The Medline Latex Foley Catheter is intended to be used as a urological catheter inserted through the urethra for the purpose of draining urine and other fluids from the urinary tract.
Device Story
Flexible tubular device; natural high-grade latex; silicone elastomer coating. Inserted through urethra into bladder; secured via balloon inflated with sterile water. Available in 2-way (drainage) and 3-way (drainage plus irrigation) configurations; various sizes (6-30 Fr) and balloon capacities (3cc, 5cc, 30cc). Used in clinical settings by healthcare professionals for patients with disrupted urination or requiring urine output monitoring. Urine drains from proximal eyelets through shaft to external collection device. Benefits include effective bladder drainage and irrigation.
Clinical Evidence
No clinical data. Substantial equivalence is based on identical design, materials, and performance specifications to the predicate device. Performance testing was conducted in accordance with relevant standards for the predicate.
Technological Characteristics
Natural high-grade latex construction; silicone elastomer coating applied via dipping. 2-way and 3-way configurations; 6-30 Fr sizes; 3cc, 5cc, 30cc balloon capacities. Sterile, single-use. Gamma irradiation sterilization. No software or electronic components.
Indications for Use
Indicated for male, female, and pediatric patients requiring drainage of urine and other fluids from the urinary tract. Contraindicated in individuals with known sensitivity or allergy to latex.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
K071423
### EXECUTIVE SUMMARY - Medline Latex Foley Catheter
#### Device Description
JUL 1 8 2007
A Foley catheter is needed when normal urination is disrupted or it is necessary to keep track of urine production. The catheter is threaded through the urinary duct (urethra) and into the bladder. It is secured by means of a balloon at the tip which is inflated with sterile water after insertion. The Medline Latex Foley Catheter offers 3 different balloon sizes (3cc, 5cc, 30cc) and a wide range of catheter sizes to accommodate pediatric and adult applications (refer to table below). The relative size of the Foley catheter is described using French units (Fr). 1 Fr is equivalent to 0.33 mm or .013".
| Catheter type | Balloon capacity | Size (Fr) | Item Number |
|-----------------|------------------|-----------|-------------|
| 2-way standard | 3cc | 6 | DYND11706 |
| 2-way standard | 3cc | 8 | DYND11708 |
| 2-way standard | 3cc | 10 | DYND11710 |
| 2-way standard | 5cc | 12 | DYND11752 |
| 2-way standard | 5cc | 14 | DYND11754 |
| 2-way standard | 5cc | 16 | DYND11756 |
| 2-way standard | 5cc | 18 | DYND11758 |
| 2-way standard | 5cc | 20 | DYND11760 |
| 2-way standard | 5cc | 22 | DYND11762 |
| 2-way standard | 5cc | 24 | DYND11764 |
| 2-way standard | 5cc | 26 | DYND11766 |
| 2-way standard | 5cc | 28 | DYND11768 |
| 2-way standard | 5cc | 30 | DYND11770 |
| 2-way coude tip | 5cc | 14 | DYND11214 |
| 2-way coude tip | 5cc | 16 | DYND11216 |
| 2-way coude tip | 5cc | 18 | DYND11218 |
| 2-way coude tip | 5cc | 20 | DYND11220 |
| 2-way coude tip | 5cc | 22 | DYND11222 |
| 2-way coude tip | 5cc | 24 | DYND11224 |
| 2-way standard | 30cc | 12 | DYND11772 |
| 2-way standard | 30cc | 14 | DYND11774 |
| 2-way standard | 30cc | 16 | DYND11776 |
| 2-way standard | 30cc | 18 | DYND11778 |
| 2-way standard | 30cc | 20 | DYND11780 |
| 2-way standard | 30cc | 22 | DYND11782 |
| 2-way standard | 30cc | 24 | DYND11784 |
| 2-way standard | 30cc | 26 | DYND11786 |
| 2-way standard | 30cc | 28 | DYND11788 |
| 2-way standard | 30cc | 30 | DYND11790 |
| 3-way standard | 30cc | 16 | DYND11800 |
| 3-way standard | 30cc | 18 | DYND11801 |
| 3-way standard | 30cc | 20 | DYND11802 |
| 3-way standard | 30cc | 22 | DYND11803 |
| 3-way standard | 30cc | 24 | DYND11804 |
| 3-way standard | 30cc | 26 | DYND11805 |
{1}------------------------------------------------
The Medline Latex Foley Catheter is a flexible tubular device manufactured from natural high grade latex. For a listing of all the raw materials used, refer to attachment 2. It is sterile as packaged and intended for single patient use. The Medline Latex Foley Catheter consists of two types (2-way & 3-way) of devices. These catheters constitute a double or triple lumen drainage tube with two opposing eyelets on the proximal tip. On the other end of the catheter shaft, the urine drains out into an appropriate collection device. The 3-way catheter has an additional set of eyelets for irrigation purposes.
