DIMENSION CLINICAL CHEMISTRY SYSTEM MPO FLEX REAGENT CARTRIDGE, CALIBRATOR,CONTROL, MODEL RF425, RC425,RC426
Applicant
Siemens Healthcare Diagnostics
Product Code
NTV · Immunology
Decision Date
Dec 10, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5600
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K071474 · Dec 10, 2008
DIMENSION CLINICAL CHEMISTRY SYSTEM MPO FLEX REAGENT CARTRIDGE, CALIBRATOR,CONTROL, MODEL RF425, RC425,RC426
Siemens Healthcare Diagnostics
Medical records; Medical database searches
Retrospective review of medical records and databases to determine the incidence of MACE in patients presenting with chest pain, used to evaluate the clinical utility of MPO measurements in predicting cardiac events.
Retrospective clinical study; Follow-up/Duration: 30 days and 180 days post-presentation
400 patients presenting to the Emergency Department or outpatient facilities with chest pain or symptoms suggestive of Acute Coronary Syndrome (ACS); Sample Size: 400
Not applicable for this study
Incidence of Major Adverse Cardiac Events (MACE) including myocardial infarction, recurrent ischemia, and death
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Myeloperoxidase (MPO) concentration
—
—
—
—
—
Clinical study of 400 EDTA plasma samples from patients presenting to the Emergency Department or outpatient facilities with chest pain or symptoms suggestive of Acute Coronary Syndrome.
—
Indications for Use
The MPO method is an in vitro diagnostic test for the quantitative measurement of myeloperoxidase (MPO) in human plasma on the Dimension® clinical chemistry system with the heterogeneous immunoassay module. Myeloperoxidase measurements may be used in conjunction with clinical history, ECG, and cardiac biomarkers to evaluate patients presenting with chest pain that are at risk for major adverse cardiac events, including myocardial infarction, need for revascularization, or death.
Device Story
The Dimension MPO Flex reagent cartridge is an in vitro diagnostic immunoassay for quantitative MPO measurement in human plasma. It utilizes a one-step sandwich enzyme immunoassay principle on the Dimension clinical chemistry system. The process involves incubating patient samples with chromium dioxide particles coated with MPO-specific monoclonal antibodies and a beta-galactosidase-labeled conjugate. Magnetic separation removes unbound conjugate; the remaining sandwich is reacted with a chromogenic substrate (CPRG). Hydrolysis releases a chromophore (CPR), with color change measured at 577 nm. The concentration of MPO is directly proportional to the rate of color change. The device is used in clinical laboratory settings by trained personnel. Results are provided to physicians to assist in risk stratification for MACE in patients presenting with chest pain, alongside other clinical data. It aids in identifying patients at higher risk for cardiac events, potentially guiding clinical management decisions.
Clinical Evidence
Clinical study of 400 patients presenting with chest pain or ACS symptoms. MACE (myocardial infarction, revascularization, or death) assessed at 30 days and 6 months. Logistic regression showed MPO concentration correlates with MACE risk; odds ratios increased with MPO quartiles. MPO provided statistically significant incremental value (p<0.05) over race, gender, hyperlipidemia, and hsCRP in predicting MACE. MPO levels >633 pmol/L identified patients at higher risk for MACE even when Troponin I was <0.07 ng/mL.
Technological Characteristics
Sandwich enzyme immunoassay; chromium dioxide particles; beta-galactosidase label; chromogenic substrate (CPRG); 577 nm photometric detection. Compatible with Dimension clinical chemistry system (heterogeneous immunoassay module). Sample types: EDTA, lithium heparin, or sodium heparin plasma. Analytical specificity: <0.15% cross-reactivity with common cardiac markers (Troponin I, CRP, etc.).
Indications for Use
Indicated for quantitative measurement of myeloperoxidase (MPO) in human plasma to evaluate patients presenting with chest pain at risk for major adverse cardiac events (MACE), including myocardial infarction, revascularization, or death. Used in conjunction with clinical history, ECG, and cardiac biomarkers. Prescription use only.
Regulatory Classification
Identification
A low-density lipoprotein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the low-density lipoprotein in serum and other body fluids. Measurement of low-density lipoprotein in serum may aid in the diagnosis of disorders of lipid (fat) metabolism and help to identify young persons at risk from cardiovascular diseases.
Predicate Devices
PrognostiX MPO
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k071474
B. Purpose for Submission:
New device
C. Measurand:
Myeloperoxidase
D. Type of Test:
Quantitative, Enzyme Immunoassay
E. Applicant:
Dade Behring, Inc.
F. Proprietary and Established Names:
Dimension MPO Flex reagent cartridge
Dimension MPO Calibrator
Dimension MPO Control
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NTV (Myeloperoxidase, Immunoassay, System, Test) | II | 21 CFR 866.5600 | 82 (immunology) |
| JIT (Calibrator, secondary) | II | 21 CFR 862.1150 | 75 (chemistry) |
| JJX (Quality Control Material) | I | 21 CFR 862.1660 | 75 (chemistry) |
{1}
H. Intended Use:
1. Intended use(s):
See indications for use statement below.
