← Product Code [NSU](/productcode/NSU) · K070597

# VERIGENE SYSTEM, F2 NUCLEIC ACID TEST, F5 NUCLEIC ACID TEST, AND MTHFR NUCLEIC ACID TEST (K070597)

_Nanosphere, Inc. · NSU · Oct 11, 2007 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K070597

## Device Facts

- **Applicant:** Nanosphere, Inc.
- **Product Code:** [NSU](/productcode/NSU.md)
- **Decision Date:** Oct 11, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.2570
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Verigene® F5 Nucleic Acid Test is an in vitro diagnostic for the detection and genotyping of a single point mutation (G to A at position 1691; also known as Factor V Leiden) of the human Factor V gene (F5; Coagulation Factor V gene) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System. The Verigene® F2 Nucleic Acid Test is an in vitro diagnostic for the detection and genotyping of a single point mutation (G to A at position 20210 of the human Factor II gene (F2; prothrombin gene)) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System. The Verigene® MTHFR Nucleic Acid Test is an in vitro diagnostic for the detection and genotyping of a single point mutation (C to T at position 677) of the human 5, 10-methylenetetrahydrofolate reductase gene (MTHFR) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System.

## Device Story

Verigene System is a bench-top molecular diagnostics workstation for automated genotyping of DNA samples. Input: isolated genomic DNA loaded into disposable Test Cartridges. Operation: hybridization of target DNA to synthetic gene-specific oligonucleotide capture strands on cartridge substrate; formation of 'sandwich' complex with mediator oligonucleotide; binding of gold nanoparticle probes; signal enhancement via two-part reagent. Reader illuminates nanoparticles; photosensor measures relative brightness of wild-type vs. mutant spots. Output: genotype result based on relative brightness. Used in clinical laboratories by trained professionals. Benefits: automates complex molecular analysis; provides rapid (approx. 1.5 hours) genotyping to aid diagnosis of hereditary thrombophilia factors.

## Clinical Evidence

Method comparison study using 287 clinical samples (68% from 'rule-out thrombophilia' patients) compared Verigene results to bi-directional DNA sequencing. Results showed 100% Positive Percent Agreement (95% CI: 98.9-100.0%) and 100% Negative Percent Agreement (95% CI: 98.9-100.0%). No clinical data provided; bench testing only.

## Technological Characteristics

System uses disposable cartridges with synthetic oligonucleotide capture strands and gold nanoparticle probes. Detection via evanescent wave light scatter. Requires isolated genomic DNA (40-400 ng/μL). Connectivity: Up to 8 processors per reader. Calibration required for temperature control system. Software performs automated on-line quality monitoring of fluidics, instrument, and reagent integrity.

## Regulatory Identification

Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems."

## Predicate Devices

- Roche Diagnostics Corporation LightCycler® Instrument ([K033734](/device/K033734.md))
- Roche Diagnostics Corporation Factor V Leiden Kit ([K033607](/device/K033607.md))
- Roche Diagnostics Corporation Factor II (Prothrombin) G20210A Kit ([K033612](/device/K033612.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k070597

B. Purpose for Submission:
New device

C. Measurand:
Factor II (Prothrombin)
Factor V Leiden
5, 10-Methylenetetrahydrofolate reductase (MTHFR)

D. Type of Test:
Genotype

E. Applicant:
Nanosphere INC

F. Proprietary and Established Names:
Verigene® F5 Nucleic Acid Test
Verigene® F2 Nucleic Acid Test
Verigene® MTHFR Nucleic Acid Test

G. Regulatory Information:
1. Regulation section:
21 CFR 864.7280 Factor V Leiden DNA mutation detection systems
2. Classification:
Class II
3. Product code:
NPR, Test, Factor II G20210A mutations, genomic DNA PCR
NPQ, Test, Factor V Leiden mutations, genomic DNA PCR
OMM, Test, 5, 10-Methylenetetrahydrofolate reductase mutations, genomic DNA PCR
NSU, Instrumentation for Clinical Multiplex Test Systems
4. Panel:
Hematology (81); Chemistry (75)

H. Intended Use:
1. Intended use(s):
The Verigene® F5 Nucleic Acid Test is an *in vitro* diagnostic for the detection and genotyping of a single point mutation (G to A at position 1691; also known as Factor V Leiden) of the human Factor V gene (F5; Coagulation Factor V gene) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System.

