The InviroSnap Safety Syringe is used to inject fluids into, or withdraw fluids from, the body. In addition, the InviroSnap Safety Syringe is designed to aid in the prevention of needle stick injuries.
Device Story
InviroSnap Safety Syringe is a disposable, plastic, retractable piston syringe designed to prevent needle stick injuries. It consists of a barrel with a luer lock, plunger, O-ring, stopper, adapter, cannula/needle, and cap. Operation: Healthcare professional draws or injects fluid; after use, fully depresses plunger to engage adapter-needle assembly; pulls back plunger to retract needle into barrel; applies lateral pressure to snap off plunger. Both syringe and plunger are discarded in a sharps container. Used in clinical settings by healthcare professionals. Benefits include reduced risk of accidental needle sticks during disposal.
Clinical Evidence
Bench testing only. Compliance with ISO 594-1:1986 (Luer taper connectors), ISO 594-2:1986 (Locked connectors), ISO 7886-1:1993 (Single use sterile syringes), and ISO 7886-3:2005 (Auto-disable syringes for fixed-dose immunization).
Technological Characteristics
Plastic disposable piston syringe; includes luer lock fitting, retractable needle mechanism, and snap-off plunger. Sterilized by Ethylene Oxide gas. Complies with ISO 594-1, ISO 594-2, ISO 7886-1, and ISO 7886-3 standards.
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from the body, while providing a mechanism to aid in the prevention of needle stick injuries.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
K070203
1 of 1
# 510(k) Summary (As required by 21 CFR 807.92(a))
## Submitter Information A.
Inviro Medical 3235 Satellite Blvd. Bldg 400, Suite 300 Duluth, GA 30096
APR 2 3 2007
| Phone Number: | 770-291-2186 |
|---------------|----------------|
| Fax Number: | 770-291-2181 |
| Contact: | Jim Barley |
| | Director RA/QA |
. .
Trade Name:
InviroSnap Safety Syringe
#### Device Information B.
| | Trade/Proprietary Name: | | InviroSnap Safety Syringe |
|----|-------------------------|----|---------------------------------------|
| | Common name of device: | | Safety Syringe |
| | Classification Name: | | Piston Syringe with Safety<br>Syringe |
| | Product Code: | | 80 MEG |
| | Regulatory Class: | II | |
| | Classification Number: | | 880.5860 |
| | Reason for 510(k): | | Special 510(k) |
| C. | Predicate Device: | | INVIRO 6 cc Syringe |
| | Predicate 510(k) #: | | K941450 |
| | Predicate product code: | | MEG |
{1}------------------------------------------------
K070203
2 & 4
#### D. Device Description
The Inviro Medical InviroSnap Safety Syringe is used to inject medicines and vaccines into, or withdraw fluids from, the body. The InviroSnap Safety Syringe is designed to aid in the prevention of needle stick injuries.
The retractable type piston syringe is a plastic disposable antineedlestick syringe made of the following components:
- l Barrel - The barrel has a scale showing the capacity of the syringe. In addition, the tip of the barrel has a luer lock fitting for the user to attach a needle.
- 2 Plunger - After using the syringe, the plunger is retracted and snapped off leaving the needle in the barrel of the syringe.
- 3 O-Ring - The O-ring minimizes the risk of leakage around the Adapter.
- Stopper The Stopper maintains the fluid in the barrel 4 between the Adapter and Plunger.
- Adapter The Adapter facilitates passage of the fluid 5 between the cannula and the barrel. In addition, the cannula/needle is bonded to the Adapter.
- 6 Cannula - The cannula/needle penetrates the patient's skin to inject/withdraw fluid from the body.
- 7 Cap - Covers the cannula/needle until the syringe is to be used.
After use, the health care professional fully depresses the plunger to engage the Adapter Needle assembly. Once the Adapter Needle Assembly is engaged, pulling back the plunger causes the Adapter and the attached needle to be withdrawn into the safety of the barrel. In this position against the flange, lateral pressure on the plunger results in a controlled fracture of the plunger. Both the syringe and plunger are discarded in a Sharps container.
{2}------------------------------------------------
K670203
3 of 4
#### Device Description D.
The InviroSnap Safety Syringes are sterilized by Ethylene Oxide Gas and supplied sterile in blister pack. One hundred blister packs are packaged in a chipboard box. Each Blister pack and chipboard box is labeled with the contents and the appropriate information per the FDA's Quality System Regulation and Labeling requirements.
#### Statement of Indications for Use E.
The InviroSnap Safety Syringe is used to inject medicine or vaccines into, or withdraw fluids from, the body. In addition, the Inviro Snap Safety Insulin Syringe is designed to aid in the prevention of needle stick injuries.
# Comparison of Required Technological Characteristics: F.
Information was submitted to demonstrate that there are no significant differences in technological characteristics between the InviroSanp 3 ml, 5 ml and 10 ml Safety Syringe and the cited predicate device.
### Summary and Conclusion of Nonclinical and Clinical Tests: G.
The intended use of the InviroSnap 3 ml, 5 ml and 10 ml Safety Syringes is identical to that of the cited predicate device. Any differences in technological characteristics were insignificant and do not raise new issues of safety or effectiveness.
Performance testing consisted of compliance to the applicable sections of the following voluntary standards:
- ISO 594-1:1986. "Syringe, Syringe Needle and Other 1. Medical Apparatus 6% (Lu-Er) Taper Connector -Part One: General Requirement"
{3}------------------------------------------------
КОЛОЗОВ
4. А4
# Summary and Conclusion of Nonclinical and Clinical Tests: G
- ISO 594-2:1986, "Syringe, Syringe Needle and Other 2. Medical Apparatus 6% (Lu-Er) Taper Connector -Part Two: Locked Connector"
- ISO 7886-1:1993, "Single Use Sterile Syringe" 3.
- ISO 7886-3:2005, "Auto-disable syringes for fixed-4. dose immunization
# Conclusion:
The InviroSnap 3 ml, 5 ml and 10 ml Safety Syringes are substantially equivalent to the Inviro 6 ml Syringe in indications for use and technological characteristics.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a stylized caduceus symbol. The caduceus is a symbol of medicine and health, and it is depicted as a staff with two snakes entwined around it and wings at the top.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. James Barley Director of Regulatory Affairs/Quality Assurance Inviro Medical Devices, Incorporated 3235 Satellite Boulevard, Building 400, Suite 300 Duluth, Georgia 30096
APR 2 3 2007
Re: K070203
Trade/Device Name: InviroSnap Safety Syringe, 3ml, 5ml and 10ml Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: March 23, 2007 Received: March 26, 2007
Dear Mr. Barley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Barley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susa Rungs
Chiu Lin, Ph.D.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K070203
# Indications For Use
510(k) Number (if known): K070203
Device Name: InviroSnap Safety Syringes, 3 ml, 5 ml and 10 ml
Indications For Use:
The InviroSnap Safety Syringe is used to inject fluids into, or withdraw fluids from, the body. In addition, the InviroSnap Safety Syringe is designed to aid in the prevention of needle stick injuries.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use __ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ants Vepre
on of Aneculestiology, General Houpliall, Jon Control, Dismal Devices
Page 1 of 1
്(k) Number:_________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.