EDWARDS AQUARIUS HEMOFILTRATION SYSTEM, MODEL ABM 01
Applicant
Edwards Lifesciences Services GmbH
Product Code
KDI · Gastroenterology, Urology
Decision Date
Jun 7, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K070320 · Jun 7, 2007
EDWARDS AQUARIUS HEMOFILTRATION SYSTEM, MODEL ABM 01
Edwards Lifesciences Services GmbH
European market surveillance and vigilance activities; Sales data of tubing sets
The sponsor provided European postmarket surveillance data to demonstrate the device's established field experience, noting approximately 1,500 units in use performing 64,000 treatments annually.
Postmarket surveillance; European market experience; Vigilance activities
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
European surveillance and vigilance activities; Postmarket surveillance / observational summary; Follow-up/Duration: Six years of European market experience; Study Period: 2006 (figures)
Patients undergoing renal replacement or plasma therapies in Europe; Sample Size: Approximately 1,500 machines
Not applicable for this study
Treatment volume (estimated 64,000 treatments per year)
Indications for Use
The Edwards Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Edwards Aquarius system may also be used in Therapeutic Plasma Exchange (TPE) therapies. The Edwards Aquarius system is indicated for use in a clinical setting and not for home use.
Device Story
Automated fluid balance monitor for renal replacement or plasma therapies; used in clinical settings by healthcare professionals. Device manages three circuits: extracorporeal (blood), substitution/dialysate, and filtrate. Inputs include blood flow, substitution/dialysate volumes, and pressure readings from arterial, return, pre-filter, and filtrate sensors. System utilizes two scales for precise fluid balancing and an anticoagulant (heparin) pump. Features include an ultrasonic air detector, blood leak detector, and integrated fluid warmer. Output consists of real-time monitoring of fluid balance and therapy parameters on a TFT color display. Healthcare providers use data to adjust therapy settings, ensuring patient safety via a 2-channel protective system. Benefits include precise fluid removal and solute clearance in critically ill patients.
Clinical Evidence
No clinical testing included. Substantial equivalence supported by comparative bench testing against the Baxter Accura System and documentation of European field experience (approx. 1,500 machines, 64,000 treatments/year).
Technological Characteristics
Automated fluid balance monitor; 10.4" TFT color display; 18-bit converter for weighing scales; ultrasonic air detection; blood leak detector (2-4 ml blood/1l filtrate); integrated plate heater (35-39°C). Connectivity includes RS232 optical connector. Operates on 115V/230V. Software-controlled pumps and pressure sensors. Materials and manufacturing processes are identical to the Baxter Accura System.
Indications for Use
Indicated for patients with acute renal failure or fluid overload requiring continuous solute/fluid removal or Therapeutic Plasma Exchange (TPE). For clinical use only; not for home use.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Center for Devices and Radiological Health
KC 7032C
Page 1 of 8
1/31/2007
Edwards Lifesciences Edisonstrasse 6 85716 Unterschleissheim Germany Telephone: +49 89 95475 203 JUN - 7 2007
Image /page/0/Picture/4 description: The image shows a logo with the word "Edwards" written in a serif font at the bottom. Above the word, there is a graphic element that appears to be a stylized letter "E" or a design composed of squares and rectangles. The logo has a simple and clean design.
# 510(k) Summary as required by 21 CFR 807.92 (c)
Date Summary Prepared:
January 31, 2007
#### Owner's Name:
Fax: +49 89 95475 301
| Edwards Lifesciences Services GmbH | |
|------------------------------------|-------------------------------------------------------------|
| Address: | Edisonstrasse 6<br>Unterschleissheim<br>D-85717<br>Germany |
| Tel: | +49 89 95475 203 |
| Fax: | +49 89 95475 301 |
| Contact Person: | Robert Madjno<br>Director of Regulatory and Quality Affairs |
# 510(k) Contact Persons:
Neil R. Armstrong Regulatory Affairs Advisor MeddiQuest Limited Business and Technology Centre Bessemer Drive Stevenage SG1 2DX United Kingdom Tel: +44 1763 222505 Fax: +44 1763 222049
or
{1}------------------------------------------------
Robert Madino Director of Regulatory and Quality Affairs Edwards Lifesciences Services GmbH Edisonstrasse 6 85716 Unterschleissheim Germany +49 89 95475 203 Tel: +49 89 95475 301 Fax:
## Type of 510(k):
Traditional
#### Classification Name:
High Permeability Hemodialysis Systems
78 KDI
## Common/Usual Name:
Hemofiltration System
## Proprietary Name:
Edwards Aquarius System
## Establishment Registration Number:
The device will be manufactured for: Edwards Lifesciences Services GmbH Edisonstrasse 6 Unterschleissheim D-85717 Germany Establishment Registration Number applied for
by
MeSys GmbH Beneckallee 30 Hannover D-30419 Germany Establishment Registration Number 9680719
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows handwritten text on a white background. The text includes "KC70320C", "Page 3 of 8", and the date "1/31/2007". The handwriting appears to be cursive and slightly slanted. The text is arranged in three lines, with the date at the bottom.
