AEMED STIMPAD ACU-TENS TRANSCUTANEOUS ELECTRICAL NERVE STIMULATOR, MODEL 021999
K071120 · Aemed, Inc. · GZJ · Mar 13, 2008 · Neurology
Device Facts
Record ID
K071120
Device Name
AEMED STIMPAD ACU-TENS TRANSCUTANEOUS ELECTRICAL NERVE STIMULATOR, MODEL 021999
Applicant
Aemed, Inc.
Product Code
GZJ · Neurology
Decision Date
Mar 13, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The StimPad™ System is a transcutaneous electrical nerve stimulator (TENS) device intended for the delivery of electrical stimulation applied through the surface of patient's skin for the treatment of chronic intractable pain, and post-traumatic acute pain. The StimPad™ can be used for self treatment as well as for the treatment of others. It is a prescription use device.
Device Story
StimPad™ is a portable, battery-operated, non-invasive TENS device for pain relief. It features self-contained electrodes snap-fitted to electronic components, eliminating separate wires or pads. Powered by a non-replaceable 3V lithium battery. Device is placed directly at the site of pain. Pre-programmed settings include 7.1 Hz frequency, 47ms pulse width, and a fixed treatment duration of 4 seconds or less. User selects between low, medium, and high intensity levels. Operated by patients for self-treatment or by others. Device ceases function upon battery depletion and is disposed of. Output performance characteristics (voltage, current, pulse frequency/width) are consistent with predicate TENS devices.
Clinical Evidence
Bench testing only. No clinical data presented. Evidence consists of electrical safety and electromagnetic compatibility testing per IEC 60601-1:1988 and IEC 60601-1-2:2001, alongside design verification testing to confirm electrical output performance consistency with predicate devices.
Technological Characteristics
Portable, hand-held, battery-operated TENS device. Powered by a 3V lithium battery. Features integrated snap-fit electrodes (no external wires). Pre-programmed settings: 7.1 Hz frequency, 47ms pulse width, 4-second max duration. Intensity levels: low, medium, high. Compliance with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for symptomatic relief of chronic intractable pain and post-traumatic acute pain in patients requiring TENS therapy.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
Gemore Technology Company Limited GM3 Series HV (High Voltage) TENS Model GM3X3HV (or equivalent GM323HV) (K032994)
Innovative Designer Products Inc. Solitens Modified (K913522)
{0}------------------------------------------------
K071120
MAR 1 3 2008
# 510(k) Summary AEMED, Inc. StimPad™ TENS System (As required by 21 C.F.R. Parts 807.87(h) and 807.92) Prepared: December 6, 2007
# A. MANUFACTURER
AEMED, Inc. 3715 Victoria Road West Palm Beach, Florida 33411
#### B. CONTACT PERSON
Ralph Jugo Submission Correspondent Senior Technical Specialist Quality First International Limited 6530 SW 49" Terrace South Miami, FL 33155 305-661-7005 Telephone 305-661-7005 Telefax 305-409-7791 Cellular ralph@qualityfirstint.com e-mail address
# C. DEVICE
| Trade Name or Proprietary name: | AEMED StimPad™ Transcutaneous Electrical Nerve Stimulator (TENS) System (Model # 021999) |
|---------------------------------|------------------------------------------------------------------------------------------|
| Common Name: | TENS Device |
| Classification Name: | Transcutaneous electrical nerve stimulator (TENS) for pain relief |
| Proposed Class/Device: | II |
| Product Code: | GZJ (21 CFR 882.5890) |
{1}------------------------------------------------
#### D. INTENDED USE
The StimPad™ System is a transcutaneous electrical nerve stimulator (TENS) device intended for the delivery of electrical stimulation applied through the surface of patient's skin for the treatment of chronic intractable pain, and post-traumatic acute pain. The StimPad™ can be used for self treatment as well as for the treatment of others. It is a prescription use device.
### E. INDICATIONS FOR USE
The StimPad™ System is indicated for the symptomatic relief of chronic intractable pain, and post-traumatic acute pain.
# F. LEGALLY MARKETED PREDICATE DEVICES
The StimPad™ is substantially equivalent to the Gemore Technology Company Limited GM3 Series HV (High Voltage) TENS Model GM3X3HV (or equivalent GM323HV) cleared via 510(k) K032994, and to the Innovative Designer Products Inc. Solitens Modified cleared via 510(k) K913522 and to the Medipoint Microcurrent Therapy device cleared via K982961.