2. Indication(s) for use:
MPO Flex® reagent cartridge:
The MPO method is an *in vitro* diagnostic test for the quantitative measurement of myeloperoxidase (MPO) in human plasma on the Dimension® clinical chemistry system with the heterogeneous immunoassay module. Myeloperoxidase measurements may be used in conjunction with clinical history, ECG, and cardiac biomarkers to evaluate patients presenting with chest pain that are at risk for major adverse cardiac events, including myocardial infarction, need for revascularization, or death.
MPO Calibrator:
The MPO Calibrator is an *in vitro* diagnostic product intended to be used to calibrate the Myeloperoxidase (MPO) method on the Dimension® clinical chemistry system with the heterogeneous immunoassay module.
MPO Control:
The myeloperoxidase control is an *in vitro* diagnostic product intended for use as an assayed quality control product to monitor the performance of the Myeloperoxidase (MPO) method on the Dimension® clinical chemistry system with the heterogeneous immunoassay module.
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
Siemens Dimension® RxL
I. Device Description:
MPO Reagent:
The MPO reagent contains 4 flex cartridges per carton. Reagents are contained in 8 segregated wells in a plastic cartridge. Wells 1 and 2 contain the liquid mouse monoclonal conjugate preparation. Wells 3 and 4 contain a tablet of chromium dioxide linked mouse monoclonal. Wells 5 and 6 contain a CPRG substrate reagent. Well 7
{2}
contains a diluent for the CPRG tablet. Well 8 contains a diluent for the chromium dioxide tablet.
## MPO Calibrator:
The MPO calibrators contain ten 2.0 mL vials (2 vials per 5 levels) of calibrators. The MPO Calibrator is a frozen bovine albumin based product containing human myeloperoxidase, stabilizers and preservatives in a synthetic matrix.
The sponsor states that all donors of human serum supplied with the kit have been tested and found negative for hepatitis B surface antigen (HBsAg) and antibodies to human immunodeficiency virus (HIV1 and HIV2) and hepatitis C virus.
## MPO Control:
The MPO Controls contain twelve 2.0 mL vials (6 vials per 2 levels) of controls. The MPO control is a frozen bovine albumin based product containing human myeloperoxidase, stabilizers and preservatives in a synthetic matrix.
The sponsor states that all donors of human serum supplied with the kit have been tested and found negative for hepatitis B surface antigen (HBsAg) and antibodies to human immunodeficiency virus (HIV1 and HIV2) and hepatitis C virus.
## J. Substantial Equivalence Information:
| Item | Predicate - PrognostiX MPO | Dimension® MPO |
| --- | --- | --- |
| Intended Use | The CardioMPO Reagent Kit is an enzyme immunoassay intended for the quantitative determination of myeloperoxidase in human plasma, to be used in conjunction with clinical history, ECG and cardiac biomarkers to evaluate patients presenting with chest pain that are at risk for major adverse cardiac events, including myocardial infarction, need for revascularization, or death. | The MPO method is an in vitro diagnostic test for the quantitative measurement of myeloperoxidase (MPO) in human plasma on the Dimension® clinical chemistry system with the heterogeneous immunoassay module. Myeloperoxidase measurements may be used in conjunction with clinical history, ECG, and cardiac biomarkers to evaluate patients presenting with chest pain that are at risk for major adverse cardiac events, including myocardial infarction, need for revascularization, or death. |
| Assay Type | Sandwich enzyme immunoassay | Sandwich enzyme immunoassay |
| Reportable Range | 13 to 5000 pmol/L | 20 to 5000 pmol/L |
{3}
4
| Item | Predicate - PrognostiX MPO | Dimension® MPO |
| --- | --- | --- |
| Hook Effect | No high dose effect up to 800,000 pmol/L | No high dose effect up to 800,000 pmol/L |
| Clinical study results | Odds ratio increases from 1.0 to a max. of 3.3 across 4 quartiles | Odds ratio increases from 1.0 to a max. of 2.3 across 4 quartiles |
| Expected Values | ≤ 539 pmol/L | <20 - 633 pmol/L |
| Antibody | PrognostiX polyclonal rabbit and goat monoclonal | Dade Behring mouse monoclonal |
| Calibration Interval | Calibration curve using six levels updated for each run. | Calibration curve updated for each lot, using five levels every 30 days with the same reagent lot. |
