← Product Code [DWM](/productcode/DWM) · K071095

# ASPIRA PLEURAL DRAINAGE SYSTEM (K071095)

_Bard Access Systems, Inc. · DWM · May 18, 2007 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K071095

## Device Facts

- **Applicant:** Bard Access Systems, Inc.
- **Product Code:** [DWM](/productcode/DWM.md)
- **Decision Date:** May 18, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5050
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The AspiraTM Pleural Drainage System is intended for long-term intermittent drainage of pleural fluid accumulated in the pleural cavity for the purpose of relieving symptoms associated with pleural effusion. The Aspira™ Pleural Drainage System is indicated for intermittent drainage of recurrent and symptomatic pleural effusions. The catheter is intended for long-term access to the pleural cavity in order to relieve symptoms such as dyspnea and chest discomfort associated with malignant pleural effusion and other recurrent effusions. The AspiraTM Drainage Bag is indicated for use only with the Aspira™ Drainage Catheter for intermittent drainage. The AspiraTM Dressing Kit is indicated for dressing of a catheter and exit site. The AspiraTM Luer Adapter is intended to provide access to the Aspira™ Drainage Catheter. It is used to drain fluid using standard wall suction, syringe or other appropriate method. The Aspira™ Valve Assembly attaches to the Aspira™ Drainage Catheter, The Aspira™ Repair Kit is for the repair of the Aspira™ Drainage Catheter and replacement of the Aspira™ Valve.

## Device Story

System provides long-term intermittent drainage of pleural fluid; relieves symptoms of pleural effusion (dyspnea, chest discomfort). Components: indwelling fenestrated silicone catheter, drainage bag, valve assembly, luer adapter, dressing kit. Catheter implanted in chest cavity; patient/caregiver performs drainage at home. Drainage bag collects fluid via gravity; activated by in-line silicone pump. Luer adapter allows connection to wall suction or syringe for maintenance. Valve assembly prevents fluid/air exchange when not in use. System enables home-based symptom management.

## Clinical Evidence

Bench testing only. Verification testing performed according to protocols based on standards including BS EN 1617, BS EN 1618, BS EN ISO 14630, ISO 10993-1/7, ASTM F 640-79, ISO 594-1/2, ISO 8669-2, AAMI ISO 14538, BS EN 550, ISO 11135, ISO 11737-1, and ISO 14971-1.

## Technological Characteristics

Indwelling silicone catheter; fenestrated design. Components include drainage bag with in-line silicone pump, valve assembly, and luer adapter. Materials evaluated per ISO 10993-1. Radiopacity per ASTM F 640-79. Conical fittings per ISO 594-1/2. Sterilization via ethylene oxide (BS EN 550, ISO 11135).

## Regulatory Identification

A patient care suction apparatus is a device used with an intrathoracic catheter to withdraw fluid from the chest during the recovery period following surgery.

## Predicate Devices

- Denver® PLEURX® Pleural Catheter Kit and Denver® PLEURX® Home Drainage Kit ([K971753](/device/K971753.md), [K010642](/device/K010642.md), [K011831](/device/K011831.md), [K051084](/device/K051084.md), [K052436](/device/K052436.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko71095

# Section 5 510(k) Summary 21 CFR 807.92(a)

MAY 1 8 2007

#### 5.1 General Information

| Submitter Name:                  | Bard Access Systems, Inc. (BAS)               |
|----------------------------------|-----------------------------------------------|
|                                  | [Wholly owned Subsidiary of C. R. Bard, Inc.] |
| Address:                         | 5425 West Amelia Earhart Drive                |
|                                  | Salt Lake City, Utah 84116                    |
| Telephone Number:                | (801) 595-0700 ext. 7175                      |
| Fax Number:                      | (801) 595-5425                                |
| Contact Person:                  | JiHyun Kim                                    |
| Date of Preparation:             | January 31, 2007                              |
| Registration Number:             | 1720496                                       |
| Additional Registration Numbers: |                                               |
| C. R. Bard:                      | 2212754                                       |

### 5.2 Subject Device Information

| Device Name:          | Aspira™ Pleural Drainage System         |
|-----------------------|-----------------------------------------|
| Trade Name:           | Aspira™                                 |
| Common/Usual Name:    | Patient Care Suction Apparatus          |
| Classification Name:  | Patient Care Suction Apparatus          |
|                       | 21 CFR 870.5050, Class II               |
|                       | 74 DWM - Patient Care Suction Apparatus |
| Classification Panel: | Cardiovascular devices                  |

