The study compared the performance of the TRU RSV device against the standard reference method (tissue culture) using 625 clinical samples collected from symptomatic patients aged 5 years or less across five laboratory sites.
Clinical/field study; Retrospective/Prospective clinical evaluation against tissue culture reference method; Study Period: 2006-07 season and earlier
Symptomatic patients aged 5 years or less; Sample Size: 625; Number of Sites: 5
Tissue culture
Sensitivity, specificity, and correlation compared to tissue culture
Indications for Use
TRU RSV is a rapid, qualitative, lateral-flow immunoassay for the detection of Respiratory Syncytial Virus (RSV) antigens (fusion protein or nucleoprotein) in human nasal wash, nasopharyngeal aspirate, and nasal and nasopharyngeal swab samples. It is designed to test specimens from symptomatic patients aged 5 years or less. A negative result does not preclude RSV infection. It is recommended that all negative test results be confirmed by cell culture.
Device Story
TRU RSV is a single-use, lateral-flow immunoassay for rapid detection of RSV antigens in respiratory specimens. Input: nasal wash, nasopharyngeal aspirate, or nasal/nasopharyngeal swab samples. Operation: sample mixed with sample diluent in a conjugate tube containing lyophilized colloidal gold-linked monoclonal anti-RSV antibodies; mixture applied to test strip with immobilized capture antibodies. Output: visual pink-red line at test position indicates presence of RSV antigen; control line indicates valid test. Used in clinical laboratories by professionals; no instrumentation required. Results assist physicians in clinical evaluation and determining patient management. Benefits include rapid identification of RSV in symptomatic pediatric patients, aiding in clinical decision-making.
Clinical Evidence
Clinical study evaluated 625 samples from symptomatic patients aged ≤5 years across five sites. Performance compared to tissue culture reference method. Overall sensitivity for fresh nasal wash/aspirate was 88.9% (95% CI: 79.3-95.1%) and specificity 81.1% (95% CI: 73.8-88.4%). Fresh swab sensitivity was 92.3% (95% CI: 64.0-99.8%) and specificity 93.5% (95% CI: 87.0-97.3%). Discrepant results were further analyzed via DSFA or PCR.
Technological Characteristics
Lateral-flow immunoassay; nitrocellulose membrane; colloidal gold-linked murine monoclonal antibodies (anti-RSV fusion/nucleoprotein). Single-use format. Components: conjugate tube, test strip, sample diluent. Visual readout via pink-red lines. No external power source. Manual operation.
Indications for Use
Indicated for the qualitative detection of RSV antigens (fusion or nucleoprotein) in nasal wash, nasopharyngeal aspirate, nasal swab, and nasopharyngeal swab specimens from symptomatic patients aged 5 years or younger. Negative results do not rule out infection; confirmation by cell culture is recommended.
Regulatory Classification
Identification
Poliovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to poliovirus in serum. Additionally, some of these reagents consist of poliovirus antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify polioviruses from clinical specimens or from tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of poliomyelitis (polio) and provides epidemiological information on this disease. Poliomyelitis is an acute infectious disease which in its serious form affects the central nervous system resulting in atrophy (wasting away) of groups of muscles, ending in contraction and permanent deformity.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K071101
B. Purpose for Submission:
This device is a modification of ImmunoCard STAT! RSV PLUS (K041445).
C. Measurand:
Respiratory Syncytial Virus antigens
D. Type of Test:
Rapid, qualitative, lateral-flow immunoassay for the detection of Respiratory Syncytial Virus (RSV) antigens (fusion protein or nucleoprotein) in human nasal wash, nasopharyngeal aspirate and nasal and nasopharyngeal swab samples.
E. Applicant:
Meridian Bioscience, Inc.
