Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Sep 27, 2007
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 884.2740
Device Class
Class 2
Indications for Use
The Philips Avalon FM20 (M2702A), FM30 (M2703A), Fetal/Maternal Monitors are intended for noninvasive monitoring of parameters of pregnant women during antepartum testing and labor and delivery. The FM30 is additionally intended for invasive monitoring. All monitors are intended for monitoring fetal and maternal heart rates, uterine activity, maternal noninvasive blood pressure, and additionally for the FM30, oxyqen saturation (SpO2). All monitors are intended for generating alarms fetal and maternal parameters, from for displaying, storing and recordinq patient data and related waves, transmitting patient data to a patient information and surveillance system on a network, and for postpartum monitoring of the mother. All monitors are intended for use by trained health care professionals. They are intended for use in labor and delivery rooms, antepartum testing areas and during postpartum recovery in the hospital environment. They are not intended for use in intensive care units or operating rooms. The FM20 and FM30 are additionally intended for use in healthcare facilities outside hospitals, for example in doctors' offices, and for home use.
Device Story
Philips Avalon FM20/FM30 are fetal/maternal monitors; input signals include fetal/maternal heart rate (ultrasonic), uterine activity, maternal non-invasive blood pressure, pulse rate, SpO2 (FM30), and ECG (FM30). Device processes physiological signals to generate real-time waveforms, numerical data, and alarms. Used by trained healthcare professionals in hospitals (labor/delivery, antepartum, postpartum) and private households. Output displayed on integrated screen; data stored, recorded, or transmitted to networked patient information/surveillance systems. Clinical decision-making supported by continuous monitoring of fetal/maternal status; enables early detection of distress. Benefits include expanded monitoring capability outside hospital settings, facilitating remote or home-based care for pregnant patients.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and intended use similarity to predicate devices.
Technological Characteristics
Fetal/maternal monitoring system; non-invasive and invasive (FM30) sensing. Includes ultrasonic transducers, pressure transducers, SpO2 sensors, and ECG leads. Connectivity via network for patient information systems. Form factor: bedside monitor. Software-based signal processing for heart rate and uterine activity. Sterilization: accessories provided non-sterile or sterile as appropriate for clinical use.
Indications for Use
Indicated for trained healthcare professionals to monitor pregnant women and fetuses (single, twins, triplets) in labor/delivery, antepartum areas, and private households. FM20 monitors uterine activity, heart rate, non-invasive blood pressure, pulse rate, and fetal heart rates. FM30 monitors the same plus ECG and oxygen saturation. Contraindicated for use during defibrillation, electro-surgery, MRI, or on patients with cardiac pacemakers/external electrical stimulators. IUP/ECG module (M2738A) contraindicated in domestic establishments connected to public low-voltage supply.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
K071800 p 1/3
## SEP 27 2007
This summary of 510(k) safety and effectiveness 5.0 510(k) information is submitted in accordance with the Summary requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. \$807.92.
> The submitter of this premarket notification is: Michael Asmalsky Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany
Tel: +49 7031 463-1277 Fax: +49 7031 463-2442 Email: michael.asmalsky@philips.com
This summary was prepared on June 26th, 2007.
5.1 Device The names of the devices are the Philips Avalon Names Fetal Monitors FM20 and FM30. Classification names are as follows.
| Device Panel | Classification | ProCode | Description |
|-----------------------------------------------------------|----------------|--------------------------|----------------------------------------------------------------|
| Obstetrical and<br>Gynecological<br>Monitoring<br>Devices | \$884.2660, II | MAA<br>HEL<br>HEK<br>KNG | Fetal ultrasonic monitor and<br>accessories |
| | \$884.2675, II | HGP | Fetal scalp circular (spiral)<br>electrode and applicator |
| | \$884.2700, II | HGS<br>KXO<br>HFO<br>HFN | Intrauterine pressure monitor<br>and accessories |
| | \$884.2720, II | HFM | External uterine contraction<br>monitor and accessories |
| | \$884.2740, II | HGM | Perinatal monitoring system and<br>accessories |
| | \$884.2960, II | HGL | Obstetric ultrasonic transducer<br>and accessories |
| Circulatory<br>System Devices | \$870.1100, II | DSJ | Alarm, Blood Pressure |
| | \$870.1110, II | DSK | Computer, Blood Pressure |
| | \$870.1130, II | DXN | System, Measurement, Blood-<br>Pressure, Non-Invasive |
| | \$870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer Signal |
| | \$870.2300, II | DRT | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm) |
| | \$870.2340, II | DPS | Electrocardiograph |
| | \$870.2600, I | DRJ | System, Signal Isolation |
| | \$870.2700, II | DQA | Oximeter |
| | \$870.2810, I | DSF | Recorder, Paper Chart |
| | \$870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector |
{1}------------------------------------------------
K071800 P²/3
5.2 Subject The subject devices Philips Avalon Fetal Monitors devices
FM20 and FM30 are substantially equivalent to previously cleared Philips Avalon Fetal Monitors FM20 and FM30 marketed pursuant to K052795 and K062137.