## Substantial Equivalence:
The Edwards Aguarius system is based on product manufactured by MeSys, which has also been used as the basis for the Baxter Accura System - K021615.
The Edwards Aquarius system has been marketed and developed in Europe for six years. The software has been significantly improved from that used in the Baxter Accura System and the hardware has been similarly improved reflecting the European field experience.
The Edwards Aquarius system is substantially equivalent in design, use and materials to the:
Baxter Accura System Gambro Prisma System B. Braun Diapact
K021615 K993064, K981681, K946279 K973322
The Edwards Aquarius system is made of the same materials as the Baxter Accura System (K021615).
The Edwards Aquarius system is manufactured by the same processes as the Baxter Accura System (K021615).
Comparison of Technological Features:
| Feature | Actual Device:<br>Edwards Aquarius | Predicate Device A:<br>Baxter Accura<br>System,<br>K021615 |
|---------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Housing | Front panel | Different - repositioning |
| | Rear panel | Same, except color |
| | Interior (including card<br>slots) | Same, except ADU and<br>new BLD |
| Cart | Trolley on rollers<br>equipped with brakes | Same, except different<br>tube surface treatment |
| Computer systems | | Same |
| Display | Solid State Color<br>monitor | Same |
| | Monitor casing | Same |
| Motherboard | | Same, except Accura<br>battery |
| Master extension | Extension for ADU | Not available |
| Feature | Actual Device:<br>Edwards Aquarius | Predicate Device A:<br>Baxter Accura<br>System,<br>K021615 |
| board | | |
| Serial connector<br>(RS232) | Optical connector | Same |
| Power supply | | Same |
| Wiring | Cable form | Same, except ADU<br>connection |
| Blood pump with<br>rotation sensor | | Same |
| Heparin pump | Hardware and Software | Different |
| Arterial pressure<br>sensor | | Same |
| Prefilter pressure<br>sensor | | Same |
| Filtrate pressure<br>sensor | Membrane unit holder<br>Pressure transducer<br>Evaluation electronics<br>Software | Same<br>Same<br>Same with later Accura<br>SW ver. 2.05<br>Same |
| Bubble trap and<br>venous tube clamp | | Same |
| Blood leak detector | Housing<br>Evaluation electronics<br>Software | Same<br>Different, change to<br>SMD technology<br>Different, changed<br>values for alarm limits |
| Predilution pump | | Same |
| Filtrate pump | | Same |
| Postdilution pump | | Same |
| Substitution fluid<br>weighing scales | Weighing scales hooks | Different, more<br>ergonomic with<br>Aquarius |
| | Weighing device<br>Evaluation electronics<br>(18 bit converter)<br>Software | Same<br>Different, change to<br>SMD technology<br>Same |
| Filtrate weighing<br>scales | Weighing scales hooks | Different, more<br>ergonomic with<br>Aquarius |
| Feature | Actual Device:<br>Edwards Aquarius | Predicate Device A:<br>Baxter Accura<br>System,<br>K021615 |
| | Weighing device | Same |
| | Evaluation electronics<br>(18 bit converter) | Different, change to<br>SMD technology |
| | Software | Same |
| ADU | Pressure monitoring<br>and pump | Not available at<br>Accura |
| Heater | Mechanical structure | Same |
| | Plate heater | Same |
| | Temperature Monitor | Different, integrated to<br>heparin pump housing |
| | Driver and evaluation<br>electronics | Different, change to<br>SMD technology |
| | Software | Same |
| AC Power | Operating Voltage | Same |
| | Input Line Current | Same |
{3}------------------------------------------------
KC 70320
Page 4 of 8
1/31/2007
{4}------------------------------------------------
KC7032C
Page 5 of 8
1/31/2007
## Description of Product:
The Edwards Aquarius System needs to be used in conjunction with a tubing set and a filter to provide Hemofiltration treatment to the patient.
