DATEX-OHMEDA S/5 COMPACT CRITICAL CARE MONITOR WITH L-CICU05 AND L-CICU05A SOFTWARE
K071020 · GE Healthcare · MHX · Jun 18, 2007 · Cardiovascular
Device Facts
Record ID
K071020
Device Name
DATEX-OHMEDA S/5 COMPACT CRITICAL CARE MONITOR WITH L-CICU05 AND L-CICU05A SOFTWARE
Applicant
GE Healthcare
Product Code
MHX · Cardiovascular
Decision Date
Jun 18, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A is intended for multiparameter patient monitoring with optional patient care documentation. The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS), and neurophysiological status of all hospital patients. The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software when using BIS is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents. The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for use by qualified medical personnel only.
Device Story
Modular multiparameter patient monitor; processes inputs from plug-in measurement modules (ECG, invasive/non-invasive blood pressure, pulse oximetry, temperature, EEG). Transforms raw signals into numeric values, traces, and trends; performs arrhythmia analysis, ST-segment calculation, heart rate, impedance/respiration rate, energy expenditure, and EEG spectrum/evoked potential averaging. Operated by clinicians (physicians/nurses) via keyboard and ComWheel™ pointing device; supports optional remote controller and barcode reader. Used in clinical settings; stand-alone or networked configuration for central station viewing and archiving. Provides audible/visual alarms for patient status; assists clinicians in real-time patient monitoring and decision-making. Benefits include continuous, integrated physiological data acquisition and documentation.
Clinical Evidence
Bench testing only. Device verified and validated against multiple international standards including IEC 60601-1, IEC 60601-2-27 (ECG), IEC 60601-2-30 (BP), IEC 60601-2-34 (Invasive BP), IEC 60601-2-40 (EEG), and ISO 9919 (Pulse Oximetry). No clinical trial data presented.
Technological Characteristics
Modular multiparameter monitor; utilizes legacy Datex-Ohmeda M-series or E-series plug-in modules. Connectivity via hardwired or wireless network. Software-based processing for arrhythmia, ST-segment, hemodynamic, respiratory, and EEG analysis. Complies with IEC 60601 series, ISO 9918, ISO 9919, ISO 7767, and EN 1060 standards.
Indications for Use
Indicated for monitoring hemodynamic, respiratory, ventilatory, gastrointestinal/regional perfusion, BIS, and neurophysiological status of all hospital patients. BIS monitoring aids in assessing brain state and anesthetic agent effects. For use by qualified medical personnel only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740)
Submission Summary (Full Text)
{0}------------------------------------------------
# JUN 1 8 2007
# Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5TM Compact Critical Care Monitor with L-CICU05 and L-CICU05A Software
# GENERAL COMPANY INFORMATION as required by 807.92(a)(1)
# COMPANY NAME/ADDRESS/PHONE/FAX:
GE Healthcare 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344
#### NAME OF CONTACT:
Mr. Joel Kent
DATE:
March 30, 2007
# DEVICE NAME as required by 807.92(a)(2)
#### TRADE NAME:
Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A Software
COMMON NAME:
Patient Monitor
#### CLASSIFICATION NAME:
The following Class II classification appears applicable:
| Product Code | Classification Name | CFR Section |
|--------------|-----------------------------|-------------|
| MHX | Arrhythmia detector & alarm | 870.1025 |
| MLD | Monitor ST-segment & alarm | 870.1025 |
| DSF | Paper Chart Recorder | 870.2810 |
{1}------------------------------------------------
### NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
The Datex-Ohmeda S/5TM Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is substantially equivalent to the predicate Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740).
#### DEVICE DESCRIPTION as required by 807.92(a)(4)
The two software options for the S/5™ Compact Critical Care Monitor are identical, except that the L-CICU05A is equipped with extended arrhythmia analysis capability.
The S/5™ Compact Critical Care Monitor with L-CICU05 and L CICU05A uses several types of plug-in measurement modules. The legacy Datex-Ohmeda M-series or E-series measurement modules are used. Modules are the subject of separate 510(k)'s and are not part of this notification.
The S/5TM Compact Critical Care Monitor with L-CICU05 and L-CICU05A is typically furnished with a module that measures ECG, invasive and non-invasive blood pressures, pulse oximetry and temperature. Modules are placed in the S/5™ Monitor frame and are automatically recognized by the monitor. The patient cables are connected to the module plug in jacks and then monitoring can begin.
The S/5TM Compact Critical Care Monitor with L-CICU05 and L-CICU05A can display measurements in the form of numeric values, traces and trends. Audible and visual alarms are used to indicate patient status. The priority profile of an aların depends on the parameter.
The S/5TM Compact Critical Care Monitor with L-CICU05 and L-CICU05A is operated by a keyboard. Typically pressing a key results in a pop up menu appearing on the screen. Selections can then be made easily from the menu using a unique ergonomically designed pointing device on the keyboard called a ComWheel™.
The software L-CICU05 and L-CICU05A perform some module related tasks like arrhythmia analysis, ST values calculation, heart rate calculation, impedance and respiration rate calculation, energy expenditure calculation and EEG spectrum analysis evoked potential response averaging calculations. All the module communication is also handled in the main software.
