FRESENIUS RENAL ACUTE CARE (RAC 100) FILTER, MODEL 0500610N
K070222 · Fresenius Medical Care North America · KDI · Apr 24, 2007 · Gastroenterology, Urology
Device Facts
Record ID
K070222
Device Name
FRESENIUS RENAL ACUTE CARE (RAC 100) FILTER, MODEL 0500610N
Applicant
Fresenius Medical Care North America
Product Code
KDI · Gastroenterology, Urology
Decision Date
Apr 24, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Intended for use in removing solutes or fluid in renal supportive therapies in an acute setting. This filter can be used for intermittent and continuous therapies. The RAC 100 Filter is intended for use in the following continuous veno-venous therapies: continuous veno-venous hemofiltration (CVVH) continuous veno-venous hemodialysis (CVVHD) The RAC 100 Filter is intended for use in the following intermittent therapies, e.g. slow low efficiency dialysis (SLED) or other renal supportive therapies of less than 8 hours duration.
Device Story
The Fresenius Renal Acute Care (RAC 100) Filter is a high-permeability dialyzer used in acute clinical settings for renal replacement therapy. It functions by removing solutes and excess fluid from patient blood during extracorporeal circulation. The device is intended for use in both continuous (CVVH, CVVHD) and intermittent (e.g., SLED) therapies. It is operated by healthcare professionals in a clinical environment as part of a renal therapy delivery system. The filter facilitates solute exchange and fluid management, aiding in the treatment of patients with acute renal failure or fluid overload. By providing effective filtration, the device helps restore fluid and electrolyte balance, potentially improving patient outcomes in acute care scenarios.
Clinical Evidence
No clinical data provided; substantial equivalence is supported by bench testing and comparison of technological characteristics to predicate devices.
Technological Characteristics
High-permeability dialyzer/hemofilter. Construction, design, and materials are equivalent to predicate devices. Operates as a passive filtration component within a renal therapy delivery system. No electronic or software components.
Indications for Use
Indicated for patients requiring solute or fluid removal in acute renal supportive therapies, including continuous veno-venous hemofiltration (CVVH), continuous veno-venous hemodialysis (CVVHD), and intermittent therapies (e.g., SLED) of less than 8 hours duration.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
APR 2 4 2007
Image /page/0/Picture/2 description: The image shows the logo and text for Fresenius Medical Care. The text below the logo reads "Fresenius Renal Acute Care (RAC 100) Filter" and "510(k) Summary". The logo is a stylized image of a filter.
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990.
### A. Submitter's Information:
| Name: | Fresenius Medical Care North America | |
|----------------------|---------------------------------------------|--|
| Address: | 920 Winter Street<br>Waltham, MA 02451-1457 | |
| Phone: | 1-781-699-4475 | |
| Fax: | (781) 699-9635 | |
| Contact Person: | Janet C. Kay RAC Regulatory Affairs Manager | |
| Date of Preparation: | January 23, 2007 | |
- B. Device Name:
Proprietary Name:
Common Name:
Fresenius Renal Acute Care (RAC 100) Filter
Dialyzer, High Permeability with or without Sealed Dialysate System
Product Code/Classification Panel:
78KDI/Gastroenterology-Urology
Classification:
Class 11 per §876.5860
510(k) for Fresenius Renal Acute Care (RAC 100) Filter 31
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with a geometric design above the text. The geometric design consists of three stacked, inverted trapezoids, creating a layered effect. The text below the design appears to be the beginning of a word, with only the letters 'us Med' visible, suggesting it is part of a larger company or brand name.
## Fresenius Medical Care Fresenius Renal Acute Care (RAC 100) Filter 510(k) Summary
#### C. Predicate Devices
The Fresenius Renal Acute Care (RAC 100) Filter is substantially equivalent to the Fresenius Optiflux F160NR, Gambro Hospal Multiflow 100 Hemofilter and the Gambro M100 Hemofilter.
Gambro Hospal Multifiow 100:
- K980386 (2/24/99) .
### Gambro M100
- · K032431 (9/15/03)
### Optiflux F160NR
- · K003498 (1/4/01)
#### D. Indications for Use/Intended Use:
Intended for use in removing solutes or fluid in renal supportive therapies in an acute setting. This filter can be used for intermittent and continuous therapies.
The RAC 100 Filter is intended for use in the following continuous veno-venous therapies:
- . continuous veno-venous hemofiltration (CVVH)
- continuous veno-venous hemodialysis (CVVHD) �
The RAC 100 Filter is intended for use in the following intermittent therapies, e.g. slow low efficiency dialysis (SLED) or other renal supportive therapies of less than 8 hours duration.
### E. Substantial Equivalence:
#### 1. Is the product a device?
YES - The Fresenius Renal Acute Care (RAC 100) Filter is a device pursuant to 21 CFR §201 [321] (h).
> 510(k) for Frescnius Renal Acute Care (RAC 100) Filter 32
{2}------------------------------------------------
## K070222 3 OF 4 PAGE
Image /page/2/Picture/1 description: The image shows the logo and text for Fresenius Medical Care. Below the logo, the text reads "Fresenius Medical Care" followed by "Fresenius Renal Acute Care (RAC 100) Filter". The last line of text says "510(k) Summary".
