The AUSCO ES-3100 Electronic Stethoscope is intended for medical diagnostic purposes. It may be used for the amplification of sounds associated with the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment.
Device Story
AUSCO ES-3100 is an electronic stethoscope for clinical auscultation. Device captures internal body sounds (heart, lungs, arteries, veins) via a sensor; processes signals through active high and low pass filtering circuits to provide two frequency modes: Bell (20-200Hz) and Diaphragm (20-1000Hz). Amplified audio is delivered to the clinician via an active speaker and passive sound tubes. Features include volume control, frequency mode selection, and automatic power-off. Operated by healthcare professionals during physical assessments to aid in diagnosis by enhancing sound clarity and reducing frictional noise. Powered by two AAA alkaline batteries.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1 for limited skin contact. Electrical safety and EMC testing performed per IEC/ISO/EN 60601-01 and 60601-01-02.
Technological Characteristics
Electronic stethoscope with active high/low pass filtering. Frequency modes: Bell (20-200Hz), Diaphragm (20-1000Hz). Power: 2x AAA alkaline batteries. Biocompatibility: ISO 10993-1 compliant. Safety/EMC: IEC/ISO/EN 60601-01 and 60601-01-02. Embedded firmware controls volume, frequency selection, and power management.
Indications for Use
Indicated for medical diagnostic purposes, specifically the amplification of heart, lung, arterial, venous, and other internal organ sounds in any person undergoing a physical assessment.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
3M™ LITTMANN® Electronic Stethoscope, Model 3000 (K041934)
3M™ LITTMANN® Electronic Stethoscope, Model 2000 (K961848)
American Diagnostic Corporation's, Model ADC 656 Electronic Stethoscope (K012304)
Submission Summary (Full Text)
{0}------------------------------------------------
# Pre-Market Notification 510(k) Summary
# 1. Sponsor Information:
Koratek Company, LTD. 307, 4th Venture Building, Hoseo University 165 Sechul-ri, Baebang-myun Asan-city, Chungnam, 336-795 Republic of Korea
Contact Person: Contact Title: Contact Phone Number: Contact Fax Number:
Bong Woo Lee President 82-41-549-5437 82-41-549-5438
NOV 0 6 2007
Date of Summary:
February 20, 2007
# 2. Device Name and Classification:
| Common and Usual Name: | Electronic Stethoscope |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | AUSCO ES-3100 |
| Classification Name: | Electronic Stethoscope<br>(21 CFR § 870.1875(b)) |
| Performance Standards: | No applicable performance standards have been issued<br>under section 514 or under section 513(b) of the Food,<br>Drug and Cosmetic Act. |
# 3. Predicate Device(s):
3MTM LITTMANN® Electronic Stethoscope, Model 3000, K041934 3MTM LITTMANN® Electronic Stethoscope, Model 2000, K961848 American Diagnostic Corporation's, Model ADC 656 Electronic Stethoscope, K012304
## 4. Description of Device:
The AUSCO ES-3100 Electronic Stethoscope is a healthcare device that clectronically filters and amplifies sounds of the heart, lungs, arteries, veins and other internal organs, and transfers them to the user's ears through an active speaker and passive sound tubes. The AUSCO ES-3100 provides two filter frequency modes for auscultation: Bell (20 to 200Hz) and Diaphragm (20 to 1000Hz).
{1}------------------------------------------------
K070550/51
P2/2
The AUSCO ES-3100 incorporates electronics and embedded software, which controls all of the various features such as volume control, frequency mode selection, and automatic powering off of the stethoscope. Active high and low pass filtering circuits are employed to produce the bell and diaphragm frequency response modes that are used to listen to the heart, lungs and other body sounds, while reducing frictional noises.
The AUSCO ES-3100 does not employ any off-the-shelf-software.
The AUSCO ES-3100 uses two (2) AAA alkaline batteries.
## Indications for Use: ನ.
The AUSCO ES-3100 Electronic Stethoscope is intended for medical diagnostic purposes. It may be used for the amplification of sounds associated with the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment.
# Comparative Data for Determining Substantial Equivalence to Predicate Device: 6.
The similarities and differences of the AUSCO ES-3100 (new device) were compared to predicate devices (Littmann® models 2000 and 3000 and ADC 656). The proposed device under this new pre-market notification is similar in characteristics, materials, features, has similar performance features, intended use and indications for use as the Littmann " Model 2000 cleared under K961848 (September 26, 1997 clearance date), the Littmann® Model 3000 cleared under K041934 (October 15, 2004 clearance date) and American Diagnostic Corporation's ADC 656 electronic stethoscope cleared under K012304 (November 1, 2001 clearance date).
# Non-Clinical Performance Summary: 7.
Components of the AUSCO ES-3100 electronic stethoscope have been reviewed for biocompatibility and conform to ISO10993-Part 1 Biological Evaluation of Medical Devices for limited (≤ 24 hour) skin contact for direct patient exposure.
The AUSCO ES-3100 electronic stethoscope was tested and conforms to applicable safety and EMC requirements found in the IEC/ISO/EN 60601-01 and IEC/ISO/EN 60601-01-02 international standards.
#### 8. Conclusions:
The AUSCO ES-3100 electronic stethoscope is similar if not identical in materials, technological characteristics, and has the same intended use as the predicate devices. Any differences in technological characteristics between the AUSCO ES-3100 and the predicate devices do not raise any new questions of safety or effectiveness.
We conclude that the AUSCO ES-3100 electronic stethoscope is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned to the right of the text, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 0 6 2007
Koratek Company, LTD. c/o Mr. Paul Summer Director, Regulatory and Quality Systems Arkin Consulting Group 1733 Canton Lane Marietta, GA 30062
Re: K070550
AUSCO ES-3100 Electronic Stethoscope Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II (two) Product Code: DQD Dated: September 24, 2007 Received: September 26, 2007
Dear Mr. Summer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
・・
{3}------------------------------------------------
Page 2 – Mr. Paul Summer
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
ls.M
COR Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 4 Indications for Use Statement
# Indications for Use Statement
510(k) Number (if known): K070550
Device Name: AUSCO ES-3100 Electronic Stethoscope
Indications For Use:
The AUSCO ES-3100 Electronic Stethoscope is intended for medical diagnostic purposes. It may be used for the amplification of sounds associated with the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cllsl
(Division Sign-Off)
Page 1 of
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Division of Cardiovascular Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.