← Product Code [KWP](/productcode/KWP) · K070742

# MODIFICATION TO VERTEX RECONSTRUCTION SYSTEM (K070742)

_Medtronic Sofamor Danek · KWP · Sep 14, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K070742

## Device Facts

- **Applicant:** Medtronic Sofamor Danek
- **Product Code:** [KWP](/productcode/KWP.md)
- **Decision Date:** Sep 14, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3050
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

When intended to promote fusion of the occipitocervical spine, cervical spine, and the thoracic spine, (Occiput-T3), the VERTEX® Reconstruction System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors. The occipitocervical plate rods, plates, rods, occipital screws, and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine. Occipitocervical construct require bilateral fixation to C2 and below. Note: Segmental fixation is recommended for these constructs. The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine. The use of multi-axial screws is limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine. Titanium ATLAS® Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine. In order to achieve additional levels of fixation, the VERTEX® Reconstruction System may be connected to the CD HORIZON® Spinal System rods with the VERTEX® rod connectors. Refer to the CD HORIZON® Spinal System package insert for a list of the CD HORIZON® Spinal System indications of use.

## Device Story

Posterior spinal fixation system; consists of plates, rods, hooks, screws, and multi-axial screws; provides stabilization to promote fusion in occipitocervical, cervical, and thoracic (C1-T3) spine. Surgeon-operated in clinical/surgical settings; components tailor-made for individual patient anatomy; rigidly locked into constructs. Multi-axial screws include Nitinol retaining rings and silicone adhesive elastomeric stakes. System allows for cable attachment via ATLAS® Cable System and connection to CD HORIZON® Spinal System rods. Provides mechanical stabilization to facilitate bone fusion; benefits patients by addressing spinal instability, pain, and deformity.

## Clinical Evidence

Bench testing only. Mechanical testing performed to verify structural integrity and performance equivalence to predicate systems.

## Technological Characteristics

Materials: medical grade titanium, titanium alloy, cobalt chromium, Nitinol (NiTi) retaining rings, silicone adhesive. Components: plates, rods, hooks, screws, multi-axial screws, connectors. Connectivity: None. Sterilization: Not specified. Software: None.

## Regulatory Identification

A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.

## Predicate Devices

- VERTEX® Reconstruction System ([K042789](/device/K042789.md))
- VERTEX® Reconstruction System ([K052402](/device/K052402.md))
- VERTEX® Reconstruction System ([K052734](/device/K052734.md))
- VERTEX® Reconstruction System ([K053483](/device/K053483.md))
- TRAVERSE™ OCT Spinal Fixation System ([K062879](/device/K062879.md))
- CD HORIZON® Spinal System ([K061304](/device/K061304.md))
- CD HORIZON® Spinal System ([K063417](/device/K063417.md))
- VERTE-STACK® Spinal System ([K062427](/device/K062427.md))
- TSRH® Spinal System ([K062807](/device/K062807.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko70742

Page 1 of 2

CONFIDENTIAL

# VERTEX® Reconstruction System 510(k) Summary

# June 2007

SEP 1 4 2007

- I. Company: Medtronic Sofamor Danek USA 1800 Pyramid Place Memphis, Tennessee 38132 Telephone: (901) 396-3133 Fax: (901) 346-9738
#### Contact: Christine Scifert Director, Regulatory Affairs

#### II. Proposed Proprietary Trade Name: VERTEX® Reconstruction System

- III. Classification Name(s): Spinal Interlaminal Fixation Orthosis; Spinal Intervertebral Body Fixation Orthosis; Pedicle Screw Spinal System; Orthosis, Spinal Pedicle Fixation, for Degenerative Disc Disease; Class: II; Product Code(s): KWP, MNI, MNH; and Regulation No .: 888.3050 and 888.3070
- IV. Description: The VERTEX® Reconstruction System is a posterior system, which consists of a variety of shapes and sizes of plates, rods, hooks, screws, multi-axial screws, and connecting components, which can be rigidly locked to the rod in a variety of configurations, with each construct being tailor-made for the individual case. Titanium ATLAS® cable may be used with this system at the surgeon's discretion. See the package inserts of both of those systems for labeling limitations.

The VERTEX® Reconstruction System is fabricated from medical grade titanium, medical grade titanium alloy, and medical grade cobalt chromium. Medical grade titanium, medical grade titanium alloy, and/or medical grade cobalt chromium may be used together. Never use titanium, titanium alloy, and or/cobalt chromium with ataless steel in the same construct. The VERTEX® Reconstruction System includes a retaining ring for the multi-axial screw made of Shape Memory Alloy (Nitinol - NiTi) - Thane Memory Alloy is compatible with titanium, titanium alloy, and cobalt chrom. Shippe only. Some multi-axial screws contain elastomeric stakes made of silicore adhesive commonly used in implantable medical devices. Do not use with stainless steel. No warranties, express or implied, are made. Implied warranties of merchantability and fitness for a particular purpose or use are specifically excluded. See the MDT Catalog or price list for further information about warranties and limitations of liability.

