K071193 · Vidar Systems Corp. · MQB · May 30, 2007 · Radiology
Device Facts
Record ID
K071193
Device Name
VIDAR VISION 3000 AND 4000 X-RAY SYSTEMS
Applicant
Vidar Systems Corp.
Product Code
MQB · Radiology
Decision Date
May 30, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
Vidar Vision 3000 and Vidar Vision 4000 X-Ray Systems are indicated for use in generating radiographic images of human anatomy. It has a Solid State X-ray Imaging system intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. The Vidar Vision 3000 and Vidar Vision 4000 is not indicated for diagnostic X-ray mammography.
Device Story
Vidar Vision 3000 and 4000 are stationary digital radiography (DR) systems for general-purpose diagnostic imaging. Systems combine X-ray generators with solid-state digital detectors (3K detector for Vision 3000; 4K detector for Vision 4000). Operated by clinicians in imaging centers, orthopedic practices, and hospitals via software-driven touch panel LCDs and remote consoles. Devices capture X-ray signals, converting them into high-resolution digital images to replace traditional film/screen systems. Output is displayed for physician review to support clinical decision-making. Benefits include rapid image acquisition, improved productivity, and high-resolution imaging.
Clinical Evidence
Bench, user, and standards testing performed. No clinical trial data provided.
Technological Characteristics
Stationary X-ray system; solid-state digital detector (160-micron pitch for 3000; 120-micron pitch for 4000); software-driven touch panel LCD control; 50-80 kW generator output; Balco B302I/A DHHS collimator.
Indications for Use
Indicated for generating radiographic images of human anatomy in general-purpose diagnostic procedures. Not indicated for diagnostic X-ray mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K071193
MAY 3 0 2007
Image /page/0/Picture/2 description: This image shows the letterhead for VIDAR systems corporation. The letterhead includes the company logo, address (365 Herndon Parkway, Herndon, VA 20170), phone numbers (800.471.7226, +1.703.471.7070), and fax number (+1.703.471.7665). The contact person is Carrie L. Brancart, Quality Manager, and the date prepared is May 13, 2007.
-1
- 1. Identification of the Device: Proprietary-Trade Name: Vidar Vision 3000 and Vidar Vision 4000 X-Ray Systems Classification Name: Stationary X-ray system, Product Codes Product Code 90 KPR and MQB Common/Usual Name: General purpose diagnostic X-ray Unit.
- 2. Equivalent legally marketed devices: This notification is for a MODIFIED device. This device COMBINES two 510(k) cleared devices, the SEDECAL Universal Radiographic Systems K012546 AND the Vidar 2000 Digital Radiographic Detector, K070563. This combination is functionally identical to a SEDECAL cleared device, Sedecal URS LP X-Ray Units with Digital Detector, K042876 ..
- 3. Indications for Use (intended use) Vidar Vision 3000 and Vidar Vision 4000 X-Ray Systems are indicated for use in generating radiographic images of human anatomy. It has a Solid State X-ray Imaging system intended to replace radiographic film/screen systems in all generalpurpose diagnostic procedures. The Vidar Vision 3000 and Vidar Vision 4000 is not indicated for diagnostic X-ray mammography.
- 4. Description of the Device: The VIDAR VISION 3000 Digital Radiography system is an affordable DR (Digital Radiographic) solution designed for smaller radiography environments, including imaging centers, orthopedic practices, small hospitals, and outpatient areas of hospitals. The VIDAR VISION 3000 is the first truly affordable, full-featured DR system. providing a total solution for all digital radiographic examinations. The VIDAR VISION 3000 features high image quality, rapid image acquisition, improved productivity, and ease of use. The flexible system features a 3K detector for the high-resolution imaging needs of busy radiology centers, orthopedic practices, and hospitals. VIDAR's new VISION DR product line also includes the VISION 4000 system, with a 4K detector that is ideal for higher volume settings.
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench, user, and standards testing indicates that the new device is as safe and effective as the predicate devices.
{1}------------------------------------------------
- 6. Substantial Equivalence Chart, Vidar Vision 3000 and Vidar Vision 4000
:
Y
| Characteristic | Sedecal URS X-Ray<br>Units with Digital<br>Detector<br>K042876 | Vidar Vision 3000<br>and Vidar Vision<br>4000<br>Combines two<br>cleared devices:<br>K070563 AND<br>K012546 |
|--------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | General purpose<br>diagnostic X-ray unit | SAME |
| User Interface | Depends on Control<br>Console option chosen.<br>Mainly dedicated touch<br>controls | Software Driven Touch<br>Panel LCD, + remote<br>control unit + remote<br>console |
| Maximum output | Depends on model of<br>generator chosen.<br>Models available from<br>30 kW to 64 kW | 50, 64, and 80 kW |
| Image Acquisition | Digital: CANON<br>CXDI-50G.<br>K031447 | Digital: Vidar Digital<br>Radiographic Detector<br>K070563 |
| Digital Panel Size | Up to 14" x 17" active<br>area | Vision 3000/4000:<br>Active image size: 16 x<br>16 inches (40cm x 40cm) |
| Digital Resolution | 160 x 160 microns<br>pixel pitch, with<br>approximately 6<br>million pixels | Vision 3000:<br>160 Micron.<br>3056 x 3056 (9<br>megapixels)<br><br>Vision 4000:<br>120 Micron<br>4096 x 4096 (16<br>megapixels) |
| Method of Control | Dedicated push button<br>Controls | Software Driven Touch<br>Panel LCD |
| Collimator | Manual R302/A | Balco B302I/A DHHS |
7. Conclusion After analyzing bench, user, and standards testing data, it is the conclusion of Vidar Systems that the Vidar Vision 3000 and Vidar Vision 4000 X-Ray Systems are as safe and effective as the predicate devices, have tew technological differences, and have no new indications for use, thus rendering them substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo includes a symbol with three blue lines and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in blue font. The symbol is on the left side of the text.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Vidar Systems Corporation % Mr. Daniel Kamm Principal Consultant Kamm & Associates P.O. Box 7007 DEERFIELD IL 60015
AUG 2 3 2013
Re: K071193
Trade/Device Name: Vidar Vision 3000 and Vidar Vision 4000 X-Ray Systems Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: April 24. 2007 Received: April 30, 2007
Dear Mr. Kamm:
This letter corrects our substantially equivalent letter of May 30, 2007.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K071193
Device Name: Vidar Vision 3000 and Vidar Vision 4000 X-Ray Systems
Indications For Use:
Vidar Vision 3000 and Vidar Vision 4000 X-Ray Systems are indicated for use in generating radiographic images of human anatomy. It has a Solid State X-ray Imaging system intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures. The Vidar Vision 3000 and Vidar Vision 4000 is not indicated for diagnostic X-ray mammography.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hul
(Division Sigh-Off)
Division of Reproductive, Abdominal, and
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Page I of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.