REMOVABLE ARYLIC HERBST, ALLESEE SNORE APPLIANCE (ASA), AND ENOCH SNORINATOR
K070327 · Sybron Dental Specialties, Inc. · LRK · May 25, 2007 · Dental
Device Facts
Record ID
K070327
Device Name
REMOVABLE ARYLIC HERBST, ALLESEE SNORE APPLIANCE (ASA), AND ENOCH SNORINATOR
Applicant
Sybron Dental Specialties, Inc.
Product Code
LRK · Dental
Decision Date
May 25, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the Removable Acrylic Herbst is for the reduction of snoring and mild to moderate obstructive sleep apnea (OSA) by moving the lower jaw into a prescribed relationship to the upper jaw. The intended use of the Enoch Snorinator is for the reduction of snoring and mild to moderate obstructive sleep apnea (OSA) by moving the lower jaw into a prescribed relationship to the upper jaw. The intended use of the Allesee Snore Appliance (ASA) is for the reduction of snoring and mild to moderate obstructive sleep apnea (OSA) by moving the lower jaw into a prescribed relationship to the upper jaw.
Device Story
Devices are intraoral appliances for snoring and mild to moderate OSA; prescribed by dentists. Removable Acrylic Herbst: upper/lower arch system; stainless steel telescoping Herbst mechanism; bilateral; supports lower jaw in dentist-determined anterior position. Enoch Snorinator: one-piece upper arch splint; pressure-formed; anterior ramp postures lower jaw forward. Allesee Snore Appliance (ASA): one-piece upper/lower pressure-formed splint; joined by orthodontic acrylic; rectangular airway; postures lower jaw forward. All devices function by mechanically repositioning the mandible to increase airway passage; patient-specific advancement determined by clinician. Benefits include reduced snoring and improved airway patency during sleep.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, intended use, and comparison to legally marketed predicate devices.
Technological Characteristics
Intraoral appliances; materials include orthodontic acrylic and stainless steel (Herbst mechanism). Form factors include one-piece pressure-formed splints or two-piece arch systems with telescoping rods. Mechanical principle: mandibular advancement via physical positioning. No software or energy source.
Indications for Use
Indicated for patients with snoring and mild to moderate obstructive sleep apnea (OSA) requiring mandibular repositioning.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
Predicate Devices
Removable Herbst Appliance (State University of New York at Buffalo, School of Dental Medicine)
{0}------------------------------------------------
K070327
Image /page/0/Picture/1 description: The image shows a logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a curved line above them. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are printed in a bold, sans-serif font. The word 'Submitter:' is printed below the company name.
MAY 2 5 2007
Allesee Orthodontic Appliances, Inc. 13931 Spring Street Sturtevant, WI 53177
Contact: Sybron Dental Specialties, Inc. 1717 W. Collins Avenue Orange, California 92867 (714) 516-7484 - Phone (714) 516-7488 - Facsimile Colleen Boswell - Contact Person Date Summary Prepared: May 2007
#### Device Name:
- Trade Name Removable Acrylic Herbst .
- . Common Name - Intraoral Device for Snoring and/or Sleep Apnea
- Classification Name Intraoral Device for Snoring and Intraoral Device for Snoring and � Obstructive Sleep Apnea, per 21 CFR § 872.5570
Devices for Which Substantial Equivalence is Claimed:
- State University of New York at Buffalo, School of Dental Medicine, Removable Herbst . Appliance
#### Device Description:
The Removable Acrylic Herbst appliance is an upper and lower arch system designed to hold the lower jaw forward which increases the lower airway passage and reduces the tendency to snore. The upper and lower appliances are connected by means of a telescoping male and female stainless steel tube (upper) and rod (lower) Herbst mechanism. The system is bilateral, left and right. While the patient can open and close their mouth, the lower jaw arcs as predetermined by the construction of the appliance. The degree to which the patient's lower jaw is supported in an anterior position is determined by the dentist.
#### Intended Use of the Device:
The intended use of the Removable Acrylic Herbst is for the reduction of snoring and mild to moderate obstructive sleep apnea (OSA) by moving the lower jaw into a prescribed relationship to the upper jaw.
#### Substantial Equivalence:
The Removable Acrylic Herbst is substantially equivalent to another legally marketed device in the United States. The Removable Acrylic Herbst is used in a manner similar to the Removable Herbst Appliance developed by the State University of NY at Buffalo, School of Dental Medicine.
{1}------------------------------------------------
# K070327
Image /page/1/Picture/1 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized, bold font, with a curved line above the letters. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are written in a smaller, sans-serif font. The logo is black and white.
