Nasopharyngoscopes are intended to examine or treat the nasal cavity and nasal pharynx. They are used between the upper respiratory tracts of the nasal passage and the vocal chords. They allow the recognition of anomalies in the tissue of the accessible organs. The EF-N 14 and EF-N 14S models with their working channel allow the insertion of an instrument to carry out a biopsy.
Device Story
Flexible nasopharyngoscope used for visualization and treatment of nasal cavity and pharynx; includes insertion tube with bendable distal tip, handle, eyepiece, and focus ring. Models EF-N 14 and EF-N 14S feature a working channel for biopsy instrument insertion. Operated by physicians in clinical settings to identify tissue anomalies. Device transmits optical images via fiber-optics to the eyepiece. Provides diagnostic visualization and enables interventional access to upper respiratory tracts.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Flexible fiber-optic endoscope. Materials: medical grade silicone (body cover), stainless steel (connectors/valves), medical grade plastics (eyepiece/focus ring), anodized/parylene-coated aluminum (control lever). Features: 2.8-4.1 mm sheath diameter, 320 mm working length, 130° up/down bending angle, 80° field of view. Connectivity: compatible with standard light guide cables (ACMI, Richard Wolf, Storz). Sterilization: not specified.
Indications for Use
Indicated for examination or treatment of the nasal cavity and nasal pharynx, including the upper respiratory tracts from the nasal passage to the vocal chords, in patients requiring visualization or biopsy (EF-N 14/14S models) of tissue anomalies.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
Flexible nasopharyngo-laryngoscope and flexible bronchoscope (K992526)
Submission Summary (Full Text)
{0}------------------------------------------------
K070580
Image /page/0/Picture/1 description: The image shows the logo for XION medical. The logo consists of the word "XION" in a sans-serif font, with the "i" dotted with a solid black circle. Below the "ION" portion of the logo is the word "medical" in a smaller font size.
## 510(k) - Summary 6
| Submitter: | XION GmbH<br>Pankstr. 8-10<br>13127 Berlin<br>Germany<br>Phone +49 (0)30 474 987 56<br>Fax +49 (0)30 474 987 11 | JUN 1 3 2007 |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Contact: | Alexander Kliem<br>Email akliem@xion-medical.com | |
| Date: | 05/11/2007 | |
| Name of Device: | Nasopharyngoscope EF-N, EF-NS, EF-N Slim, EF-N 14, EF-N 14 S | |
| Common Name: | Naso-pharyngo-laryngo-fiberscope | |
| Classification Name of Device: | Nasopharyngoscope a) product code: EOB<br>b) regulation number: 874.4760 | |
| Legally Marketed Device to which Equivalence is Claimed: | Flexible nasopharyngo-laryngoscope and flexible bronchoscope<br>Richard Wolf Medical Instruments Corp.<br>(Premarket notification K992526) | |
| Description: | Flexible nasopharyngoscopes with or without a working channel are built of the insertion tube with its bendable distal tip, the handle, the eyepiece and the focus ring. The handle incorporates the control lever to bend the distal tip and the connectors for the leakage tester and the fibre-optic light- guide cable. | |
| Indications for Use: | Nasopharyngoscopes are intended to examine or treat the nasal cavity and nasal pharynx. They are used between the upper respiratory tracts of the nasal passage and the vocal chords. They allow the recognition of anomalies in the tissue of the accessible organs. The EF-N 14 and EF-N 14S models with their working channel allow the insertion of an instrument to carry out a biopsy. | |
.
.
.