F. Proprietary and Established Names:
TRU RSV
G. Regulatory Information:
1. Regulation section:
21 CFR section 866.3480, Respiratory syncytial virus serological reagents.
2. Classification:
Class I
3. Product code:
GQG
4. Panel:
Microbiology (83)
{1}
H. Intended Use:
1. Intended use(s):
TRU RSV is a rapid, qualitative, lateral-flow immunoassay for the detection of Respiratory Syncytial Virus (RSV) antigens (fusion protein or nucleoprotein) in human nasal wash, nasopharyngeal aspirate, and nasal and nasopharyngeal swab samples. It is designed to test specimens from symptomatic patients aged 5 years or less. A negative result does not preclude RSV infection. It is recommended that all negative test results be confirmed by cell culture.
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
For prescription use.
4. Special instrument requirements:
Not applicable.
I. Device Description:
TRU RSV is a single use capture immunoassay to detect RSV antigen in human respiratory samples. The test consists of a Conjugate Tube, a Test Strip and Sample Diluent. The Conjugate Tube contains a lyophilized bead of colloidal gold-linked monoclonal antibodies to RSV fusion protein and nucleoprotein (detector antibodies). Test Strip carries a nitrocellulose membrane with dried capture antibodies placed at a designated Test Line for RSV. The Test Strip holder caps the Conjugate Tube during testing and subsequent disposal to reduce exposure to potential pathogens.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ImmunoCard STAT! RSV PLUS.
2. Predicate K number(s):
(K041445)
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | TRU RSV | ImmunoCard STAT! RSV PLUS |
| Device Type | | |
| Technology | Single use, rapid, lateral flow immunoassay | Single use, rapid, lateral flow immunoassay |
| In vitro diagnostic device | Yes | Yes |
| Control | Excludes external control reagent (purchased separately) | Excludes external control reagent (purchased separately) |
| Calibrator | No | No |
| Intended Use | | |
| Detection of RSV antigens | Yes | Yes |
| Screening test | No | No |
| Diagnostic test | Yes | Yes |
| Identification test | No | No |
| Monitoring therapy | No | No |
| Acceptable Samples | | |
| Swab -- Nasal | Yes | Yes |
| Swab -- Nasopharyngeal | Yes | Yes |
| Wash -- Nasal | Yes | Yes |
| Aspirate -- Nasopharyngeal | Yes | Yes |
| Reagents/Components | | |
| Nitrocellulose test strip | Yes | Yes |
| Conjugate reagent | Yes | Yes |
| Reading Guide | Yes | Yes |
| Sample Diluent/Negative Control (external) | Yes | Yes |
| Internal procedural control | Yes | Yes |
| External positive control | No (purchased separately) | No (purchased separately) |
| Similarities (continued) | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | TRU RSV | ImmunoCard STAT! RSV PLUS |
| Source of RSV antibodies | Capture: Murine monoclonal anti-RSV (antibodies to fusion protein and nucleoprotein)Detector: Murine monoclonal anti-RSV (antibodies to fusion protein and nucleoprotein) | Capture: Murine monoclonal anti-RSV (antibodies to fusion and internal proteins)Detector: Murine monoclonal anti-RSV (antibodies to fusion and internal proteins |
| Assay steps | | |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | TRU RSV | ImmunoCard STAT! RSV PLUS |
| Reagents/Components Provided | | |
| Nitrocellulose test strip | Attached to plastic holder/tube closure | Enclosed in plastic frame |
| Conjugate reagent | Supplied as dried bead in Conjugate Tube | Supplied in conjugate pad attached to test strip |
| Reading Guide | Part of plastic holder/tube closure | Part of plastic frame |
| Test Device | Nitrocellulose membrane with immobilized capture antibody. Top end is inserted into plastic frame or holder. Conjugate Tube containing antibody- colloidal gold conjugate (lyophilized bead) | Test Card with nitrocellulose membrane with immobilized capture antibody, conjugate pad with colloidal gold particle-linked detector antibody, plastic frame with reading/reaction window and sample port. |
| Level of skill required | Moderate | CLIA Waived |
{4}
5
| Assay steps | 1. Add 100 μL Sample Diluent to the Conjugate Tube.
