K071418 · Sooil Development Co., Ltd. · MRZ · Sep 4, 2007 · General Hospital
Device Facts
Record ID
K071418
Device Name
DANA 3CC SYRINGE INSULIN RESERVOIR
Applicant
Sooil Development Co., Ltd.
Product Code
MRZ · General Hospital
Decision Date
Sep 4, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DANA 3cc Syringe Insulin Reservoir is an accessory to the DANA Diabecare® Insulin Pumps. The DANA 3cc Syringe Insulin Reservoir is intended for the delivery of insulin from the DANA Diabecare® Insulin Pumps using specified insulin administration sets. The DANA 3cc Syringe Insulin Reservoir is intended exclusively for use with the DANA Diabecare® Insulin Pumps. The DANA Diabecare® Insulin Pumps external programmable syringe infusion pumps used for the are subcutaneous delivery of insulin for the treatment of diabetes mellitus. The pumps are not intended for use with blood or blood products.
Device Story
DANA 3cc Syringe Insulin Reservoir is a sterile, single-use, 3 ml polypropylene syringe; 300-unit insulin capacity; includes 22 gauge, 12.7 mm needle. Designed as accessory for DANA Diabecare® insulin pumps; features silicone ring-shaped gasket. User fills reservoir like standard insulin syringe; removes air; inserts into pump; attaches to pump linking screw. Used in clinical or home settings by patients or caregivers. Provides insulin storage and delivery interface for pump; ensures compatibility with specific DANA pump models. Benefits patient by enabling precise, programmable subcutaneous insulin administration.
Clinical Evidence
Bench testing only. Design verification activities included evaluation of physical and functional characteristics and shelf-life testing to confirm compatibility with DANA Diabecare® insulin pumps.
Technological Characteristics
3 ml piston syringe; 300-unit insulin capacity; polypropylene body; silicone ring-shaped gasket; 22 gauge, 12.7 mm needle. Sterile, single-use, non-pyrogenic. Mechanical interface with DANA Diabecare® pump via linking screw.
Indications for Use
Indicated for patients with diabetes mellitus requiring subcutaneous insulin delivery via DANA Diabecare® external programmable syringe infusion pumps. Not for use with blood or blood products.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
## 510(k) Summary for DANA 3cc Syringe Insulin Reservoir
### 1. SPONSOR
- 4 2007
Sooil Development Co., Ltd. 196-1, Dogok-dong, Kangnam-gu Seoul, 135-270, KOREA
Soo Bong Choi, CEO Contact: (82) 2-3463-0041 Telephone: (82) 2-3463-7077 Facsimile:
Date Prepared: July 9, 2007
#### 2. DEVICE NAME
| Proprietary Name: | DANA 3cc Syringe Insulin Reservoir |
|----------------------|------------------------------------|
| Common/Usual Name: | Insulin reservoir |
| Classification Name: | Accessories, infusion pump |
#### 3. PREDICATE DEVICES
- SOOIL Development Co., Ltd., DANA Diabecare® (K001604)*
- Spectrx, Inc., SimpleChoice 3.0 ml Reservoir Pro (K051045)
- Applied Diabetes Research, Inc., SureSet 3.0 ml Reservoir (K024056)
- Sterling Medivations, Inc., Simplicity Infusion Reservoir (K013767)
- Medtronic MiniMed, Inc., 3.0 ml Reservoir Model 103/193 (K991936)
- *Insulin reservoir accessory included in K001604
## DEVICE DESCRIPTION
The DANA 3cc Syringe Insulin Reservoir is a sterile, single-use, non-pyrogenic, syringe that is designed for use with all versions of the DANA Diabcare® Insulin pump that have been cleared for marketing in the U.S. (K001604, K022317, and K063126). The syringe reservoir is a 3 ml polypropylene syringe with a 300 unit insulin capacity that is supplied with a 22 gauge, 12.7 mm straight needle. The
SOOIL Development Co., Ltd. Additional Information - K071418
August 20, 2007
510(k) Summary, Page 1 of 3
{1}------------------------------------------------
reservoir is filled similarly to filling an insulin syringe. Once filled and inspected for reservoir is filled similariy to fining an insume syringer esservor is inserted into the distal entrapped air, the needle is Tentoved and the syringe has groves cut into the distal.
DANA Diabecare® Insulin pump. The syringe thas grooves cut into the DANA Diabecare® Insulin punip. The syrings pringer has g
end for attachment to the linking screw that connects the syringe reservoir to the end Tor attachmont - A Diabecare® Insulin Pump.
The proposed DANA 3cc Syringe Insulin Reservoir is a modification of the insulin
t the insults the first to 1001604 the wriving 510(k) premarket notification The proposed DANA 3CC Syringe Insumi Reserver and 510(k) premarket notification
reservoir accessory described in K001604, the woldfied to a change reservoir accessory described in Rocroor, the modification was limited to a change
for the DANA Diabecare® Insulin Pump. The moresal DANA 3cc Syringe Insulin for the DANA Diabecare@ Insumil Tamp. The monneed DANA 3cc Syringe Insulin
in the gasket configuration and material. The proposed DANA 3cc Syringe Insulin
Systems on igurati in the gasket contriguration and material.
Reservoir has a ring shape gasket that is manufactured from silicone.