{1}------------------------------------------------
| | SE Device | | Predicate Device |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| | XION Device | | Richard Wolf Device |
| Name | Nasopharyngoscope<br>(flexible or rigid) | | Flexible nasopharyngo-<br>laryngoscope and flexible<br>bronchoscope,<br>Richard Wolf Medical Instruments<br>Corp. |
| Identification | EF-N<br>EF-NS<br>EF-N Slim<br>EF-N14<br>EF-N14S | 130 400 034<br>130 400 134<br>130 400 028<br>130 414 040<br>130 414 140 | 7222, 7223, 7224, 7265, 7325,<br>7330 |
| Performance specification | Nasopharyngoscope | | Nasopharyngoscope |
| Sheath diameter | 2.8 - 4.1 mm | | 3.5 mm |
| Working length | 320 mm | | 300 mm |
| Bending angle<br>Up | 130° | | 130° |
| Down | 130° | | 130° |
| Bending radius | 8 mm | | |
| Optical system<br>Field of view | 80° | | 95° |
| Depth of field | 1 - 50 mm | | 3 - 50 mm |
| Weight | 230 - 320 g | | |
| Forceps channel diameter | 1.4 mm | | 1.1 mm |
| SE Device | | Predicate Device | |
| | XION Device | Richard Wolf Device | |
| Appearance, shape | | | |
| Control lever | Handling by turning, rotation axis<br>across to longitudinal axis of the<br>endoscope;<br>Milled of aluminium, anodized (for<br>mechanical resistance), parylene<br>coated (for chemical resistance) | Handling by turning, rotation axis<br>across to longitudinal axis of the<br>endoscope;<br>Milled of aluminium, black coated | |
| Light guide connector | Compatible to ACMI, Richard Wolf<br>and STORZ light guide cable;<br>Manufactured of stainless steel | Compatible to Richard Wolf light<br>guide cable;<br>Manufactured of stainless steel | |
| Leak tester connector | XION-standard;<br>Manufactured of stainless steel | Richard Wolf-standard;<br>Manufactured of stainless steel | |
| Biopsy channel entry | Compatible to FUJINON;<br>Manufactured of stainless steel | ISO594-2.2(Luer-Lock-innercone);<br>Manufactured of stainless steel | |
| Eyepiece | DIN 58105 (Medical endoscopes);<br>Manufactured of medical grade<br>plastics | DIN 58105 (Medical endoscopes);<br>Manufactured of medical grade<br>plastics | |
| Body cover | Manufactured of medical grade<br>silicone | Milled / drilled of aluminium, black<br>coated | |
| Suction valves<br>(bronchoscope only) | XION construction, cleanable by<br>removing;<br>Manufactured of stainless steel,<br>PTFE and medical grade silicone | Richard Wolf construction,<br>cleanable by removing;<br>Manufactured of stainless steel | |
| Focus ring | Handling by turning, rotation axis<br>parallel to longitudinal axis of the<br>endoscope;<br>Manufactured of medical grade<br>plastics | Handling by turning, rotation axis<br>parallel to longitudinal axis of the<br>endoscope:<br>Manufactured of medical grade<br>plastics | |
| Labelling | Image: XION Device Labelling | Image: Richard Wolf Device Labelling | |
## Technological Characteristics in Comparison to Predicate Device
.
{2}------------------------------------------------
:
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for XION medical. The logo consists of the word "XION" in a bold, sans-serif font, with a small dot above the "I". Below the word "XION" is the word "medical" in a smaller, italicized font. The logo is simple and modern, and the use of black and white gives it a professional look.
| Similarities to predicate<br>device: | The submitted devices are equivalent to the flexible endoscopes referred<br>to in K992526 flexible nasopharyngo-laryngoscope and flexible<br>bronchoscope. The fiberscopes in the submission use the same basic<br>design and device material as submitted in K992526.<br>Some flexible nasopharyngoscopes have a fixed light cable instead of a<br>removable one; some of them have a working channel. |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Differences to predicate<br>device: | There are only differences regarding diameters and lengths of the<br>insertion tube and the outer appearance and shaping of the shell parts<br>and control elements. |
The submitted devices pose the same type of questions about safety and effectiveness as the compared devices. The new technological characteristics have no influence on safety or effectiveness. The submitted devices are substantially equivalent to the 510(k) devices sold by Richard Wolf Medical Instruments Corp.
{4}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Xion GmbH c/o Alexander Kliem Quality Management Pankstraße 8-10 D-13127 Berlin Germany
JUN 1 3 2007
Re: K070580
Trade/Device Name: Nasophyngoscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: May 11, 2007 Received: May 14, 2007
Dear Mr. Kliem:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
## Page 2 – Alexander Kliem
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Egelman S. MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for XION medical. The logo consists of the word "XION" in bold, sans-serif font, with a small circle above the "I". Below the word "XION" and slightly offset to the right is the word "medical" in a smaller, italicized font.
## 5 Indications for Use
Nasopharyngoscope
510(k) Number (if known):
Device Name:
Indications for Use:
Nasopharyngoscopes are intended to examine or treat the nasal cavity and nasal pharynx. They are used between the upper respiratory tracts of the nasal passage and the vocal chords.
They allow the recognition of anomalies in the tissue of the accessible organs. The EF-N 14 and EF-N 14S models with their working channel allow the insertion of an instrument to carry out a biopsy.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anabolut
(Division Sign-Off)
Division of Ophthalmic Ear,
Nose and Throat Devises
**510(k) Number** K070580
Prescription Use (Per 21 CFR 801.109)
510 (k) Notification 061204 - 5 Indications for Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.