← Product Code [IPF](/productcode/IPF) · K071320

# EVERLIFE POWERED MUSCLE STIMULATOR, MODEL E-100204, E-100504, E-100509, E-100804, E-201111, E-201211, E-201311 (K071320)

_Everlife Medical Equipment Co., Ltd. · IPF · Aug 17, 2007 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K071320

## Device Facts

- **Applicant:** Everlife Medical Equipment Co., Ltd.
- **Product Code:** [IPF](/productcode/IPF.md)
- **Decision Date:** Aug 17, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Relaxation of muscle spasms Prevention or retardation of disuse atrophy Muscle re-education Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Maintaining or increasing range of motion

## Device Story

Sequenced transcutaneous muscle stimulation system; consists of stimulator, sequencer for channel selection, patient cable, and skin-applied electrodes. Device outputs various electrical waveforms to muscles. Patient controls signal intensity for safety/comfort. Multi-channel output allows bilateral operation or simultaneous/serial treatment of multiple regions. Used in clinical or home settings; operated by patient or clinician. Output provides electrical stimulation to induce muscle contraction; aids in muscle rehabilitation and thrombosis prevention.

## Clinical Evidence

No clinical data. Bench testing only, including electrical safety, electromagnetic compatibility, and system-level waveform testing.

## Technological Characteristics

Transcutaneous electrical nerve/muscle stimulator; multi-channel sequencer; patient-controlled intensity; shrouded connector patient cables; 510(k) cleared electrodes. Complies with medical device electrical safety and electromagnetic compatibility standards.

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- APEX Powered Muscle Stimulator ([K002336](/device/K002336.md): MS-104A)
- APEX Powered Muscle Stimulator ([K002339](/device/K002339.md): AEMSV)
- APEX Powered Muscle Stimulator ([K021754](/device/K021754.md): TS-1311, TS-1312)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## SUMMARY OF SAFETY AND EFFECTIVENESS for Powered Muscle Stimulator

| 510(k) NUMBER:                    | K071320                                                                                                                                       | AUG 17 2007 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| DATE OF<br>SUBMISSION:            | May 3, 2007                                                                                                                                   |             |
| SUBMITTER:                        | EVERLIFE MEDICAL EQUIPMENT CO., LTD.<br>NO 58, FU-CHIUN ST.<br>HSIN-CHU CITY, CHINA (TAIWAN) 30067<br>TEL: 886-3-5208829<br>FAX:886-3-5209783 |             |
| ESTABLISHMENT<br>REGISTRATION NO: | 3004753827                                                                                                                                    |             |
| OFFICIAL<br>CONTACT:              | Dr. JEN, KE-MIN<br>NO 58, FU-CHIUN ST.<br>HSIN-CHU CITY, CHINA (TAIWAN) 30067<br>TEL: 886-3-5208829<br>FAX:886-3-5209783                      |             |
| TRADE NAME:                       | EVERLIFE Powered Muscle Stimulator,<br>E-100204, E-100504, E-100509, E-100804,<br>E-201111, E-201211, E-201311                                |             |
| COMMON/USUAL<br>NAME:             | Powered Muscle Stimulator                                                                                                                     |             |
| CLASSIFICATION<br>NAME:           | Powered Muscle Stimulator                                                                                                                     |             |
| REGULATION<br>NUMBER:             | 890.5850, Class II                                                                                                                            |             |
| PREDICATED<br>DEVICE:             | APEX Powered Muscle Stimulator,<br>K002336: MS-104A; K002339: AEMSV;<br>K021754: TS-1311, TS-1312                                             |             |

:

:

.

:

:

{1}------------------------------------------------

| INTENDED USE:                 | <ul><li>Relaxation of muscle spasms</li><li>Prevention or retardation of disuse atrophy</li><li>Muscle re-education</li><li>Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis</li><li>Maintaining or increasing range of motion</li></ul>                                                                                                                                                                                                                                                                                          |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description of Device:        | A sequenced system for transcutaneous muscle stimulation consists of a stimulator, a sequencer for channel selection, patient cable, and electrodes applied to the skin.<br>Various types of waveforms may be output to generate the desired effect on the muscle(s) to be treated, and the patient is given control of the signal intensity for personal safety and comfort. Sequenced system may have more than on output channel in order to operate bilaterally on the body or to treat multiple regions simultaneously or serially in a prescribed sequence. |
| Non-Clinical Tests Submitted: | The EVERLIFE Powered Muscle Stimulator has been tested in accordance with applicable standards for medical device electrical safety, electromagnetic compatibility, and the particular requirements for safety of nerve and muscle stimulators.<br>Accessories also meet safety requirements: 510(k) electrodes are specified, and the patient cable utilizes shrouded connectors to meet lead wire safety requirements.<br>System level testing including waveform testing was performed in combination the EVERLIFE Powered Muscle Stimulator.                  |
| Clinical Tests Submitted:     | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Conclusion:                   | As the product description and tests as above, the new device: EVERLIFE Powered Muscle Stimulator: E-100204, E-100504, E-100509, E-100804, E-201111, E-201211, and E-201311 are as safe and effective as, and the function in a manner equivalent to the predicate devices: APEX Powered Muscle Stimulator: K002336: MS-104A; K002339: AEMSV; K021754: TS-1311, TS-1312.<br>Thus the new device is substantially equivalent to the predicate devices in this aspect.                                                                                              |

..............................................................................................................................................................................

.

:

.

…

.

.

:

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is facing to the right.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Everlife Medical Equipment Co., Ltd. % Dr. Jen, Ke-Min No. 58, Fu Chiun Street IIsin Chu City, 30067 Taiwan, ROC

AUG 1 7 2007

Re: K071320

Trade/Device Name: Everlife Powered Muscle Stimulator Models E-100204, E-100504, E-100509, E-100804, E-201111, E-201211, E-201311 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF Dated: July 5, 2007 Received: July 11, 2007

Dear Dr. Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Dr. Jen, Ke-Min

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark Melkerson Director Division of General Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

| 510(k) Number: | K071320                                                                                                        |
|----------------|----------------------------------------------------------------------------------------------------------------|
| Device Name:   | EVERLIFE Powered Muscle Stimulator,<br>E-100204, E-100504, E-100509, E-100804,<br>E-201111, E-201211, E-201311 |

## Indications for Use :

- Specific indications: used to apply an electrical current to electrodes on patient's skin to function as:
	- Relaxation of muscle spasms
	- Prevention or retardation of disuse atrophy ●
	- . Muscle re-education
	- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
	- . Maintaining or increasing range of motion

| Prescription Use            | V |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (Division Sign-Off) |             |
|----------------------------------------------------------------------|-------------|
| Division of General, Restorative,                                    |             |
| and Neurological Devices                                             |             |
| 510(k) Number                                                        | k071320     |
|                                                                      | Page 1 of 1 |

---

**Source:** [https://fda-staging.innolitics.com/device/K071320](https://fda-staging.innolitics.com/device/K071320)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).