The tubing set has already been the subject of a PreMarket Notification, K063293 Edwards Aqualine Sterile Tubing Sets, and the filter will be the subject of a separate PreMarket Notification, currently in preparation.
The Aquarius system is an Automated Fluid Balance Monitor, designed to be used with various extracorporeal treatments in the field of renal replacement therapies or plasma therapies. All therapies must be prescribed by a physician.
The Aquarius system is divided into three circuits: the extracorporeal (blood) circuit, the substitution/dialysate circuit and the filtrate circuit.
Toxic substances are removed by filters and clean blood is returned to the patient.
The Aquarius system allows the patient to be positioned left or right of the instrument.
{5}------------------------------------------------
KC70320
Page 6 of 8
1/31/2007
The Aquarius system uses two scales to accurately measure and precisely balance filtration and substitution volumes.
Heparin may be supplied to the extracorporeal circuit via an anticoagulant pump (Heparin pump). The prescribing physician may select continuous or intermittent options.
The Aquarius protective system is designed as a 2-channel system to protect the patient from foreseeable danger.
At the back of the scale system a removable hand-crank is mounted. This can be used to manually turn the blood pump in case a pump stops.
The Aquarius system is portable. It has a wheeled base connected with a handle to move or carry the Aquarius,
| Fluids Management | Low Blood Flow | Regular Blood Flow |
|-------------------------|-------------------------------------|------------------------|
| Pump accuracy | ±5% | ±5% |
| Blood | 10 to 200 ml/min | 30 to 450 ml/min |
| | by 2 ml/min | by 10 ml/min |
| Pre-dilution | 0-100 to 6000 ml/h | 0-100 to 10000 ml/h |
| Post-dilution | 0-100 to 4000 ml/h | 0-100 to 10000 ml/h |
| Dialysate | 0-100 to 10000 ml/h<br>(CVVHD only) | 0-100 to 10000 ml/h |
| Filtrate | 0-100 to 11000 ml/h | 0-100 to 12000 ml/h |
| Net fluid loss | 0-10 to 1000 ml/h<br>by 1 ml | 0-10 to 2000 ml/h |
| Heparin pump accuracy | ± 0.2 ml/h | ± 0.2 ml/h |
| Heparin pump settings | 0-15 ml/h by 0.5 ml/h | 0-15 ml/h by 0.5 ml/h |
| Heparin syringe size | 30 or 50 ml | 30 or 50 ml |
| Bolus function | 0.5 to 2.5 ml by | 0.5 to 2.5 ml by |
| | 0.5 ml bolus increment | 0.5 ml bolus increment |
| Scales | Low Blood Flow | Regular Blood Flow |
| Fluid balance alarm | 20g | 50g |
| Substitution scale | 0-30kg | 0-30kg |
| capacity | (max. load: 23kg) | (max. load: 23kg) |
| Filtrate scale capacity | 0-30kg | 0-30kg |
| | (max. load: 23kg) | (max. load: 23kg) |
| Fluid warmer | Low Blood Flow | Regular Blood Flow |
Significant Physical and Performance Characteristics:
{6}------------------------------------------------
| Accuracy | $\pm$ 0.3 °C | $\pm$ 0.3 °C |
|---------------|---------------------------------|---------------------------------|
| Working range | Off or 35°C<br>to 39°C by 0.5°C | Off or 35°C<br>to 39°C by 0.5°C |
| Pressure Monitoring | |
|--------------------------------|-------------------------------------------------|
| Sensor accuracy | $\pm$ 5 mmHg |
| Access | -250 to +200 mmHg |
| Return | -50 to +350 mmHg |
| Pre-filter | -150 to +500 mmHg |
| Filtrate | -250 to +400 mmHg |
| Degassing unit | |
| Pressure | -300 to +30 mmHg |
| Gas removal Max. | 10 ml/min |
| Weight and Dimensions | |
| H x W x D | 170 cm x 50 cm x 60 cm |
| Floor space | Approx. 55 cm (W) x 65 cm (D) |
| Weight | Approx. 75 kg |
| Power Requirements | |
| European voltage | 230 V (alternating voltage) $\pm$ 10%,<br>50 Hz |
| Power consumption | 500 VA |
| North American voltage | 115 V (alternating voltage) $\pm$ 10%,<br>60 Hz |
| North American consumption | 500 VA |
| Monitor / Detection Parameters | |
| Air detector | Ultrasonic measurement |
| Blood leak detector | 2 or 4 ml blood/1l filtrate at<br>HCT 32% |
| Display monitor | 10.4" TFT color |
## Intended use:
The Edwards Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Edwards Aquarius system may also be used in Therapeutic Plasma Exchange (TPE) therapies.