| Sample Volume | 5 uL | 30 uL |
| Sample | Lithium Heparin Plasma | EDTA, Li or Na heparin plasma |
| Controls | 3 levels; human MPO in lithium heparin plasma | 2 levels; human MPO in BSA |
## K. Standard/Guidance Document Referenced (if applicable):
| STANDARDS |
| --- |
| Title and Reference Number |
| Stability Testing of In Vitro Diagnostic Reagents (13640) |
| Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP09-A2) |
| Interference Testing in Clinical Chemistry; Approved Guideline (EP 7-A) |
| Medical devices – Application of risk management to medical devices (14971:2000) |
| Medical Devices - Symbols to be used with medical device labels, labeling and information to be supplied (15223) |
| Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition (EP5-A2) |
| GUIDANCE | | |
| --- | --- | --- |
| Document Title | Office | Web Page |
| User Fees and Refunds for Premarket Notification Submissions (510(k)s) – Guidance for Industry and FDA Staff | CDRH | http://www.fda.gov/cdrh/mdufma/guidance/1511.html |
| Guidance for Industry and FDA Staff - Use of Symbols on Labels and | CDRH | http://www.fda.gov/cdrh/ocd/guidance/4444.html |
{4}
5
in Labeling of In Vitro Diagnostic Devices Intended for Professional Use
Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff
CDRH
http://www.fda.gov/cdrh/oivd/guidance/1588.html
L. Test Principle:
The Dimension MPO reagent kit is a one step enzyme immunoassay based on the "sandwich" principle. The sample is incubated with chromium dioxide particles coated with mouse monoclonal antibodies specific for MPO, and conjugate reagent (B-galactosidase labeled mouse monoclonal antibodies specific for MPO). A particle/MPO/conjugate forms during the incubation period. Unbound conjugate is removed by magnetic separation and washing. The sandwich bound B-galactosidase is combined with the chromogenic substrate chlorophenol red-B-D-galactopyranoside (CPRG). Hydrolysis of CPRG releases a chromophore (CPR). The concentration of MPO present in the patient sample is directly proportional to the rate of color change due to formation of CPR measured at 577 nm.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
A precision study was conducted based on a modified version of a 20 day AN OVA protocol (EP5-A2) using 3 levels of plasma pools (low, mid and high) and 2 control levels with 7 reagent lots at three sites. Samples 1, 2, 3, 5, 6 and 8 were conducted at the sponsor's internal site. Samples 4, 7, 9 and 10 were conducted at two external sites. Samples 8-10 below were tested using one lot with two plasma pools run in triplicate at three sites over twenty days. The compiled table below shows the results from both studies and a multi-site ANOVA summary.
| Plasma Pool Low | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 | Sample 6 | Sample 7 | Sample 8 | Sample 9 | Sample 10 | Multi-Site ANOVA Summary Stats |
| Mean: | 400.5 | 485.9 | 441.7 | 464.5 | 538.9 | 426.1 | 494.1 | 559.1 | 525.0 | 540.6 | 542.0 |
| SD_{R}: | 8.5 | 11.9 | 11.5 | 11.0 | 21.5 | 13.3 | 10.9 | 16.2 | 9.8 | 14.9 | 13.9 |
| %CV_{R}: | 2.1 | 2.4 | 2.6 | 2.4 | 4.0 | 3.1 | 2.2 | 2.9 | 1.9 | 2.8 | 2.7 |
| SD_{T}: | 13.6 | 14.1 | 14.0 | 11.3 | 22.3 | 14.7 | 17.2 | 24.2 | 21.3 | 22.4 | 27.9 |
| %CV_{T}: | 3.4 | 2.9 | 3.2 | 2.4 | 4.1 | 3.4 | 3.5 | 4.3 | 4.1 | 4.2 | 5.1 |
{5}
| Plasma Pool Mid | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 | Sample 6 | Sample 7 | Sample 8 | Sample 9 | Sample 10 |
| Mean: | 1318 | 1469 | 1463 | 1529 | 1450 | 1307 | 1465 | 803.8 | 775.2 | 782.0 |
| \( SD_R \): | 31.9 | 30.2 | 32.9 | 59.1 | 35.9 | 46 | 43.6 | 17.2 | 19.0 | 16.0 |
| \( \%CV_R \): | 2.4 | 2.1 | 2.2 | 3.9 | 2.5 | 3.5 | 3.0 | 2.1 | 2.5 | 2.1 |
| \( SD_T \): | 48.5 | 44.5 | 49.9 | 59.1 | 38.5 | 50.8 | 53.3 | 25.0 | 28.4 | 28.5 |
| \( \%CV_T \): | 3.7 | 3.0 | 3.4 | 3.9 | 2.7 | 3.9 | 3.6 | 3.1 | 3.7 | 3.7 |
| Plasma Pool High | | | | | |
| --- | --- | --- | --- | --- | --- |
| | Sample 1 | Sample 2 | Sample 3 | Sample 4 | Sample 5 |
| Mean: | 2608 | 2831 | 2728 | 2846 | 2639 |
| \( SD_R \): | 65.7 | 67.4 | 49.4 | 68.2 | 76.2 |
| \( \%CV_R \): | 2.5 | 2.4 | 1.8 | 2.4 | 2.9 |
| \( SD_T \): | 109.4 | 82.4 | 74.9 | 76.1 | 89.6 |
| \( \%CV_T \): | 4.2 | 2.9 | 2.7 | 2.7 | 3.4 |
| Dim MPO QC Lev 1 | | | | | |
