# CERADRILL (K071352)

_Gebr. Brasseler GmbH & Co. KG · EJL · Jun 29, 2007 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K071352

## Device Facts

- **Applicant:** Gebr. Brasseler GmbH & Co. KG
- **Product Code:** [EJL](/productcode/EJL.md)
- **Decision Date:** Jun 29, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3240
- **Device Class:** Class 1
- **Review Panel:** Dental
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The drill is used to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. The drills are used with a handpiece. CeraDrill is a ceramic drill for the dental implantology made of a modern high performance ceramic. The drills are intended to be used with a handpiece. The intended use for this device is to cut into the maxilla or mandible to create an osteotomy for endosseous dental implant placement.

## Device Story

CeraDrill is a ceramic pilot drill used in dental implantology to create an initial osteotomy in the maxilla or mandible. The device is operated by a dentist using an electrical powered handpiece. It features a high-performance ceramic (ZrO2/Al2O3/Y2O3) construction with specific twist drill blade geometry and lasered depth markings for precise penetration control. Unlike metal drills, it is metal-free, biocompatible, and corrosion-resistant. The device provides high cutting efficiency and improved thermal safety compared to traditional steel burs. It is designed for repeated use, capable of being cleaned, disinfected, and sterilized for at least 10 cycles without significant loss of mechanical stability. The output is a prepared bone site ready for subsequent implant placement, facilitating positive osseointegration.

## Clinical Evidence

No clinical data. Substantial equivalence is based on non-clinical performance testing, including physical and biological characterization, thermal safety analysis, and mechanical stability testing after repeated reprocessing cycles.

## Technological Characteristics

Material: ZrO2/Al2O3/Y2O3 ceramic. Design: Twist drill with 2 or 3 cutting flutes; external irrigation. Dimensions: Various diameters and lengths. Energy: Powered by electrical handpiece. Biocompatibility: ISO 7405 compliant. Reprocessing: Sterilizable, validated for at least 10 cycles. Features: Lasered depth markings at 2 mm intervals.

## Regulatory Identification

A dental bur is a rotary cutting device made from carbon steel or tungsten carbide intended to cut hard structures in the mouth, such as teeth or bone. It is also intended to cut hard metals, plastics, porcelains, and similar materials intended for use in the fabrication of dental devices.

## Predicate Devices

- Amorphous Diamond Coated Drill ([K990846](/device/K990846.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Premarket notification 510(k) submission Gebr. Brasseler Gmbh & Co KG CeraDrill K:

JUN 2 9 100

# 10(k) Summarv

Gebr. Brasseler GmbH & Co. KG owner's name: address: Trophagener Weg 25 32631 Lemgo Germany phone: + 49 5261 701 467 fax numbers: + 49 5261 701 289 Mr. Olaf Brand name of contact person: date the summary was prepared: February 2007 Establishment Registration number: 8010468 name of the device: CeraDrill trade or proprietary name: CeraDrill the classification name: Bone cutting instrument and accessories (21 CFR 872.4120 Product Code DZI)

> Date: 2007-02-01 Revision: 3 Chapter 05 1/8 Page:

{1}------------------------------------------------

Nobel Biocare USA, Inc.

Device:

Amorphous Diamond Coated Drill

510(k) No.:

K990846

#### Description of the device:

CeraDrills within the new "CeraLine" is intended for use in implantology. These are made of high efficiency ceramics featuring an excellent cutting performance. The new CeraDrill offers the possibility of working without any metal, thus guaranteeing a biocompatible operation. Damages, as they occur in particular when disinfecting and cleaning steel burs with inappropriate agents, are now a thing of the past. The multifunctional ceramic drills are suited for initial preparation of the implant site axis and depth. The drills feature high initial sharpness and optimal cutting efficiency, thus achieving an effective material reduction. Due to their high efficiency toothing and the distinct tip-transversing blade, the drill has very good axial drilling properties for precise cutting without the need for prior use of a centering bur. The special twist drill blade geometry ensures smooth and precise operation. Moreover, the depth markings, which are lasered on to the working part in 2 mm intervals guarantee precise and safe control of the penetration depth.

