Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal-OB/GYN, Abdominal, Small Organs (breast, thyroid, testicle), Trans-Vaginal, Trans-Rectal. Typical examinations performed using the system are: General abdominal and pelvic studies including organ surveys, assessment, and retroperitoneal cavity studies. Study of small parts and superficial structures including breasts, shoulders, thyroid, and the abdominal wall. First, second and third trimester pregnancy studies. Prostate, bladder and rectum visualization.
Device Story
Mobile black and white diagnostic ultrasound imaging system; utilizes piezoelectric transducers to transmit ultrasound waves into body tissue; receives reflected echoes; converts echoes to electrical signals; processes signals to display 2D and M-mode images on a monitor. Supports linear and convex probes (3.5 MHz to 7.5 MHz). Used in clinical settings for visualization and assessment of organs, small parts, and pregnancy. Provides high-resolution imaging and measurement functions. Healthcare providers use the displayed images to perform diagnostic assessments and clinical decision-making. Benefits include portable diagnostic imaging capabilities.
Clinical Evidence
Bench testing only. No clinical data required.
Technological Characteristics
Mobile ultrasound system; piezoelectric transducer sensing; B, B/B, B/M, and M operating modes; 3.5-7.5 MHz frequency range. Conforms to EN60601-1 (safety), EN60601-1-2 (EMC), EN60601-1-4 (programmable electric), EN60601-2-37 (ultrasonic safety), EN61157 (acoustic output), and ISO 14971 (risk management).
Indications for Use
Indicated for fetal, abdominal, small organ (breast, thyroid, testes), transrectal, transvaginal, neonatal cephalic, peripheral vascular, and superficial structure ultrasound imaging in adult and pediatric populations.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
SonoAce 600 Diagnostic Ultrasound System (K000030)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) Summary
JUL - 3 2007
This summary of 510(k) safety and effectives information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _ Ko71156
Submitter / Distributor:
Sounmed, Inc. 6800 N.W. 77th Court Miami, FI 33166 Establishment Registration # 3004483577
Telephone: 305-477-0986 Fax: 305-477-5351
- Contact Person: .
Jhoana Lores Email: jhoana@sounmed.com
- Manufacturer: .
Xuzhou Kaixin Electronic Instrument Co., LTD Kaixin Mansion C-01, Economic Development Zone 221004 Xuzhou, Jiangsu REPUBLIC OF CHINA
- Date Prepared: .
April 20, 2007
# Name of the device:
- Sounmed SD-2000 Plus Portable Ultrasound scanner Trade/Proprietary Name: . and Transducers.
- Ultrasonic Imaging Systems and Transducers. . Common Name:
- . Classification:
Regulatory Class: II
FR Number Product Code 90-IYO 892.1560 Ultrasonic pulsed echo imaging system Diagnostic ultrasonic transducer 892.1570 90-ITX
# Legally Marketed Predicate Device:
SonoAce 600 Diagnostic Ultrasound System. K000030
{1}------------------------------------------------
## Device Description:
The Sounmed SD-2000 Plus (KX2000G) is a black and white mobile diagnostic ultrasound imaging system. This device is designed to project ultrasound waves into body tissue and to present the returned echo information on the monitor. Operating Modes of SD-2000 Plus (KX2000G) are B, B/B, B/M, and M .
The Sounmed SD-2000 Plus (KX2000G) supports a selection of Linear and Convex probes for wide variety of application. It is an ultrasound scanner, which provides high resolution, high penetration, and various measurement functions. Probes are supported in frequencies from 3.5 MHz to 7.5 MHz.
All transducers used with SD the SD-2000 Plus ultrasonic diagnostic imaging system are track 1
## Statement of Intended Use:
- . Fetal-OB/GYN
- . Abdominal
- . Small Organs (breast, thyroid, testicle)
- Trans-Vaqinal .
- . Trans-Rectal
Typical examinations performed using the system are:
- . General abdominal and pelvic studies including organ surveys, assessment, and retroperitoneal cavity studies.