| | Sample 6 | Sample 7 | Sample 8 | Sample 9 | Sample 10 |
| Mean: | 416.5 | 428.4 | 465.4 | 443.7 | 454.2 |
| \( SD_R \): | 11.9 | 16.3 | 12.7 | 17.6 | 16.9 |
| \( \%CV_R \): | 2.9 | 3.8 | 2.7 | 4.0 | 3.7 |
| \( SD_T \): | 15.7 | 20.4 | 16.1 | 22.4 | 21.8 |
| \( \%CV_T \): | 3.8 | 4.8 | 3.5 | 5.1 | 4.8 |
| Dim MPO QC Lev 2 | | | | | |
| | Sample 6 | Sample 7 | Sample 8 | Sample 9 | Sample 10 |
| Mean: | 3455 | 3644 | 2584.2 | 2556.1 | 2556.8 |
| \( SD_R \): | 122.3 | 121.0 | 61.3 | 59.9 | 49.3 |
| \( \%CV_R \): | 3.5 | 3.3 | 2.4 | 2.3 | 1.9 |
| \( SD_T \): | 125.1 | 131.6 | 80.0 | 94.6 | 111.0 |
| \( \%CV_T \): | 3.6 | 3.6 | 3.1 | 3.7 | 4.3 |
| Notes: | | | | | |
| 1. \( SD_R \) and \( \%CV_R \) are estimates of repeatability | | | | | |
| 2. \( SD_T \) and \( \%CV_T \) are estimates of within-lab precision | | | | | |
| 3. Different plasma pools were made for different studies. They were targeted for similar analyte concentrations but had different means. As such, precision estimates may be compared across the studies but the mean values would be expected to vary. | | | | | |
{6}
b. Linearity/assay reportable range:
Linearity was assessed in a recovery study according to CLSI EP6-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach. A 7-level dilution series (concentrations shown below) run in replicates of five were prepared by mixing low and high patient samples. A linear regression equation was conducted for expected results versus observed results. The resulting linear regression equation is $y = 1.018x - 1.833$ with a $R^2$ of .999. The results of the recovery study shown below and the functional sensitivity study shown in section d. below support the sponsor claimed range of 20 to 5000 pmol/L.
| Sample # | Expected (pmol/L) | Observed Mean (pmol/L) | % Recovery |
| --- | --- | --- | --- |
| 1 (Low) | 26 | 26 | |
| 2 | 892 | 848 | 95.0 |
| 3 | 1759 | 1783 | 101.4 |
| 4 | 2574 | 2636 | 102.4 |
| 5 | 3440 | 3550 | 103.2 |
| 6 | 4307 | 4432 | 102.9 |
| 7 (High) | 5122 | 5122 | |
The sponsor also conducted a hook effect study to assess samples above the range of the test. Samples were prepared by spiking purified human MPO into a normal human plasma base. The concentrations were verified by making a dilution with human pooled plasma of known low analyte concentration so that the expected value would fall within the MPO assay range. The samples were tested to determine if high samples would give erroneous results. The MPO assay upper range is 5000 pmol/L. Samples whose signals either exceed the upper assay range limit or cause the instruments protected feature are flagged as "Above Assay Range". Based on the results from the study and the additional features of the instrument, the sponsor reports that there was no observed hook effect for samples up to 800,000 pmol/L.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability
No reference standard is available for myeloperoxidase. An anchor pool was prepared from patient reference samples which consisted of native analyte in human serum. A master pool was derived from the anchor pool which is a 5-level liquid working standard of purified analyte in BSA. The initial master pool is assigned from the anchor pool during a patient sample method comparison spanning the assay range. All subsequent master pools are assigned using a set protocol on multiple instruments. The working calibrator materials are assigned values based on the master pool using a set protocol on multiple instruments.
{7}
The two MPO controls have the same formulation and approximate values as the MPO calibrator levels 2 and 4 (500 and 3000 pmol/L) and are made from independent production runs.
## Stability
The sponsor conducted a closed vial calibrator/control stability study in which 3 lots of five levels (approx. 0, 500, 1000, 3000, 5375 pmol/L) of calibrators (stored at -20 C) were tested in replicates at time increments between 0 and 13 months. The study results support the sponsor's calibrator/control stability claim of 13 months.
The sponsor conducted an open vial stability study in which 2 vials of each level were assayed in triplicate for 15 days. The open vial results were compared with an unopened vial for 15 days. The study results support the sponsor's calibrator/control stability claim of 14 days when stored at 2-8 C°.
