← Product Code [MOB](/productcode/MOB) · K070458

# MODIFICATION TO FIDIS VASCULITIS, MODEL MX007 (K070458)

_Biomedical Diagnostics (Bmd) SA · MOB · Dec 21, 2007 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K070458

## Device Facts

- **Applicant:** Biomedical Diagnostics (Bmd) SA
- **Product Code:** [MOB](/productcode/MOB.md)
- **Decision Date:** Dec 21, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5660
- **Device Class:** Class 2
- **Review Panel:** Immunology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K070458 · Dec 21, 2007 | MODIFICATION TO FIDIS VASCULITIS, MODEL MX007 | Biomedical Diagnostics (Bmd) SA | Routine clinical laboratory samples; Previously characterized clinical samples (predicate device comparison) | Retrospective clinical samples were used to assess potential biological interferences, establish assay threshold values, and validate the performance of the modified device against the predicate device. | Retrospective clinical samples; Interference study; Comparative performance study |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective analysis of clinical samples | Samples with potential biological interferences (e.g., cryoglobulinemia, rheumatoid factor, hemolyzed sera); Sample Size: 27 | Not applicable for this study | Detection of interference in MPO, PR3, and GBM assays |
| Retrospective comparative study | Human serum samples; Sample Size: 219 | Predicate FIDIS™ VASCULITIS (K053012) | Positive/Negative percent agreement |

## Indications for Use

The FIDIS™ VASCULITIS® kit is a semi-quantitative homogeneous fluorescent-based microparticles immunoassay using flow cytometry readings. The test system is used to detect in patient serum samples the presence of anti-neutrophil cytoplasm antibodies (ANCA) directed against Myeloperoxidase (MPO) and Serine Proteinase 3 (PR3); and anti-glomerular basement membrane (GBM) antibodies. The results of the FIDIS™ VASCULITIS® test are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of various primary systemic small vessel vasculitis. The presence of anti-MPO and anti-PR3 antibodies associated primary systemic small vessel vasculitis: Wegener's granulomatosis, Churg Strauss syndromes, microscopic periarteritis and idiopathic crescentic glomerulonephritis; and the presence of anti-GBM antibodies is associated with Goodpasture's syndrome. FIDIS™ VASCULITIS® kit is used on the FIDIS Analyser, MLX-BOOSTER Software and FIDIS™ Washer. FIDIS™ VASCULITIS* kit could be used with CARIS™ system (diluting and dispensing device). This test is for in vitro diagnostic use.

## Device Story

In vitro diagnostic kit; detects anti-MPO, anti-PR3, and anti-GBM antibodies in patient serum. Uses color-coded microspheres covalently coupled to specific antigens; multiplexed assay format. Patient serum incubated with microspheres; bound antibodies detected via phycoerythrin-labeled anti-human IgG conjugate. Benchtop flow cytometer (Luminex 100/200) uses red laser for microsphere identification and green laser for fluorescence quantification. MLX-BOOSTER software processes signals; calculates antibody titers via interpolation against calibration system. Used in clinical laboratories; operated by trained personnel. Results aid clinicians in diagnosing systemic small vessel vasculitis and Goodpasture's syndrome. Optional CARIS system automates sample dilution and dispensing.

## Clinical Evidence

Comparative study on 219 clinical samples (135 positive, 84 negative) against predicate K053012. Results: MPO (97.18% positive agreement, 100% negative agreement), PR3 (95.92% positive, 99.41% negative), GBM (100% positive, 100% negative). Additional study on 117 samples compared manual vs. CARIS™ automated preparation, showing high agreement (98.29-100%). Precision studies (intra-assay and inter-assay) met acceptance criteria (CV% ≤15-20%).

## Technological Characteristics

Multiplexed fluorescent-based microparticle immunoassay. Components: color-coded microspheres coupled to MPO, PR3, and GBM antigens; phycoerythrin-labeled anti-human IgG conjugate. Platform: flow cytometry (Luminex 100/200). Energy: laser-based excitation. Connectivity: PC-based software (MLX-BOOSTER). Sample matrix: serum. Format: 96-well microplate with filtering membrane.

