediatrics can be displayed. The multi-gas module continuously measures the content of CO2, N2O, O2 and one

{1}------------------------------------------------

510(k) Number: _ K 0 718 0 2

Date: _

Page 2 of 2

of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any of the areasiness agence, teal time and derived gas information to the TeslaGuard Patient Monitor.

The other intended uses of the TeslaGuard remain unchanged.

#### Predicate Devices

MIPM TeslaGuard® as cleared under 510(k) # K052119 Draeger Infinity Gamma XL as cleared under 510(k) # K053484

#### Non Clinical Testing

Laboratory bench testing was conducted to validate performance specifications of the multi-gas module. Simulated use testing was conducted to establish safety and effectiveness of the TeslaGuard multi-gas module under maximum intended MRI conditions.

#### Risk Management

This device has been designed to either completely eliminate or mitigate known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program.

#### Substantial Equivalence

Mammendorfer Institut für Physik und Medizin GmbH (MIPM) believes that the TeslaGuard multi-gas module is safe and effective when used as instructed by knowledgeable and trained personnel, and is substantially equivalent to the legally marketed predicate devices.

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized symbol that resembles a human figure in profile, with three overlapping faces or profiles. The symbol is rendered in black and has a flowing, abstract design.

NOV 0 9 2007

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mammendorfer Institut für Physik und Medizin GmbH c/o Mr. Robert N. Clark President and Senior Consultant Medical Device Regulatory Advisors, Inc. 13605 West 7th Ave. Golden, CO 80401

Re: K071802

Trade/Device Name: Tesla Guard Model EFT G04 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (two) Product Code: MWI Dated: October 7, 2007 Received: October 10, 2007

Dear Mr. Clark:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Robert N. Clark

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be actived that i Dr unstion that your device complies with other requirements of the Act that I Drehan statutes and regulations administered by other Federal agencies. You must or any I oderal statutes and requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and h your e FDA finding of substantial equivalence of your device to a legally premative included in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Blyminar for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use

510(k) Number (if known):

Tesla Guard @ Device Name: _

Indications for Use:

The TeslaGuard® Patient Monitor is capable of monitoring:

- SpO2 (Arterial Oxygen Saturation)
- ECG (3-Lead)
- IBP (Invasive Blood Pressure) 0
- NIBP (Non-invasive Blood Pressure)
- CO2 and Anesthetic Agents (with optional multi-gas module)

This device will produce visual and audible alarms if any of these parameters vary beyond preset limits and produce timed or alarm recordings.

With the optional multi-gas module installed, sampled breathing gases from adults and pediatrics can be vinti the optional motile gas module continuously measures the content of CO2, N2O, O2 and one of the ansthetic displayed. The hibiti-gas model continues and Desflurane in any mixtere, and communicates with and communicates real
agents, Halothane, Isoflurane, Sevolurane and Desfluran

The device is intended to be used in the environment where patient care is provided by Healthcare The device is intended to be assum the enternicians, trained on the use of the device, who will determine when r rolodionally, no pryliciance based upon their professional assessment of the patient's medical condition.

The device is intended for use in the Adult, Pediatric and Neonatal populations.

#### MRI Compatibility Statement

The Testa Suard® Patient Monitor is designed for use in an MRI-environment at a maximum magnetic field strength of 20mT.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

B.A. Summer

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K071802

Page 1 of 1

MIPM TeslaGuard Special 510(k) Pace 16 of 215

---

**Source:** [https://fda-staging.innolitics.com/device/K071802](https://fda-staging.innolitics.com/device/K071802)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).