The Model 960 Series is a desktop monitor intended for use as a continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, non invasive blood pressure and temperature. The intended patient population comprises adult, and pediatric weighing more than 5Kg. The intended environment of use is hospital environment. Hospital use typically covers areas such as general care floors, operating rooms, special procedure areas, intensive and critical care areas within the hospital plus hospital-type facilities such as surgical centers, sub-acute centers, special nursing facilities and sleep labs. The Model 900 is an Pulse Oximeter only. WARNING: The Model 900 and 960 Series is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.
Device Story
Mediaid M960/M900 series are desktop vital signs monitors. M960 monitors SpO2, pulse rate, NIBP, and temperature; M900 monitors SpO2 and pulse rate. SpO2 measurement uses red/infrared light transmission through pulsating vascular beds; photodiode detects light changes; non-pulsatile signals are electronically removed for calculation. NIBP uses Suntech Medical OEM module; temperature uses YSI-compatible probe. Devices are used in hospital environments (OR, ICU, general care) by clinicians. Output is displayed in real-time for patient assessment. Devices serve as adjuncts to clinical signs/symptoms to assist in monitoring patient status.
Clinical Evidence
Clinical hypoxia studies conducted in independent lab compared SpO2 readings to co-oximeter blood samples. NIBP accuracy verified via clinical study on Suntech OEM module per IEC 60601-2-30 and AAMI SP10. Thermometer accuracy verified in hospital setting per ASTM E 1112:00 and EN 12470-4. Bench testing performed for pulse rate accuracy. Biocompatibility, electrical safety, and EMC testing performed.
Technological Characteristics
Desktop monitor; SpO2 via red/infrared light photoplethysmography; NIBP via Suntech OEM module; temperature via YSI-compatible probe. Standards: IEC 60601-1-2 (EMC), IEC 60601-2-30 (NIBP), AAMI SP10 (NIBP), ASTM E 1112:00 (Thermometer), EN 12470-4 (Thermometer).
Indications for Use
Indicated for continuous noninvasive monitoring of SpO2, pulse rate, NIBP, and temperature in adult and pediatric patients weighing >5kg in hospital and hospital-type settings. Adjunct for patient assessment; must be used with clinical signs/symptoms.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Tango+Automatic Blood pressure and Oxygen Saturation Measurement System of Suntech Medical, Inc (K053209)
Submission Summary (Full Text)
{0}------------------------------------------------
page 1 of 4
MAY - 2 2008
# Section 5 - 510(k) Summary of Safety and Effectiveness
K07/642
1. Submitter Information:
Mediaid, Inc. 17517 Fabrica Way #H Cerritos, CA 90703 Registration # 2087439
2. Contact:
Jayesh Patel, CEO Telephone: 714-367-2848 Fax: 714-367-2852
3. Date Prepared: April 1st, 2008
#### 4. Description:
Mediaid M960 Series, Vital Signs Monitor and Mediaid M900 Series, Pulse Oximeter Trade Name: Common Name: Vital signs measurement devices Name: Non-invasive pulse Oximeter, SpO2 (21CFR870.2700), also contains NIBP measurement system (21CFR870.1130) and clinical electronic thermometer (21CFR880.2910) Classification: Class II
Mediaid M900 Series, Pulse Oximeter Trade Name: Common Name: Pulse Oximeter Name: Non-invasive pulse Oximeter, SpO2 (21CFR870.2700) Classification: Class II
The Mediaid Vital Signs monitor Pulse Oximeter Model M960P provides real time monitoring and display of ox vgen saturation of arteriolar hemoglobin (SpO2), Non Invasive Blood Pressure, pulse rate and body temperature. Where as the Mediaid Pulse Oximeter Model M900 provides real time monitoring and display of oxygen saturation of arteriolar hemoglobin (SpO2) and pulse rate only.
Noninvasive arterial oxygen saturation measurement is obtained by directing red and infra red light through a pulsating vascular bed. The pulsating arterioles in the path of the light beam cause a change in the amount of light detected by a photodiode. The pulse oximeter determines the oxygen saturation of arterial blood by measuring the ratio of transmitted red to infrared light within the pulse waveform. The non-pulsatile signal is removed electronically for the purpose of calculation. Therefore, skin, bone, and other non-pulsating substances do not interfere with the measurement of arterial oxygen saturation.
{1}------------------------------------------------
Non invasive blood pressure is intended to non invasively measure systolic, diastolic and mean arterial pressures (MAP).
Temperature- Thermometer takes patient temperature in oral, axillary or rectal mode.
## 5. Substantial Equivalence:
The Mediaid Pulse Oximeter Model 900 is substantially equivalent to the following devices:
- Model 305 of Mediaid Palco marketed under K943842 .
- Model 340 of Mediaid Palco marketed under K920066 .