## Regulatory Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

## Predicate Devices

- FIDIS™ VASCULITIS (k053012)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k070458

B. Purpose for Submission:
New device

C. Measurand:
Anti-Myeloperoxidase Antibodies (MPO)
Anti-Serine Protease 3 Antibodies (PR3)
Anti-Glomerular Membrane Antibodies (GBM)

D. Type of Test:
Semi-quantitative

E. Applicant:
Biomedical Diagnostics, S.A. (bmd)

F. Proprietary and Established Names:
FIDIS™ VASCULITIS, MX 007

G. Regulatory Information:
1. Regulation section:
21 CFR § 866.5660, Multiple autoantibodies, immunological test system
2. Classification:
Class II
3. Product codes:
MOB, Test System, antineutrophil antibodies (ANCA)
MVJ, Devices, Measure, Antibodies to Glomerular Basement Membrane (GBM)
4. Panel:
Immunology (82)

H. Intended Use:
1. Intended use(s):
The FIDIS™ VASCULITIS® kit is a semi-quantitative homogeneous fluorescent-based microparticles immunoassay using flow cytometry readings. The test system is used to detect in patient serum samples the presence of anti-neutrophil cytoplasm antibodies (ANCA) directed against Myeloperoxidase (MPO) and Serine Proteinase 3 (PR3); and anti-glomerular basement membrane (GBM) antibodies.
The results of the FIDIS™ VASCULITIS® test are to be used in conjunction with the clinical findings and the other laboratory tests to aid in the diagnosis of various primary systemic small vessel vasculitis.
The presence of anti-MPO and anti-PR3 antibodies associated primary systemic small vessel vasculitis: Wegener's granulomatosis, Churg Strauss syndromes, microscopic periarteritis and idiopathic crescentic glomerulonephritis; and the presence of anti-GBM antibodies is associated with Goodpasture's syndrome.
FIDIS™ VASCULITIS® kit is used on the FIDIS Analyser, MLX-BOOSTER Software and FIDIS™ Washer.

{1}

FIDIS™ VASCULITIS* kit could be used with CARIS™ system (diluting and dispensing device).

This test is for in vitro diagnostic use.

* Detection of the serologic markers for primary systemic small vessel vasculitis (ANCA) and for Goodpasture syndrome (GBM)

2. Indication(s) for use:

Same as Intended use.

3. Special conditions for use statement(s):

For prescription only.

4. Special instrument requirements:

FIDIS™ 100 or FIDIS™ 200 Analyzer (Luminex 100™ or 200™)

FIDIS™ MLX-007 Software

FIDIS™ XY Platform

FIDIS™ Sheath Delivery System (SD)

Computer Central Processing Unit (PC)

Computer Monitor Screen

FIDIS™ Washer (k041002)

Ultrasonic bath

I. Device Description:

Each device contains the following: distinct uniform size color-coded microspheres (each microsphere set is covalently coupled to one of the following antigens: MPO, PR3, and GBM) (ready to use); calibrator (ready to use); positive control (to be diluted); negative control (to be diluted); goat anti-human IgG coupled to phycoerythrin (ready to use); sample dilution buffer (ready to use); wash buffer PBS-Tween (ready to use); one 96 wells microplate including a filtering membrane and a lid.

J. Substantial Equivalence Information:

1. Predicate device name(s):

FIDIS™ VASCULITIS

2. Predicate K number(s):

k053012

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | New Device | Predicate Device  |
|   | FIDIS™ VASCULITIS (MX 007) | FIDIS™ VASCULITIS  |
|  Intended use | Individual determination of IgG antibodies to MPO, PR3, GBM | Same  |
|  Antigen | Purified MPO, PR3, GBM | Same  |
|  Sample type | Serum | Same  |
|  Type of test | Semi-quantitative | Same  |
|  Platform | 96 well plates | Same  |
|  Technology | Flow Cytometry | Same  |
|  Assay Format | Multiplexed | Same  |

{2}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | New Device | Predicate Device  |
|  Sample dilution | 1:200 | Same  |
|  Substrate | None | Same  |
|  Enzyme-Conjugate | Phycoerythrin | Same  |
|  Detection method | Fluorescence | Same  |
|  Solid surface (antigen coated) | Color-coded microspheres | Same  |
|  Result Interpretation for each specificity: Anti-MPO, Anti-PR-3, Anti-GBM Antibody | Negative: < 20 AU/mL Borderline: 20-25 AU/mL Positive: >25 AU/mL | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Instrument | Luminometer (Luminex 100™ and/or Luminex 200™) | Luminometer (Luminex 100™)  |
|  Calibrator | 2 Wells/Test | 1 Well/Test  |
|  Software version | MLX Booster version 2.2 | MLX Booster version 1.35  |

# K. Standard/Guidance Document Referenced (if applicable):

None provided.