The basic design features of StimPad™ are similar to the predicate devices. The StimPad™ device and each of the predicate devices is a portable hand held battery operated micro current device that delivers electrical current by means of a direct contact point for the purpose of alleviating or relief of pain. The electrical output performance characteristics of the StimPad™ device consisting of current and voltage output, pulse frequency, pulse width, pulse duration, maximum current amplitude, time setting, and waveform are comparable and consistent with those of the predicate devices. The user controls and displays are operated in a comparable manner by the user, and the operating modes are analogous.
# G. DEVICE DESCRIPTION
The StimPad™ is a non-invasive nerve stimulation therapy device that can be placed directly at the site of pain for direct electrical stimulation.
{2}------------------------------------------------
Stimulation is achieved through self-contained electrodes that are snap fit connected to the electronic components. There are no separate wires or electrode pads. The StimPad™ is powered by a self-contained three-volt lithium battery.
The StimPad™ device has pre-programmed settings for frequency (i.e. 7.1 Hz), pulse width [i.e. 47 milliseconds for the low, medium, and high intensity settings at 500 ohms, 2000 ohms, and 10,000 ohms] and the timed setting for treatment duration (4 seconds or less). The treatment duration is limited to four seconds or less of stimulation per application/treatment, and the user is not allowed to modify this setting. The intensity levels can be adjusted between low (stimulation is usually imperceptible), medium (usually the most comfortable), and high (maximum level for comfort).
A three-volt lithium battery powers the StimPad™. The life of the battery varies with the frequency of activation, repetition and ambient temperature conditions. The battery is not replaceable. Once the battery is depleted, the device ceases to function and should be disposed or returned to the manufacturer.
#### TECHNOLOGICAL SUMMARY OF CHARACTERISTICS AND ન. SUBSTANTIAL EQUIVALENCE
The StimPad™ is substantially equivalent to the predicate device for the following reasons:
• It has essentially the same indications for use and intended use as the predicate devices consisting of the Gemore Technology Company Limited GM3 Series HV (High Voltage) TENS Model GM3X3HV (or equivalent GM323HV) cleared via 510(k) K032994, and the Innovative Designer Products Inc. Solitens Modified cleared via 510(k) K913522, and the Medipoint Microcurrent Therapy device cleared via K982961.
· The StimPad™ also uses the same fundamental technology and operating principles as the Gemore Technology Company Limited GM3 Series HV (High Voltage) TENS Model GM3X3HV (or equivalent GM323HV) cleared via 510(k) K032994, and the Innovative Designer Products Inc. Solitens Modified cleared via 510(k) K913522, and the Medipoint Microcurrent Therapy device cleared via K982961. The StimPad™ device and each of the predicate
{3}------------------------------------------------
devices are portable hand held battery operated micro current devices that delivers electrical current by means of a direct contact point for the purpose of alleviating or relief of pain.
· The basic design features of StimPad™ are similar to the predicate devices. The electrical output performance characteristics of the StimPad™ device are comparable and consistent with those of the predicate devices and there is overlap in the performance characteristics of the StimPad™ device with those same performance characteristics of the predicate devices. The user controls and displays are operated in a comparable manner by the user, and the operating modes are analogous.
## I. SUMMARY OF TESTING PERFORMED
As is the case with the predicate device, the StimPad™ device was subjected to electrical safety and electromagnetic compatibility testing and was found to be in compliance with the requirements of the standards IEC 60601-1:1988 and IEC 60601-1-2:2001 respectively. Conformity with the requirements of these main standards supports the claim and demonstrates that the StimPad™ device is substantially equivalent to the predicate devices. Design Verification testing to characterize and assure consistent electrical output performance and comparability to the predicate devices was also successfully performed in association with the software validation testing activities.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized eagle or bird-like symbol with three curved lines forming its body and wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AEMED, Inc. % Quality First International Limited Mr. Ralph Jugo Senior Technical Specialist 6530 SW 49th Terrace South Miami, FL 33155
MAR 1 3 2008
K071120 Re:
Trade/Device Name: AEMED StimPadTM Transcutaneous Electrical Nerve Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ Dated: December 13, 2007 Received: December 14, 2007
Dear Mr. Jugo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Ralph Jugo
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Exhibit 1(a) Revised Indication for Use Page
#### STATEMENT OF INDICATIONS FOR USE
510(k) Number: K071120
Device Name: AEMED StimPad™ TENS System
Indications For Use:
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
The StimPad™ TENS System is indicated for the symptomatic relief of chronic intractable pain, and post-traumatic acute pain.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*Neil N. Ayh for BXM*
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| Prescription Use<br>(Per 21 CFR 801.109) | <u>X</u> |
|------------------------------------------|----------------------|
| 510(k) Number | K071120<br>OR |
| | Over-The-Counter Use |
______________________________________________________________________________________________________________________________________________________________________________
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.