The Edwards Aquarius system is indicated for use in a clinical setting and not for home use.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows handwritten text on a white background. The text includes "K070320", "Page 8 of 8", and "1/31/2007". The handwriting is cursive and slightly slanted. The text appears to be part of a document or record.
# Testing Summary:
## Electrical Safety and Electromagnetic Compatibility:
Testing has been conducted following the IEC 60601 series of standards, which are FDA recognized voluntary consensus standards.
#### Bench Testing:
Comparative bench testing of the Edwards Aquarius System and Baxter Accura System, using an Edwards Aqualine Sterile Tubing Set K063293, is included. This concludes that the two machines have substantially equivalent performance.
#### Animal Testing:
No animal testing is included.
There are no patient or patient's blood contacting materials.
## Clinical Testing:
No clinical testing is included.
#### Other Information:
A summary of European surveillance and vigilance activities is included.
Currently approximately 1,500 machines have been placed on the European market and it is estimated these now (2006 figures) perform approximately 64,000 treatments per year (based on sales of tubing sets).
| - H . TIL<br>Onco<br> | SATE<br>ANNO AND AN E & . | ------------<br>9000 | 00 | | A CHINE A M<br>Career<br>C | |
|-----------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| P. L. 1 . Ph.<br>and the delivery of Children | 1<br>M-Added and see a from the commend of the commend of the | CHANGE OF AN | | CALLER ARA | SALES CONSULER AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND | A PRODUCT OF COLL CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLAIM OF CLA<br>Status Controllers of Children<br>1 |
| www.go.government.com/shoulers | A PROPERTY COLLECTION OF THE | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A A A C C Status Concess of | MARK AND ANNAL BARREN BARREN BALLERY<br>Parcellers of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution of the contribution of | | Controllation Company of | <br>Phone Charmore<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Robert Madjno Director of Regulatory and Quality Affairs Edwards Lifesciences Services GmbH
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a seal for the Department of Health & Human Services (USA). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Edwards Lifesciences Services GmbH c/o Neil R. Armstrong Regulatory Affairs Advisor MeddiQuest Limited Business and Technology Centre Bessemer Drive Stevenage SG1 2DX UNITED KINGDOM
JUN - 7 2007
Re: K070320
Trade/Device Name: Edwards Aquarius System Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: May 14, 2007 Received: May 21, 2007
Dear Mr. Armstrong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/8/Picture/10 description: The image shows a circular logo with the text "1906-2006 FDA Centennial" in the center. The letters "FDA" are prominently displayed in a bold, stylized font. Three stars are arranged below the word "Centennial". The text around the circle reads "U.S. Department of Health and Human Services".
Protecting and Promoting Public Health
{9}------------------------------------------------
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21. CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours, .
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use
4070320 510(k) Number (if known):
Edwards Aquarius System Device Name:
Indications for Use:
The Edwards Aquarius System is indicated for continuous solute and/or fluid removal in patients with acute renal failure or fluid overload. The Edwards Aquarius system may also be used in Therapeutic Plasma Exchange (TPE) therapies.
The Edwards Aquarius system is indicated for use in a clinical setting and not for home use.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Posted November 13, 2003)
Page 1 of 1
Nancy Brogdon
Division of Repr ductive, J and Radiological Devi
Radiological Devices
510(k) Number K070320
Indications for Use Statement 4
Page 4.2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.