The Verigene® F2 Nucleic Acid Test is an *in vitro* diagnostic for the detection and genotyping of a single point mutation (G to A at position 20210 of the human Factor II gene (F2; prothrombin gene)) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System.

{1}

The Verigene® MTHFR Nucleic Acid Test is an in vitro diagnostic for the detection and genotyping of a single point mutation (C to T at position 677) of the human 5, 10-methylenetetrahydrofolate reductase gene (MTHFR) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System.

2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For Prescription Use Only.
4. Special instrument requirements:
Verigene® System

I. Device Description:
The Verigene® System consists of two instruments, the Verigene® Processor and the Verigene® Reader. It utilizes single-use, disposable test Cartridges to process and genotype multiple genes in a DNA sample in ~1 ½ hrs. Up to 8 Verigene® Processors may be connected to a single Verigene® Reader. Each processor contains 4 hybridization modules, and the modules can simultaneously run different tests.

J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche Diagnostic Corp. Factor V Leiden Kit
Roche Diagnostic Corp. Factor II (Prothrombin) G20210A Kit.
2. Predicate 510(k) number(s):
k033607, k033612
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | The Verigene® F5 Nucleic Acid Test is an in vitro diagnostic for the detection and genotyping of a single point mutation (G to A at position 1691; also known as Factor V Leiden) of the human Factor V gene (F5; Coagulation Factor V gene) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System. | The Factor V Leiden Kit is an in vitro diagnostic test for the detection and genotyping of a single point mutation (G to A at position 1691) of the human Factor V gene, from DNA isolated from human whole peripheral blood. The Factor V Leiden Kit is indicated as an aid to diagnosis in the evaluation of patients with suspected thrombophilia  |
|  Intended use | The Verigene® F2 Nucleic Acid Test is an in vitro diagnostic for the detection and genotyping of a single point mutation (G to A at position | The Factor II(Prothrombin) G20210OA Kit is an in vitro diagnostic test for the detection and genotyping  |

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | 20210of the human Factor II gene (F2; prothrombin gene) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System. | of a single point mutations (G to A at position 202 10) of the human Factor II gene, from DNA isolated from human whole peripheral blood. The Factor 1I (Prothrombin) G20210A Kit is indicated as an aid to diagnosis in the evaluation of patients with suspected thrombophilia  |
|  Specimen Type | Peripheral whole blood | same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Intended use | The Verigene® MTHFR Nucleic Acid Test is an in vitro diagnostic for the detection and genotyping of a single point mutation (C to T at position 677) of the human 5, 10-methylenetetrahydrofolate reductase gene (MTHFR) in patients with suspected thrombophilia, from isolated genomic DNA obtained from whole blood samples. The test is intended to be used on the Verigene® System. | None  |
|  Detection method | Signal amplification Chemical amplification of reporter signal | DNA amplification (PCR)  |
|  Sample size | 25 μL | 10-20 μL in glass capillaries  |
|  Detection procedure | Single-image sensor where nanoparticles are illuminated using a fixed-wavelength light source | Optical detection of stimulated fluorescence  |
|  Detection Chemistry | SNP discrimination via oligonucleotide probes; diction via evanescent wave light scatter with nanoparticles | Paired hybridization probes using fluorescence resonance energy transfer (FRET)  |

{3}

4

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Analysis time | 1.5 hrs | Detection occurs at defined intervals during PCR cycle and can e viewed in real-time  |

K. Standard/Guidance Document Referenced (if applicable):
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
Class II Special Controls Guidance Document: Factor V Leiden DNA Mutation Detection Systems
CLSI EP5-A2; Evaluation of Precision Performance of Quantitative Measurement Methods

L. Test Principle:
The analysis sequence is the same for each of the three tests. After extracted and purified DNA, mixed with hybridization buffer, is loaded into the sample well of the test cartridge, it is ready for processing. An internal barcode reader reads the cartridge ID and sends the information to the Verigene® Reader. The reader establishes the hybridization parameters and starts the hybridization process.