## 5. INTENDED USE
The DANA 3cc Syringe Insulin Reservoir is an accessory to the DANA Dibecare@
in the DANA 3cc Syringe In Decemeir is interveir is intended for the delivery The DANA 3cc Syringe Insumit Reservor is intended for the delivery
Insulin Pumps. The DANA 3cc Syringe Insulin Reservor is unsuling specified, insulin Insulin Pumps. The DANA See Syrings Insulin Pumps using specified insulin administration sets.
The DANA 3cc Syringe Insulin Reservoir is intended exclusively for use with the The DANA 3cc Syringe Insulin Reserver 10 mm Diabecare® Insulin Pumps are DANA Diabecare® Insular Fumps. "The Subcutaneous delivery of
external programmable syringe infusion pumps are not intended for use external programmable syringe midslow painfor and intended for use with blood or blood products.
# 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The material components and technological characteristics of the proposed DANA The material colliponents and the predicate devices are similar. The proposed devices are 3.0 ml piston syringes with a 300 unit insulin capacity. The proposed devices are 3.0 In piston syringes what a comments are supplied sterile and are indicated for single use.
The proposed DANA 3cc Syringe Insulin Reservoir and the predicate devices are indicated for use only with a specified insulin pump. As stated elsewhere, the indicated for use only with a spection on the predicate insulin reservoir reservoir proposed DANA SCC Syringe modial research for use with the DANA Diabecare®
described in K001604 are designed specifically for use with the Bearner The other predicate reservoirs (SimpleChoice 3.0 ml Reservoir Insulin Pump. (K051045), Sureset 3.0 ml Reservoir (K024056), Simplicity Infusion Reservoir (K031043), Surcher 5.0 ml Reservoir (K991936)) are designed exclusively for use with the Medtronic MiniMed® pumps.
SOOIL Development Co., Ltd. Additional Information - K071418 August 20, 2007
510(k) Summary, Page 2 of 3
{2}------------------------------------------------
| Item for Comparison | DANA 3cc Syringe<br>Insulin Reservoir<br>(Proposed) | DANA<br>Insulin Reservoir<br>(K001604) | Simplicity Infusion<br>Reservoir<br>(K013767) | MiniMed 3.0 ml<br>Reservoir<br>(K991936) |
|-----------------------------|------------------------------------------------------------------------------|----------------------------------------|-----------------------------------------------|------------------------------------------|
| Intended Use | Infusion reservoirs for use to infuse insulin from an external infusion pump | | | |
| Insulin Pump | DANA Diabecare®<br>Insulin Pumps | DANA Diabecare®<br>Insulin Pumps | Medtronic<br>MiniMed® Pumps | Medtronic<br>MiniMed® pumps |
| Technical<br>Specifications | • 3 ml syringe<br>• 300 unit capacity | • 3 ml syringe<br>• 300 unit capacity | • 3 ml syringe<br>• 300 unit capacity | • 3 ml syringe<br>• 300-unit capacity |
| Sterility Status | • Supplied sterile<br>• Single use | • Supplied sterile<br>• Single use | • Supplied sterile<br>• Single use | • Supplied sterile<br>• Single use |
# Comparison Chart for Determination of Substantial Equivalence
## 7. PERFORMANCE TESTING
Design verification activities for the DANA 3cc Syringe Insulin Reservoir included Design vermeanon avon evaluation of physical and functional characteristics, and shelf-life testing. The results confirm that the DANA 3cc Syringe Insulin Reservoir showing to may insulin reservoir for all legally marketed versions of the DANA Diabecare® Insulin Pumps and support the claimed shelf life.
Deleted: a three-yea
SOOIL Development Co., Ltd. Additional Information - K071418
August 20, 2007
510(k) Summary, Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Sooil Development Company, Limited C/O Cynthia J.M. Nolte, Ph.D., RAC Senior Regulatory Consultant Medical Device Consultants, Incorporated 49 Plain Street North Attleboro, Massachusetts 02760
SEP - 4 2007
Re: K071418
Trade/Device Name: DANA 3cc Syringe Insulin Reservoir Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MRZ Dated: July 9, 2007 Received: July 10, 2007
Dear Dr. Nolte:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Dr. Nolte
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K071418
Device Name: DANA 3cc Syringe Insulin Reservoir
Indications for Use:
The DANA 3cc Syringe Insulin Reservoir is an accessory to the DANA Diabecare® Insulin Pumps. The DANA 3cc Syringe Insulin Reservoir is intended for the delivery of insulin from the DANA Diabecare® Insulin Pumps using specified insulin administration sets.
The DANA 3cc Syringe Insulin Reservoir is intended exclusively for use with the DANA Diabecare® Insulin Pumps. The DANA Diabecare® Insulin Pumps external programmable syringe infusion pumps used for the are subcutaneous delivery of insulin for the treatment of diabetes mellitus. The pumps are not intended for use with blood or blood products.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C) хаанаа
מחלק מערב משמעות המועד המונח של המועד המועד המועד המועד המועד המועד המועד המועד המועד המועד המונח של המועד המועד המועד המועד המוזיקה המועד המוזיקה המועד המוזיקה המועד המוזיק
מאמיני
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
id. OQ. for Chulin
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: Lo 7 148
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.