The sponsor conducted a sample stability study to determine the specimen types and the temperature required for sample storage. 20 EDTA plasma samples (10 collected in house and 10 from a blood bank) and 20 lithium heparin plasma samples (10 collected in house and 10 from a blood bank) were frozen at -20 C. Calibration was conducted daily and the samples were tested in replicates of 5 at day 0, month 1, month 3 and month 6 with the MPO reagent. The results support the sponsor's 6 month sample stability claim when frozen at -20°C.
The sponsor conducted a further sample stability study to determine the number of storage days for a sample if stored at 4°C. Ten paired EDTA and Lithium Heparin samples were stored at 4°C and were run in triplicate with the reagent. The results support the sponsor's 6 day sample stability claim when stored at 4°C.
The sponsor conducted a freeze-thaw stability study to determine the number of times a sample containing myeloperoxidase could be thawed. The paired lithium heparin and EDTA plasma samples from the anticoagulant study in the matrix comparison section 2b below were assayed fresh and then frozen at -70°C. The samples were thawed, assayed in replicates of five and then refrozen. The results support the sponsor's freeze thaw statement within the package insert of "MPO levels in EDTA and heparinized plasma have been shown to be stable through three freeze/thaw cycles."
## d. Detection limit:
The sponsor conducted a limit of the blank study according to CLSI EP17 in which calibrator 1 (0 pmol/L) was assayed 20 consecutive times. The limit of the blank is 13 pmol/L which is the mean value (n=20) plus two standard deviations of the level 1 (0 pmol/L) MPO Calibrator.
8
{8}
The sponsor conducted a functional sensitivity study (one run per testing day for 20 days). Four separate studies (3 in-house sites and 1 external site) were conducted with five human plasma pools with low MPO concentrations. The within-lab precision versus the mean MPO concentration was plotted. The functional sensitivity estimate for each study was read as the MPO concentration giving a 20% CV within-lab precision. The results per site ranged from 7 pmol/L to 16 pmol/L which support the sponsor's functional sensitivity claim of 20% CV at 20 pmol/L.
e. Analytical specificity:
The sponsor conducted an interference study according to CLSI EP7-A2. Bias was defined as the difference in the results between the control sample (without the interferent) and the test sample (contains the interferent) expressed in percent. The sponsor's acceptance criterion was that a bias exceeding 10% was considered as significant interference. Myeloperoxidase samples with concentrations between 481 and 2000 pmol/L were tested for bilirubin (conjugated and non-conjugated), lipemia and hemoglobin. The results showed no significant interference for hemoglobin up to 1000 mg/dL, bilirubin (conjugated and non-conjugated) up to 60 mg/dL and lipemia (Intralipid) up to 3000 mg/dL
Cross reactivity was studied using potentially cross-reacting substances commercially purchased. The substances were spiked into samples containing 0 and approximately 1500-2000 pmol/L myeloperoxidase. Percent cross-reactivity was defined as (Mean Test Result - Mean Control Result)/Cross Reactant Compound Concentration in pmol/L *100. The cross-reactivity was less than 0.15% for all cross reactants except Eosinophil Peroxidase which was 0.43%. The results for the cross-reactivity studies are shown in the tables below
| Test Sample: DM MPO Calibrator Level 1 (0 pmol/L) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | [Test Substance] | | Mean [MPO] | | Bias | X-Reactivity |
| Substance | nmol/L | pmol/L | Control | Test | (pmol/L) | (%) |
| α-1 Antitrypsin | 1250 | 1,250,000 | -4.8 | -5.7 | -0.9 | 0.00 |
| C-reactive protein | 550 | 550,000 | -4.8 | -5.3 | -0.5 | 0.00 |
| Lysozyme | 4500 | 4,500,000 | -4.8 | 1.8 | 6.6 | 0.00 |
| Immunoglobin A | 400 | 400,000 | -5.8 | -5.5 | 0.3 | 0.00 |
| Elastase | 2500 | 2,500,000 | -5.0 | 499.2 | 504.2 | 0.02 |
