K071271 · Biomet Manufacturing, Inc. · JDQ · Oct 23, 2007 · Orthopedic
Device Facts
Record ID
K071271
Device Name
BIOMET CABLE SYSTEM
Applicant
Biomet Manufacturing, Inc.
Product Code
JDQ · Orthopedic
Decision Date
Oct 23, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Biomet® Cable System is indicated for use whenever a conservative or non-implant surgery is deemed insufficient to improve the medical condition of the patient. The System can be utilized anywhere monofilament wire has been previously indicated. 1. Spinal applications include spinal degenerative surgery, as an adjunct to spinal fusions and sublaminar and intraspinous process wiring for trauma applications. The Biomet® Cable System may also be used with instrumentation involving the correction of scoliotic, kyphotic, and lordotic deformities. The titanium system is compatible with other titanium implants and the stainless steel system with stainless steel implants, wherever "wiring" may help secure the attachment of other implants. 2. Trochanteric reattachment after trochanteric osteotomy, following total hip arthroplasty. 3. Sternotomy indications include the "re-wiring" of sternums following osteotomy. 4. Trauma surgery indications include olecranon, ankle, patella, and some shoulder fracture rewiring.
Device Story
Biomet Cable System provides bone fixation via flexible, multi-strand cables; available in titanium alloy (Ti-6Al-4V) or stainless steel (316LVM). System includes single and double cable styles, secured by pre-loaded crimps. Used by surgeons in OR settings for orthopedic and spinal procedures; functions as an alternative to traditional monofilament wire. Instruments include tensioner, crimper, and cutter. Provides mechanical stabilization of bone segments or attachment of other implants; benefits patient by facilitating secure fixation during spinal, trauma, or reconstructive surgeries.
Clinical Evidence
No clinical data provided. Substantial equivalence based on non-clinical laboratory testing confirming device functionality.
Technological Characteristics
Materials: Titanium alloy (ASTM F-136) or 316LVM stainless steel (ASTM F-138). Design: Multi-strand cable with pre-loaded crimp locking mechanism. Manual instrumentation: tensioner, crimper, cutter. Compatible with corresponding metal-type implants.
Indications for Use
Indicated for patients requiring bone fixation where conservative surgery is insufficient. Applications include spinal surgery (fusion, deformity correction, trauma), trochanteric reattachment post-hip arthroplasty, sternal re-wiring post-osteotomy, and fracture fixation (olecranon, ankle, patella, shoulder).
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
{0}------------------------------------------------
OCT 2 3 2007
# 510(k) Summary
| Preparation Date: | October 17, 2007 |
|----------------------|------------------------------------------------------------------------------------------------------------|
| Applicant/Sponsor: | Biomet Spine (aka EBI, names may be used interchangeably)<br>100 Interpace Parkway<br>Parsippany, NJ 07054 |
| Contact Person: | Becky Earl/Debra L. Bing |
| Proprietary Name: | Biomet® Cable System |
| Common Name: | Cable System |
| Classification Name: | Cerclage, Bone Fixation (CFR 888.3010) |
Legally Marketed Devices To Which Substantial Equivalence Is Claimed: Medtronic/Sofamor Danek Atlas™ Cable System, K920201, K925812, K030816; Songer Spinal Cable System, K922952.
Device Description: The Biomet® Cable System is available in either titanium or stainless steel, consisting of a flexible, multi-strand, medical grade titanium alloy (Ti-6Al-4V, ASTM F-136) or a medical grade stainless steel (316LVM, ASTM F-138) cable. The systems are available in both single and double styles. The cable is locked into place via a pre-loaded crimp. Instruments for the Biomet® Cable System include a Cable Tensioner, a Cable Crimper, and a Cable Cutter. The titanium system is compatible with other medical grade implants made of titanium and the stainless steel system is compatible with other stainless steel implants, whenever "wiring" may help secure the attachment of other implants.
Indications for Use: The Biomet® Cable System is indicated for use whenever a conservative or nonimplant surgery is deemed insufficient to improve the medical condition of the patient. The System can be utilized anywhere monofilament wire has been previously indicated.
- 1. Spinal applications include spinal degenerative surgery, as an adjunct to spinal fusions and sublaminar and intraspinous process wiring for trauma applications. The Biomet® Cable System may also be used with instrumentation involving the correction of scoliotic, kyphotic, and lordotic deformities. The titanium system is compatible with other titanium implants and the stainless steel system with stainless steel implants, wherever "wiring" may help secure the attachment of other implants.
- 2. Trochanteric reattachment after trochanteric osteotomy, following total hip arthroplasty.
- 3. Sternotomy indications include the "re-wiring" of sternums following osteotomy.
- 4. Trauma surgery indications include olecranon, ankle, patella, and some shoulder fracture rewiring.
### Summary of Technologies:
The technological characteristics (materials, design, sizing, indications) of the Biomet® Cable System are similar or identical to the predicate devices.
Non-Clinical Testing: Non-clinical laboratory testing was performed to determine substantial equivalence. The results indicated that the device was functional within its intended use.
Clinical Testing: None provided as a basis for substantial equivalence.
All trademarks are property of Biomet, Inc., except for Atlas™, which belongs to Medtronic/Sofamor Danek
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure or a caduceus, with three wavy lines representing the body and a stylized head.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## OCT 2 3 2007
Biomet Manufacturing Corp % Ms. Becky Earl Regulatory Specialist P. O. Box 587 Warsaw, IN 46581
Re: K071271
> Trade/Device Name: Biomet® Cable Ssystem Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: JDQ Dated: October 17, 2007 Received: October 18, 2007
Dear Ms. Earl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 -- Ms. Becky Earl
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Milliken
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known): ______________
Device Name: Biomet® Cable System
Indications for Use: The Biomet® Cable System is indicated for use whenever a conservative or non-implant surgery is deemed insufficient to improve the medical condition of the patient. The System can be utilized anywhere monofilament wire has been previously indicated.
- 1. Spinal applications include spinal degenerative surgery, as an adjunct to spinal fusions and sublaminar and intraspinous process wiring for trauma applications. The Biomet® Cable System may also be used with instrumentation involving the correction of scoliotic, kyphotic, and lordotic deformities. The titanium system is compatible with other titanium implants and the stainless steel system with stainless steel implants, wherever "wiring" may help secure the attachment of other implants.
- 2. Trochanteric reattachment after trochanteric osteotomy, following total hip arthroplasty.
- 3. Sternotomy indications include the "re-wiring" of sternums following osteotomy.
- 4. Trauma surgery indications include olecranon, ankle, patella, and some shoulder fracture rewiring.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
[signature]
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K07127
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.