K070999 · Rjl Systems, Inc. · MNW · Aug 2, 2007 · Cardiovascular
Device Facts
Record ID
K070999
Device Name
BC (BODY COMPOSITION SOFTWARE)
Applicant
Rjl Systems, Inc.
Product Code
MNW · Cardiovascular
Decision Date
Aug 2, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
BC is a software accessory for RJL Systems Quantum-X, Quantum Desktop, and Quantum-III. It requires a separate personal computer running the Windows Operating System. To automatically retrieve data from the Quantum Desktop and Quantum-III analyzers, the computer must have an available serial port. Indications: Calculation and Historical Tracking of: Actual Impedance . Actual Phase Angle (PA) . . Estimated Body Fat (FAT) . . Estimated Fat Free Mass (FFM) . . Estimated Total Body Water (TBW) . Estimated Intra-Cellular Water (ICW) . . Estimated Extra-Cellular Water (ECW) . Estimated Basal Metabolic Rate (BMR) . . Estimated Daily Energy Expenditure (DEE) . . Actual Body Mass Index (BMI)
Device Story
BC software accessory for RJL Systems bio-impedance analyzers (Quantum-X, Quantum Desktop, Quantum-III); runs on Windows PC. User inputs resistance, reactance, and demographic data (age, height, weight, gender, activity level, frame size). Software stores data in database for historical tracking; applies clinical prediction equations to estimate body composition parameters (fat, water, metabolic rates). Non-diagnostic; provides no medical opinions. Used by clinicians or patients for tracking body composition metrics over time.
Clinical Evidence
No clinical data provided. Substantial equivalence based on comparison of intended use, technological characteristics, and performance of prediction equations derived from clinical literature.
Technological Characteristics
Software accessory for bio-impedance analyzers. Operates on Windows-based PC. Connectivity via serial port for data retrieval. Uses prediction equations derived from clinical studies to estimate body composition. Non-diagnostic.
Indications for Use
Indicated for healthy adults and adolescents aged 9-94. Used for calculation and historical tracking of body composition parameters including impedance, phase angle, body fat, fat-free mass, total body water, intra-cellular water, extra-cellular water, BMR, DEE, and BMI.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
### 510(k) Summary
RJL Systems, Inc. 33939 Harper Avenue Clinton Township, Michigan 58035, USA Phone : 586-790-0200 Fax : 586-790-0205 Homepage : http://www.rjlsystems.com
1) Identification of the Device: Proprietary - Trade Name: BC body composition software Common Name: Body Composition Analyzer Classification Number: 21 CFR 870.2770 Classification Name: Impedance Plethysmograph Product Code: MNW
- 2) Equivalent Legally Marketed Devices: RJL Systems; BIA-103 (K862383 ) Impedimed; DF50 ( K050395 ) BodyStat; QuadScan 4000 ( K002835 )
#### 3) Indications for use
BC is a software accessory for RJL Systems Quantum-X, Quantum Desktop, and Quantum-III. It requires a separate personal computer running the Windows Operating System. To automatically retrieve data from the Quantum Desktop and Quantum-III analyzers, the computer must have an available serial port.
Indications:
Calculation and Historical Tracking of:
- Actual Impedance .
- Actual Phase Angle (PA) .
- . Estimated Body Fat (FAT)
- . Estimated Fat Free Mass (FFM)
- . Estimated Total Body Water (TBW)
- . Estimated Intra-Cellular Water (ICW)
- . Estimated Extra-Cellular Water (ECW)
- Estimated Basal Metabolic Rate (BMR) .
- Estimated Daily Energy Expenditure (DEE) .
- . Actual Body Mass Index (BMI)
AUG - 2 2007
{1}------------------------------------------------
#### Device Description 4)
The BC application is intended to be used as a software accessory to RJL Systems' existing line of Bio-Impedance Analyzers ( K830292 and K862383 ). The user obtains values for the Resistance and Reactance of an individual, and enters these numbers, along with the individual's name, age, height, weight, gender, activity level, frame size, and optionally, subject ID and desired target weight. These values are stored in a database to enable historical tracking, and are then used in a series of prediction equations to estimate the parameters listed above.