A silicon elastomer coating is applied to the catheter through a dipping process.
The catheters are supplied in individually packaged, sealed single use poaches. Packaging Materials: Paper/Poly/Mylar film Sealing Method: Heat sealed
#### Substantial Equivalence
| Feature | Proposed Device | Predicate Device |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| Name | Medline Latex Foley Catheter | Foleycath NR Latex Balloon<br>Catheter (K903777) |
| Intended Use | The Medline Latex Foley<br>Catheter is intended to be used as<br>a urological catheter inserted<br>through the urethra for the<br>purpose of draining urine and<br>other fluids from the urinary tract. | The Foleycath is intended to be<br>used for the drainage of urine and<br>other fluids from the urinary<br>bladder. |
| Patient Populations | Male, Female & Pediatric | Male, Female & Pediatric |
| Types | • 2-way<br>• 3-way | • 2-way<br>• 3-way |
| Size Range (Fr) | • Pediatric (6-10)<br>• Female / Male (12-30) | • Pediatric (6-10)<br>• Female / Male (12-30) |
| Balloon Capacity | 3cc, 5cc, 30cc | 3cc, 5cc, 30cc |
| Material | Natural rubber latex | Natural rubber latex |
| Coating | Silicone | Silicone |
| Labeling | See attachment I | See attachment I |
#### Discussion of similarities and differences:
The proposed Medline Latex Foley Catheter is identical in intended uses, function and mode of operation to the predicate device which received marketing approval under K903777.
There are no significant differences between the two products. As reflected in the table above, they include identical product lines which support the same patient populations. They are fabricated from the same materials and designed in a same fashion.
{2}------------------------------------------------
## Biocompatibility / Sterilization
Per the Guidance for Industry and FDA Staff, Format for Traditional and Abbreviated 510(k)s, August 12,2005, section 15 – "If identical materials are used in a predicate with the same type and duration of patient contact, you may identify the predicate in lieu of performing biocompatibility testing and state that your device is comprised of identical materials and that are processed by identical manufacturing methods."
As noted, this device is identical to that of WRP Foleycath (K903777). The manufacturing and sterilization methods remain unchanged. The history of the biocompatibility of these products is therefore well established.
The product is sterilized by a Gamma Irradiation method. Refer to attachment 2 for details of this process.
## Summary of Performance Testing
The Medline Latex Foley Catheter referenced in this submission is held to the same design, manufacture, and performance specifications as the currently marketed predicate device (K903777). The predicate device was tested in accordance with relevant standards for performance and function.
## Product Labeling
The labeling, dimensional drawings & promotional materials for this device can be found in Attachment 1.
### Conclusions
In accordance with 21 CFR Part 807, and based on the information provided in this premarket notification, Medline Industries, Inc. concludes that the Medline Latex Foley Catheter is safe, effective and substantially equivalent as described herein.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
JUL 1 8 2007
Mr. Matt Clausen Regulatory Affairs Medline Industries. Inc. One Medline Place MUNDELEIN IL 60060-4486
Re: K071423
Trade/Device Name: Medline Latex Foley Catheter Regulation Number: 21 CFR §876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: May 18, 2007 Received: May 25, 2007
Dear Mr. Clausen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image shows a circular seal with the text "1906-2006" at the top. The letters "FDA" are prominently displayed in the center of the seal. Below the letters, the word "Centennial" is written in a curved font. Three stars are arranged below the word "Centennial", and the entire seal is surrounded by a ring of small stars.
Protecting and Promoting Public Health
{4}------------------------------------------------
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
> 21 CFR 876.xxxx 21 CFR 884.xxxx 21 CFR 892 xxxxx Other
(Gastroenterology/Renal/Urology) (Obstetrics/Gynecology) (Radiology);
240-276-0115 240-276-0115 240-276-0120 240-276-0100
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours, .
Nancy C. Brogdon
Nancy C. Brogdon Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K07/422
Device Name: Medline Latex Foley Catheter
Indications for Use:
The Medline Latex Foley Catheter is intended to be used as a urological catheter inserted through the urethra for the purpose of draining urine and other fluids from the urinary tract.
Precautions:
Do not use petroleum based ointments or lubricants.
Contraindications:
Those individuals with a known sensitivity or allergy to latex are excluded from the use of this device. Product use should be discontinued should signs of sensitization occur.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
currence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number. K071423
(3)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.