To achieve best results, do not use any of the VERTEX® Reconstruction System implant components with components from any other system or manufacturer unless specifically labeled to do so in this or another MEDTRONIC document. As with all orthopedia and neurosurgical implants, none of the VERTEX® Reconstruction System components should ever be reused under any circumstances.

{1}------------------------------------------------

K070742

Page 2 of 2

# CONFIDENTIAL

- V. Indications for Use: When intended to promote fusion of the occipitocervical spine, cervical spine, and the thoracic spine, (Occiput-T3), the VERTEX® Reconstruction System is indicated for the following:
DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors.

## Occipitocervical Components: Plate Rod/Plates/Rods/Occipital Screws/Hooks

The occipitocervical plate rods, plates, rods, occipital screws, and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical junction and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine.

Occipitocervical construct require bilateral fixation to C2 and below. Note: Segmental fixation is recommended for these constructs.

# Hooks and Rods

The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.

## Multi-axial Screws/Connectors

The use of multi-axial screws is limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine.

Titanium ATLAS® Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.

In order to achieve additional levels of fixation, the VERTEX® Reconstruction System may be connected to the CD HORIZON® Spinal System rods with the VERTEX® rod connectors. Refer to the CD HORIZON® Spinal System package insert for a list of the CD HORIZON® Spinal System indications of use.

- VI. Substantial Equivalence: Documentation was provided demonstrating that the VERTEX® Reconstruction System is substantially equivalent to other commercially available posterior fixation systems and other pre-enactment devices including the VERTEX® Reconstruction System (K042789, SE 12/21/04; K052402, SE 09/23/05; K052734, SE 10/21/05; K053483, SE 01/05/06), the TRAVERSE™ OCT Spinal Fixation System (K062879, SE 12/22/06), the CD HORIZON® Spinal System (K061304, SE 07/18/06; K063417, SE 01/18/2007), the VERTE-STACK® Spinal System (K062427, SE 09/12/06) and the TSRH® Spinal System (K062807, SE 11/06/06). The results of mechanical testing performed for the subject VERTEX® Reconstruction System components were equivalent to or better than the testing performed for the predicate VERTEX® Reconstruction System components.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The seal is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Medtronic Sofamor Danek USA, Inc. % Ms. Christine Scifert Supervisor, Regulatory Affairs 1800 Pyramid Place Memphis. TN 38132

SEP 1 4 2007

Re: K070742

> Trade/Device Name: VERTEX® Reconstruction System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: KWP, MNI, MNH Dated: August 6, 2007 Received: August 9, 2007

Dear Ms. Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Christine Scifert

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,
Barbara Boehm

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known): K070742

Device Name: __ VERTEX® Reconstruction System

### Indications for Use:

When intended to promote fusion of the occipitocervical spine, and the thoracic spine, (Occiput-T3), the VERTEX® Reconstruction System is indicated for the following: DDD (neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, spinal stenosis, fracture, dislocation, failed previous fusion and/or tumors.

# Occipitocervical Components: Plate Rod/Plates/Rods/Occipital Screws/Hooks

The occipitocervical plate rods, plates, rods, occipital screws, and hooks are intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the occipitocervical innetion and the cervical spine. When used to treat these occipitocervical and cervical conditions, these screws are limited to occipital fixation only. The screws are not intended to be placed in the cervical spine.

Occipitocervical constructs require bilateral fixation to C2 and below. Note: Segmental fixation is recommended for these constructs.

#### Hooks and Rods

The hooks and rods are also intended to provide stabilization to promote fusion following reduction of fracture/dislocation or trauma in the cervical/upper thoracic (C1-T3) spine.

#### Multi-axial Screws/Connectors

The use of multi-axial screws are limited to placement in T1-T3. The screws are not intended to be placed in the cervical spine.

Titanium ATLAS® Cable System to be used with the VERTEX® Reconstruction System allows for cable attachment to the posterior cervical or thoracic spine.

In order to achieve additional levels of fixation, the VERTEX® Reconstruction System may be connected to the CD HORIZON® Spinal System rods with the VERTEX® rod connectors. Refer to the CD HORIZON® Spinal System package insert for a list of the CD HORIZON® Spinal System indications of use.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Huber Buellm
Division Sign Off

and Neurological Devices

510(k) Number K070746

---

**Source:** [https://fda-staging.innolitics.com/device/K070742](https://fda-staging.innolitics.com/device/K070742)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
