The Durom Hip Resurfacing system femoral component, when used for hemi-hip arthroplasty, has been designed for cemented use in patients who are likely to outlive a conventional hip prosthesis. Femoral hip resurfacing hemi arthroplasty is most appropriate for patients with good bone quality in the femoral head and acetabulum, where the bearing surface and supportive bone structure of the acetabulum is normal, and where acetabular replacement is neither required nor desirable. Such patients will generally be under the age of 65. - . Non inflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis if remaining bone stock is adequate, and osteoarthritis - . Inflammatory joint disease (IJD), e.g. rheumatoid arthritis - . Joint replacements may be considered for younger patients if any unequivocal indication outweighs the risks associated with the age of the patient and modified demands regarding activity and hip joint loading are assured. This includes severely crippled patients with multiple joint involvement for whom an immediate need of hip mobility leads to an expectation of significant improvement in the quality of their lives Additional indications include other abnormalities where major pathology affects the femoral head: where the bearing surface and supportive bone structure of the acetabulum is normal; and where acetabular replacement is neither required nor desirable * WARNING: The Durom Femoral component is not approved for use with the Durom Acetabular cup for total hip surface replacement arthroplasty in the USA.
Device Story
Durom Hip Resurfacing System femoral component; monoblock head with slim-line femoral guide pin; extended hemisphere design (>180°); 12 sizes (38mm-60mm OD). Used in hemi-hip resurfacing arthroplasty; implanted by orthopedic surgeons in clinical settings. Underside features recesses for cement pressurization into cancellous bone; provides rotational stability. Device replaces femoral head bearing surface while preserving acetabulum; intended for younger, active patients likely to outlive conventional prostheses. Benefits include hip mobility restoration and improved quality of life for patients with femoral head pathology.
Clinical Evidence
Bench testing only. Non-clinical laboratory performance testing demonstrated safety and effectiveness; no clinical data provided.
Technological Characteristics
Monoblock femoral head; extended hemisphere geometry; 12 sizes (38-60mm OD); sphericity controlled to <10 microns. Features recesses on underside for cement pressurization and rotational stability. Metallic construction. Cemented fixation.
Indications for Use
Indicated for patients generally under 65 requiring hemi-hip resurfacing arthroplasty with good femoral head/acetabular bone quality and normal acetabular bearing surfaces. Conditions include NIDJD (avascular necrosis, osteoarthritis) and IJD (rheumatoid arthritis). Applicable to younger patients with multiple joint involvement requiring improved mobility. Contraindicated for total hip surface replacement (not approved for use with Durom Acetabular cup).
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic resurfacing prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral resurfacing component made of alloys, such as cobalt-chromium-molybdenum.
Predicate Devices
Cemented Femoral Head Resurfacing Device (K021799)
Cormet 2000 Hemi Hip Metallic Resurfacing Prostheses (K994153)
Submission Summary (Full Text)
{0}------------------------------------------------
K070292 Pg 1 of 3
# Summary of Safety and Effectiveness
. . .
# APR 2 6 2007
| Submitter: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Anthony Francalancia, RAC<br>Senior Associate, Regulatory Affairs<br>Telephone: (574) 372-4570<br>Fax: (574) 372-4605 |
| Date: | April 9, 2007 |
| Trade Name: | Durom® Hip Resurfacing System, Femoral<br>Components |
| Common Name: | Prosthesis, hip, femoral, resurfacing |
| Classification Name<br>and Reference | Hip joint femoral (hemi-hip) metallic resurfacing<br>prosthesis 21 CFR § 888.3400 |
| Device Product Code and Panel Code: | Orthopedics/87/KXA |
| Predicate Devices: | Cemented Femoral Head Resurfacing Device,<br>manufactured by Biomet Orthopedics, K021799,<br>cleared June 26, 2002.<br>DePuy ASRTM Resurfacing Femoral Heads,<br>manufactured by DePuy Orthopaedics, K032659,<br>cleared December 4, 2003.<br>BIOPRO TARA Femoral Resurfacing Component,<br>manufactured by BioPro, Inc., K043542, cleared<br>May 18, 2005.<br>The Cormet 2000 Hemi Hip Metallic Resurfacing<br>Prostheses, manufactured by Corin Medical,<br>K994153, cleared February 25, 2000. |
| Device Description: | The Durom® Hip Resurfacing System femoral<br>component consists of a monoblock head and slim-<br>line femoral guide pin. The head is an extended<br>hemisphere (greater than 180°). It is available in 12<br>sizes with Outside Diameters (ODs) ranging from<br>38mm to 60mm. The spherecity of the head is<br>tightly controlled to less than 10 microns. The |
{1}------------------------------------------------
underside of the head has recesses for controlled pressurization of cement into the cancellous bone, and for rotational stability after implantation.
