K071666 · Bio-Medical Research Corp. · NGX · Mar 12, 2008 · Physical Medicine
Device Facts
Record ID
K071666
Device Name
SLENDERTONE ENERVIVE, MODEL 561
Applicant
Bio-Medical Research Corp.
Product Code
NGX · Physical Medicine
Decision Date
Mar 12, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EnerVive™ is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. EnerVive™ is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the EnerVive™ training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated. The EnerVive™ electrical impulses allow triggering action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles. The various types of muscle work that EnerVive™ can impose on the stimulated muscles are able to improve or facilitate muscle performance. EnerVive™ may therefore be considered a technique of muscle training. EnerVive™ is intended for over the counter sales.
Device Story
Portable, two-channel, battery-operated neuromuscular electrical stimulation (NMES) system. Device delivers low-frequency, constant-current, symmetric biphasic pulses to motor nerves via adhesive skin electrodes. User selects from 6 pre-programmed exercise modes (warm-up, performance, recovery). Stimulation triggers action potentials in motor nerves, inducing mechanical muscle fiber contraction. Used by healthy adults in home/non-clinical settings for muscle conditioning. Output is physical muscle contraction; user monitors intensity/program via device interface. Benefits include improved muscle performance through targeted training patterns.
Clinical Evidence
Bench testing and literature review provided. Clinical performance study (n=6) evaluated the dual electrode design, confirming effective muscle contractions in healthy subjects with no adverse events. Effectiveness of NMES technology for muscle conditioning supported by extensive referenced literature.
Technological Characteristics
Portable 2-channel stimulator; ABS plastic housing; 9V battery powered. Uses constant-current, symmetric biphasic pulses. Fixed-distance dual adhesive electrodes. Complies with EN 60601-1, EN 60601-1-2, and IEC 60601-2-10 safety standards.
Indications for Use
Indicated for healthy adults to stimulate healthy muscles to improve or facilitate muscle performance. Contraindicated for use on injured or ailing muscles or as therapy for medical conditions.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the letters "BMR" in bold, sans-serif font. To the left of the letters is a stylized, crescent-shaped graphic that resembles an eye or a curved bracket. The entire logo has a slightly distressed or textured appearance, as if it were printed with a worn stamp or brush.
K071666
-Medical Research
Tel: +353 (0)91 774300 - Fax: +353 (0)91 774301
This 510k summary is being submitted in accordance with the requirements of 21 CFR 807.92.
MAR 1 2 2008
# 1. Contact Details
| Name: | Anne-Marie Keenan |
|------------|-------------------------------------------------------------------------------|
| Address: | Bio-Medical Research Ltd.,<br>Parkmore Business Park, West<br>Galway, Ireland |
| Telephone: | +353 91 774300 |
| Fax: | +353 91 774301 |
| E-Mail: | akeenan@bmr.ie |
| Prepared: | 13th June, 2007, Revised 11th March 2008 |
# 2. Device Name
| Trade Name of Device: | Slendertone EnerVive™, Type 561 |
|-----------------------|---------------------------------------------------------------|
| Common Name: | Muscle Stimulator |
| Classification Name: | Stimulator, muscle, powered, for muscle conditioning<br>(NGX) |
# 3. Identification of Equivalent Legally Marketed Device
| Name: | Compex Sport, Model Sport 3 |
|----------------|-----------------------------|
| Manufacturer: | Compex S.A. |
| 510(k) Number: | K011880 |
### 4. Description of Device
EnerVive™ is the latest product offering from Slendertone, a division of Bio-Medical Research Ltd. It is a portable two-channel; battery operated neuromuscular electronic stimulation system. The device uses the principles of Neuromuscular Electrical Nerve Stimulation (NMES) and is intended for over the counter use by healthy adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance. It is not intended as a therapy for any medical condition.
{1}------------------------------------------------
The two channels may be operated independently when each lead wire is attached to the unit and to the supplied butterfly-shaped electrode pads. Detailed diagrams for correct pad placement are available as part of the instructions for use.
EnerVive™ stimulates action potentials in the motor nerves supplying muscle. Different patterns of muscle activity can be imposed on the target muscle, depending on the timing and intensity of the stimulation signal. EnerVive™ contains 6 programs for exercise warm-up, muscle performance improvement and exercise recovery.
Included in the product pack are an ergonomically designed EnerVive™ control unit, 4 adhesive pads (2 pairs), 1 x 9 volt battery (6F22), a belt clip, 2 lead wires, travel pouch and instructions for use.
#### 5. Statement of Intended Use/Indications for Use
Indications for use are the same as the listed predicate device. EnerVive™ is intended for use by healthy persons to apply transcutaneous electrical muscle stimulation (EMS) through skin contact electrodes. Indications for use are as follows:
"Ener Vive™" is intended to stimulate healthy muscles in order to improve or facilitate muscle performance.
"EnerVive™" is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the "EnerVive™" training programs are designed for injured or ailing muscles and its use on such muscles is contraindicated.
The "EnerVive™" electrical impulses allow triggering action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.
The various types of muscle work that "EnerVive™" can impose on the stimulated muscles are able to improve or facilitate muscle performance. "EnerVive™" may therefore be considered a technique of muscle training.
EnerVive™" is intended for over the counter sales.