The CeraDrill family is available in diverse dimensions and shapes for the individual preparation on the osteotomy. The following table shows the REF numbers, diameters, penetration depth, shank types, number of cutting flutes:

05

{2}------------------------------------------------

Remark:

further explanations concerning the columns of the following

table can be found in chapter 11 of this submission

| Model            | REF /<br>"Bezeichnung" | ID Nr. /<br>Ident<br>Nr. | Description                         | Diameter /<br>SØK Kopf-<br>durch-<br>messer" | penetration<br>depth / KL<br>"Kopflänge" | Total<br>length/<br>Gesamt-<br>länge | Shank<br>type<br>ISO<br>1797 | Shank<br>type<br>US | Cutting<br>flutes/<br>Schnei-<br>denzahl |
|------------------|------------------------|--------------------------|-------------------------------------|----------------------------------------------|------------------------------------------|--------------------------------------|------------------------------|---------------------|------------------------------------------|
| K210L16.RA.020   | K210L16.204.020        | 033792                   | CeraDrill<br>pilot drill<br>2,0 L16 | 2,0 mm                                       | 16,0 mm                                  | 30,5 mm                              | 204                          | RA                  | 2                                        |
| K210L19.RA.020   | K210L19.204.020        | 033793                   | CeraDrill<br>pilot drill<br>2,0 L19 | 2,0 mm                                       | 19,0 mm                                  | 33,5 mm                              | 204                          | RA                  | 2                                        |
| K210L16.RA.028   | K210L16.204.028        | 033794                   | CeraDrill<br>pilot drill<br>2,8 L16 | 2,8 mm                                       | 16,0 mm                                  | 32,0 mm                              | 204                          | RA                  | 3                                        |
| K210L19.RA.028   | K210L19.204.028        | 033795                   | CeraDrill<br>pilot drill<br>2,8 L19 | 2,8 mm                                       | 19,0 mm                                  | 35,0 mm                              | 204                          | RA                  | 3                                        |
| K210L20.RASL.020 | K210L20.205.020        | 034899                   | CeraDrill<br>pilot drill<br>2,0 L20 | 2,0 mm                                       | 20,6 mm                                  | 41,0 mm                              | 205                          | RASL                | 2                                        |
| K210L20.RASL.028 | K210L20.205.028        | 034900                   | CeraDrill<br>pilot drill<br>2,8 L20 | 2,8 mm                                       | 20,84 mm                                 | 41,0 mm                              | 205                          | RASL                | 3                                        |
| K210L20.RASL.035 | K210L20.205.035        | 035211                   | CeraDrill<br>pilot drill<br>3,5 L20 | 3,5 mm                                       | 21,05 mm                                 | 41,0 mm                              | 205                          | RASL                | 3                                        |
| K210L20.RASL.042 | K210L20.205.042        | 035212                   | CeraDrill<br>pilot drill<br>4,2 L20 | 4,2 mm                                       | 21,26 mm                                 | 41,0 mm                              | 205                          | RASL                | 3                                        |

Date: 2007-02-01 Revision: 3 Chapter

{3}------------------------------------------------

## Intended use:

CeraDrill is a ceramic pilot drill for the dental implantology made of a modern high performance ceramic. The drills are intended to be used with a handpiece.

- Indications for use .
The drill is used to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. The drills are used with a handpiece.

- Target population
The target population for the CeraDrill are Patients applicable for dental treatment with endosseous implants CeraDrill is used in the following anatomical sites: maxilla or mandible

CeraDrill is intended to be used at the Dentist

The device is an integral part of any implantological preparation.

Due to the fact that the device is intended to be used in the framework of the dental implantology the patients must be applicable for dental treatment with endosseous implants.

05

{4}------------------------------------------------

# Summary of technological Characteristics

# (compared to the predicate device)

The following table shows the technological characteristics of both

devices:

.

| Aspect                          | New device                                                                                                                                                                | Predicate device                                               |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
|                                 | CeraDrill                                                                                                                                                                 | Amorpheus Diamond<br>coated drill                              |
| energy used and/or<br>delivered | The device is intended to<br>be used in electrical<br>powered hand pieces                                                                                                 | same                                                           |
| design                          | two / three cutting flutes<br>external irrigation                                                                                                                         | two / three cutting flutes<br>internal /external<br>irrigation |
| performance                     | Ordinary concentricity                                                                                                                                                    | Ordinary concentricity                                         |
| corrosion                       | Corrosion resistive<br>Advantage:<br>not leading to decolouri-<br>sation while reprocessing                                                                               | Limited corrosion<br>resistance for stainless<br>steel         |
| materials                       | The CeraDrill is made of<br>ZrO2/Al2O3/Y2O3<br><br>Advantage:<br>- just one material<br>- no coating<br>- limited probability of<br>abrasion due to<br>inadequate bonding | stainless steel<br>coated with diamond                         |
| Biocompatibility                | According ISO 7405                                                                                                                                                        | same                                                           |

Date: 2007-02-01 Revision: 3 Chapter 05 Page: 5/8

:

:

:

:

: ﺗﺒ ......... :

:

{5}------------------------------------------------

#### Premarket notification 510(k) submission Gebr. Brasseler Gmbh & Co KG K: ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------CeraDrill ..