2. Add 100 μL sample to the Conjugate Tube and mix.
3. Insert Test Strip to Conjugate Tube.
4. Press down on cap of Test Strip to seal Conjugate Tube.
5. Incubate 15 min, 20-25 C.
6. Read at end of incubation using guide on holder. | 1. Add 4 drops Sample Diluent to a test tube.
2. Add 150 μL sample and mix.
3. Add 150 μL diluted specimen to Test Device.
4. Incubate 15 min, 20-25 C.
5. Read at end of incubation using guide at reaction window. |
| --- | --- | --- |
K. Standard/Guidance Document Referenced (if applicable):
Not applicable.
L. Test Principle:
The conjugate bead is first re-hydrated in the Conjugate Tube with Sample Diluent. Patient sample is then added, the contents mixed and the Test Strip added. If RSV antigens are present, they first bind to the monoclonal antibody-colloidal gold conjugate. When the sample migrates up the Test Strip to the Test Line, the antigen-conjugate complex is bound to the capture antibody, yielding a pink-red line. When no antigen is present, no complexes are formed and no pink-red line appears at the Test Line. An internal control line helps determine whether adequate flow has occurred through the Test Strip during a test run. A visible pink-red line at the Control position of the Test Strip should be present each time a specimen or control is tested. If no pink-red control line is seen, the test is considered invalid.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
i. Assay precision, intra-assay variability and inter-assay variability were assessed with a reference panel prepared from pools of negative samples spiked with specific virus. The reproducibility panel consisted of high positive (n=2), low negative (n=2), and low positive (n=3) and high negative specimens (n=3). The latter were prepared near the assay limit of sensitivity. Each sample was evaluated twice per day for three consecutive days by three different laboratories.
{5}
ii. Reproducibility was 100% with no intra-assay and inter-assay variability for samples prepared above or below the limit of analytical sensitivity.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Not applicable.
d. Detection limit:
The analytical sensitivity of this assay was established in tests with dilutions of RSV strains VR-26 (RSV A), VR-1302 (RSV A), VR-1540 (RSV A), VR-955 (RSV B), VR-1400 (RSV B) and VR-1401 (RSV B). The lower limit of detection is dependent on factors such as the cell culture lines used, the number of passages performed and the efficiencies of the isolation methods. For this reason, assay limit of detection levels may vary if other strains or samples are tested.
| Strain ID | Strain Type | Limit of Detection (LOD) TCID_{50}/ml |
| --- | --- | --- |
| VR-26 | A | 2.49 x 10^{2} |
| VR-1302 | A | 4.47 |
| VR-1540 | A | 5.52 x 10^{1} |
| VR-955 | B | 4.47 |
| VR-1400 | B | 1.10 x 10^{1} |
| VR-1401 | B | 2.47 |
e. Analytical specificity:
The specificity of TRU RSV was tested utilizing the following bacterial, viral and yeast strains. Adenovirus Types 1, 5 and 7A, Coxsackie Type A9, Human Coronavirus Types 229E and OC43, Cytomegalovirus, Influenza A (2 strains), Influenza B (1 strain), Human metapneumovirus, Measles, Parainfluenza Types 1, 2 and 3, Rhinovirus Type 39, Bacillus cereus, Bacillus subtilis, Bordetella parapertussis, Bordetella pertussis, Branhamella catarrhalis, Candida albicans, Candida glabrata, Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Haemophilus influenzae, Klebsiella oxytoca, Klebsiella pneumoniae, Listeria monocytogenes, Legionella pneumophila, Neisseria cinerea, Neisseria gonorrhoeae, Neisseria meningitidis, Nocardia asteroides, Proteus vulgaris, Pseudomonas aeruginosa, Pseudomonas fluorescens,
6
{6}
Serratia liquifaciens, Staphylococcus aureus, Staphylococcus aureus (Cowan I), Staphylococcus epidermidis, Streptococcus (not typed), Streptococcus Groups A, B, D, F, and G, Streptococcus pneumoniae.
RSV positive and negative respiratory specimens were spiked with ≥ 4 × 10⁷/mL bacteria or yeast. Virus inoculations were performed at ≥ 6.7 × 10⁴ TCID₅₀/mL. None of the microorganisms tested yielded a positive result in the RSV-negative sample or interfered with detection of the RSV-positive sample. The RSV-negative respiratory sample was positive when spiked with RSV strain VR-26.
