The IMPAX SE SigmaCom Client Workstations are intended to perform operations relating to the display, digital processing, review, transfer, storage, printing, measurements, teleradiology exchange of medical images and patient demographic information and to be integrated with separately cleared third party applications. They are intended for use by the physician to aid diagnosis and by medical professionals whenever they require access to medical images and patient demographic information.
Device Story
Multifunction PACS workstation for display and interpretation of medical images; inputs include DICOM images from various modalities (CT, MR, MG, general radiography). Device performs digital processing, 3D rendering (MIP, MPR), and orthopedic measurements; supports teleradiology for remote consultation and quality control. Used in clinical settings by physicians, radiologists, and technologists. Output displayed on workstation monitors for clinical review and diagnostic aid. Facilitates workflow by enabling image access, analysis, and sharing across healthcare environments.
Clinical Evidence
Bench testing only. Device tested for conformance to specifications.
Technological Characteristics
Software application installed on standard computers. Supports DICOM connectivity. Functions include 3D rendering (MIP, MPR), orthopedic measurement tools, and teleradiology exchange. No specific hardware materials or sterilization methods applicable as it is a software-based workstation.
Indications for Use
Indicated for storage, reading, interpretation, clinical review, analysis, annotation, distribution, printing, editing, and processing of DICOM digital images/data by healthcare professionals (radiologists, physicians, technologists, clinicians). Includes teleradiology for remote consultation, mammography screening/diagnosis (MG 'for presentation' images), orthopedic measurements (hip, knee, spine), 3D rendering (MIP, MPR, series synchronization), and quality control activities.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K07/297
## 510(k) Summary
JUN 2 2 2007
## Impax SE SigmaCom Client (PACS Medical Imaging Display Workstation)
Common/Classification Name: Picture Archiving and communications system (PACS), 21 CFR 892.2050 Proprietary Name: Impax SE SigmaCom Client (PACS Medical Imaging Display Workstation)
Agfa HealthCare Corporation 10 South Academy Street Greenville, SC 29602-9048
Contact: Tom Holbrook, Prepared: DATE Telephone: (519) 746-6210 ext. 3297 Facsimile: (519) 746-3745
### A. LEGALLY MARKETED PREDICATE DEVICES
This is a 510(k) for Agfa's Impax SE SigmaCom Client (PACS Medical Imaging Display Workstation).
The predicate devices are Agfa's previous earlier workstations, the OT3000 (K050751), the Embrace (mammography) workstation (K040555) and the Web1000 remote access workstation (K053458). This new device includes most of the clinical functions of the predicate devices but was developed by Agfa HealthCare Enterprise Solutions, Lognes, France.
#### DEVICE DESCRIPTION B.
The new device is a multifunction PACS workstation for displaying and interpretation of medical images. It includes functions similar to the predicates. It can be used as a general radiography or orthopedic workstation. It displays "for presentation" (MG) mammography images. Image processing is handled by the mammography modalities. It can also display 3D images acquired from CT and MR modalities. It provides user teleradiology functionalities to share information and consult with remote users and can be used to perform quality control activities related to patient images and data.
The principles of operation and technological characteristics of the new and predicate devices are the same.
1
{1}------------------------------------------------
#### C. INTENDED USE
The IMPAX SE SigmaCom Client Workstations are intended to perform operations relating to the display, digital processing, review, transfer, storage, printing, measurements, teleradiology exchange of medical images and patient demographic information and to be integrated with separately cleared third party applications. They are intended for use by the physician to aid diagnosis and by medical professionals whenever they require access to medical images and patient demographic information.
#### D. SUBSTANTIAL EQUIVALENCE SUMMARY
Agfa's Impax SE SigmaCom Client (PACS Medical Imaging Display Workstation) has the same indications for use as the legally marketed predicate devices with the exception that it does not offer users access to electronic orthopedic device templates.
This indication difference does not modify the intended diagnostic effect: To assist users performing measurements common to the orthopedic specialty.
The new device has the same control methods and operating principles as the predicates. Descriptive characteristics and data provided in this submission are sufficiently precise to assure substantial equivalence.
#### TECHNOLOGICAL CHARACTERISTICS ய்
The technological characteristics are the same in the proposed and predicate devices. Both the new device and the predicates are software applications installed on standard computers.
### F. TESTING
Agfa's Impax SE SigmaCom Client (PACS Medical Imaging Display Workstation) has been tested for conformance to specifications.
### G. CONCLUSIONS
This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes an abstract symbol of three wavy lines. The text "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the perimeter of the circle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
## 2 2 2007
Mr. Tom Holbrook Ouality Assurance & Regulatory Affairs Manager Agfa Healthcare Corporation 10 South Academy Street GREENVILLE SC 29601
Re: K071297
Trade/Device Name: Impax SE SigmaCom Client (PACS Medical Imaging Display Workstation) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 4, 2007 Received: May 9, 2007
Dear Mr. Holbrook:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a black and white circular logo. The logo contains the text "1906-2006" at the top, "FDA" in the center, and "Centennial" at the bottom. There are also three stars at the very bottom of the logo. The logo is surrounded by a dotted border.
Proloting and Promoting Publio Hoalth
{3}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlvindustry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K07 297
## Device Name: Impax SE SigmaCom Client (PACS Medical Imaging Display Workstation)
#### Indications for Use:
The system is indicated for the storage, reading, interpretation, clinical review, analysis, annotation, distribution, printing, editing and processing of digital images and data acquired from any DICOM device by healthcare professionals, including radiologists, physicians, technologists and clinicians.
- With the Teleradiology option it can be used to consult and share ﮯ information with remote users.
- With the Mammography option it can be used for screening and ా diagnosis (with MG, "For presentation" images only) from FDA approved modalities in softcopy and printed formats.
- With the Orthopedic option it can be used to perform common ﺖ orthopedic measurements of the hip, knee, spine (Coxometry, Gonometry, Lipman Cobb).
- With the Slice Imaging option it can be used to process basic 3D rendering: MIP, MPR or series synchronization.
- With the Quality Assurance option it can be used by PACS administrators or technologists to perform quality control activities related to patient images and data.
Nancy C. Brogdon
(Division Sign-Off)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K071297
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
1
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.