K070584 · Duoject Medical Systems, Inc. · LHI · Apr 2, 2008 · General Hospital
Device Facts
Record ID
K070584
Device Name
SMART-ROD
Applicant
Duoject Medical Systems, Inc.
Product Code
LHI · General Hospital
Decision Date
Apr 2, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The Smart-Rod™ is a single use reconstitution device intended for the transfer of a diluent contained in a prefilled syringe to a vial containing a drug product, and the return transfer of the admixture to the syringe. The Smart-Rod™ is intended for use by physicians, nurses, and other practitioners who routinely administer injections, or patients and other individuals authorized by a physician to administer injections of prescribed medication.
Device Story
Smart-Rod™ is a single-use reconstitution device comprising a vial socket and plunger rod; replaces standard plunger rod of pre-filled syringe. Device facilitates fluid transfer between pre-filled syringe and drug vial. Integral needle perforates syringe plunger to create fluid path; vial socket attaches to drug vial. After reconstitution, admixture is transferred back to syringe; needle retracted into plunger rod; vial socket and vial discarded. Used in clinical or home settings by physicians, nurses, or authorized patients. Simplifies reconstitution process; maintains sterility; enables injection using syringe's own needle.
Clinical Evidence
Bench testing only. 16 qualification tests performed to verify operational performance, including attachment to vials and syringes, interaction with vial stoppers and syringe plungers, and overall device integrity during use.
Technological Characteristics
Materials: polycarbonate, polyethylene, ABS, and steel. Single-use, sterile, manual fluid transfer device. Consists of vial socket and plunger rod with integral needle. Mechanical operation; no energy source or software.
Indications for Use
Indicated for patients requiring reconstituted injectable medications; intended for use by physicians, nurses, or authorized patients/individuals for drug reconstitution and transfer between prefilled syringes and vials.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
#### 510(k) SUMMARY 5.
K070.584
APR - 2 2008
| Submitter's Name: | Duoject Medical Systems Inc. |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Address: | 50 chemin de Gaspe, Complex B-5 |
| | Bromont, Quebec |
| | Canada J2L 2N8 |
| Telephone: | (450) 534-3666 ext.24 |
| Fax: | (450) 534-3700 |
| Contact Person: | Maud Richard |
| Date of Summary: | April 1, 2008 |
| Trade Name: | Smart-Rod™ Reconstitution System |
| Common Name: | Reconstitution system |
| Classification Name: | Fluid Transfer Set (includes vial adapter) per 21 CFR |
| | 880.5440 |
| Predicate Device: | Clip'n'Ject® Reconstitution System (K041691)<br>Bioject® Needle-Free Vial Adapter (K963012 &<br>K043304) |
#### Device Description:
The Smart-Rod™ is a single-use reconstitution device consisting of a preassembled vial socket and plunger rod. The Smart-Rod™ is designed to be used together with a pre-filled syringe. The pre-filled syringe would be legally authorized through its own clearance or customer New Drug Application and is not intended to be cleared as part of this premarket notification. The Smart-Rod™ plunger rod replaces the usual plunger rod of the pre-filled syringe and has an integral needle designed to perforate the plunger of the pre-filled syringe for fluid transfer from the syringe to the drug vial (connected using the vial socket) through the internal fluid path in the rod. Once reconstitution is complete, the device permits the transfer of the reconstituted drug product back to the The integral needle is retracted into the plunger rod by unscrewing the syringe. The vial socket and vial are discarded, and the filled syringe vial socket. containing reconstituted drug is ready for use and the drug product may be injected by means of the yet unexposed needle equipping the syringe.
pg 1 + 2
{1}------------------------------------------------
## Intended Use:
The Smart-Rod™ is a single use reconstitution device intended for the transfer of a diluent contained in a prefilled syringe to a vial containing a drug product, and the return transfer of the admixture to the syringe.
The Smart-Rod™ is intended for use by physicians, nurses, and other practitioners who routinely administer injections, or patients and other individuals authorized by a physician to administer injections of prescribed medication.
## Substantial Equivalence:
The Smart-Rod™ has the same intended use as the predicate devices, with the primary purpose of permitting the transfer of a diluent contained in a syringe to a vial containing a drug product, in order to allow drug reconstitution prior to injection. All devices are single-use and intended to be marketed sterile in individual packaging.
It combines design concepts found in the predicate devices for vial attachment, vial closure piercing and use of the syringe plunger to open a fluid path.
The Smart-Rod™ device is similar in design to Bioject®'s Needle-Free Vial Adapter and West's Clip'n'iect® with respect to the mechanism of attachment to a vial. Its vial closure piercing mechanism is identical to that of Bioject®'s Needle-free Vial Adapter, however unlike the Bioject® Needle-Free vial adapter, the Smart-Rod™ is intended for use with a syringe equipped with a needle, like the West Clip'n'ject® device.
Construction materials include polycarbonate, polyethylene, ABS and steel. Predicate devices use polycarbonate, polyethylene and steel in the operation of the device. ABS is a well established material for other devices of similar intended use.
## Performance testing:
16 qualification tests were conducted to demonstrate the operational performance of the device, including attachment to vial and syringe, interaction with the vial stopper and the syringe plunger, operational performance and integrity during use.
070584 pg 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in a simple, bold design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 2 2008
Ms. Maud Richard Compliance Engineer Duoject Medical Systems, Incorporated 50 rue de Gaspe, Complex B-5 Bromont, (Quebec) CANADA J2L 2N8
Re: K070584
Trade/Device Name: Smart-Rod Reconstitution System Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: LHI Dated: March 12, 2008 Received: March 17, 2008
Dear Ms. Richard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Richard
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 4. INDICATIONS FOR USE STATEMENT - -------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number (if known): K070584 Device Name: Smart-Rod™
Indications for Use:
The Smart-Rod™ is a single use reconstitution device intended for the transfer of a diluent contained in a prefilled syringe to a vial containing a drug product, and the return transfer of the admixture to the syringe.
The Smart-Rod™ is intended for use by physicians, nurses, and other practitioners who routinely administer injections, or patients and other individuals authorized by a physician to administer injections of prescribed medication.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Prescription Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Ams 20
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control. Dental Devices
510(k) Number: 1 1 < 4 20194
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.