{2}------------------------------------------------
## 6. Summary of Technological Characteristics
A summary of the technological characteristics of the EnerVive™ device compared to the predicate devices in terms of design, material and energy source is given below:
| | EnerVive™ | Compex Sport |
|---------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Type/Model | 561 | Sport 3 |
| Design | Stimulator unit, leads/cables (2),<br>adhesive electrodes (4), user<br>instructions & pouch.<br>2 channels | Stimulator unit, battery charger,<br>leads/cables (4), adhesive<br>electrodes, user instructions & case.<br>4 channels |
| Material | Unit Housing - ABS Plastic | Unit Housing - Plastic |
| Energy Source | 1 x 9 volt battery | 7.2 V NiMH rechargeable battery |
Comparison in basic unit characteristics and output specifications for the device and the predicate have been compiled using the "Guidance Document for Powered Muscle Stimulator 510(k)s", June 1999 and have been submitted as part of this 510(k) Premarket Notification.
Slendertone EnerVive™ uses the same technology as the predicate device i.e. low frequency pulsed electrical stimulation, using constant-current controlled, symmetric biphasic pulses. Both products use almost identical timing parameters, and it is the frequency ranges of the applied pulses that determine the pattern of activity imposed on the muscle fibers.
### 7. Substantial Equivalence
Bio-Medical Research Ltd (BMR) has over 30 years experience in the research, design, manufacture and marketing of medical grade products for both muscle strengthening and pain relief. Bio-Medical Research Ltd. complies with 21 CFR 820 and is registered to I.S. EN ISO 13485:2003, Medical Device Quality Management System for the design, manufacture and distribution of electro-medical devices.
EnerVive™ complies with the following international safety standards:
- EN 60601-1-2:2001 Medical electrical equipment Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and Tests.
- EN 60601-1 (1990) + A1 (1993) + A2 (1995) +A13 (1996) + Corrigenda (July 1994) -Medical Electrical Equipment - Part 1: General Requirements for Safety
- D IEC 60601-2-10, 1st ed., 1987 Medical Electrical Equipment Part 2: Particular Requirements For The Safety Of Nerve And Muscle Stimulators.
{3}------------------------------------------------
A risk management plan has been carried out to I.S. EN ISO 14971 2001 AMD 1 2003.
As with the predicate device, the evidence for effectiveness of this technology in facilitating or improving the performance of skeletal muscle relies on the large body of research literature on EMS extending over several decades. The following research articles have been referenced for this 510(k) Premarket Notification.
- 1. Alon G. "Principles of Electrical Stimulation" in Nelson RM., Currier DP. " Clinical Electrotherapy", 2nd Ed, pp 35-103 Appleton and Lange 1991
- 2. Baker LL, Wederich CL, McNeal DR, Newsam C, Waters RL. "Neuromuscular Electrical Stimulation. A Practical Guide. 4th Ed. Rancho Los Amigos. 2000
- 3. Bax Leon, Staes Filip, Verhagen Arianne. "Does Neuromuscular Electrical Stimulation Strengthen the Quadriceps Femoris, A Systematic Review of Randomised Controlled Trials" Published Sports Med 2005:35(3):191-212 0112-1642/05/0003-0191/S34.95/0
- 4. Hainault Karl, Duchateau Jacques "Neuromuscular Electrical Stimulation and Voluntary Sports Medicine 14(2):100-113.1992 0112-1642/92/0008-Exercise" Published 0100/\$7.00/0
- S. C. Small and M. J. Stokes, "Stimulation frequency and force potentiation in the human 5. adductor pollicis muscle" European Journal of Applied Physiology, vol. 65, p. 5, 1992.
- G. M. Lyons, G. E. Leane, and P. A. Grace, "The effect of electrical stimulation of the 6. calf muscle and compression stocking on venous blood flow velocity," Eur J Vasc Endovasc Surg, vol. 23, pp. 564-6, Jun 2002.
- 7. J. E. Sherry, K. M. Oehrlein, K. S. Hegge, and B. J. Morgan, "Effect of burst-mode transcutaneous electrical nerve stimulation on peripheral vascular resistance," Phys Ther, vol. 81, pp. 1183-91, Jun 2001.
Further clinical data [8] was presented to detail the effectiveness of a dual electrode design in producing adequate muscle contractions. The dual electrode design, unlike that of the predicate, maintains a fixed distance between the electrodes. The study was performed in
{4}------------------------------------------------
compliance with Good Clinical Practices (GCP) and includes the archiving of essential documents.
- 8. Ciaran Byrne, MSc., Eta Tierney, "Quantitative Performance Measures of Slendertone Enervive," Bio-Medical Research Institute, Galway, Ireland.
Data was collected on 6 subjects who were representative of the intended use population. All participants provided informed consent prior to participation.
The study was conducted in compliance with the protocol, applicable FDA Regulations, applicable local Regulations, Good Clinical Practice Guidance, and the Declaration of Helsinki.
The results showed that effective muscle contractions were elicited when a group of healthy adults used EnerVive™. There were no adverse events during this study.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, with a human profile visible in the negative space between the wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Bio-Medical Research Ltd. % Ms. Anne-Marie Keenan Parkmore Business Park West Galway Ireland
MAR 1 2 2008
Re: K071666 Trade/Device Name: Slendertone EnerVive™, Type 561 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: December 13, 2007 Received: December 17, 2007
Dear Ms. Keenan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 – Ms. Anne-Marie Keenan
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Mark N Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name:
EnerVive™, Type 561
Indications for Use:
"EnerVive™" is intended to stimulate healthy muscles in order to improve or facilitate muscle performance.
"EnerVive™" is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the "EnerVive™ training programs are designed for injured or ailing muscles and it's use on such muscles is contraindicated.
The "EnerVive™' electrical impulses allow triggering action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the muscle fibers via the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration of contraction, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.
The various types of muscle work that "EnerVive™ can impose on the stimulated muscles are able to improve or facilitate muscle performance. "EnerVive™ may therefore be considered a technique of muscle training.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
> Concurrence of CD Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative. and Neurological Devices
510(k) Number K071666
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.