# L. Test Principle:

FIDIS™ VASCULITIS is based on the use of distinct uniform size color-coded microspheres and a bench top flow cytometer interfaced to digital signal processing hardware and software. A red diode laser beam in the flow cytometer classifies each set of microspheres on the basis of its unique fluorescence intensity (red to orange) thus identifying which analyte is being tested. At the same time, a green laser beam illuminates the external second molecule fluorescence to quantify the reaction related to the specific analyte.

Each antigen required for the assay is covalently coupled to an individual set of microspheres through its surface functional groups. The different antigen coupled microspheres are mixed together, constituting the final microspheres reagent.

The test is performed in a 96 wells blank microplate including a filtering membrane at the bottom of the wells.

- In the first step, the sample is distributed in each well containing the microspheres mixture. If this sample contains one or more of the suspected antibodies, this(ese) antibody(ies) bind to the corresponding antigen(s) on the various sets of microspheres.
After incubation, the unbound antibodies are removed by a wash step using a filtration process.

{3}

- A phycoerythrin labeled anti-human IgG conjugate is then added that binds to the previously bound antibodies.
A final wash step stops the reaction.
- The reaction is then directly measured by the flow cytometer, which differentiates each set of microspheres according to its fluorescence color while simultaneously measures the average fluorescence emitted by the conjugate.

A calibration system permits the determination of the antibody titer of each sample by interpolation for each antigenic specificity.

# M. Performance Characteristics (if/when applicable):

# 1. Analytical performance:

# a. Precision/Reproducibility:

For the intra-assay study, six samples for MPO, six samples for PR3, seven samples for GBM (weak, moderate and high titers) were analyzed 10 times (for each specificity) in one run. For the inter-assay study, the same samples were analyzed 3 times in 5 different runs. The intra-assay CV ranges for Anti-MPO, Anti-PR3, and Anti-GBM were:  $4.9\%$  to  $12.4\%$ ;  $3.3\%$  to  $11.6\%$ ; and  $2.3\%$  to  $11.3\%$  respectively. The inter-assay CV ranges on Anti-MPO, Anti-PR3, and Anti-GBM were:  $5.4\%$  to  $11.2\%$ ;  $10.9\%$  to  $19.6\%$ ; and  $3.8\%$  to  $14.6\%$  respectively (see table below).

|  Antigen | Sample # | Within-run |   | Between-run  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Mean value | CV (%) | Mean value | CV (%)  |
|  MPO | 1 | 15 | 5.3 | 18 | 11.2  |
|   |  2 | 28 | 7.5 | 28 | 6.2  |
|   |  3 | 33 | 12.4 | 35 | 10.9  |
|   |  4 | 54 | 8.0 | 56 | 10.5  |
|   |  5 | 261 | 10.8 | 261 | 10.1  |
|   |  6 | 391 | 4.9 | 392 | 5.4  |
|  PR3 | 7 | 16 | 11.3 | 20 | 19.6  |
|   |  8 | 30 | 9.7 | 31 | 12.0  |
|   |  9 | 37 | 11.6 | 43 | 14.2  |
|   |  10 | 73 | 3.3 | 76 | 10.9  |
|   |  11 | 198 | 6.1 | 222 | 13.8  |
|   |  12 | 266 | 7.2 | 303 | 13.4  |
|  GBM | 13 | 10 | 11.3 | 11 | 14.6  |
|   |  14 | 36 | 10.6 | 38 | 12.0  |
|   |  15 | 40 | 5.8 | 46 | 6.8  |
|   |  16 | 59 | 2.3 | 60 | 6.1  |
|   |  17 | 120 | 7.3 | 122 | 6.9  |
|   |  18 | 215 | 5.6 | 215 | 5.7  |
|   |  19 | 321 | 2.7 | 326 | 3.8  |

The reproducibility performance of FIDISTM VASCULITIS with the CARIS system (diluting and dispensing device) was determined as follows: Six samples for MPO, seven samples for PR3, six samples for GBM were analyzed 10 times in one run (on each specificity) for the intra-assay study.

