iMagic v2.0™ device is software intended for viewing and diagnostic interpretation of images acquired from Ultrasound and other DICOM compliant medical imaging systems (CT, MR, CR, DR), when installed on suitable commercial standard hardware. iMagic v2.0™ receives Ultrasound images and other modality imaging studies over a network from servers, directly from the imaging modality or from an archive (including media) utilizing both lossless (reversible) and lossy (irreversible) compression. iMagic v2.0™ does not use lossy (irreversible) compression during image handling, manipulation, or storage. Only DICOM, for presentation, images will be captured for display and diagnosis. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
Device Story
iMagic v2.0 is a PACS software application for medical imaging centers. It receives DICOM-compliant images from imaging modalities, servers, or archives via network or media. The software allows clinicians to view, retrieve, and compare images across patient visits, edit AVI/cine loops, crop images, and generate reports for distribution via print, email, or CD. It includes features for patient record management, statistical data generation, and image tagging. The device is operated by healthcare professionals in a clinical environment. It does not contact the patient or control life-sustaining equipment; a physician provides human intervention for all diagnostic interpretations. The system facilitates clinical decision-making by providing organized access to patient imaging history and diagnostic reports.
Clinical Evidence
No clinical data provided. The submission relies on bench testing and hazard analysis. The device is classified as having a 'minor' level of concern for potential hazards.
Technological Characteristics
Software-based PACS; DICOM compliant; operates on commercial standard hardware; supports lossless/reversible compression; no patient contact; no life-sustaining control; minor level of concern for hazards.
Indications for Use
Indicated for viewing and diagnostic interpretation of medical images (Ultrasound, CT, MR, CR, DR) in clinical settings. Contraindicated for primary interpretation of lossy compressed mammographic images and digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K071602
l
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared: May 19, 2007
Submitter's Information: 21 CFR 807.92(a)(1) Mr. Sunder Natrajan, CEO Ashva Technologies Pvt. Ltd., 15 ARK Colony, Urrnilla House Eldams RD. Alwarpet, Chennai India 600018 +11 91 44 420 88897 (fax) sunder@ashvatech.com
JUL 2 3 2007
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
| Trade Name: | iMagic v2.0™ |
|------------------------|-----------------------------------------|
| Common Name: | Picture Archiving Communications System |
| Classification Name: | system, image processing, radiological |
| Product code: | LLZ |
| Device Classification: | 892.2050 |
Predicate Device: 21 CFR 807. 92(a)(3)
| iMagic v2.0™ is substantially equivalent to: | |
|--------------------------------------------------------------------|----------------------|
| Manufacturer: | Voyager Imaging |
| Device Name: | Voyager PACS Stystem |
| 510(k) Number: | K062062 |
| Product Code: | LLZ |
| Device Classification Name: SYSTEM, IMAGE PROCESSING, RADIOLOGICAL | |
| Regulation Number: | Class II - 892.2050 |
### Device Description: 21 CFR 807 92(a)(4)
iMagic, a software application for medical imaging centers, that allows to easily record patient details with study information along with their images and to prepare reports that can be distributed to the patient either in paper printout or through the Internet (email). Moreover, it allows the user to retrieve images easily and to compare the images of the patient between different visits or to distribute the patient records in CD's. It also allows generating statistical data from the available details. Overall features include:
- . Prevent unauthorized access of Patient Records
- Easy search of Patient details .
- . Flexibility to design auto patient ID
- Report transmission by mail .
- . Cropping of live images
{1}------------------------------------------------
- Editing of AVI/cine loops �
- Distribution of Reports and Images in CD .
- . Comparison of images between different visits of patient
- . Tagging of significant images/cineloops for future references
- . Generation of statistical data from available details
### Indications for Use: 21 CFR 807 92(a)(5)
iMagic v2.0™ device is software intended for viewing and diagnostic interpretation of images acquired from Ultrasound and other DICOM compliant medical imaging systems (CT, MR, CR, DR), when installed on suitable commercial standard hardware.
iMagic v2.0™ receives Ultrasound images and other modality imaging studies over a network from servers, directly from the imaging modality or from an archive (including media) utilizing both lossless (reversible) and lossy (irreversible) compression. iMagic v2.0™ does not use lossy (irreversible) compression during image handling, manipulation, or storage.
Only DICOM, for presentation, images will be captured for display and diagnosis.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images mav only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
### Technological Characteristics: 21 CFR 807 92(a)(6)
The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.
### Conclusion: 21 CFR 807 92(b)(1)
The 510(k) Pre-Market Notification for iMagic v2.0™ contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.
iMagic v2.0™ has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the "Level of Concern for potential hazards has been classified as "minor".
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and a wavy line representing its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged in a circular fashion around the eagle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
# JUL 2 3 2007
Ashva Technologies, Pvt. Ltd. c/o Mr. Carl Alletto Consultant OTech, Inc. 1600 Manchester Way CORINTH TX 76210
Re: K071602
> Trade/Device Name: iMagic v2.0TM Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 11, 2007 Received: June 11, 2007
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906-2006 FDA Centennial" in it. The logo has a star at the top and three stars at the bottom. The text is arranged in a circular fashion around the central FDA logo.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation mumber at the top of this letter.
> 21 CFR 876.xxxx 21 CFR 884.xxxx 21 CFR 892.xxxxx Other
240-276-0115 (Gastroenterology/Renal/Urology) (Obstetrics/Gynecology) 240-276-0115 (Radiology) 240-276-0120 240-276-0100
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### (Indications for Use Form)
510(k) Number:
Device Name:
iMagic v2.0
Indications for Use:
iMagic v2.0™ device is software intended for viewing and diagnostic interpretation of images acquired from Ultrasound and other DICOM compliant medical imaging systems (CT, MR, CR, DR), when installed on suitable commercial standard hardware.
iMagic v2.0™ receives Ultrasound images and other modality imaging studies over a network from servers, directly from the imaging modality or from an archive (including media) utilizing both lossless (reversible) and lossy (irreversible) compression. iMagic v2.0™ does not use lossy (irreversible) compression during image handling, manipulation, or storage.
Only DICOM, for presentation, images will be captured for display and diagnosis.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use <span style="text-decoration: overline;">✓</span><br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use <span style="text-decoration: overline;"></span><br>(21 CFR 807 Subpart C) |
|---------------------------------------------------------------------------------------------------|--------|-------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------|--------|-------------------------------------------------------------------------------------------------|
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
| 510(k) Number | <span style="text-decoration: overline;">2071602</span> |
|---------------|---------------------------------------------------------|
|---------------|---------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.