- . Study of small parts and superficial structures including breasts, shoulders, thyroid, and the abdominal wall.
- . First, second and third trimester pregnancy studies.
- . Prostate, bladder and rectum visualization.
# Comparison of Technological Characteristics:
The Sounmed SD-2000 Plus (KX2000G) ultrasonic imaging system with added transducers operates identical to the predicate device in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as a 2D and M-mode images.
{2}------------------------------------------------
#### Applicable Standards:
The Sounmed SD-2000 Plus (KX2000G) Portable Ultrasound Device with added transducers conforms to the following standards:
EN60601-1, Medical electrical equipment. General requirements for safety
EN60601-1-2, Medical electrical equipment. General requirements for safety. Collateral standard. Electromagnetic compatibility.
EN60601-1-4, Medical electrical equipment. General requirements for safety. Collateral standard. General requirements for programmable electric.
EN60601-2-37, Medical electrical equipment. Particular requirements for the safety of ultrasonic diagnostic and monitoring equipment.
EN61157, Requirements for the declaration of the acoustic output of medical diagnostic ultrasonic equipment
EN980, medical device labeling standard for CE mark compliance
ISO 14971, Medical devices - Application of risk management to medical devices
ISO 13485:2003, Medical devices - Quality management systems
ISO 9001:2000, Quality management systems
CE Mark 93/42/EEC
Clinical Tests:
Not Required
#### Conclusion:
The conclusions drawn from testing of the The Sounmed SD-2000 Plus (KX2000G) ultrasonic imaging system with added transducers demonstrates that the device is as safe, as effective as well as the legally marketed predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes representing the three branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL - 3 2007
Ms. Jhoana Lores Regulatory Affairs Soundmed, Inc. 6800 NW 77th Court MIAMI FL 33166
Re: K071156
Trade Name: Soundmed SD-2000 Plus Portable Ultrasound Scanner Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYO and ITX Dated: April 23, 2007 Received: April 26, 2007
Dear Ms. Lores:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Soundmed SD-2000 Plus Portable Ultrasound Scanner, as described in your premarket notification:
#### Transducer Model Number
# 3.5MHz Convex Array (3.5C60A2) 6.5MHz Endocavity (6.5C13B2) 7.5MHz Linear Array (7.5L40B2)
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21
{4}------------------------------------------------
Page 2 - Ms. Lores
CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to first device, you submit a postclearance special report. This report should contain complete information. including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Andrew Kang at (240) 276-3666.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
510(k) Number : Device Name :
Sounmed SD-2000 Plus Portable Ultrasound Scanner
Diagnostic ultrasound imaging or fluid flowd analyisis of the human body as follows: Intended Use:
| | Mode of Operation | | | | | | | | |
|--------------------------|-------------------|---|---|-----|-----|------------------|------------------------------|-------------------|-------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(B-M) | Other<br>(specify |
| Ophtalmic | | | | | | | | | |
| Fetal | | N | N | | | | | N | |
| Abdominal | | N | N | | | | | N | |
| Intraoperative (specify) | | | | | | | | | |
| Intraoperative | | | | | | | | | |
| Neurological | | | | | | | | | |
| Pediatric | | | | | | | | | |
| Small Organ (specify) | | N | N | | | | | N | Note 1 |
| Neonatal Cephalic | | N | N | | | | | N | |
| Adult Cephalic | | | | | | | | | |
| Cardiac | | | | | | | | | |
| Trasesophageal | | | | | | | | | |
| Transrectal | | N | N | | | | | N | |
| Transvaginal | | N | N | | | | | N | |
| Transurethral | | | | | | | | | |
| Intravascular | | | | | | | | | |
| Peripheral Vascular | | N | N | | | | | N | |
| Laparoscopic | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | |
| Conventional | | | | | | | | | |
| Musculo-skeletal | | | | | | | | | |
| Superficial | | N | N | | | | | N | |
| Other (specify) | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E Note 1: Small Organ: breast, thyroid, testes.