#### 2. Does the new device have the same intended use?
YES -- The intended use for the Fresenius Renal Acute Care Filter is equivalent to that of the Gambro Hospal Multiflow 100 Hemofilter, Gambro M100 Hemofilter.
#### Fresenius Renal Acute Care Filter - Intended Use
Intended for use in removing solutes or fluid in renal supportive therapies in an acute setting. This filter can be used for intermittent and continuous therapies.
The RAC 100 Filter is intended for use in the following continuous veno-venous therapies:
- continuous veno-venous hemofiltration (CVVH) .
- continuous veno-venous hemodialysis (CVVHD) .
The RAC 100 Filter is intended for use in the following intermittent therapies, e.g. slow low efficiency dialysis (SLED) or other renal supportive therapies of less than 8 hours duration.
#### Gambro Hospal Multiflow 100 - Intended Use
The Hospal Multiflow 100 is indicated for use whenever continuous veno-venous hemofiltration or hemodialysis is indicated. This can be used for acute hemodialysis and hemofiltration. In these therapies, monitoring of patient vital signs, the therapy delivery system, heparin administration, and clotting times should be performed under the direction of a physician.
#### Gambro M100 - Intended Use
Indicated for use with the Prisma Control Unit in providing continuous fluid management and renal replacement therapies for patients who have acute renal failure, fluid overload, or both.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo and text for Fresenius Medical Care. The text below the logo reads "Fresenius Renal Acute Care (RAC 100) Filter" and "510(k) Summary". The logo is a stylized image of a filter.
#### 3. Does the device have technological characteristics that raise new types of safety or effectiveness questions?
NO -The technological characteristics of the Fresenius Renal Acute Care (RAC 100) Filter are equivalent to those of the Fresenius Optiflux F160NR, Gambro Hospal Multiflow 100 Hemofilter and the Gambro M100 Hemofiter and raise no new types of safety or effectiveness questions.
#### 4. Does descriptive or performance information demonstrate equivalence?
YES - Fresenius Medical Care North America believes that the information provided in this submission clearly describes the Fresenius Renal Acute Care (RAC 100) Filter and that the performance characteristics and other information supplied demonstrates that it is substantially equivalent to the Fresenius Optiflux F160NR, Gambro Hospal Multiflow 100 Hemofilter and the Gambro M100 Hemofilter.
#### F. Safety Summary
The Fresenius Renal Acute Care (RAC 100) Filter is substantially equivalent in construction, design, materials, and intended use to the commercially available Fresenius Optiflux F160NR, Gambro Hospal Multiflow 100 Hemofilter and the Gambro M100 Hemofilter. In addition, testing of the Fresenius Renal Acute Care (RAC 100) Filter indicates that it is safe and effective for its intended use.
### G. General Safety and Effectiveness Concerns
The device labeling contains a package insert, which includes indications for use, cautions and warnings, as well as the general operating instructions required for proper use of the device. This information promotes safe and effective use of the Fresenius Renal Acute Care (RAC 100) Filter.
Janet C Kay
Janet C. Kav RAC Regulatory Affairs Manager
4/19/07
Date
Date
510(k) for Fresenius Renal Acute Care (RAC 100) Filter 34
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Ms. Janet C. Kay Regulatory Affairs Manager Fresenius Medical Care North America 920 Winter Street WALTHAM MA 02451-1457
Re: K070222
Trade/Device Name: Fresenius Renal Acute Care (RAC 100) Filter Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: January 23, 2007 Received: January 24, 2007
Dear Ms. Kay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/4/Picture/10 description: The image shows a circular logo with the text "FDA Centennial" and the years "1906-2006". The logo also features three stars and the letters "FDA" in a stylized font. The text around the circle reads "Product of Indonesia Protected Trademark".
Protecting and Promoting Public Health
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all . the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a geometric design of nested inverted trapezoids. The trapezoids are stacked on top of each other, decreasing in size from top to bottom. The design creates a sense of depth and perspective, with the smallest trapezoid forming the bottom point of the overall shape. The lines are bold and black, contrasting with the white background.
# Fresenius Medical Car Fresenius Renal Acute Care (RAC 100) Filter
Device Name:
Fresenius Renal Acute Care (RAC 100) Filter
#### Indications for Use:
Intended for use in removing solutes or fluid in renal supportive therapies in an acute setting. This filter can be used for intermittent and continuous therapies.
The RAC 100 Filter is intended for use in the following continuous veno-venous therapies:
- continuous veno-venous hemofiltration (CVVH) .
- continuous veno-venous hemodialysis (CVVHD) ●
The RAC 100 Filter is intended for use in the following intermittent therapies, e.g. slow low efficiency dialysis (SLED) or other renal supportive therapies of less than 8 hours duration.
PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Nancy C. Brogdon
(Division Sign Off) Division of Reproductive, Abdominal, and Radiological Devic 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.