← Product Code [FMF](/productcode/FMF) · K070856

# ICU MEDICAL SINGLE-USE SYRINGE (K070856)

_Icu Medical, Inc. · FMF · Oct 5, 2007 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K070856

## Device Facts

- **Applicant:** Icu Medical, Inc.
- **Product Code:** [FMF](/productcode/FMF.md)
- **Decision Date:** Oct 5, 2007
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 880.5860
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

ICU Medical piston syringes are single use syringes that are intended for injecting fluids into or withdrawing fluids from the body.

## Device Story

Single-use, sterile, disposable piston syringe; consists of calibrated hollow barrel and moveable plunger. Available in 1ml, 3ml, 6ml, 10ml, 20ml, and 30ml sizes. Operates via manual advancement or withdrawal of plunger to express or aspirate fluids. Features tapered nozzle and male luer hub for connection to needles or other mating luer devices. Used in clinical settings for fluid administration or withdrawal. Benefits include precise fluid measurement and delivery via calibrated barrel. Biocompatible materials; latex-free.

## Clinical Evidence

Bench testing only. Device conforms to ISO 7886-1 for performance and ISO 10993-1 for biocompatibility. Sterility Assurance Level (SAL) meets 10^-6.

## Technological Characteristics

Materials: Polycarbonate barrel, polyisoprene plunger tip. Energy source: Manual. Connectivity: None. Sterilization: Ethylene Oxide (EtO). Standards: ISO 7886-1, ISO 10993-1. Form factor: Calibrated hollow barrel with male luer hub.

## Regulatory Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

## Predicate Devices

- Merit Medical Systems, Inc. ([K024052](/device/K024052.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 5 2007

K070856

# ICU MEDICAL INC.

4455 Atherton Drive Salt Lake City, Utah (801) 264 - 1332, Phone (801) 264 - 1755, Fax Tracy S. Best, Sr. Regulatory Affairs Specialist Preparation Date: March 23, 2007

# 510(K) Summary of Safety and Effectiveness for the:

|  | Trade Name: | ICU Medical Syringes, multiple sizes |
|--|-------------|--------------------------------------|
|--|-------------|--------------------------------------|

Common Name: Syringe, Hypodermic

Piston Syringe, 21 CFR 880.5860, Class II Device Classification Name:

### Legally Marketed Predicate Devices for Substantial Equivalence:

*K024052 - Merit Medical Systems, Inc. (MMS)

### Rationale for SE:

The MMS syringe is a Class II device pre-sterilized device that is intended for single use. Its components are: a calibrated hollow barrel and a moveable piston or plunger. At the end of the barrel, there is a tapered nozzle and a male luer hub. These are universal style connectors that are used for connecting a needle or connecting to other approved devices with a mating (female) luer connector. This predicate device, along with the submitted device operates similarly. The device works with a small amount of silicone oil that provides a lubricant for the moving piston shaft. The materials used in the submitted device and in the predicate devices are equivalent. The materials have previously been tested and accepted for biocompatibility and are widely accepted as medical industry grade material. Both the submitted device and the predicate are latex free.

## Description of Submitted Device:

The ICU Medical Single-use Syringes is a pre-sterilized, single-use disposable that is equivalent to the predicate device. It is made of a calibrated hollow barrel, a moveable piston or plunger. Only two of the components ever touch the fluids or blood: the barrel and piston tip. All components are made from existing biocompatible materials that are routinely used in the medical device industry. This device works with a small amount of lubricant for moving the piston shaft. The connection luer on the end is identical to the predicate device. Syringe sizes include: 1ml; 10ml; 20ml; and 30ml. The barrel is visually clear, and the plungers are available in multiple colors (for indication of contents per hospital protocol).

## Intended Uses of the ICU Medical Syringe:

ICU Medical piston syringes are single use syringes that are intended for injecting fluids into or withdrawing fluids from the body.

{1}------------------------------------------------

## Technological Characteristics and Substantial Equivalence Table:

| Component:                | ICU Medical, Inc. | Merit Medical, Inc. |
|---------------------------|-------------------|---------------------|
| Syringe Barrel:           | Polycarbonate     | Polycarbonate       |
| Plunger Tip:              | Polyisoprene      | Polyisoprene        |
| Calibrated Barrel Volume: | YES               | YES                 |
| Sterilization method:     | EtO               | EtO                 |
| 510(k) Approval           | This submission   | K024052             |

The operational characteristics are identical in that by manually advancing or withdrawing the plunger in the barrel to express or withdraw fluids. Operation is the same for all piston syringes, including the predicate device.

### Safety and Performance:

ICU Medical syringes will conform to the requirements of ISO 7886-1, an FDA recognized standard, prior to marketing the devices. Additionally, ICU Medical's Sterility Assurance Level, (SAL) has an established history of meeting the 10° level. The single use syringes will be packaged in a way as to ensure conformity with ISO 10993-1, including minimizing residual gases.

#### Conclusion:

The materials, performance, and operational features of both the submitted device and the predicate device are substantially equivalent and are safe and effective for their intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Tracy S. Best Senior Regulatory Affairs Specialist ICU Medical, Incorporated 4455 South Atherton Drive Salt Lake City, Utah 84123

OCT 5 ° 2007

Re: K070856

Trade/Device Name: ICU Medical Sterile Disposable Piston Syringes, Sizes: Iml; 3ml; 6ml; 10ml; 20ml; & 30ml Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: September 24, 2007 Received: September 25, 2007

Dear Mr. Best:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Best

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known):

#### Device Name:

ICU Medical sterile disposable piston syringes, Sizes: 1ml; 3ml; 10ml; 20ml; & 30ml

#### Indications for Use:

ICU Medical piston syringes are single use syringes that are intended for injecting fluids into or withdrawing fluids from the body.

Prescription Use _ XX (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

#### Concurrence of CDRH, Office of Device Evaluation (ODE)

Cinta Lanto

(Tision Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

510(k) Number:________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K070856](https://fda-staging.innolitics.com/device/K070856)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).