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign Division of Reproductive. Abo and Radiological Devices 510(k) Number
Page 1 of 4
{6}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form 07115
510(k) Number : Device Name : Transducer:
Sounmed SD-2000 Plus Portable Ultrasound scanner 3.5Mhz Convex Array Probe Model : 3.5C60A2
ర
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|-----|-----|---------------|------------------------|------------------|------------------------|------------------|-----------------|-----------|--|--|--|--|--|--|--|--|--|-------|--|---|---|--|--|--|--|---|--|-----------|--|---|---|--|--|--|--|---|--|--------------------------|--|--|--|--|--|--|--|--|--|----------------|--|--|--|--|--|--|--|--|--|--------------|--|--|--|--|--|--|--|--|--|-----------|--|--|--|--|--|--|--|--|--|-----------------------|--|--|--|--|--|--|--|--|--|-------------------|--|--|--|--|--|--|--|--|--|----------------|--|--|--|--|--|--|--|--|--|---------|--|--|--|--|--|--|--|--|--|----------------|--|--|--|--|--|--|--|--|--|-------------|--|--|--|--|--|--|--|--|--|--------------|--|--|--|--|--|--|--|--|--|---------------|--|--|--|--|--|--|--|--|--|---------------|--|--|--|--|--|--|--|--|--|---------------------|--|--|--|--|--|--|--|--|--|--------------|--|--|--|--|--|--|--|--|--|------------------|--|--|--|--|--|--|--|--|--|--------------|--|--|--|--|--|--|--|--|--|------------------|--|--|--|--|--|--|--|--|--|-------------|--|--|--|--|--|--|--|--|--|-----------------|--|--|--|--|--|--|--|--|--|
| | <table><tr><th></th><th>A</th><th>B</th><th>M</th><th>PWD</th><th>CWD</th><th>Color Doppler</th><th>Color Velocity Imaging</th><th>Combined (B - M)</th><th>Other (specify)</th></tr><tr><th>Ophtalmic</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Fetal</th><td></td><td>N</td><td>N</td><td></td><td></td><td></td><td></td><td>N</td><td></td></tr><tr><th>Abdominal</th><td></td><td>N</td><td>N</td><td></td><td></td><td></td><td></td><td>N</td><td></td></tr><tr><th>Intraoperative (specify)</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Intraoperative</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Neurological</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Pediatric</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Small Organ (specify)</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Neonatal Cephalic</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Adult Cephalic</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Cardiac</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Trasesophageal</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Transrectal</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Transvaginal</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Transurethral</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Intravascular</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Peripheral Vascular</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Laparoscopic</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Musculo-skeletal</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Conventional</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Musculo-skeletal</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Superficial</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr><tr><th>Other (specify)</th><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td><td></td></tr></table> | | A | B | M | PWD | CWD | Color Doppler | Color Velocity Imaging | Combined (B - M) | Other (specify) | Ophtalmic | | | | | | | | | | Fetal | | N | N | | | | | N | | Abdominal | | N | N | | | | | N | | Intraoperative (specify) | | | | | | | | | | Intraoperative | | | | | | | | | | Neurological | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Cardiac | | | | | | | | | | Trasesophageal | | | | | | | | | | Transrectal | | | | | | | | | | Transvaginal | | | | | | | | | | Transurethral | | | | | | | | | | Intravascular | | | | | | | | | | Peripheral Vascular | | | | | | | | | | Laparoscopic | | | | | | | | | | Musculo-skeletal | | | | | | | | | | Conventional | | | | | | | | | | Musculo-skeletal | | | | | | | | | | Superficial | | | | | | | | | | Other (specify) | | | | | | | | | |
| | A | B | M | PWD | CWD | Color Doppler | Color Velocity Imaging | Combined (B - M) | Other (specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Ophtalmic | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Fetal | | N | N | | | | | N | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Abdominal | | N | N | | | | | N | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Neurological | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Trasesophageal | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Transrectal |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.