### 5.3 Predicate Device Information

| Device Name:          | Denver® PLEURX® Pleural Catheter Kit and Denver®<br>PLEURX® Home Drainage Kit                                                                                                                                           |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name:           | Denver® PLEURX®                                                                                                                                                                                                         |
| Common/Usual Name:    | Patient Care Suction Apparatus                                                                                                                                                                                          |
| Classification Name:  | Patient Care Suction Apparatus<br>21 CFR 870.5050 - Class II<br>74 DWM -Patient Care Suction Apparatus                                                                                                                  |
| Classification Panel: | Cardiovascular devices                                                                                                                                                                                                  |
| 510(k) Clearance:     | K971753, concurrence date June 27, 1997.<br>K010642, concurrence date March 20, 2001<br>K011831, concurrence date June 28, 2001<br>K051084, concurrence date June 13, 2005<br>K052436, concurrence date October 6, 2005 |

# 000012

: 上一篇:

:

:

:

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

---------

{1}------------------------------------------------

#### 5.4 Intended Use

The AspiraTM Pleural Drainage System is intended for long-term intermittent drainage of pleural fluid accumulated in the pleural cavity for the purpose of relieving symptoms associated with pleural effusion.

#### 5.5 Indications for Use

The Aspira™ Pleural Drainage System is indicated for intermittent drainage of recurrent and symptomatic pleural effusions. The catheter is intended for long-term access to the pleural cavity in order to relieve symptoms such as dyspnea and chest discomfort associated with malignant pleural effusion and other recurrent effusions.

The AspiraTM Drainage Bag is indicated for use only with the Aspira™ Drainage Catheter for intermittent drainage.

The AspiraTM Dressing Kit is indicated for dressing of a catheter and exit site.

The Aspira™ Luer Adapter is intended to provide access to the Aspira™ Drainage Catheter. It is used to drain fluid using standard wall suction, syringe or other appropriate method.

The Aspira™ Valve Assembly attaches to the Aspira™ Drainage Catheter, The Aspira™ Repair Kit is for the repair of the Aspira™ Drainage Catheter and replacement of the Aspira™ Valve.

#### 5.6 Device Description

The Aspira™ Pleural Drainage System is designed for long-term intermittent drainage of recurrent and symptomatic pleural effusions. The primary components of the system are the AspiraTM Pleural Drainage Catheter and the Aspira™ Drainage Bag.

The Aspira™ Pleural Drainage Catheter is a long-term indwelling silicone catheter used to drain accumulated fluid from the pleural cavity to relieve symptoms associated with pleural effusion. The fenestrated catheter is implanted in the patient's chest cavity enabling the patient or caregiver to perform intermittent drainage of their pleural effusion at home.

The Aspira™ Drainage Bag is used to collect pleural fluid by gravity. The drainage bag attaches to the implanted catheter and is activated using an in-line silicone pump.

The AspiraTM Luer Adapter is designed to access the AspiraTM Drainage Catheter. The luer adapter is connected to wall suction or a syringe to perform intermittent drainage or catheter maintenance.

The Aspira™ Valve assembly attaches to the proximal end of the Aspira™ Pleural Drainage Catheter to prevent fluid or air exchange through the catheter when not in use.

The Aspira™ Pleural Drainage System provides patients with a convenient method to relieve pleural effusion symptoms at home.

{2}------------------------------------------------

#### 5.7 Technological Comparison to Predicate Device

The technological characteristics of the Aspira™ Pleural Drainage System is substantially equivalent to the predicate device, Denver® PLEURX® Pleural Catheter Kit and Home Drainage Kit, in terms of intended use, application, user population, basic design, performance and labeling.

#### 5.8 510(k) Substantial Equivalence Decision Tree

### New device is compared to Marketed Device?

- > Yes. Aspira™ Pleural Drainage System was compared to the legally marketed predicate device, Denver® PLEURX®.
### Does the new device have the same indication statement and intended use as the predicate?

- > Yes. The Indications for Use and the intended use are the same as that of the predicate with minor verbiage changes.
### Do the differences alter the intended therapeutic/diagnostic/etc. effect (i.e. deciding may consider impact on safety and effectiveness)?

- A No. The differences do not alter the intended use of the device.
### Does the new device have the same technological characteristics (e.g. design, materials, etc.)?

- a Not in all regards. The principles of operation and basic design are the same as the predicate device. The main differences of the Aspira™ Pleural Drainage system hour the Denver® PLEURX® are a pump chamber initiating pleural fluid flow and a drainage bag collecting pleural fluid drained by gravity.
### Could the new characteristics affect safety or effectiveness?

