K071310 · Siemens AG · JAK · Jun 8, 2007 · Radiology
Device Facts
Record ID
K071310
Device Name
SYNGO CT ONCOLOGY SOFTWARE PACKAGE
Applicant
Siemens AG
Product Code
JAK · Radiology
Decision Date
Jun 8, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
syngo CT Oncology is a self-contained, non invasive image analysis software package designed to support the physician in routine diagnostic oncology, staging and follow-up. Flexible layouts and automated image registration facilitate the synchronous display and navigation of multiple datasets for viewing multi-phase CT data and easy follow-up comparison. The application provides a range of interactive tools specifically designed for segmentation and volumetric analysis of lesions and includes dedicated tools to support the physician in lung, liver and lymph node assessment in CT data. Dedicated workflowsupport and integrated, accumulative reporting help to track lesions and their changes in e.g. size, shape and enhancement pattern over time. syngo CT Oncology also facilitates functional imaging offering fusion with other modalities such as PET data. It also features syngo LungCAD for detecting small lung nodules (PMA-approved).
Device Story
syngo CT Oncology is a non-invasive image analysis software package for diagnostic oncology, staging, and follow-up. Input: multi-phase CT datasets; PET data for fusion. Operation: software provides automated image registration, segmentation, and volumetric analysis tools for lung, liver, and lymph node lesions. Workflow support includes accumulative reporting to track lesion changes in size, shape, and enhancement patterns over time. Includes syngo LungCAD for lung nodule detection. Used in clinical settings by physicians to view, navigate, and compare datasets. Output: visual displays, volumetric measurements, and reports. Benefit: facilitates efficient lesion tracking and diagnostic assessment.
Clinical Evidence
No clinical data provided. Safety and effectiveness are supported by hazard analysis, software verification, and validation testing.
Technological Characteristics
Software-based image analysis package. Features automated image registration, segmentation, and volumetric evaluation tools. Supports multi-modality fusion (e.g., PET/CT). Operates on standard workstation hardware. Software verification and validation performed per industry standards.
Indications for Use
Indicated for use by physicians for routine diagnostic oncology, staging, and follow-up of patients undergoing CT imaging. Supports assessment of lung, liver, and lymph node lesions. Contraindications not specified.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Summary
K071310
JUN - 8 2007
# SECTION 9
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### GENERAL INFORMATION l.
#### 1. Device Name and Classification
syngo CT Oncology - Software Package Product Name: Classification Name: Accessory to Computed Tomography System Classification Panel: Radiology 21 CFR §892.1750 CFR Section: Device Class: Class II Product Code: 90 JAK
#### 2. lmporter/Distributor Establishment:
Registration Number: 2240869
Siemens Medical Solutions, Inc. 51 Valley Stream Pkwy Malvern, PA 19355
#### Manufacturing Facility: 3.
Siemens AG Wittelsbacherplatz 2 D-80333 Muenchen, Germany
#### 4. Contact Person:
Dr. Kristin Frowein Regulatory Submissions Specialist Siemensstr.1; D-91301 Forchheim +49 9191 18-9638 Phone: +49 9191 18-9782 Fax:
#### 5. Date of Preparation of Summary: January 25, 2007
{1}------------------------------------------------
# II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUBSTANTIAL EQUIVALENCE DETERMINATION
#### General Safety and Effectiveness Concerns 6.
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
### 7. Substantial Equivalence
The syngo CT Oncology software package that is addressed in this premarket notification, is substantially equivalent to the following commercially available software package:
| Manufacturer | Product | 510(k) | Clearance date |
|--------------------------|---------------------------|---------|----------------|
| 1. Siemens AG | LungCARE CT<br>SW Package | K022013 | 07/16/2002 |
| 2. Edda Technology, Inc. | IQQA-Liver | K061696 | 11/13/2006 |
| 3. GE Medical Systems | Advantage Windows | K963345 | |
| | Tissue Volume Option | | 10/18/1996 |
| 4. Siemens AG | syngo Multimodality | K010938 | 06/26/2001 |
| | Workstation | | |
| | (feature: Image Fusion) | | |
#### Device Description and Intended Use: 8.
syngo CT Oncology is a self-contained, non invasive image analysis software package designed to fast-track routine diagnostic oncology, staging and follow-up. Flexible layouts and automated image registration facilitate the synchronous display and navigation of multiple datasets for viewing multi-phase CT data and easy follow-up comparison. The application provides a range of automated tools specifically designed to support physicians in the segmentation and volumetric evaluation of suspicious lesions including dedicated tools for lung, liver and lymph node assessment in CT data. Dedicated workflow-support and integrated, accumulative reporting allow to track lesions and their changes in e.g. size, shape and enhancement pattern over time. syngo CT Oncology also facilitates functional imaging offering fusion with other modalities such as PET data. It also features syngo LungCAD for detecting small lung nodules (PMA-approved).
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows a circular logo with text around the perimeter and a symbol in the center. The text reads "HEALTH & HUMAN SERVICES - USA". The symbol in the center appears to be three stylized lines or curves, possibly representing a stylized bird or abstract design.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Siemens AG Medical Solutions % Mr. Stefan Preiss Responsible Third Party Official TÜV SÜD America, Inc. 1775 Old Highway 8 N.W., Suite 104 NEW BRIGHTON MN 55112-1891
Re: K071310
Trade/Device Name: syngo CT Oncology - Software Package Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 22, 2007 Received: May 25, 2007
### Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/9 description: The image shows a circular logo with the text "1896-1996" at the top and "Centennial" at the bottom. In the center of the logo, there are the letters "PA" in a stylized font. Below the word "Centennial" are four stars arranged in a horizontal line. The logo appears to be a commemorative emblem for a centennial celebration.
noting Public Health
{3}------------------------------------------------
Page 2 -
Enclosure
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlvindustry/support/index.html.
Sincerely yours,
Nancy Cbrogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
SIEMENS
Indication for Use
### SECTION 3
## INDICATION FOR USE
510(k) Number (if known):
Device Name:
syngo CT Oncology - Software Package
### Indications for Use:
syngo CT Oncology is a self-contained, non invasive image analysis software package designed to support the physician in routine diagnostic oncology, staging and follow-up. Flexible layouts and automated image registration facilitate the synchronous display and navigation of multiple datasets for viewing multi-phase CT data and easy follow-up comparison. The application provides a range of interactive tools specifically designed for segmentation and volumetric analysis of lesions and includes dedicated tools to support the physician in lung, liver and lymph node assessment in CT data. Dedicated workflowsupport and integrated, accumulative reporting help to track lesions and their changes in e.g. size, shape and enhancement pattern over time. syngo CT Oncology also facilitates functional imaging offering fusion with other modalities such as PET data. It also features syngo LungCAD for detecting small lung nodules (PMA-approved).
| Prescription Use | X |
|-----------------------------|----|
| (Part 21 CFR 801 Subpart D) | AN |
ND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
concurrence of CDRH, Office of Device Evaluation (ODE)
Helstener
on Sign-Off Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.