| Eosinophil peroxidase | 922 | 922,000 | 2.3 | 3960.5 | 3958.2 | 0.43 |
| Lactoperoxidase | 800 | 800,000 | -4.8 | -5.1 | -0.3 | 0.00 |
| Lactoferrin | 800 | 800,000 | -4.8 | -1.7 | 3.1 | 0.00 |
| COX1 | 900 | 900,000 | -4.1 | -2.5 | 1.6 | 0.00 |
| COX2 | 900 | 900,000 | -2.7 | -2.5 | 0.2 | 0.00 |
| Thyroid peroxidase | 600 | 600,000 | -3.4 | 284.7 | 288.1 | 0.05 |
| Troponin I | 2150 | 2,150,000 | -3.1 | 6.2 | 9.3 | 0.00 |
{9}
10
| Test Sample: Sodium Heparin Human Plasma Pool (~1800 pmol/L) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | [Test Substance] | | Mean [MPO] | | Bias | X-Reactivity |
| Substance | (nmol/L) | (pmol/L) | Control | Test | (pmol/L) | (%) |
| α-1 Antitrypsin | 1250 | 1,250,000 | 1815.2 | 1831.1 | 15.9 | 0.00 |
| C-reactive protein | 550 | 550,000 | 1815.2 | 1834.0 | 18.8 | 0.00 |
| Lysozyme | 4500 | 4,500,000 | 1815.2 | 1869.9 | 54.7 | 0.00 |
| Immunoglobin A | 400 | 400,000 | 1840.3 | 1763.8 | -76.5 | -0.02 |
| Elastase | 2500 | 2,500,000 | 1837.0 | 2113.5 | 276.5 | 0.01 |
| Eosinophil peroxidase | 922 | 922,000 | 1809.6 | 3436.5 | 1626.9 | 0.18 |
| Lactoperoxidase | 800 | 800,000 | 1815.2 | 1694.6 | -120.6 | -0.02 |
| Lactoferrin | 800 | 800,000 | 1815.2 | 1790.7 | -24.5 | 0.00 |
| COX1 | 900 | 900,000 | 1868.7 | 1762.7 | -106.0 | -0.01 |
| COX2 | 900 | 900,000 | 1541.4 | 1457.9 | -83.5 | -0.01 |
| Thyroid peroxidase | 600 | 600,000 | 1884.5 | 2010.0 | 125.5 | 0.02 |
| Troponin I | 2150 | 2,150,000 | 1859.1 | 1899.7 | 40.6 | 0.00 |
An extensive list of other compounds was evaluated for interference and was found to have no significant interference. The list can be found in the package insert.
f. Assay cut-off:
Not applicable.
2. Comparison studies:
a. Method comparison with predicate device:
The sponsor conducted a method comparison study according to CLSI EP9-A2. One hundred and thirty-nine lithium heparin samples ranging from 154 to 4869 pmol/L were analyzed using the Dimension MPO on the Dimension RxL and the Prognostix MPO assay. The Passing-Bablok linear regression results are shown in the table below.
| Slope | Intercept (pmol/L) | Correlation Coefficient | N |
| --- | --- | --- | --- |
| 1.03
(95%CI: 0.97 to 1.12) | 91.8
(95% CI: -10.0 to 180.6) | 0.88 | 139 |
b. Matrix comparison:
The sponsor conducted a matrix comparison study to evaluate the effects of EDTA
{10}
and Sodium Heparin samples with the Dimension MPO assay. The Passing-Bablok linear regression (x=Li heparin values) results are shown in the table below.
| Comparative Specimen | Slope | Intercept (pmol/L) | Correlation Coefficient | Range (pmol/L) | N |
| --- | --- | --- | --- | --- | --- |
| Lithium Heparin vs. EDTA | 1.01 | -19.3 | 0.997 | 165-3812 | 59 |
| Lithium Heparin vs. Sodium Heparin | 0.96 | -7.2 | 0.999 | 167-3401 | 49 |
The frequencies for the $\%$ differences ranged from $< -20\%$ to $>20\%$ as shown in the plots below.


# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
A clinical study was conducted which included 400 EDTA samples collected from patients presenting to the Emergency Department or acutely to outpatient facilities with chest pain or equivalent symptoms suggestive of Acute Coronary Syndrome (ACS). Approximately $85\%$ of the patient population presented within 24 hours of the onset of symptoms. On electrocardiogram (ECG), $15.5\%$ of patients had ST
{11}
segment elevation; $6.75\%$ of patients had ST segment depression; $33\%$ had one of the following: new Q-waves, old Q-waves, or non-specific T wave abnormality; and $44.75\%$ had none of the aforementioned changes on ECG. Patients were assessed for major adverse cardiac events (MACE) which includes myocardial infarction, recurrent ischemia associated with ECG changes, recurrent ischemia requiring hospitalization or revascularization, and death. Incidence of MACE was assessed by a review of medical records and/or medical database searches. Patients were excluded: if the chest pain was due to trauma or if they were apparently healthy individuals. The demographics of the patients are shown in the table below.