- . Model 120 of Mediaid Palco marketed under K994372
The Mediaid Vital Signs Monitor Model 960 series is substantially equivalent to the following devices:
For Pulse Oximeter :
- Model 305 of Mediaid Palco marketed under K943842 ●
- Model 340 of Mediaid Palco marketed under K920066 �
- Model 120 of Mediaid Palco marketed under K994372 �
For Ni-BP Monitor:
- Welch Allyn Spot Ultra Vital Signs marked under K040490 .
- Tango+Automatic Blood pressure and Oxygen Saturation Measurement System of Suntech Medical . Inc Marketed under K053209.
For Clinical Thermometer:
- Welch Allyn Spot Ultra Vital Signs marked under K040490 .
12
#### Intended Use:
## Indications for Use
The Model 960 Series is a desktop monitor intended for use as a continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, non invasive blood pressure and temperature. The intended patient population comprises adult, and pediatric weighing more than 5Kg. The intended environment of use is hospital environment. Hospital use typically covers areas such as general care floors, operating rooms, special procedure areas, intensive and critical care areas within the hospital plus hospital-type facilities such as surgical centers, sub-acute centers, special nursing facilities and sleep labs. The Model 900 is an Pulse Oximeter only.
{2}------------------------------------------------
WARNING: The Model 900 and 960 Series is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.
## 510(k) Number (if known): K071642
Device Name: Mediaid M960, and M960P Vital Signs Monitor and Mediaid M900, Mediad 900P Pulse Oximeter
#### Comparison to Predicate Device:
Mediaid Pulse Oximeter Model 900 series and pulse Oximeter module of Model 960 series uses the same theory and principle of operation as the predicate device. Design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by product testing and accuracy claims.
Mediaid uses Suntech Medical Inc's Advantage OEM BP Module in Model 960 series. This module is compared with Tango+Automatic Blood Pressure and oxygen saturation measurement system of Suntech Medical Inc. NiBP module of Welch Allyn Spot Ultra Vital Signs monitor. It is observed that design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by product testing and accuracy claims.
Mediaid use YSI compatible temperature probe for temperature measurement and its performance is compared with the thermometer module of Welch Allyn Spot Ultra Vital Signs monitor. It is observed that design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by product testing and accuracy claims.
## Technological Characteristics
A comparison of the technological characteristics of Model 960 Vital Signs Monitor and the predicate devices has been performed. The results of this comparison demonstrate that the Model 960 series is substantially equivalent to the marketed predicate devices in technological characteristics.
#### Environmental and non-clinical testing
Applicable environmental and non-clinical testing was performed per IEC60601-1-2 as well as other applicable standards and procedures. The Mediaid Pulse Oximeter Model 960 series passed all tests as per the requirements of various applicable international standards.
#### Performance Data & Conclusions:
{3}------------------------------------------------
Performance testing was conducted during clinical hypoxia studies conducted in an independent research lab. Mediaid Pulse Oximeter Model 900 scries readings were compared to arterial blood samples analyzed on a laboratory co-oximeter and found to be equivalent to predicate device accuracy claims. Bench testing was performed to verify pulse rate accuracy.
For Ni-BP, clinical study was conducted by Suntech Medical Inc on their OEM module and the results show that the claimed accuracy is met as per the requirement of IEC 60601-2-30 and AAMI SP10.
Clinical Tests for Clinical thermometer is conducted on an external hospital and the results shows that the thermometer module of Model 960 series meets the claimed accuracy as per ASTM E 1112:00 and EN 12470-4
Biocompatibility, electrical safety, and EMC testing were also performed to demonstrate conformance with established industry standards.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three wing-like shapes, oriented towards the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 2 2008
Mr. Jayesh Patel Chief Executive Officer Mediaid, Incorporated 17517 Fabrica Way, Suite H Cerritos, California 90703
Re: K071642
Trade/Device Name: Mediaid M960, and M960P Vital Signs Monitor and Mediaid M900, Mediaid 900P Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, DXN Dated: April 24, 2008 Received: April 28, 2008
Dear Mr. Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - Mr. Patel
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Section 4 - Indications for Use Statement
Detailed instructions are given in the user's manual.
#### Indications for Use
The Model 960 Series is a desktop monitor intended for use as a continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, non invasive blood pressure and temperature. The intended patient population comprises adult, and pediatric weighing more than 5Kg. The intended environment of use is hospital environment. Hospital use typically covers areas such as general care floors, operating rooms, special procedure areas, intensive and critical care areas within the hospital plus hospital-type facilities such as surgical centers, sub-acute centers, special nursing facilities and sleep labs. The Model 900 is an Pulse Oximeter only.
WARNING: The Model 900 and 960 Series is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.
510(k) Number (if known): K071642
Device Name: Mediaid M960, and M960P Vital Signs Monitor and Mediaid M900, Mediad 900P Pulse Oximeter
Prescription Use XX · (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mi Zhang
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K071692 510(k) Number: ___
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.