{4}

The same samples were analyzed 3 times in 5 different runs for the inter-assay study. The intra-assay CV ranges on Anti-MPO, Anti-PR3, and Anti-GBM were:  $2.6\%$  to  $10.8\%$ ;  $3.8\%$  to  $7.2\%$ ; and  $3.8\%$  to  $7.1\%$  respectively. The inter-assay CV ranges on Anti-MPO, Anti-PR3, and Anti-GBM were:  $6.7\%$  to  $13.6$ ;  $5.9\%$  to  $11.7\%$ ; and  $5.9\%$  to  $9.7\%$  respectively (see table below).

|  Antigen | Sample # | Within-run |   | Between-run  |   |
| --- | --- | --- | --- | --- | --- |
|   |   |  Mean value | CV (%) | Mean value | CV (%)  |
|  MPO | 1 | 17 | 6.4 | 17 | 13.6  |
|   |  2 | 21 | 4.9 | 22 | 6.7  |
|   |  3 | 29 | 3.8 | 30 | 7.6  |
|   |  4 | 37 | 4.3 | 38 | 8.4  |
|   |  5 | 59 | 2.6 | 57 | 9.3  |
|   |  6 | 94 | 10.8 | 96 | 9.4  |
|  PR3 | 7 | 16 | 6.9 | 16 | 6.9  |
|   |  8 | 21 | 5.0 | 20 | 7.8  |
|   |  9 | 24 | 5.9 | 24 | 7.4  |
|   |  10 | 36 | 3.8 | 34 | 5.9  |
|   |  11 | 63 | 6.5 | 60 | 11.7  |
|   |  12 | 83 | 7.2 | 80 | 7.6  |
|   |  13 | 386 | 7.2 | 371 | 6.8  |
|  GBM | 14 | 13 | 4.9 | 12 | 8.2  |
|   |  15 | 42 | 7.0 | 42 | 5.9  |
|   |  16 | 55 | 3.8 | 53 | 6.5  |
|   |  17 | 90 | 6.4 | 87 | 7.9  |
|   |  18 | 115 | 7.1 | 117 | 9.7  |
|   |  19 | 392 | 4.8 | 400 | 7.5  |

# Lot to lot reproducibility:

Two lots were analyzed using 3 positive samples (2 near cut off) and 9 negative samples. The acceptable criterion is  $25\%$ . The lowest variability was  $2\%$  and highest variability was  $15\%$ . Both lots were within the  $25\%$  variability.

# b. Linearity/assay reportable range:

Linearity is not claimed in this assay.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): There are no reference standards for anti-MPO, anti-PR3 and anti-GBM. The calibrators and controls (positive and negative) were prepared in-house and assigned arbitrary units (AU/mL) during the development process.

# d. Detection limit:

Not applicable.

# e. Analytical specificity:

Interference study: Twenty seven samples were selected for evaluation of potential interference and crossreactivity: 2 Cryoglobulinemia, 8 Complement, 2 IgG monoclonal immunoglobulins, 3 IgM monoclonal immunoglobulins, 7 Rheumatoid Factor, 3 citrated plasmas, and 2 hemolyzed samples. Twenty six samples were negative and one RF sample was positive

{5}

for PR3. To ensure appropriate samples are used in the test, the package insert states to avoid grossly hemolyzed, lipemic, microbially contaminated, heat-treated, specimens with visible particulates, or samples with abnormal concentration of IgG and/ or complement levels, or samples with rheumatoid factor.

f. Assay cut-off:

The cut-off value of &gt;25 AU/mL was based on testing 65 samples consisted of 37 normal blood donor sera and 28 selected samples with potential biologic interferences. With this cut-off value, 37 normal donor and 28 selected samples (100%) were negative for MPO and GBM; and 37 normal donor and 27 selected samples (98.5%) were negative for PR3.

2. Comparison studies:

a. Method comparison with predicate device:

Testing was performed on 219 samples which included 135 positive for one or more ANCA antibodies and/or positive for GBM; and 84 negative samples. No information about age, gender, and clinical status were available.