MPO Quartiles and Selected Demographics by Quartile
| | | Quartile (Range of MPO in pM) | | | |
| --- | --- | --- | --- | --- | --- |
| Variable | N* | Q1 (94-581) | Q2 (582-894) | Q3 (895-1657) | Q4 (1658-5000) |
| Mean Age (years) | 400 | 54 | 54 | 57 | 59 |
| WBC (x 10^9/L) | 386 | 6.76 | 8.09 | 9.05 | 12.09 |
| Male Sex | 400 | 59% | 61% | 55% | 55% |
| Caucasian | 400 | 42% | 41% | 50% | 52% |
| HX Diabetes | 400 | 28% | 33% | 27% | 41% |
| HX Hypertension | 400 | 65% | 73% | 68% | 69% |
| HX Hyperlipidemia | 400 | 37% | 36% | 36% | 39% |
| Current Smoker | 400 | 38% | 39% | 39% | 32% |
| Prior MI | 400 | 21% | 28% | 23% | 19% |
| Prior PCI | 400 | 14% | 14% | 13% | 12% |
| Prior CABG | 400 | 5% | 7% | 8% | 12% |
| HX CHF | 400 | 6% | 11% | 14% | 21% |
*Number of patients from whom data was available; Q1=99 patients, Q3 =101 patients, Q2 & Q4=100 patients
Odds ratios were determined using univariate and multivariate logistic regression models. Each MPO quartile was compared versus MPO quartile 1 for the outcomes of MACE at the 30 day and 180 day intervals post-presentation. The multivariate analysis included the following variables in addition to MPO: age, sex, race, history of hypertension, history of hyperlipidemia, smoking status, history of diabetes, white blood cell count (WBC), and the following cardiac markers: Dimension® troponin I (CTNI), NTproBNP (PBNP) and hsCRP (CCRP). Results of the univariate and multivariate analysis are displayed in the following two tables below.
Univariate Analysis (N=400)
| Analysis | Odds ratio (Confidence Interval) | p-value |
| --- | --- | --- |
| 30 days | | |
| MPO Q2 vs. Q1 | 1.36 (0.55-3.40) | 0.827 |
| MPO Q3 vs. Q1 | 2.63 (1.14-6.06) | 0.024 |
| MPO Q4 vs. Q1 | 4.29 (1.91-9.61) | 0.000 |
| | | |
{12}
| 180 days | | |
| --- | --- | --- |
| MPO Q2 vs. Q1 | 0.99 (0.41-2.40) | 0.799 |
| MPO Q3 vs. Q1 | 2.10 (0.95-4.63) | 0.066 |
| MPO Q4 vs. Q1 | 4.12 (1.95-8.73) | 0.000 |
Multivariate Analysis (N=386; 14 patients excluded due to missing WBC)
| Analysis | Odds ratio (Confidence Interval) | p-value |
| --- | --- | --- |
| 30 days | | |
| MPO Q2 vs. Q1 | 1.41 (0.51-3.89) | 0.510 |
| MPO Q3 vs. Q1 | 3.03 (1.19-7.76) | 0.021 |
| MPO Q4 vs. Q1 | 4.31 (1.62-11.50) | 0.004 |
| Age | 1.01 (0.98-1.03) | 0.684 |
| Sex | 2.69 (1.40-5.16) | 0.003 |
| History of hypertension | 1.07 (0.53-2.19) | 0.843 |
| History of hyperlipidemia | 3.40 (1.80-6.44) | 0.000 |
| Smoking status | 0.92 (0.49-1.74) | 0.796 |
| History of diabetes | 0.99 (0.52-1.87) | 0.970 |
| White blood cell count | 1.06 (1.00-1.13) | 0.044 |
| CTNI | 1.37 (1.09-1.74) | 0.008 |
| PBNP | 1.00 (1.00-1.00) | 0.606 |
| CCRP | 0.99 (0.98-1.00) | 0.072 |
| 180 days | | |
| MPO Q2 vs. Q1 | 0.98 (0.37-2.56) | 0.960 |
| MPO Q3 vs. Q1 | 2.21 (0.93-5.26) | 0.073 |
| MPO Q4 vs. Q1 | 3.66 (1.50-8.94) | 0.004 |
| Age | 1.01 (0.99-1.03) | 0.421 |
| Sex | 2.19 (1.19-4.03) | 0.011 |
| History of hypertension | 0.98 (0.50-1.91) | 0.943 |
| History of hyperlipidemia | 2.24 (1.23-4.09) | 0.009 |
| Smoking status | 0.79 (0.43-1.46) | 0.455 |
| History of diabetes | 0.70 (0.39-1.28) | 0.251 |
| White blood cell count | 1.04 (0.99-1.10) | 0.117 |
| CTNI | 1.32 (1.05-1.66) | 0.016 |
| PBNP | 1.00 (1.00-1.00) | 0.928 |
| CCRP | 1.00 (0.99-1.00) | 0.321 |
Logistic-regression models were used to calculate odds ratios and 95th percentile confidence intervals. Odds ratios (shown below) were calculated for MPO separately and after adjustment for age, gender, Troponin I, NT-proBNP, C-reactive protein, white blood cell count, ST-segment depression, history of hypertension, history of hypercholesterolemia, history of diabetes, and smoking status. In each analysis the first quartile served as the reference group for calculation of odds ratio.