There are various optional types of keyboards, some are like standard keyboards and another is a hand-held Remote controller (K-CREMCO) which is still directly connected to the S/5™ Compact Critical Care Monitor via a long cord but provides more flexibility in controlling the monitor while the doctor or nurse is handling other patient care needs. To facilitate quick access to menus, a bar code reader is also supported, although the bar code reader is not manufactured anymore. The S/5TM Compact Critical Care Monitor can be in a stand-alone or networked configuration. If networked, measurements are sent to the network for central station or monitor-to-monitor viewing. Trends as well as the patient care documentation can be sent via a network to a central computer for archiving. Networking can be hardwired or wireless.
The S/5 Compact Critical Care Monitor can also be upgraded to L-CICU05(A) software using the S/5 I...F.E. upgrade program that offers a means to continuously keeping products up-to-date, by upgrading modular anesthesia and critical care monitors and network products dating from back to 2000 to the latest S/S software level. Upgrading of modular monitors and network products is performed with one of the available U-LIFE upgrade kits. The kit includes all hardware and software components needed to make the monitor or network product compatible with the latest main software being delivered. The currently available U-LIFE upgrade kit codes for S/5 Compact Critical Care Monitor are:
ULIFE4-00-03 with U-CICU05 and User's Guide ULIFE4-00-03 with U-CICU05A and User's Guide
{2}------------------------------------------------
#### INTENDED USE as required by 807.92(a)(5)
#### Intended Use:
The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A is intended for multiparameter patient monitoring with optional patient care documentation.
#### Indications for use:
The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS), and neurophysiological status of all hospital patients.
The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software when using BIS is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents.
The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for use by qualified medical personnel only.
### SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is substantially equivalent to the predicate Datex-Ohmeda S/5TM Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740). The S/57M Compact Critical Care Monitor with L-CICU05and L-CICU05A is a modular multiparameter patient monitor providing connections to measurement modules. The general construction, intended use of the S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A are the same as for the predicate S/57M Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740). The indications for use for the Datex-Ohmeda S/5TM Compact Critical Care Monitor with L-CICU05 and L-CICU05A software are the same as for the predicate. The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS) and neurophysiological status of all hospital patients. The S/5TM Compact Critical Care Monitor with L-CICU05 and L-CICU05A software when using B1S is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents. The S/5TM Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for use by qualified medical personnel only. Based on the above and a detailed analysis in Tab 5 Comparison and other documentation
included in this 510(k) notification and attachments it is evident that the main features and indications for use of the S/5TM Compact Critical Care Monitor with
L-CICU05 and L-CICU05A software is substantially equivalent to the predicate S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740).
{3}------------------------------------------------
# SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
The Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A Software has been assessed against the standards below and details of conformity are presented in the attached 510(k) notification. The device has been thoroughly tested through validation and verification of specifications.
- IEC 60601-1:1988+ Amdt.:1:1991 + Amdt. 2:1995 .
- EN 60601-1: 1990 + A1:1993+A2:1995+A13:1996 .
- CAN/CSA-C22.2 No.601.1-M90 +S1:1994+Amdt. 2:1998 .
- IEC 60601-2-27:1994/EN 60601-2-27:1994 .
- IEC 60601-2-30:1999/EN 60601-2-30:2000 .
- IEC 60601-2-34:2001/EN 60601-2-34:2000 .
- IEC 60601-2-40:1998 ●
- IEC 60601-2-49:2001 .
- IEC 60601-1-2(2001)/EN 60601-1-2 .
- IEC 60601-1-4: 1996+Amdt. 1:1999/EN 60601-1-4 .
- ISO 9918:1993/EN 864:1996 .
- ISO 9919:1992/EN865:1997 ●
- ISO 7767:1997/EN12598:1999 .
- IEC 601-2-10:1987/EN 60601-2-10:2000 + Amd.1:2001 .
- IEC 60601-2-26:2002/EN60601-2-26 .
- EN 1060-1:1995 / EN-1060-3:1997 .
- . EN 12470-4:1992
- IEC 60068-2 .
- UL 2601-1.1997 .
- ANSI/AAMI ES-1:1993 .
- ANSI/AAMI EC57:1998 .
- FDA 21 CFR 898.12 .
#### CONCLUSION:
The summary above shows that the Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A Software is substantially equivalent in safety and effectiveness to the legally marketed (predicate) Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU02 and L-CICU02A software (K022740).
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird with three curved lines representing its body and wings. The bird is facing to the right. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 1 8 2007
GE Healthcare c/o Mr. Joel C. Kent. RAC Manager, Quality and Regulatory Affairs 86 Pilgrim Road Needham, MA 02492
Re: K071020
Datex-Ohmeda S/5TM Compact Critical Care Monitor with L-CICU05 and CICU05A Software Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: May 21, 2007 Received: May 21, 2007
Dear Mr Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Joel C. Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Blfemmimara for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): _(07 1020
Device Name: Datex-Ohmeda S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A Software.
## Indications for use:
The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated for monitoring of hemodynamic (including arrhythmia and ST-segment analysis), respiratory, ventilatory, gastrointestinal/regional perfusion, Bispectral index (BIS), and neurophysiological status of all hospital patients.
The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software when using BIS is for monitoring the state of the brain by data acquisition and processing of electroencephalograph signals and may be used as an aid in monitoring the effects of certain anesthetic agents.
The S/5™ Compact Critical Care Monitor with L-CICU05 and L-CICU05A software is indicated or use by qualified medical personnel only.
cription Use Over-The-Counter Use AND/OR tt 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) E.Do not write below this line-continue on another page of needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Bhammamo
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K071020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.