A clinical sample containing Epstein Barr virus at 2.32 × 10⁸ genome equivalents/mL was non-reactive with TRU RSV.
f. Assay cut-off:
Not applicable.
g. Interfering Substances:
The following substances, when introduced directly into nasal samples, do not interfere with testing at the concentrations identified: Acetaminophen (10 mg/mL), Acetylsalicylic acid (20 mg/mL), Albuterol (9.1% v/v), Halls® Throat Drops (20 mg/mL), Ludens® Throat Drops (20 mg/mL), Chlorpheniramine maleate (1.7 mg/mL), Clemastine fumarate (5 mg/mL), Diphenhydramine HCl (5 mg/mL), Dextromethorphan (9.1% v/v), Naproxen sodium (10 mg/mL), Phenylephrine hydrochloride (9.1% v/v), Oxymetazoline (9.1% v/v), Guaifenesin (9.1% v/v), Pseudoephedrine HCl (20 mg/mL), Listerine® Mouthwash (9.1% v/v).
Whole blood at concentrations greater than 2.9% interfered with test interpretation. Chlorpheniramine maleate at concentrations greater than 1.7 mg/mL may cause false-positive test results.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
{7}
8
3. Clinical studies:
a. b. Clinical Sensitivity and specificity
Clinical studies were performed during the 2006-07 season to evaluate the performance of TRU RSV against tissue culture in the laboratory setting. Four independent laboratories (in three different geographic regions of the US) and the manufacturer's laboratory tested a total of 625 samples submitted for RSV testing. 304 of the samples were tested prospectively, while 321 were tested as frozen/thawed samples. Samples were equally distributed between male and female patients.
| Fresh Wash/Aspirate | TRU RSV | | |
| --- | --- | --- | --- |
| Tissue Culture | Positive | Negative | Total |
| Positive | 64 | 8 | 72 |
| Negative | 21 | 90 | 111 |
| Total | 85 | 98 | 183 |
| | | | 95% CI |
| Sensitivity | 64/72 | 88.9% | 79.3-95.1% |
| Specificity | 90/111 | 81.1% | 73.8-88.4% |
| Correlation | 154/183 | 84.2% | 78.9-89.4% |
| Fresh Swab | TRU RSV | | |
| --- | --- | --- | --- |
| Tissue Culture | Positive | Negative | Total |
| Positive | 12 | 1 | 13 |
| Negative | 7 | 100 | 107 |
| Total | 19 | 101 | 120 |
| | | | 95% CI |
| Sensitivity | 12/13 | 92.3% | 64.0-99.8% |
| Specificity | 100/107 | 93.5% | 87.0-97.3% |
| Correlation | 112/120 | 93.3% | 87.3-97.1% |
| Frozen Wash/ Aspirate | TRU RSV | | |
| --- | --- | --- | --- |
| Tissue Culture | Positive | Negative | Total |
| Positive | 79 | 9 | 88 |
| Negative | 12 | 149 | 161 |
| Total | 91 | 158 | 249 |
| | | | 95% CI |
| Sensitivity | 79/88 | 89.8% | 81.5-95.2% |
| Specificity | 149/161 | 92.5% | 87.3-96.1% |
| Correlation | 228/249 | 91.6% | 87.4-94.7% |
{8}
| Frozen Swab | TRU RSV | | |
| --- | --- | --- | --- |
| Tissue Culture | Positive | Negative | Total |
| Positive | 33 | 13 | 46 |
| Negative | 1 | 25 | 26 |
| Total | 34 | 38 | 72 |
| | | | 95% CI |
| Sensitivity | 33/46 | 71.7% | 56.5-84.0% |
| Specificity | 25/26 | 96.2% | 80.4-99.9% |
| Correlation | 58/72 | 80.6% | 69.5-88.9% |
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
Labeling is sufficient and it satisfies the requirements of 21 CFR section 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.