{13}
14
MACE at 30 Days
| | Q1* | Q2 | Q3 | Q4 |
| --- | --- | --- | --- | --- |
| MPO (pmol/L) | 94-581 | 582-894 | 895-1657 | 1658-5000 |
| Odds Ratio | 1.00 | 1.36 | 2.63 | 4.29 |
| 95% CI | NA | 0.55-3.40 | 1.14-6.06 | 1.91-9.61 |
| Adjusted Odds Ratio | 1.00 | 1.41 | 3.03 | 4.31 |
| 95% CI | NA | 0.51-3.89 | 1.19-7.76 | 1.62-11.50 |
MACE at 6 Months
| | Q1* | Q2 | Q3 | Q4 |
| --- | --- | --- | --- | --- |
| MPO (pmol/L) | 94-581 | 582-894 | 895-1657 | 1658-5000 |
| Odds Ratio | 1.00 | 0.99 | 2.10 | 4.12 |
| 95% CI | NA | 0.41-2.40 | 0.95-4.63 | 1.95-8.73 |
| Adjusted Odds Ratio | 1.00 | 0.98 | 2.21 | 3.66 |
| 95% CI | NA | 0.37-2.56 | 0.93-5.26 | 1.50-8.94 |
ROC analysis was conducted to determine if MPO is a significant predictor of the ROC curve (or area under the curve (AUC) after adjustment for covariates such as race, gender, hyperlipidemia, CRP and troponin (chart shown below). MPO did not have a statistically significant added effect over Troponin I, which could be due to competing effect of MPO and Troponin I or simply a difference in their biological variation.
| ROC Curve Results for Predictive Probability of Covariates With and Without Addition of MPO | | | |
| --- | --- | --- | --- |
| Model† | 0 days | 30 days | 180 days |
| | | | |
| Race | 0.62 (AUC) | 0.61 (AUC) | 0.62 (AUC) |
| Race & MPO | 0.72 (AUC) | 0.71 (AUC) | 0.71 (AUC) |
| Difference (95% CI) | -0.10 (-0.15 to -0.04) | -0.10 (-0.15 to -0.04) | -0.09 (-0.14 to -0.04) |
| p-value | 0.0010 | 0.0011 | 0.0004 |
| | | | |
| Sex | 0.59 (AUC) | 0.60 (AUC) | 0.59 (AUC) |
| Sex & MPO | 0.73 (AUC) | 0.74 (AUC) | 0.73 (AUC) |
| Difference (95% CI) | -0.14 (-0.21 to -0.07) | -0.14 (-0.20 to -0.08) | -0.14 (-0.20 to -0.08) |
| p-value | <0.0001 | <0.0001 | <0.0001 |
{14}
| hsCRP | 0.51 (AUC) | 0.51 (AUC) | 0.54 (AUC) |
| --- | --- | --- | --- |
| hsCRP & MPO | 0.69 (AUC) | 0.68 (AUC) | 0.68 (AUC) |
| Difference (95% CI) | -0.17 (-0.26 to -0.09) | -0.17 (-0.26 to -0.08) | -0.14 (-0.22 to -0.06) |
| p-value | 0.0001 | 0.0002 | 0.0003 |
| Hyperlipidemia | 0.63 (AUC) | 0.64 (AUC) | 0.62 (AUC) |
| Hyperlipidemia & MPO | 0.73 (AUC) | 0.73 (AUC) | 0.71 (AUC) |
| Difference (95% CI) | -0.10 (-0.15 to -0.04) | -0.09 (-0.14 to -0.04) | -0.09 (-0.14 to -0.04) |
| p-value | 0.0004 | 0.0003 | 0.0003 |
†Values for covariate and covariate & MPO correspond to area under the ROC curve (AUC)
Further analysis of the clinical data was made by stratifying patients using the upper limit of the MPO reference interval (633 pmol/L) and Dimension Troponin I at the 99th percentile (0.07 ng/mL). The analysis shows that patients at risk for MACE within 30 and 180 days were identified significantly more often for MPO>633 pmol/L than for MPO ≤633 pmol/L when Troponin ≤0.07 ng/mL.
Improved Stratification for MACE When Dimension Troponin I is 0.07 ng/mL.*

* p = 0.008 vs. MPO ≤ 633 pmol/L. One sided p values were calculated using Fisher's Exact test
+ Data shown represents incidence of MACE at 30 days; incidence of MACE at 180 days was similar
The results showed that MPO is a significant predictor of MACE at 30 days and at 6 months and supports a substantial equivalence decision.
4. Clinical cut-off:
Not applicable.
{15}
16
5. Expected values/Reference range:
The sponsor conducted a reference range study using EDTA samples per CLSI protocol C28-A2 How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline—Second Edition. The sponsor selected the lower, one-sided 95% region of results obtained with the assay in a healthy population as the Dimension® MPO reference interval.
The sponsor includes the following in the labeling:
“In a study of 286 EDTA plasma samples from apparently healthy individuals, the reference interval encompassing the lower 95% region for the Dimension® MPO method was as follows:
| N | Range encompassing the lower 95% region |
| --- | --- |
| 286 | 51–633 pmol/L |
Each laboratory should establish its own expected values for MPO as performed on the Dimension® system.”
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.