Equivocal results were considered negative. The positive, negative and total percent agreements for Anti-MPO, Anti-PR3, and Anti-GBM are shown in tables below.

|  Anti-MPO | FIDISTM VASCULITIS MPO  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  FIDISTM VASCULITIS (MX 007) MPO | Positive | 69 | 0 | 69  |
|   |  Negative | 2 | 148 | 150  |
|   |  Total | 71 | 148 | 219  |

Positive percent agreement: 97.2% (69/71)

Negative percent agreement: 100% (148/148)

Overall percent Agreement: 99.1% (217/219)

|  Anti-PR3 | FIDISTM VASCULITIS PR3  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  FIDISTM VASCULITIS (MX 007) PR3 | Positive | 47 | 1 | 48  |
|   |  Negative | 2 | 169 | 171  |
|   |  Total | 49 | 170 | 219  |

Positive percent agreement: 95.9% (47/49)

Negative percent agreement: 99.4% (169/170)

Overall percent Agreement: 98.6% (216/219)

|  Anti-GBM | FIDISTM VASCULITIS GBM  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  FIDISTM VASCULITIS (MX 007) GBM | Positive | 25 | 0 | 25  |
|   |  Negative | 0 | 194 | 194  |
|   |  Total | 25 | 194 | 219  |

{6}

Positive percent agreement: 100% (25/25)

Negative percent agreement: 100% (194/194)

Overall percent Agreement: 100% (219/219)

Comparison of manual preparation and the automated CARISTM System: Testing was performed on 117 samples: 75 positive for one or more ANCA parameters and/or positive for GBM; and 42 negative samples including some samples evaluated for their potential biological interferences. Results are summarized in tables below.

|  Anti-MPO | Manual FIDIS  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  CARISTM FIDIS | Positive | 34 | 2 | 36  |
|   |  Negative | 0 | 81 | 81  |
|   |  Total | 34 | 83 | 117  |

Positive percent agreement: 100% (34/34)

Negative percent agreement: 97.6% (81/83)

Overall agreement: 98.3% (115/117)

|  Anti-PR3 | Manual FIDIS  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  CARISTM FIDIS | Positive | 25 | 2 | 27  |
|   |  Negative | 0 | 90 | 90  |
|   |  Total | 25 | 92 | 117  |

Positive percent agreement: 100% (25/25)

Negative percent agreement: 97.8% (90/92)

Overall agreement: 98.3% (115/117)

|  Anti-GBM | Manual FIDIS  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Positive | Negative | Total  |
|  CARISTM FIDIS | Positive | 23 | 0 | 23  |
|   |  Negative | 0 | 94 | 94  |
|   |  Total | 23 | 94 | 119  |

Positive percent agreement: 100% (23/23)

Negative percent agreement: 100% (94/94)

Overall agreement: 100% (117/117)

b. Matrix comparison:

Serum is the only recommended matrix.

3. Clinical studies:

a. Clinical Sensitivity:

Not applicable.

b. Clinical specificity:

Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

{7}

Not applicable.

4. Clinical cut-off:
Same as assay cut-off.

5. Expected values/Reference range:
Expected values in the normal population should be negative.

Table of incidence on Diseases with PR3, MPO and GBM antibodies

|  Diseases | PR3 | MPO | GBM | Reference  |
| --- | --- | --- | --- | --- |
|  Wegener's granulomatosis | 30-90% |  |  | 14  |
|   |  50-90% |  |  | 15  |
|  Churg Strauss Syndrome | 35% | 35% |  | 14  |
|   |  31% | 7% |  | 15  |
|  Microscopic periarteritis | 20-40% | 50% |  | 14  |
|   |  25-30% | 36% |  | 15  |
|  Idiopathic glomerulonephritis | 20-40% | 50% |  | 14  |
|   |   | 80% |  | 15  |
|  Goodpasture's syndrome or anti-GBM nephritis |  |  | 98% | 16  |

(14) Molloy P. ANCA and Associated disease: Update. PSA Consult, 2000, vol III, 5.
(15) Sanchez-Lallyoyer N. ANCA. Spectra Biologie, 1993; 93/3: 38-42.
(16) Rossert J. Goodpasture's disease. Orphanet encyclopedia, 2002: 1-4.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda-staging.innolitics.com/device/K070458](https://fda-staging.innolitics.com/device/K070458)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