- A Yes. The design changes may affect safety or effectiveness of the device.
### Do the new characteristics raise new types of safety or effectiveness questions?

- No. Safety and effectiveness questions are the same as for the predicate device. A
Do accepted scientific methods exist for assessing effects of the new characteristics? > Yes. The device performance was evaluated based on the following standards.

| BS EN 1617      | Sterile drainage catheters and accessory devices for single use                         |
|-----------------|-----------------------------------------------------------------------------------------|
| BS EN 1618      | Catheters Other Than Intravascular Catheters - Test Methods for Common Properties       |
| BS EN ISO 14630 | Non-Active Surgical Implants – General Requirements                                     |
| ISO 10993-1     | Biological Evaluation of Medical Devices Part 1: Evaluation and Testing                 |
| ISO 10993-7     | Biological Evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization Residuals |
| ASTM F 640-79   | Standard Test methods for Radiopacity of Plastics for Medical Use                       |

{3}------------------------------------------------

| ISO 594-1      | Conical Fittings with 6% (Luer) Taper for Syringes, Needles<br>and Certain Other Medical Equipment - Part 1: General<br>Requirements                            |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 594-2      | Conical Fittings with 6% (Luer) Taper for Syringes, Needles<br>and Certain Other Medical Equipment – Part 2: Lock Fittings                                      |
| ISO 8669-2     | Urine Collection Bags Part 2                                                                                                                                    |
| AAMI ISO 14538 | Biological Evaluation of Medical Devices - Establishment of<br>Permissible Limits for Sterilization and process Residues<br>Using Health -Based Risk Assessment |
| BS EN 550      | Sterilization of Medical Devices - Validation and Routine<br>Control of Ethylene Oxide Sterilization                                                            |
| ISO 11135      | Medical Devices - Validation and Routine Control of<br>Ethylene Oxide Sterilization                                                                             |
| ISO 11737-1    | Sterilization of Medical Devices -- Microbiological Methods -<br>Part 1: Estimation of Population of Microorganisms on<br>Products                              |
| ISO 14971-1    | Medical Devices - Risk Management - Part 1: Application of<br>Risk Analysis                                                                                     |

## Are performance data available to assess effects of new characteristics?

- > Yes. Verification testing was performed according to the protocols developed based on the standards listed above.
### Performance data demonstrate equivalence?

- A Yes. The Aspira™ Pleural Drainage System met performance criteria of the safety and effectiveness tests performed and, based on FDA's decision tree, is substantially equivalent to the predicate device Denver® PLEURX Pleural Catheter Kit and Home Drainage Kit covered by K971753, K010642, K011831, K051084, K052436.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of several curved lines. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 8 2007

Bard Access Systems, Incorporated C/O Mr. Bradley J. Bonnette Responsible Third Party Official CITECH 5200 Butler Pike Plymouth Meeting, Pennsylvania 19462-1298

Re: K071095

Trade/Device Name: Aspira™ Pleural Drainage System Regulation Number: 870.5050 Regulation Name: Patient Care Suction Appartus Regulatory Class: II Product Code: DWM Dated: April 17, 2007 Received: April 18, 2007

Dear Mr. Bonnette:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 -Mr. Bonnette

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sutte y. Michau Om.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Section 4 Indications for Use

510(k) Number (if known):

Device Name: Aspira™ Pleural Drainage System

Indications for Use:

The Aspiral Pleural Drainage System is indicated for intermittent drainage of recurrent and symptomatic pleural effusions. The catheter is intended for long-term access to the pleural cavity in order to relieve symptoms such as dyspnea and chest discomfort associated with malignant pleural effusion and other recurrent effusions.

The Aspira™ Drainage Bag is indicated for use only with the Aspira™ Drainage Catheter for intermittent drainage.

The Aspira™ Dressing Kit is indicated for dressing of a catheter and exit site.

The Aspira™ Luer Adapter is intended to provide access to the Aspira™ Drainage Catheter. It is used to drain fluid using standard wall suction, syringe or other appropriate method.

The Aspira™ Valve Assembly attaches to the Aspira™ Drainage Catheter. The Aspira™ Repair Kit is for the repair of the Aspira™ Drainage Catheter and replacement of the Aspira™ Valve.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ിന്ന Sign-Off)
Nision of Anesthesiology, General Hospital, ് സ്ട്രീഗ് of Affesthoolorogy

510(k) Number: k071093

000010

---

**Source:** [https://fda-staging.innolitics.com/device/K071095](https://fda-staging.innolitics.com/device/K071095)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