#### Submitter:
Allesee Orthodontic Appliances, Inc. 13931 Spring Street Sturtevant, WI 53177
Contact:
Sybron Dental Specialties, Inc. 1717 W. Collins Avenue Orange, California 92867 (714) 516-7484 - Phone (714) 516-7488 - Facsimile Colleen Boswell - Contact Person Date Summary Prepared: May 2007
#### Device Name:
- . Trade Name - Enoch Snorinator
- . Common Name - Intraoral Device for Snoring and/or Sleep Apnea
- . Classification Name - Intraoral Device for Snoring and Intraoral Device for Snoring and Obstructive Sleep Apnea, per 21 CFR § 872.5570
Devices for Which Substantial Equivalence is Claimed:
- . Hays & Meade Inc., Snore Guard
#### Device Description:
The Enoch Snorinator is a one-piece appliance designed for the upper arch only. It is a pressureformed anterior repositioning splint which postures the lower jaw forward to alleviate snoring. The anterior portion is formed into a "ramp" shape to engage the lower incisors. This ramp postures the lower jaw into the prescribed position as determined by the dentist.
#### Intended Use of the Device:
The intended use of the Enoch Snorinator is for the reduction of snoring and mild to moderate obstructive sleep apnea (OSA) by moving the lower jaw into a prescribed relationship to the upper jaw.
#### Substantial Equivalence:
The Enoch Snorinator is substantially equivalent to another legally marketed device in the United States. The Enoch Snorinator is used in a manner similar to the Snore Guard appliance developed by Hays & Meade Inc.
> 1717 West Collins Avenue, Orange, CA 92867 800-537-7824 714-516-7400
{2}------------------------------------------------
# 070327
Image /page/2/Picture/1 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a curved line above them. Below the letters, the words 'SYBRON DENTAL SPECIALTIES' are printed in a bold, sans-serif font. The word 'Submitter:' is printed below the company name.
Allesee Orthodontic Appliances, Inc. 13931 Spring Street Sturtevant, WI 53177
Contact:
Sybron Dental Specialties, Inc. 1717 W. Collins Avenue Orange, California 92867 (714) 516-7484 - Phone (714) 516-7488 - Facsimile Colleen Boswell - Contact Person Date Summary Prepared: May 2007
#### Device Name:
- . Trade Name - Allesee Snore Appliance (ASA)
- . Common Name - Intraoral Device for Snoring and/or Sleep Apnea
- Classification Name Intraoral Device for Snoring and Intraoral Device for Snoring and ● Obstructive Sleep Apnea, per 21 CFR § 872.5570
#### Devices for Which Substantial Equivalence is Claimed:
- . Great Lakes Orthodontic Laboratory, NAPA Appliance
#### Device Description:
The Allesee Snore Appliance (ASA) is a one-piece upper and lower, pressure-formed splint. The upper and lower jaw plastic appliances are joined together with orthodontic acrylic. The appliance postures the patient's lower jaw in a forward or anterior position. The amount of advancement is patient specific and determined by the dentist. The appliance has a rectangular airway between the upper and lower anterior teeth.
#### Intended Use of the Device:
The intended use of the Allesee Snore Appliance (ASA) is for the reduction of snoring and mild to moderate obstructive sleep apnea (OSA) by moving the lower jaw into a prescribed relationship to the upper jaw.
#### Substantial Equivalence:
The Allesee Snore Appliance (ASA) is substantially equivalent to another legally marketed device in the United States. The Allesee Snore Appliance (ASA) is used in a manner similar to the NAPA Appliance produced by Great Lakes Orthodontic Laboratory.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name arranged in a circular pattern around a central emblem. The emblem consists of a stylized caduceus-like symbol, with three human profiles facing right, stacked on top of each other. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Colleen Boswell Vice President, Regulatory Affairs Sybron Dental Specialties, Incorporated 1717 West Collins Avenue Orange, California 92867
MAY 2 5 2007
Re: K070327
Trade/Device Name: Removable Acrylic Herbst, Allesee Snore Appliance (ASA), and Enoch Snorinator Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: May 11, 2007 Received: May 14, 2007
Dear Ms. Boswell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Boswell
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Suyite y. Michie Omp.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## K070327
### Indications for Use
#### 510(k) Number (if known): K070327
Device Name: Removable Acrylic Herbst
#### Indications for Use:
The Removable Acrylic Herbst is intended to be used for the reduction of snoring and mild to moderate obstructive sleep apnea (OSA) by repositioning the lower jaw into a prescribed relationship to the upper jaw.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
RSBer DOS for Dr. Susan Runner
ospital
{6}------------------------------------------------
K070327
### Indications for Use
510(k) Number (if known): K070327
Device Name: Allesee Snore Appliance (ASA)
Indications for Use:
The Allesee Snore Appliance (ASA) is intended to be used for the reduction of snoring and mild to moderate obstructive sleep apnea (OSA) by repositioning the lower jaw into a prescribed relationship to the upper jaw.
× Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert Betz DDS for Dr Susan Runner
Hospital
{7}------------------------------------------------
# 70327
### Indications for Use
510(k) Number (if known): K070327
Device Name: Enoch Snorinator
#### Indications for Use:
The Enoch Snorinator is intended to be used for the reduction of snoring and mild to modcrate obstructive sleep apnea (OSA) by repositioning the lower jaw into a prescribed relationship to the upper jaw.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robert Betz DDS for Dr. Susan Rummer
Hospital
310(k): K070327
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.