K070965 · Alicia International Pty , Ltd. · HHE · Mar 5, 2008 · Obstetrics/Gynecology
Device Facts
Record ID
K070965
Device Name
ALICIA MENSTRUAL CUP
Applicant
Alicia International Pty , Ltd.
Product Code
HHE · Obstetrics/Gynecology
Decision Date
Mar 5, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5400
Device Class
Class 2
Indications for Use
The Alicia™ Menstrual Cup is a receptacle placed in the vagina to collect blood and cellular debris that is extruded from the uterus via the cervix during menstruation. The Alicia'10 Menstrual Cup is placed low enough in the vagina to be retrieved readily and to prevent it's toughing the cervix or interfering with menstrual flow through it.
Device Story
The Alicia™ Menstrual Cup is a reusable, internally worn silicone receptacle designed to collect menstrual flow. It is inserted into the vagina, positioned to avoid contact with the cervix, and can be worn for up to 12 hours. The device features a stem for retrieval. It is available in four sizes (XS, S, M, L) to accommodate different anatomical needs. The device is intended for over-the-counter use by women. It functions as a passive collection vessel, requiring no external power or complex operation. Users manually insert and remove the device, cleaning it between uses. By collecting rather than absorbing fluid, it provides a non-absorbent alternative for menstrual hygiene management.
Clinical Evidence
Bench testing only. Biocompatibility testing was performed in accordance with ISO 10993-1, ISO 10993-5, and ISO 10993-10 standards to ensure safety of the silicone elastomer material.
Technological Characteristics
Material: Silicone elastomer. Form factor: Reusable, internally worn cup with retrieval stem. Available in four sizes (XS, S, M, L). Biocompatibility: Compliant with ISO 10993-1, ISO 10993-5, and ISO 10993-10. Non-powered, passive mechanical device.
Indications for Use
Indicated for use by women during menstruation to collect blood and cellular debris. Available in four sizes: XS (teenagers/small-framed), S (under 30, no childbirth), M (vaginal childbirth/C-section or over 30), and L (larger build, history of childbirth).
Regulatory Classification
Identification
A menstrual cup is a receptacle placed in the vagina to collect menstrual flow.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
K070965
## 510(k) SUMMARY
MAR - 5 2011
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
1. Submitter's Identifications: Establishment: Alicia™ International Pty, Ltd. 84A Skyline Dr, Burleigh Heads, Gold Coast, Queensland, Australia 4220 Operations: Foreigner importer
#### Owner/Operator:
Alicia™ International Pty, Ltd. 84A Skyline Dr, Burleigh Heads, Gold Coast, Queensland, Australia 4220
Date of Summary Preparation: April 2, 2007.
- 2. Name of the Device: Alicia110 Menstrual Cup / 99ALXS, 100ALSM, 101ALMD, 102ALLG
- 3. Information of device classification.
| Device | Cup, Menstrual |
|------------------------------|-----------------------|
| Regulation Description | Menstrual cup |
| Regulation Medical Specialty | Obstetrics/Gynecology |
| Review Panel | Obstetrics/Gynecology |
| Product Code | HHE |
| Regulation Number | 884.5400 |
| Device Class | II |
### 4. Device Description:
The Alicia™ menstrual cup is a soft, small internally worn reusable silicone menstrual cup that holds (instead of absorbing) monthly menstrual flow. It may remain in the body for up to 12 hours. It holds an ounce of fluid. It is available four sizes:
- · L size: 102ALLG; for larger built women that have had children.
- · M size: 101ALMD; for women who have had vaginal childbirth or C-Section; and for women who are over 30 years old and never had childbirth.
- S size: 100ALSM; for women under 30 years old who have never had childbirth or C-Section.
- · XS size: 99ALXS; for teenagers and very small framed women. ned women.
Eonch Eouch
The cup remains entirely within the vagina and does not tough the cervix but the stem remains outside the body to ensure retrieval of the cup. See the user's manual for information about how to use and care for the Alicia™ menstrual cup.
The Alicia™ menstrual cup is manufactured from a soft silicone elastomer. The properties of the silicone are described in a master file at FDA. The master file provides data on all of the biocompatibility or toxicity testing required according to FDA's memorandum to guidance G95-1. shich
Page 1 of 2
{1}------------------------------------------------
Alicia™ International Pty., Ltd.
- · 5. Intended Use: The Alicia™ Menstrual Cup is a receptacle placed in the vagina to collect blood and cellular debris that is extruded from the uterus via the cervix during menstruation. The Alicia'10 Menstrual Cup is placed low enough in the vagina to be retrieved readily and to prevent it's toughing the cervix or interfering with menstrual flow through it.
- 6. Comparison to the 510(k) Cleared Device (Predicate Device): MoonCup® Menstrual Cup (K040335)
- 7. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ISO 10993-1 for biocompatibility as well as the specified testing standard of ISO 10993-5 and ISO 10993-10.
- 8. Conclusions
The Alicia™ Menstrual Cup / Model: 99ALXS, 100ALSM, 101ALMD and 102ALLG have the same intended use and technological characteristics as the cleared devices, Menstrual Cup (K040338). Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device. Therefore, the new device Alicia Menstrual Cup / Model: 99ALXS, 100ALSM. 101ALMD and 102ALLG is considered substantial equivalent to the chosen 510(k) cleared device MoonCup® Menstrual Cup (K040335).
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of an abstract symbol, possibly representing a human figure or a flame.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 5 2008
Alicia International PTY LTD c/o Mr. Tom Stone Official Correspondent PO Box 1471 Oxenford 4210 OLD AUSTRALIA
Re: K070965
> Trade/Device Name: Alicia™ Menstrual Cup / Models 99ALXS, 100ALSM 101ALMD, 102ALLG Regulation Number: 21 CFR 8884.5400 Regulation Name: Menstrual cup
Regulatory Class: II Product Code: HHE Dated: February 21, 2008 Received: February 21, 2008
Dear Mr. Stone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA' may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
Device Name: Alicia™ Menstrual Cup / Models 99ALXS, 100ALSM, 101ALMD, 102ALLG
Indications For Use:
The Alicia 100 Menstrual Cup / Models 99ALXS, 100ALSM, 101ALMD, 102ALLG is a receptacle placed in the vagina to collect blood and cellular debris that is extruded from the uterus via the cervix during menstruation.
Prescription Use (Part 21 CFR 801 Subpart D) OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hushin Leun
sion of Reproductive, Abdon Radiological Devices 510(k) Number
Page 1 of _ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.