For estimating body composition parameters, the BC application relies on prediction equations developed as the result of clinical studies. The user is provided the opportunity to select from several different sets of equations, each assembled from one or more studies. References are provided for every set of equations, except for those contained in the original RJL BIA-103 device software.
The BC application is non-diagnostic in nature and does not express any opinions with regard to any specific disease or medical condition.
| Company<br>Device Name | RJL Systems<br>BC | RJL Systems<br>BIA-103 | Predicate Devices<br>Impedimed<br>DF-50 | BodyStat<br>QuadScan 4000 |
|-----------------------------|-------------------|------------------------|-----------------------------------------|---------------------------|
| 510(k) number | | K862383 | K050395 | K002835 |
| OTC | YES | YES | YES | YES |
| Indication | | | | |
| Impedance | YES | | YES | YES |
| Phase Angle | YES | | YES | |
| Body Fat | YES | YES | YES | YES |
| Fat Free Mass | YES | YES | YES | YES |
| Total Body Water | YES | YES | YES | YES |
| Intra-Cellular Water | YES | | YES | YES |
| Extra-Cellular Water | YES | | YES | YES |
| Body Mass Index | YES | | YES | YES |
| Basal Metabolic Rate | YES | YES | YES | YES * |
| Daily Energy<br>Expenditure | YES | | | YES * |
#### 5) Substantial Equivalence Chart
( * ) The Indications for Use statement for the BodyStat QuadScan 4000 simply states "Metabolic Rates". However, the QuadScan 4000 reports "Basal Metabolism" and "Activity Metabolism". Activity Metabolism is an accepted synonym for "Daily Energy Expenditure".
- 6) Conclusion
The BC application represents an evolution of the software provided with RJL Systems' original BIA-103 device and, when used in conjunction with RJL Systems' existing line of Bio-Impedance Analyzers ( K830292 and K862383 ) is equivalent to the Impedimed DF-50 and BodyStat QuadScan 4000.
The product comparison chart establishes that the BC software application does not raise any new questions concerning safety and effectiveness and is equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The text is in all capital letters and is written in a simple, sans-serif font. The logo is black and white.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
- 2 2007
Mr. Barry Callahan IT Director RJL Systems 33939 Harper Avenue CLINTON TOWNSHIP MI 48035
K070999 Re:
> Trade/Device Name: BC (Body Composition Software) Regulation Number: 21 CFR §870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW Dated: July 17, 2007 Received: July 18, 2007
Dear Mr. Callahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a circular seal with the text "1906-2006" at the top. The letters "FDA" are in the center of the seal in a bold, sans-serif font. Below the letters, the word "Centennial" is written in a cursive font. Three stars are arranged in a horizontal line below the word "Centennial". The text around the perimeter of the seal is difficult to read due to the image quality.
dina and 9 m
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with allthe Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K070999
Device Name: ___BC ( Body Composition Software )___
BC is a software accessory for RJL Systems Quantum-X, Quantum-X, Quantum Desktop, and Quantum-III. It requires a separate personal computer running the Windows Operating System. To automatically retrieve data from the Quantum Desktop and Quantum-III analyzers, the computer must have an available serial port.
#### Indications for Use:
Calculation and Historical Tracking of:
- Actual Impedance .
- Actual Phase Angle (PA) .
- Estimated Body Fat (FAT) .
- Estimated Fat Free Mass (FFM) .
- . Estimated Total Body Water (TBW)
- Estimated Intra-Cellular Water (ICW) ●
- Estimated Extra-Cellular Water (ECW) ●
- Estimated Basal Metabolic Rate (BMR) .
- Estimated Daily Energy Expenditure (DEE) .
- Actual Body Mass Index (BMI) .
BC is intended only for use on normally healthy adults and adolescents age 9-94.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert Remmer
vision of Reproductive, Abdominal and Radiological Devic 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.