GAME READY PROFESSIONAL THERAPY SYSTEM, MODEL 550100, GAME READY PRE-PROGRAMMED PROFESSIONAL THERAPY SYSTEM
K071050 · Coolsystems, Inc. · IRP · Jun 29, 2007 · Physical Medicine
Device Facts
Record ID
K071050
Device Name
GAME READY PROFESSIONAL THERAPY SYSTEM, MODEL 550100, GAME READY PRE-PROGRAMMED PROFESSIONAL THERAPY SYSTEM
Applicant
Coolsystems, Inc.
Product Code
IRP · Physical Medicine
Decision Date
Jun 29, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Game Ready System combines cold and compression therapies. It is intended to treat post-surgical acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. It is intended to be used by or on the order of healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings.
Device Story
Game Ready System provides intermittent pneumatic compression and cold therapy; utilizes DC-powered control unit, connector hose, and body-specific wraps (heat exchanger + sleeve). Control unit circulates ice water for cooling and pumps air for compression. User adjusts compression, cooling, and treatment duration. Used in hospitals, clinics, athletic settings, or homes; operated by healthcare professionals or patients. System includes safety monitoring: software detects insufficient fluid flow, over-pressure, or under-pressure, triggering alarms or system shutdown to prevent injury. Benefits include reduction of edema, swelling, and pain post-surgery or injury.
Clinical Evidence
No clinical testing data are submitted, referenced, or relied on to determine substantial equivalence. Bench testing was performed at sub-assembly and assembly levels to verify design requirements.
Technological Characteristics
DC-powered system; includes control unit, connector hose, and wraps. Cooling via circulating ice water; compression via intermittent pneumatic pump. Software-controlled safety monitoring for fluid flow and pressure. Available in standard and pre-programmed configurations.
Indications for Use
Indicated for patients with post-surgical or acute injuries requiring cold and compression therapy to reduce edema, swelling, and pain. Suitable for use in clinical, athletic, or home settings under professional guidance.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
## Game Ready System
JUN 2 9 2007
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
- A. Submitter: CoolSystems, Inc. 2201 Dwight Way Berkeley, CA 94704
Phone: 510-868-5378 Fax: 510-984-5330 Email: mbaldwin@gameready.com Contact: Marianne Baldwin, VP, QA/RA/CA
June 11, 2007 Date Prepared:
## B. Device Names:
| Trade Name: | Game Ready System |
|----------------------|----------------------------------------------------------------|
| Common/usual Name: | Cold/compression therapy system |
| Classification Name: | Cold water circulating pack / Powered inflatable tube massager |
## C. Predicate Device:
The Game Ready System is substantially equivalent to the following predicate devices:
| Device, K# | Manufacturer | Product Code(s) |
|-------------------------|------------------|-----------------|
| CRYOPress, K951769 | Grimm Scientific | ILO/IRP |
| Polar Care 500, K961855 | Breg, Inc. | ILO |
### D. Device Description:
The Game Ready System is comprised of the following components:
- . Control Unit, with AC Adapter (to convert line power to DC input power)
- Connector Hose (connects Control Unit to Wrap) .
- Wrap (Heat Exchanger + Sleeve) ●
- . Optional Carry Bag
- · Optional Battery Pack
The Game Ready System is a DC-powered device that provides intermittent compression and controlled cold therapy by controlling fluid flow and air pressure through a Wrap that is specially designed for a specific body part or location (e.g., shoulder, elbow, knee, ankle, back). A Wrap is comprised of a Sleeve (the outer covering for the Heat Exchanger) and a Heat Exchanger. Cooling is provided by
{1}------------------------------------------------
circulating ice water from the Control Unit through the Connector Hose to the Wrap. Compression is provided by an intermittent-cycling pneumatic pump that pumps air from the Control Unit through the Connector Hose to the Wrap. The user can adjust the amount of compression, the amount of cooling, and the treatment time. The Control Unit software is designed to protect both the user and the Control Unit from harm by shutting off the system and sounding an alarm if there is insufficient fluid flow or in case of an over-pressure or under-pressure situation. The software detects other. less-serious error conditions and sounds a warning beep to alert the user, so that the error condition can be corrected.
The Game Ready System is available in two configurations: the Game Ready Professional Therapy System (P/N 550100) and the Pre-Programmed Game Ready Professional Therapy System (P/N 550350)
#### E. Intended Use:
The Game Ready System combines cold and compression therapies. It is intended to treat post-surgical acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. It is intended to be used by or on the order of healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings.
#### F. Comparison with the Predicate Device:
The Game Ready System is substantially equivalent to the predicate devices in design, function, and indications for use/intended use.
#### G. Non-Clinical Testing
Bench testing was performed at both the sub-assembly and assembly level. The testing showed that the Game Ready System met its requirements. Design Verification Testing showed that the Game Ready System performed according to specification.
#### H. Clinical Testing
No clinical testing data are submitted, referenced, or relied on to determine substantial equivalence.
#### -Conclusions
There are no significant differences between the Game Ready System and the predicate devices that would adversely affect the use of the proposed device. The Game Ready System is substantially equivalent to the predicate devices in design, function, and indications for use/intended use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with three lines extending from its head.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
CoolSystems, Inc % Ms. Marianne Baldwin Vice President, RA/QA/CA 2201 Dwight Way Berkeley, California 94704
JUN 2 9 2007
Re: K071050
Trade/Device Name: Game Ready Professional Therapy System, Model 550100, Game Ready Pre-programmed Professional Therapy System Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: II Product Code: IRP, ILO Dated: May 30, 2007 Received: May 31, 2007
Dear Ms. Baldwin
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
## Page 2 - Ms. Marianne Baldwin
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
| 510(k) Number: | K071050 |
|----------------|---------------------------------|
| Device Name: | Game Ready <sup>TM</sup> System |
#### Indications for Use:
The Game Ready System combines cold and compression therapies. It is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. It is intended to be used by or on the order of healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.