K071802 · Mammendorfer Institut Fur Physik Und Medizin GmbH · MWI · Nov 9, 2007 · Cardiovascular
Device Facts
Record ID
K071802
Device Name
TESLA GUARD, MODEL EFT G04
Applicant
Mammendorfer Institut Fur Physik Und Medizin GmbH
Product Code
MWI · Cardiovascular
Decision Date
Nov 9, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The TeslaGuard® Patient Monitor is capable of monitoring: SpO2 (Arterial Oxygen Saturation) ECG (3-Lead) IBP (Invasive Blood Pressure) NIBP (Non-invasive Blood Pressure) CO2 and Anesthetic Agents (with optional multi-gas module) This device will produce visual and audible alarms if any of these parameters vary beyond preset limits and produce timed or alarm recordings. With the optional multi-gas module installed, sampled breathing gases from adults and pediatrics can be displayed. The multi-gas module continuously measures the content of CO2, N2O, O2 and one of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any mixture, and communicates real time and derived gas information to the TeslaGuard Patient Monitor. The device is intended to be used in the environment where patient care is provided by Healthcare Professionals, such as physicians, nurses, and technicians, trained on the use of the device, who will determine when to use the device based upon their professional assessment of the patient's medical condition. The device is intended for use in the Adult, Pediatric and Neonatal populations.
Device Story
TeslaGuard is an MRI-compatible patient monitor; operates in magnetic fields up to 20mT; requires 1.5m distance from magnet core. Inputs: SpO2, 3-lead ECG, IBP, NIBP; optional multi-gas module samples breathing gases (CO2, N2O, O2, anesthetic agents). Device processes signals/gas concentrations; displays real-time data; triggers visual/audible alarms for out-of-limit parameters; generates recordings. Used by physicians, nurses, and technicians in clinical settings. Output informs clinical decision-making regarding patient status during MRI procedures; benefits patient by enabling continuous monitoring in high-field environments.
Clinical Evidence
Bench testing only. Laboratory bench testing validated performance specifications of the multi-gas module. Simulated use testing established safety and effectiveness under maximum intended MRI conditions.
Technological Characteristics
MRI-compatible patient monitor; multi-gas module for CO2, N2O, O2, and anesthetic agents (Halothane, Isoflurane, Enflurane, Sevoflurane, Desflurane). Operates in magnetic fields up to 20mT. Connectivity: communicates real-time and derived gas information to the monitor. Software-based alarm and recording system.
Indications for Use
Indicated for monitoring SpO2, ECG, IBP, NIBP, and (with optional module) CO2, N2O, O2, and anesthetic agents in adult, pediatric, and neonatal patients. Used by trained healthcare professionals in MRI environments (up to 20mT).
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
| 510(k) Number: | K071802 |
|----------------|---------|
|----------------|---------|
Date: Date: Date:
Page 1 of 2
# 510(k) Summary
#### Introduction
This summary is intended to comply with requirements of the SMDA and 21CFR§807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence.
## 510(k) Submitter
Mammendorfer Institut für Physik und Medizin GmbH (MIPM) Oskar-von-Miller-Strasse 6 82291 Mammendorf, Germany
NOV 0 9 2007
## 510(k) Correspondent
Robert N. Clark, President and Senior Consultant Medical Device Regulatory Advisors, Inc. 13605 West 7th Ave., Golden, CO 80401 USA Tel: +1 (303) 463-0900 / Fax: +1 (303) 234-9413
## Date Prepared
June 25, 2007
#### Trade Name of Device
TeslaGuard®
## Common Name of Device
MRI Compatible Patient Monitor
## Classification
Name: Monitor, Physiological, Patient (without arrhythmia detector or alarms) Product Code: MWI Regulation Number: 21CFR§870.2300 Device Class: II
## Device Description and Intended Use
The TeslaGuard design allows monitoring of intensive care patients while in an MRI-scanner. During use, the unit must be positioned in a way that the maximum field strength is not higher than 20 mT, and the distance to the magnet core is at least 1.5m.
This 510(k) has been filed to establish substantial equivalence for an optional multi-gas module. With the multi-gas module installed, sampled breathing gases from adults and pediatrics can be displayed. The multi-gas module continuously measures the content of CO2, N2O, O2 and one
{1}------------------------------------------------
510(k) Number: _ K 0 718 0 2
Date: _
Page 2 of 2
of the anesthetic agents, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane in any of the areasiness agence, teal time and derived gas information to the TeslaGuard Patient Monitor.
The other intended uses of the TeslaGuard remain unchanged.
#### Predicate Devices
MIPM TeslaGuard® as cleared under 510(k) # K052119 Draeger Infinity Gamma XL as cleared under 510(k) # K053484
#### Non Clinical Testing
Laboratory bench testing was conducted to validate performance specifications of the multi-gas module. Simulated use testing was conducted to establish safety and effectiveness of the TeslaGuard multi-gas module under maximum intended MRI conditions.
#### Risk Management
This device has been designed to either completely eliminate or mitigate known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program.
#### Substantial Equivalence
Mammendorfer Institut für Physik und Medizin GmbH (MIPM) believes that the TeslaGuard multi-gas module is safe and effective when used as instructed by knowledgeable and trained personnel, and is substantially equivalent to the legally marketed predicate devices.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is a stylized symbol that resembles a human figure in profile, with three overlapping faces or profiles. The symbol is rendered in black and has a flowing, abstract design.
NOV 0 9 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mammendorfer Institut für Physik und Medizin GmbH c/o Mr. Robert N. Clark President and Senior Consultant Medical Device Regulatory Advisors, Inc. 13605 West 7th Ave. Golden, CO 80401
Re: K071802
Trade/Device Name: Tesla Guard Model EFT G04 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (two) Product Code: MWI Dated: October 7, 2007 Received: October 10, 2007
Dear Mr. Clark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Robert N. Clark
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease be actived that i Dr unstion that your device complies with other requirements of the Act that I Drehan statutes and regulations administered by other Federal agencies. You must or any I oderal statutes and requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and h your e FDA finding of substantial equivalence of your device to a legally premative included in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Blyminar for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Indications for Use
510(k) Number (if known):
Tesla Guard @ Device Name: _
Indications for Use:
The TeslaGuard® Patient Monitor is capable of monitoring:
- SpO2 (Arterial Oxygen Saturation)
- ECG (3-Lead)
- IBP (Invasive Blood Pressure) 0
- NIBP (Non-invasive Blood Pressure)
- CO2 and Anesthetic Agents (with optional multi-gas module)
This device will produce visual and audible alarms if any of these parameters vary beyond preset limits and produce timed or alarm recordings.
With the optional multi-gas module installed, sampled breathing gases from adults and pediatrics can be vinti the optional motile gas module continuously measures the content of CO2, N2O, O2 and one of the ansthetic displayed. The hibiti-gas model continues and Desflurane in any mixtere, and communicates with and communicates real
agents, Halothane, Isoflurane, Sevolurane and Desfluran
The device is intended to be used in the environment where patient care is provided by Healthcare The device is intended to be assum the enternicians, trained on the use of the device, who will determine when r rolodionally, no pryliciance based upon their professional assessment of the patient's medical condition.
The device is intended for use in the Adult, Pediatric and Neonatal populations.
#### MRI Compatibility Statement
The Testa Suard® Patient Monitor is designed for use in an MRI-environment at a maximum magnetic field strength of 20mT.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B.A. Summer
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K071802
Page 1 of 1
MIPM TeslaGuard Special 510(k) Pace 16 of 215
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.