← Product Code [MNH](/productcode/MNH) · K070530

# MODIFICATION TO PASSMED SPINAL SYSTEM (K070530)

_Medicrea Technologies · MNH · Oct 18, 2007 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K070530

## Device Facts

- **Applicant:** Medicrea Technologies
- **Product Code:** [MNH](/productcode/MNH.md)
- **Decision Date:** Oct 18, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The PASSmed which includes the subject components, is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis). As a pedicle screw system PASSmed is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. It also includes hooks and a sacral plate indicated for degenerative disc disease (ddd) defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.

## Device Story

PASSmed Spinal System is a posterior pedicle screw system for spinal stabilization; components include pedicle screws, hooks, sacral plates, clamps, rods, nuts, and crosslink members. System extension adds 6mm diameter rods for in-situ bending and domino connectors for rod-to-rod coupling. Used by surgeons in clinical settings to treat thoracic, lumbar, and sacral spine instabilities. Components provide mechanical immobilization of spinal segments to facilitate fusion. Device benefits include structural support for spinal segments during healing of acute/chronic deformities or instabilities.

## Clinical Evidence

Bench testing only. Mechanical performance of additional components was evaluated according to ASTM F1717 and ASTM F1798 standards to demonstrate equivalence to commercially available PASSmed devices.

## Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136 and unalloyed titanium (T40) per ASTM F67. Components: Pedicle screws, hooks, sacral plates, clamps, rods, nuts, crosslink members, and domino connectors. Mechanical testing performed per ASTM F1717 and ASTM F1798.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Medicrea PASSMed Spinal System ([K032094](/device/K032094.md))
- Medtronic CD Horizon® ([K014296](/device/K014296.md))
- EBI® Array™ Spinal Fixation System ([K042772](/device/K042772.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# SPECIAL 510(K) SUMMARY

CT 1 8 2007

### 1. GENERAL INFORMATION

| Trade Name                         | PASSmed Spinal System                                                                                                    |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Common Name                        | ✓ Posterior pedicle screw system<br>✓ Sacral plate<br>✓ Hooks                                                            |
| Classification Name                | ✓ Pedicle screw spinal system<br>✓ Spinal interlaminal fixation orthosis                                                 |
| Class                              | Class II                                                                                                                 |
| Product Code                       | MNH / MNI / KWP                                                                                                          |
| CFR section                        | 888.3070 / 888.3050                                                                                                      |
| Device panel                       | Orthopedic                                                                                                               |
| Legally marketed predicate devices | Medicrea PASSMed Spinal System: K032094<br>Medtronic CD Horizon®: K014296<br>EBI® Array™ Spinal Fixation System: K042772 |
| Reason for special 510(k)          | Product range extension and additional components                                                                        |
| Submitter                          | MEDICREA® Technologies<br>Z.I. Chef de Baie<br>17000 La Rochelle, France                                                 |
| Contact                            | J.D. Webb<br>1001 Oakwood Blvd<br>Round Rock, TX 78681<br>512-388-0199<br>E-Mail: ortho.medix@sbcglobal.net              |

### PREDICATE DEVICE DESCRIPTION 2.

The PASSmed Spinal System consists of pedicle screws, hooks, sacral plates, clamps, rods, nuts, rod plates and crosslink members. It can be used for single or multiple level fixations. All components are manufactured from titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136 and titanium unalloyed (T40) that conforms to ASTM F67.

### DESCRIPTION OF DEVICE MODIFICATION 3.

The purpose of this submission is to include the following components in PASSmed Spinal System:

- · Ø6mm rod in length 500mm for in-situ bending: same design as rods cleared in K032094. This rod is manufactured from unalloyed titanium that conforms to ASTM F67
- " Connectors (dominoes) to connect a rod Ø6mm to another rod Ø6mm

## 4. INTENDED USE

The PASSmed which includes the subject components, is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudoarthrosis).

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the word "MEDIOREA" in a bold, black font. The letters are large and fill the frame. The word is set against a white background, and the overall impression is one of prominence and emphasis.

As a pedicle screw system PASSmed is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

It also includes hooks and a sacral plate indicated for degenerative disc disease (ddd) defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.

#### PERFORMANCE DATA 5.

When applicable, the tests performed on the additional components according to ASTM F1717 or ASTM F1798, indicate that the products are as mechanically sound as other PASSmed devices commercially available

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image contains the words "Public Health Service" in a simple, sans-serif font. The text is arranged vertically, with "Public Health" on the first line and "Service" on the second line. The words are in black color and the background is white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MEDICREA Technologies % Mr. J. D. Webb 1001 Oakwood Blvd. Round Rock, TX 78641

OCT 1 8 2007

Re: K070530 Trade/Device Name: PASSmed Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI, KWP Dated: September 16, 2007 Received: September 19, 2007

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. J.D. Webb

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbary Muelm

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if known): __ KOJOS 30

Device Name: PASSmed Spinal System

## PASSmed Spinal System

### Indications for Use

The PASSmed is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor and failed previous fusion (pseudoarthrosis).

As a pedicle screw system PASSmed is indicated for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

It also includes hooks and a sacral plate indicated for degenerative disc disease (ddd) defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, and/or lordosis), tumor, pseudoarthrosis and failed previous fusion.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

Barbara Buchner

and Neurological Devices

510(k) Number K076530

---

**Source:** [https://fda-staging.innolitics.com/device/K070530](https://fda-staging.innolitics.com/device/K070530)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).