← Product Code [LLZ](/productcode/LLZ) · K071492

# FIRSTVIEW (K071492)

_Riverain Medical Group,Llc · LLZ · Aug 28, 2007 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K071492

## Device Facts

- **Applicant:** Riverain Medical Group,Llc
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Aug 28, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The FirstView™ is intended to display and manipulate x-ray images.

## Device Story

FirstView™ is a radiological image processing system consisting of a dedicated server and client software installed on existing workstations. It receives chest x-ray images or CAD results from digital sources or PACS networks. Upon receipt, the server queries the PACS for related study images and manages them within a proprietary database. The system acts as an interface for CAD components and displays DICOM header information. Used in clinical settings by physicians, radiologists, and technicians, the device facilitates the display and manipulation of images to assist in clinical review and decision-making.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

System consists of a dedicated server and client software for existing workstations. Operates as a Picture Archiving and Communication System (PACS) interface. Manages images via a proprietary database. Supports DICOM header display and CAD component integration. Class II device under 21 CFR 892.2050.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- R2 Breast Imaging Workstation ([K051743](/device/K051743.md))
- Centricity PACS System ([K043415](/device/K043415.md))
- Platinum Reading, Review, and Analysis Workstation ([K981217](/device/K981217.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows a signature or handwritten text that appears to read "KD71492". The text is written in a cursive style with varying stroke thicknesses, giving it a dynamic and personal feel. The background is plain white, which helps to emphasize the dark, bold strokes of the signature.

## 510(K) SUMMARY 3.0

| Submission Date:       | AUG 28 2007<br>May 25, 2007                                                                                                                                                                                                                                                                                                                                                                            |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Information: |                                                                                                                                                                                                                                                                                                                                                                                                        |
| Company Name:          | Riverain Medical Group, LLC.                                                                                                                                                                                                                                                                                                                                                                           |
| Company Address:       | 3020 South Tech Blvd., Miamisburg, OH 45342-4860                                                                                                                                                                                                                                                                                                                                                       |
| Contact Person:        | Jennifer Steinke<br>Director, Regulatory Affairs and Quality Assurance<br>Riverain Medical<br>800.990.3387<br>937.425.6493<br>jsteinke@riverainmedical.com                                                                                                                                                                                                                                             |
| Device Information:    |                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade Name:            | FirstView™                                                                                                                                                                                                                                                                                                                                                                                             |
| Common Name:           | System, Image Processing, Radiological                                                                                                                                                                                                                                                                                                                                                                 |
| Classification Name:   | Picture archiving and communications system                                                                                                                                                                                                                                                                                                                                                            |
| Device Class:          | Class II, 21 CFR §892.2050                                                                                                                                                                                                                                                                                                                                                                             |
| Predicate Devices:     | R2 Breast Imaging Workstation (K051743)<br>R2 TECHNOLOGY, INC<br>Class II                                                                                                                                                                                                                                                                                                                              |
|                        | Centricity PACS System (K043415)<br>GE Medical Systems<br>Class II                                                                                                                                                                                                                                                                                                                                     |
|                        | Platinum Reading, Review, and Analysis Workstation<br>(K981217)<br>GE Medical Systems<br>Class II                                                                                                                                                                                                                                                                                                      |
| Device Description:    | The FirstView™ consists of a dedicated server that has<br>been programmed with a database and server software, as<br>well as client software that is loaded onto an existing<br>workstation. New images, which may be a chest x-ray or<br>CAD result, are sent to the FirstView™ server from a                                                                                                         |
|                        | Digital or from the PACS network, prompting FirstView to<br>query the PACS for additional images related to that study.<br>The FirstView™ server manages the images within its<br>Riverain proprietary database.                                                                                                                                                                                       |
| Intended Use:          | The FirstView™ is intended to display and manipulate x-<br>ray images.                                                                                                                                                                                                                                                                                                                                 |
| Indications for Use:   | The FirstView™ is indicated for displaying and<br>manipulating chest x-ray images. It acts as an interface<br>with the CAD component, if installed or available at the<br>clinical site, and provides display of information contained<br>within the DICOM header of the images. This device is<br>intended to be used by trained professionals, such as<br>physicians, radiologists, and technicians. |

{1}------------------------------------------------

## Comparison to Predicate Device:

|             | FirstViewTM is substantially equivalent to the cited<br>predicate devices. Differences in the design and<br>performance from the cited predicate devices do not affect<br>either the safety or effectiveness of the FirstViewTM for its<br>intended use.                     |
|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | FirstViewTM has the same intended use and technology as<br>the legally-marketed predicate devices. Riverain Medical<br>Group, LLC, has determined that the FirstViewTM is as safe<br>and effective as the predicate devices that have been<br>identified in this submission. |

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows a circular logo with text around the perimeter and a stylized graphic in the center. The text, which appears to be in a non-English language, curves along the top and bottom edges of the circle. The central graphic consists of three curved lines that converge at the bottom, resembling a stylized bird in flight or a flowing water design. The logo has a simple, monochrome design.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Ms. Jennifer Steinke Director, Regulatory Affairs and Quality Assurance Riverain Medical Group, LLC 3020 South Tech Blvd. MIAMISBURG OH 45342-4860

AUG 2 8 2007

Re: K071492

Trade/Device Name: FirstView™M Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: August 1, 2007 Received: August 2, 2007

Dear Ms. Steinke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top. The letters "PA" are prominently displayed in the center of the logo. Below the letters, the word "Centennial" is written in a smaller font. Three stars are arranged in a horizontal line beneath the word "Centennial".

cling and Promoting Public Shoath

{3}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)         | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                     | 240-276-0120 |
| Other          |                                 | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours.

Nancy C Hodgon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

## INDICATIONS FOR USE STATEMENT 2.0

510(k) Number (if known): K071492

FirstView™ Device Name:

Indications for Use:

The FirstView™ is indicated for displaying and manipulating chest x-ray images. It acts as an interface with the CAD component, if installed or available at the clinical site, and provides display of information contained within the DICOM header of the images. This device is intended to be used by trained professionals, such as physicians, radiologists, and technicians.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

TWhang

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K071492](https://fda-staging.innolitics.com/device/K071492)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).