V 070247
#### Intended Use:
## FOR HEMI-HIP RESURFACING ARTHROPLASTY:
The Durom Hip Resurfacing system femoral component, when used for hemi-hip arthroplasty, has been designed for cemented use in patients who are likely to outlive a conventional hip prosthesis. Femoral hip resurfacing hemi arthroplasty is most appropriate for patients with good bone quality in the femoral head and acetabulum, where the bearing surface and supportive bone structure of the acetabulum is normal, and where acetabular replacement is neither required nor desirable. Such patients will generally be under the age of 65.
- . Non inflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis if remaining bone stock is adequate, and osteoarthritis
- . Inflammatory joint disease (IJD), e.g. rheumatoid arthritis
- . Joint replacements may be considered for younger patients if any unequivocal indication outweighs the risks associated with the age of the patient and modified demands regarding activity and hip joint loading are assured. This includes severely crippled patients with multiple joint involvement for whom an immediate need of hip mobility leads to an expectation of significant improvement in the quality of their lives
Additional indications include other abnormalities where major pathology affects the femoral head: where the bearing surface and supportive bone structure of the acetabulum is normal; and where acetabular replacement is neither required nor desirable
* WARNING: The Durom Femoral component is not approved for use with the Durom Acetabular cup for total hip surface replacement arthroplasty in the USA.
The femoral components of the Durom® Hip Resurfacing System have the same intended use,
Comparison to Predicate Device:
{2}------------------------------------------------
ROTOATI py 2 of 3
similar performance characteristics, are made of the same material and are similar in design to the predicate devices.
## Performance Data (Non-clinical)
:
. . . .
The results of non-clinical (laboratory) performance testing demonstrate that the device is safe and effective.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Zimmer, Inc. % Mr. Anthony Francalancia P.O. Box 708 Warsaw, Indiana 46581-4605
APR 2 6 2007
Re: K070292 Trade/Device Name: Durom® Hip Resurfacing System, femoral components Regulation Number: 21 CFR 888.3400 Regulation Name: Hip joint femoral (hemi-hip) metallic resurfacing prosthesis Regulatory Class: II Product Code: KXA Dated: March 20, 2007 Received: March 21, 2007
Dear Mr. Francalancia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation
{4}------------------------------------------------
(21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 510(k) Number (if known): ¥ 070297
#### Device Name:
Durom® Hip Resurfacing System, femoral components
### Indications for Use:
#### FOR HEMI-HIP RESURFACING ARTHROPLASTY:
The Durom Hip Resurfacing system femoral component, when used for hemi-hip arthroplasty, has been designed for cemented use in patients who are likely to outlive a conventional hip prosthesis. Femoral hip resurfacing hemi arthroplasty is most appropriate for patients with good bone quality in the femoral head and acetabulum, where the bearing surface and supportive bone structure of the acetabulum is normal, and where acetabular replacement is neither required nor desirable. Such patients will generally be under the age of 65.
- Non inflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis if . remaining bone stock is adequate, and osteoarthritis
- . Inflammatory joint disease (IJD), e.g. rheumatoid arthritis
- Joint replacements may be considered for younger patients if any unequivocal . indication outweighs the risks associated with the age of the patient and modified demands regarding activity and hip joint loading are assured. This includes severely crippled patients with multiple joint involvement for whom an immediate need of hip mobility leads to an expectation of significant improvement in the quality of their lives
Additional indications include other abnormalities where major pathology affects the femoral head; where the bearing surface and supportive bone structure of the acetabulum is normal; and where acetabular replacement is neither required nor desirable
## * WARNING: The Durom Femoral component is not approved for use with the Durom Acetabular cup for total hip surface replacement arthroplasty in the USA.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
**(Division Sign-Off)**Page 1 of 1
Division of General, Restorative,
and Neurological Devices
510(k) Number K070292
0007
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.