The genotyping process occurs with a hybridization of the target analyze to a synthetic gene-specific oligonucleotide capture strand on the test cartridge’s substrate. A synthetic mediator target-specific oligonucleotide is included with the test-specific sample buffer to form a hybridization” sandwich” with the gene sequence of interest. After washing, a probe composed of a gold nanoparticle with covalently bound oligonucleotide complementary to a sequence on the intermediate oligonucleotide, is introduced. A signal enhancement reagent is added to the hybridization chamber and reacts with the gold nanoparticle to amplify the signal for the Verigene® Reader scanning and analysis.

M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The reproducibility testing included 4 studies, each using a different DNA samples extracted from whole blood. In the first study, each of the three test sites ran the same sample in duplicate.

|  Sites | F5 | F2 | MTHFR  |
| --- | --- | --- | --- |
|  1 | HET | HET | HET  |
|   |  HET | HET | HET  |
|  2 | HET | HET | HET  |
|   |  HET | HET | HET  |
|  3 | HET | HET | HET  |
|   |  HET | HET | HET  |

In the second study, one operator at Site 1 analyzed the same sample in duplicate each

{4}

day for three days.

|  Days | F2 | F5 | MTHFR  |
| --- | --- | --- | --- |
|  1 | WT | HET | WT  |
|   |  WT | HET | WT  |
|  2 | WT | HET | WT  |
|   |  WT | HET | WT  |
|  3 | WT | HET | WT  |
|   |  WT | HET | WT  |

In the third study, three operators at Site 2 each analyzed the same sample in duplicate.

|  Operators | F2 | F5 | MTHFR  |
| --- | --- | --- | --- |
|  1 | WT | HET | HET  |
|   |  WT | HET | HET  |
|  2 | WT | HET | HET  |
|   |  WT | HET | HET  |
|  3 | WT | HET | HET  |
|   |  WT | HET | HET  |

In the fourth study, one operator at Site 3 analyzed the same sample in duplicate using three lots of reagents.

|  Reagent Lots | F2 | F5 | MTHFR  |
| --- | --- | --- | --- |
|  1 | HET | WT | HET  |
|   |  HET | WT | HET  |
|  2 | HET | WT | HET  |
|   |  HET | WT | HET  |
|  3 | HET | WT | HET  |
|   |  HET | WT | HET  |

b. Linearity/assay reportable range: NA
c. Traceability, Stability, Expected values (controls, calibrators, or methods): NA
d. Detection limit: In 40 genomic samples diluted to  $40\mathrm{ng} / \mu \mathrm{L}$  the total call rate for all three assays was  $82\%$ . There were no mis-calls.
e. Analytical specificity: To simulate sample cross-contamination, a sample of known genotype was spiked with various amounts of a sample with a known but different genotype. The first sample consisted of a F2 (wild-type); F5 (mutant); MTHFR (wild type) and the

{5}

contaminating sample's genotype was: F2 (mutant); F5 (wild-type); MTHFR (hetero). The percent contaminating sample ranged from 0-50%. The results indicated that neither the call rate nor accuracy in a homozygous sample was affected in the presence of up to 10% contamination by volume of a sample with the opposite homozygous genotype. At least 20% contamination by volume of a sample with a heterozygous genotype is required to affect the call rate.

