DATATHERM II CONTINUOUS TEMPERATURE MONITOR, MODEL KD-2300
K070878 · Rg Medical Diagnostics · FLL · Jun 5, 2007 · General Hospital
Device Facts
Record ID
K070878
Device Name
DATATHERM II CONTINUOUS TEMPERATURE MONITOR, MODEL KD-2300
Applicant
Rg Medical Diagnostics
Product Code
FLL · General Hospital
Decision Date
Jun 5, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The DataTherm II Continuous Temperature Monitor is intended to continuously measure, display and monitor human body temperature updating every 4 seconds. Additionally, it stores up to 140 temperature/time/day data sets that may be stored at a user-selected frequency range from once per minute to once every 24 hours in one minute intervals. The current temperature and stored information may be viewed on the LCD screen that is integrated into the instrument or, from a PC computer screen after installation of the DataTherm software to a PC. The DataTherm II is intended to be used on patients of all ages.
Device Story
DataTherm II is a continuous temperature monitor for patients of all ages. System components: control/display unit with LCD and 5 keys; disposable flexible thermistor probe; USB interface cable; proprietary PC software. Device measures body temperature, updating every 4 seconds; stores up to 140 temperature/time/day data sets at user-defined intervals (1 minute to 24 hours). Data viewed on integrated LCD or transferred to PC via USB for archiving. Includes visual LED alert for temperature thresholds. Used in clinical or home settings to track patient temperature trends; assists healthcare providers or users in monitoring fever or temperature stability.
Clinical Evidence
No clinical data. Bench testing only, including compliance with EN 60601-1-2 (Electromagnetic Compatibility) and ASTM E1112-00 (Electronic Thermometer for Intermittent Determination of Patient Temperature).
Technological Characteristics
Electronic clinical thermometer using thermistor sensor technology. Components: LCD display, 5-key control interface, buzzer, LED alert, USB connectivity. Powered by 2 AAA batteries. Complies with ASTM E1112-00 and EN 60601-1-2. Data storage capacity: 140 sets. Connectivity: USB interface for PC data transfer.
Indications for Use
Indicated for patients of all ages for continuous measurement and monitoring of human body temperature and storage of temperature data at adjustable, preset time intervals.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the letters 'RG' stacked on top of each other in a bold, sans-serif font. Below the letters is a zig-zag line. Underneath the zig-zag line, the word 'DIAGNOSTICS' is printed in a smaller, sans-serif font. The logo appears to be for a diagnostics company.
# RG Medical Diagnostics
K070878
Pa 1 of 3
JUN - 5 2007
# Premarket 510(k) Summary
DataTherm® II
# Submitter information:
Sponsor:
RG Medical Diagnostics 21130 Bridge Street Southfield, MI 48034 1-248-750-0181 ph 1-248-750-0187 fax Contact Person: Ronald G. Le Tourneau Date Prepared: 26 February, 2007
#### Device Name:
| 1. Classification: | General Hospital |
|---------------------------|---------------------------------------------------|
| 2. Common Name: | Clinical Electronic Thermometer |
| 3. Proprietary Name: | DataTherm® II Continuous Temperature<br>Monitor |
| 4. Classification: | FLL - CFR 880.2910 |
| 5. Performance Standards: | ASTM E1112-00 (2006),<br>EN60601-1-2 (harmonized) |
### 6. Predicate Device Information:
K001281, Geratherm Babywatch Temperature Monitor
Manufactured by: Geratherm Medical AG Fahrenheitstrasse 1 D-98716 Geschwenda, Germany
7. Device Description: 21130 Bridge Street • Southfield, MI 48034 1 of 4 Phone: 888-596-9498 • Fax: 248-750-0187 • Website: www.rgmd.com
{1}------------------------------------------------
The DataTherm II Continuous Temperature Monitor (model HD-2300) consists of four components.
- 1. A control / display assembly measuring 3.16'' X 2.37' X 0.71" with an LCD screen and 5 control keys (On/Off, Mode Selection, &(increase), (decrease) and keyboard Lock). The interior electronic components consist of a circuit board, a battery compartment housing 2 - AAA batteries, a buzzer and LED alert light.
- 2. A disposable flexible temperature probe having a plug-in connector at one end and thermistor sensor at the other.
- 3. A USB interface connector cable that permits the transfer of stored and real-time temperature/time/day information to a PC computer screen.
- 4. Proprietary software that permits an individual patient's current temperature to be viewed and temperature/time/date information to be displayed and archived, if desired.
# 5. Intended Use:
The DataTherm II Continuous Temperature Monitor is intended to continuously measure, display and monitor human body temperature updating every 4 seconds. Additionally, it stores up to 140 temperature/time/day data sets that may be stored at a user-selected frequency range from once per minute to once every 24 hours in one minute intervals. The current temperature and stored information may be viewed on the LCD screen that is integrated into the instrument or, from a PC computer screen after installation of the DataTherm software to a PC. The DataTherm II is intended to be used on patients of all ages.
#### 6. Comparison to Predicate Device
The subject device is substantially equivalent to the predicate device, the Babywatch (marketed as DataTherm®) K001281. Both the subject and predicate devices are designed and engineered by the same company, Geratherm AG. The two devices employ the same technology, display information, and use the same controls. The subject device differences include disposable temperature probes, larger batteries, added visual temperature alert LED, improved water repellency, expanded memory, expanded storage interval alternatives and the ability to view and archive time/temperature/date data on a PC computer in user named files.
{2}------------------------------------------------
Ko708 78
Py 3 of B
# 7. Discussion of Non-clinical Tests Performed for Determination of Substantial Equivalence are as follows:
The DataTherm II (model KD-2300) complies with EN 60601-1-2 "Electromagnetic Compatibility" and ASTM E1112-00 " Electronic Thermometer for Intermittent Determination of Patient Temperature" standards. In addition, the DataTherm II conforms to the harmonized standard for electronic safety established by the US FCC Part 15.
# 8. Discussion of Clinical Tests Performed:
Not Applicable
#### 9. Conclusions:
The subject device and predicate devices provide the same functions and controls through employment of the same technology. Moreover, bench testing contained in this submission demonstrate that any differences in their characteristics do not present any new questions of safety or effectiveness. Thus, the DataTherm® II Continuous Temperature Monitor is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health and Human Services - USA. The logo features a stylized image of a human figure, with three overlapping profiles facing to the right. The profiles are simple and abstract, with no facial details. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Ronald G. Le Tourneau President RG Medical Diagnostics 21130 Bridge Street Southfield, Michigan 48034
JUN -5 2007
Re: K070878
Trade/Device Name: DataTherm® II Continuous Temperature Monitor Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: May 04, 2007 Received: May 07, 2007
Dear Mr. Le Tourneau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. Le Tourneau
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K070878
Device Name: _ DataTherm® II Continuous Temperature Monitor
Indications For Use:
The DataTherm® II Continuous Temperature Monitor is used to measure and monitor human body temperature and for storing temperature data at adjustable, preset time intervals.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Sion Sign-Off) | |
|---------------------------------------------|-----------|
| vision of Anesthesiology, General Hospital, | |
| Infection Control, Dental Devices | Page 1 of |
| 510(k) Number: | k470878 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.