- The modification of the Philips Avalon Fetal 5.3 Monitors FM20 and FM30 only introduces the Modifications capability for healthcare professionals of monitoring pregnant women at private households
{2}------------------------------------------------
K071800. P 3/3
5.4 Intended Use
Excludinq the added intended use location of private households, the intended use is the same as previously cleared for the legally marketed predicate device Philips Avalon Fetal Monitors FM20(M2702A)and FM30(M2703A) (K052795 and K062137) .
The Philips Avalon FM20 (M2702A), FM30 (M2703A), Fetal/Maternal Monitors are intended for noninvasive monitoring of parameters of pregnant women during antepartum testing and labor and delivery. The FM30 is additionally intended for invasive monitoring.
All monitors are intended for monitoring fetal and maternal heart rates, uterine activity, maternal noninvasive blood pressure, and additionally for the FM30, oxyqen saturation (SpO2).
All monitors are intended for generating alarms fetal and maternal parameters, from for displaying, storing and recordinq patient data and related waves, transmitting patient data to a patient information and surveillance system on a network, and for postpartum monitoring of the mother.
All monitors are intended for use by trained health care professionals.
They are intended for use in labor and delivery rooms, antepartum testing areas and during postpartum recovery in the hospital environment.
They are not intended for use in intensive care units or operating rooms. The FM20 and FM30 are additionally intended for use in healthcare facilities outside hospitals, for example in doctors' offices, and for home use.
Contraindications: All monitors are NOT intended for :
- use during defibrillation, electro-surgery, or magnetic resonance imaging (MRI).
ECG measurements on patients connected to
external electrical stimulators or with cardiac pacemakers.
- use with the IUP/ECG patient module (M2738A) in domestic establishments and those connected directly to the public low-voltage supply network that supplies buildings used for domestic purposes .
The subject devices Philips Avalon Fetal Monitors FM20 and FM30 have the same technoloqical characteristics as the legally marketed predicate devices Avalon Fetal Monitors FM20 and FM30.
5.5 Technoloqical Characteristics
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the perimeter. Inside the circle is a symbol consisting of three stylized human figures with flowing lines connecting them. The figures are arranged in a vertical stack, with the top figure being the largest and the bottom figure being the smallest.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
SFP 2 7 2007
Mr. Michael Asmalsky Regulatory Affairs Engineer Philips Medizin Systeme Boblingen GmbH Ultrasound and Monitoring - Patient Monitoring Hewlett-Packard Str.2, D-71034 Boeblingen, Germany
Re: K071800
> Trade Name: Phillips Avalon Fetal Monitors FM20 (M2702A) and FM30 (M2073A) Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: June 26, 2007 Received: July 2, 2007
Dear Mr. Asmalsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amerodments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Concessible Act (Act) that do not require approval of a premarket approval application (PMA). You COSIII COSIII therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FPA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows a logo with the text "FDA Centennial" and the years "1906-2006" inside a circular border. Below the logo, there is a cursive text that reads "Protecting and Promoting Public Health". The logo appears to be a commemorative emblem for the FDA's centennial anniversary. The text is slightly blurry, suggesting it may be a scan or a low-resolution image.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its. toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address hitp://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:
Philips Avalon Fetal Monitors FM20 (M2702A) and FM30 (M2703A) .
#### Indications for Use:
### Avalon Fetal Monitor FM20:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.
#### Avalon Fetal Monitor FM30:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.
Prescription Use Yes (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use No (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K071807
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.