A study was performed to assess the interference of heparin, hemoglobin or magnetic beads in a purified DNA sample. Interference was observed in the 5X magnetic bead sample where the F5 signal was slightly lower than the control. The genotyping ration results showed slight variations but none led to an increase in no-calls or a mis-call.

f. Assay cut-off: NA

2. Comparison studies:

a. Method comparison with predicate device:

Accuracy (percent agreement) was determined by comparison to bi-directional DNA sequencing:

A total of 287 samples, sixty-eight percent (68%) from patients undergoing "rule-out thrombophilia" testing, were analyzed at three sites using the Verigene® F5 / F2 / MTHFR Nucleic Acid Tests and by bi-directional sequencing analysis at an independent reference laboratory. All purified DNA samples were from whole blood collected using EDTA as the anticoagulant. Data comparison details for F5, F2, and MTHFR are shown in tables 3-5 below:

Table 3. Verigene® F5 Nucleic Acid Test method comparison results

|   | Verigene® F5 Nucleic Acid Test  |   |   |   |
| --- | --- | --- | --- | --- |
|  Sequence analysis* |  | Wild-type (wt) | Heterozygous (het) | Mutant (mut)  |
|   |  wt | 253 | 0 | 0  |
|   |  het | 0 | 22 | 0  |
|   |  mut | 0 | 0 | 7  |

Table 4. Verigene® F2 Nucleic Acid Test method comparison results

|   | Verigene® F2 Nucleic Acid Test  |   |   |   |
| --- | --- | --- | --- | --- |
|  Sequence analysis* |  | Wild-type (wt) | Heterozygous (het) | Mutant (mut)  |
|   |  wt | 258 | 0 | 0  |
|   |  het | 0 | 9 | 0  |
|   |  mut | 0 | 0 | 5  |

Table 5. Verigene® MTHFR Nucleic Acid Test method comparison results

{6}

|   | Verigene® MTHFR Nucleic Acid Test  |   |   |   |
| --- | --- | --- | --- | --- |
|  Sequence analysis* |  | Wild-type (wt) | Heterozygous (het) | Mutant (mut)  |
|   |  wt | 125 | 0 | 0  |
|   |  het | 0 | 117 | 0  |
|   |  mut | 0 | 0 | 26  |

*Two DNA samples did not return complete bi-directional sequencing results and, therefore, were excluded from the data. Specifically, one sample did not produce results for any genotype reads (i.e.,  $F2$ ,  $F5$ , and MTHFR) and one sample returned no results for the  $F2$  genotype. These four genotype reads represent  $0.7\%$  of the total possible calls in the method comparison study. There were no samples excluded due to discrepancies between sequenced results.

The Verigene® F5 / F2 / MTHFR Nucleic Acid Tests demonstrated  $100\%$  Positive Percent Agreement  $(95\% \mathrm{CI} = 98.9$  to  $100.0\%)$  and  $100\%$  Negative Percent Agreement  $(95\% \mathrm{CI} = 98.9$  to  $100.0\%)$  in these studies based on data from calls made.

b. Matrix comparison: NA

3. Clinical studies:

a. Clinical Sensitivity: NA
b. Clinical specificity: NA
c. Other clinical supportive data (when a. and b. are not applicable): NA

4. Clinical cut-off: NA
5. Expected values/Reference range: NA

N. Instrument Name:

Verigene® System

O. System Descriptions:

1. Modes of Operation:

Closed system

2. Software:

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes ☐ or No ☑

3. Specimen Identification:

Barcode reader

4. Specimen Sampling and Handling:

DNA should be extracted using a DNA extraction method that provides DNA with the following characteristics:

Average purity: Optimal mean: 1.89 (typical range: 1.60 to 2.00  $(\mathrm{A}_{260} / \mathrm{A}_{260})$

{7}

Concentration: 40 ng/μL to 400 ng/μL

DNA elution: water or Tris-EDTA

5. Calibration:
The temperature control system on the Verigene® Processor device is the only component that requires calibration. This is performed at the time of manufacture or by a Nanosphere service technician.

6. Quality Control:
1. Automated on-line quality misprints that monitor instrument functionality, software performance, fluidics, test conditions, reagent integrity, and procedural steps in each assay each time a test is performed
2. Liquid materials

P. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above:

Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

8

---

**Source:** [https://fda-staging.innolitics.com/device/K070597](https://fda-staging.innolitics.com/device/K070597)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
