← Product Code [OBO](/productcode/OBO) · K071250

# CA, MODEL: CAM (K071250)

_Optovue, Inc. · OBO · Sep 28, 2007 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K071250

## Device Facts

- **Applicant:** Optovue, Inc.
- **Product Code:** [OBO](/productcode/OBO.md)
- **Decision Date:** Sep 28, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1570
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The CA, an auxiliary lens adapter, when used in conjunction with RTVue, is indicated for in vivo imaging and measurement of the cornea and other ocular structures of the anterior segment of the eye.

## Device Story

The CA is an auxiliary lens adapter for the RTVue OCT system; it enables imaging of the cornea and anterior segment in addition to the RTVue's existing retinal imaging capabilities. The system uses low-coherence interferometry to scan a beam into the patient's eye, measuring tissue reflectivity to generate cross-sectional images composed of A-scans. Operated by a clinician in a clinical setting, the device is mounted on a motorized table with a chin/forehead rest; the operator uses a joystick for alignment. The software GUI allows the operator to select corneal landmarks for labeling. Output consists of cross-sectional images and corneal thickness measurements. These outputs assist clinicians in visualizing ocular structures and performing pachymetry, providing diagnostic information for anterior segment conditions.

## Clinical Evidence

Clinical comparison study evaluated corneal images and thickness measurements. RTVue/CA images were compared to Visante OCT, showing improved visibility of anterior segment anatomic features. Corneal thickness measurements were compared against ultrasound pachymetry (r=0.9889), Orbscan II (r=0.9995), and Visante OCT (r=0.9924-0.9939). Reproducibility of corneal measurements was 1.2 um root-mean-square.

## Technological Characteristics

Optical Coherence Tomography (OCT) system; auxiliary lens adapter for anterior segment imaging. Axial resolution 5.1± 0.4um. Motorized patient table with chin/forehead rest. Joystick-controlled alignment. Software-based GUI for image acquisition, analysis, and display. Energy levels and safety design identical to predicate RTVue.

## Regulatory Identification

An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- RTVue ([K062552](/device/K062552.md))
- Visante OCT ([K051789](/device/K051789.md))
- Bioptigen ([K063343](/device/K063343.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "opt.vue" in a simple, sans-serif font. A curved line arches over the word, connecting the "o" and "e". The dot between "opt" and "vue" is a solid black circle.

Optovue RTVue/CA 510(K) Premarket Notification

Ko 71250

# 510(k) Summary

# Optovue, Incorporated CA Model: CAM

SEP 2 8 2007

This 510(k) summary for the CA is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

### General Information

| Manufacturer:   | Optovue, Inc.<br>41752 Christy Street,<br>Fremont, CA 94538<br>Phone: (510) 623-8868<br>Fax: (510) 623-8668<br>Registration No.: |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jay Wei<br>CEO/President<br>Optovue, Inc.<br>Phone: (510)623-8868 x102<br>e-mail: jay_wei@optovue.com                            |

## Device Information

Classification: Trade Name:

Common Name: Classification Name: Class II CA Model: CAM Optical Coherence Tomography (OCT) Ophthalmoscope, a-c powered (21 CFR§ 886.1570)

### Predicate Devices

RTVue (K062552) – Optovue Visante OCT (K051789) - Carl Zeiss Meditec Bioptigen (K063343) - Bioptigen Inc.

### Intended Use

The CA, an auxiliary lens adapter, when used in conjunction with RTVue, is indicated for in vivo imaging and measurement of the cornea and other ocular structures of the anterior segment of the eye.

{1}------------------------------------------------

# Device Description

The CA device uses the same Optical Coherence Tomography (OCT) technology that was previously reviewed by FDA in its clearance of premarket notification K062552 for RTVue. The CA adapter gives the user an option to use the RTUve device as previously approved for retina scans, or to use it for cornea and anterior eye scans. Aside from the CA auxiliary attachment, the RTVue is virtually unchanged for the CA use except the CA software module provides for menu selections in the graphical user interface, which are selected by the operator to label corresponding corneal landmarks instead of those of the retina. The system scans a beam into patient's eye and uses a low coherence interferometer to measure the reflectivity of the ocular tissue. The cross-sectional ocular tissue structure is composed of sequence of A-scans. RTVue has a traditional patient and instrument interface like most ophthalmic devices. The device is mounted on a motorized patient table. The patient will rest their head on the forehead and chin rest. The operator uses joystick to align the device to patient's eye. The computer has a graphic user interface for acquiring, analyzing and displaying the acquired image. The RTVue image acquisition speed and image resolution remain the same when used in conjunction with CA.

### Safety

The energy level and all safety related design of the RTVue remains the same when used in conjunction with CA. The only difference is that the CA allows RTVue system to image the cornea instead of the retina. The addition of the CA does not change the safety.

## Effectiveness

The image resolution of RTVue when used in conjunction with CA is 5um which is about 3 times better than the predicate device Visante OCT. The scan speed of RTVue when used in conjunction with CA is also 13 times faster than Visante OCT.

# Substantial Equivalence

The RTVue/CA is substantially equivalent to the predicate devices presented within this premarket notification with regard to intended use, operating principle, function, material, and energy source, all of which are similar. The RTVue/CA can perform imaging on cornea and anterior segment by simply adding a lens in front the existing RTVue ocular lens. The indication of use on the cornea and anterior segment of the eye are equivalent to another predicate device, Visante OCT marketed by Carl Zeiss Meditec and to that of Bioptigen Inc. All these devices are based on the same Optical Coherence Tomography technology.

# Performance Test Data

### Non-Clinical:

### Image Resolution Test:

Objective: To measure the axial resolution of RTVue/CA

Result: The axial resolution of the tested RTVue/CA system is 5.1± 0.4um

Optovue, Inc.

{2}------------------------------------------------

# Clinical:

# 1. Corneal Imaging Comparison:

Objective: To compare images of cornea obtained with the RTVue/CA and with the Visante OCT.

Result: The RTVue/CA is capable of providing images that compare well with the Visante OCT. Anatomic features of the anterior segment of the eye were visible to the physicians that assessed the RTVue/CA images; the same physicians could not identify these features in the Visante OCT images.

# 2. Cornea Thickness Measurement Comparison:

Objective: To compare the cornea thickness measurement with other commercial available pachymeters.

Result: The cornea thickness measurement made by RTVue/CA is highly correlated with the ultrasound pachymetry system (Pearson correlation r = 0.9889), Orbscan II (Pearson correlation r = 0.9995), and Visante OCT (Pearson correlation r = 0.9924 at the vertex, r == 0.9939 over the central zone). The reproducibility of the RTVue/CA corneal measurement was 1.2 um root-mean-square for the difference between repeat pachymetry maps.

# Conclusion

As described in this 510(k) Summary, all testing and analysis were conducted on the RTVue/CA to ensure that the device is safe and effective for its intended use when used in accordance with its instructions for use.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Optovue, Inc. c/o Jay Wei 41752 Christy Street Freemont, CA 94538

SEP 2 8 2007

Re: K071250

Trade/Device Name: CA Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: OBO Dated: August 22, 2007 Received: August 23, 2007

Dear Mr. Wei:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

M.B. Egerlton, mD

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: CA Model:CAM

Indications for Use:

The CA, an auxiliary lens adapter, when used in conjunction with RTVue, is indicated for in vivo imaging and measurement of the cornea and other ocular structures of the anterior segment of the eye.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Clay R. Buttem

(Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises

510(k) Number K071250

---

**Source:** [https://fda-staging.innolitics.com/device/K071250](https://fda-staging.innolitics.com/device/K071250)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