:

| Aspect                                                                | New device                                                                                                                                                                                                                                          | Predicate device                              |
|-----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
|                                                                       |                                                                                                                                                                                                                                                     |                                               |
| Ability to be<br>reprocessed                                          | CeraDrill                                                                                                                                                                                                                                           | Amorpheus Diamond<br>coated drill             |
|                                                                       | The CeraDrill can be<br>used, cleaned,<br>disinfected and sterilised<br>according to the<br>specifications as given in<br>the instructions for use<br>for at least 10 cycles.<br>(see test report# 06m1166 in<br>chapter 14A of this<br>submission) | No limitation for<br>reprocessing indicated   |
| Mechanical stability<br>after reprocessing                            | CeraDrill may be<br>reprocessed in 100<br>cycles without a<br>significant loss of<br>mechanical stability<br>(details see chapter 18 d)                                                                                                             | No further explanation /<br>information given |
|                                                                       | thermal safety                                                                                                                                                                                                                                      | Improved thermal safety                       |
|                                                                       |                                                                                                                                                                                                                                                     |                                               |
| Advantage:                                                            |                                                                                                                                                                                                                                                     |                                               |
| It was observed that the                                              |                                                                                                                                                                                                                                                     |                                               |
| temperature rise when                                                 |                                                                                                                                                                                                                                                     |                                               |
| using ceramic burs was<br>lower than that observed<br>with steel burs |                                                                                                                                                                                                                                                     |                                               |
| (details see chapter 18 a)                                            |                                                                                                                                                                                                                                                     |                                               |

:

:

: .

:

: 上一篇:

:

. : :

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

 Page: ୧ାଃ

{6}------------------------------------------------

# Substantial equivalence determination

#### Non-clinical performance data:

The determination of substantial equivalence is based on several nonclinical tests, most of which are repeated in the body of this 510(k) submission.

This includes especially those tests showing the physical and biological characteristics. The test results can be summarised as follows:

- General technological equivalence between predicate and current device
- improved thermal safety -
- improved safety concerning the risk of inadequate bonding of the coating
- the assumable lower resistive ness against bending forces during operation is negligible, since those bending force being necessary for fracture would lead to unacceptable osteotomies (=misuse)

(Details on non-clinical test can be found in chapter 18 of this submission)

### Clinical performance data

The determination of substantial equivalence is not based on clinical data.

{7}------------------------------------------------

# Conclusion:

The main intention of the CeraDrill is to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. As such the CeraDrill can be concluded to be substantial equivalent to the predicate device:

Amorphous Diamond Coated Drill, Nobel Biocare, K990846

The only technical difference is the material used: whereas the predicate device is made of steel and coated with amorphous diamond the CeraDril is made of a single material which is Zirconium oxide.

This technological difference leads to the following advantages of the CeraDrill device:

+ Metalfree preparation

+ positve Osteointgration

+ non allgergenic compared to CrNi Steal

There is no known disadvantage; the known difference in the bending strength of the materials does not have an impact in the intended use as identified in the chapter 18 of the submission.

The safety and performance data submitted in this file support a finding of substantial equivalence between the CeraDrill and the predicate device as specified above.

#### Any other information:

none

{8}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Gebr. Brasseler GmbH & Company KG C/O Mr. Olaf Brand TUV Rheinland of North America, Incorporated 12 Commerce Road Newtown, Connecticut 06470

JUN 2 9 2007

Re: K071352

Trade/Device Name: CeraDrill Regulation Number: 21 CFR 872.4130 Regulation Name: Intraoral Dental Drill Regulatory Class: I Product Code: EJL Dated: June 25, 2007 Received: June 27, 2007

Dear Mr. Brand:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{9}------------------------------------------------

#### Page 2 - Mr. Brand

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Susan Runore

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{10}------------------------------------------------

K0713352

# Indications for Use

510(k) Number (if known): 1007 12 50

Device Name: CeraDrill

- . Indications for use
The drill is used to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. The drills are used with a handpiece.

- specific indications:
CeraDrill is a ceramic drill for the dental implantology made of a modern high performance ceramic. The drills are intended to be used with a handpiece. The intended use for this device is to cut into the maxilla or mandible to create an osteotomy for endosseous dental implant placement.

- clinical settings / target population: · .
Patients must be applicable for dental treatment with endosseous implants.

- anatomical sites: .
The products are used in the framework of dental implantology for the preparation of the upper jaw (Maxilla) or the lower jaw

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susa Roose

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page I of 1

510(k) Number: K071352

---

**Source:** [https://fda-staging.innolitics.com/device/K071352](https://fda-staging.innolitics.com/device/K071352)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).